Index: firmware/App/Services/StateServices/TubeSetInstall.c =================================================================== diff -u -rb314f68347ca0720d370fd5ace385a2d8d92fabe -r05e43e76b6ea947bf51326c43aae77582ad70943 --- firmware/App/Services/StateServices/TubeSetInstall.c (.../TubeSetInstall.c) (revision b314f68347ca0720d370fd5ace385a2d8d92fabe) +++ firmware/App/Services/StateServices/TubeSetInstall.c (.../TubeSetInstall.c) (revision 05e43e76b6ea947bf51326c43aae77582ad70943) @@ -184,7 +184,7 @@ else if ( bpTorque > BP_TORQUE_LIMIT_MNM ) { signalBloodPumpHardStop(); - SET_ALARM_WITH_1_F32_DATA( ALARM_ID_TD_AUTO_LOAD_EJECT_BLOOD_TUBE_SET_FAILURE, bpTorque ); + SET_ALARM_WITH_1_F32_DATA( ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE, bpTorque ); setBloodPumpTargetFlowRate( AUTO_LOAD_BLOOD_FLOW_RATE_ML_MIN, MOTOR_DIR_REVERSE, PUMP_CONTROL_MODE_OPEN_LOOP ); bloodPumpTimerCounter = 0; state = TUBE_SET_INSTALL_STATE_AUTO_LOAD_BACK_OFF; @@ -226,9 +226,9 @@ else if ( TRUE == isPeristalticPumpHome() ) { signalBloodPumpHardStop(); - clearAlarmTD( ALARM_ID_TD_AUTO_LOAD_EJECT_BLOOD_TUBE_SET_FAILURE ); + clearAlarmTD( ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE ); - if ( FALSE == isAlarmActive( ALARM_ID_TD_AUTO_LOAD_EJECT_BLOOD_TUBE_SET_FAILURE ) ) + if ( FALSE == isAlarmActive( ALARM_ID_TD_AUTO_LOAD_EJECT_HIGH_BP_TORQUE ) ) { bloodPumpTimerCounter = 0; state = TUBE_SET_INSTALL_STATE_AWAIT_TUBE_SET_CONFIRMATION;