Index: shared/scripts/configuration/config.py =================================================================== diff -u -r535df219a0721d69e25f36e9a24c98e7b59394b3 -r22fa52de3108551fe5f5c9a4d7246eeb9d103fa7 --- shared/scripts/configuration/config.py (.../config.py) (revision 535df219a0721d69e25f36e9a24c98e7b59394b3) +++ shared/scripts/configuration/config.py (.../config.py) (revision 22fa52de3108551fe5f5c9a4d7246eeb9d103fa7) @@ -55,485 +55,6 @@ POSITIVE_BUFFER = 10 NEGATIVE_BUFFER = -10 -ALARM_ID_AND_MESSAGES = { - 1: "Stuck button POST failure.\nStop or Off button detected to be pressed for at least 1 second during test shortly after power up.", - 2: "HD FPGA POST failure.", - 3: "DG FPGA POST failure.", - 4: "HD Watchdog POST failure.", - 5: "DG Watchdog POST failure.", - 6: "UI communication POST failure.\nUI failed to communicate within a reasonable time after power up.", - 7: "Real-time clock configuration error.", - 8: "HD accelerometer failed POST.", - 9: "DG accelerometer failed POST." , - 10: "RTC or MCU timer inaccurate.", - 11: "DG heaters failed POST.", - 12: "HD firmware image integrity POST test failed.", - 13: "DG firmware image integrity POST test failed.", - 14: "HD blood flow invalid calibration.", - 15: "HD dialysate flow invalid calibration.", - 16: "HD alarm audio failed POST.", - 17: "HD UI POST failed.", - 18: "DG load cells invalid calibration record.", - 19: "DG flow sensors invalid calibration record.", - 20: "DG conductivity sensors invalid calibration record.", - 21: "DG drain line volume invalid calibration record.", - 22: "DG reservoirs invalid calibration record.", - 23: "DG acid concentrate calibration record.", - 24: "DG bicarb concentrate calibration record.", - 25: "DG accelerometer invalid calibration record.", - 26: "HD accelerometer invalid calibration record.", - 27: "HD blood flow invalid calibration record.", - 28: "HD dialyzer flow invalid calibration record.", - 29: "HD heparin force sensor invalid calibration record.", - 30: "HD Software fault.\nSoftware found itself in an unexpected state.", - 31: "Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", - 32: "Blood pump failed motor off check.\nMeasured speed while commanded off.", - 33: "Blood pump failed motor direction check.\nMeasured vs commanded.", - 34: "Blood pump failed rotor speed check.\nMismatch with rotor and motor speeds.", - 35: "Dialysis inlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", - 36: "Dialysis inlet pump failed motor off check.\nMeasured speed while commanded off.", - 37: "Dialysis inlet pump failed motor direction check.\nMeasured vs commanded.", - 38: "Dialysis inlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", - 39: "Dialysis outlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", - 40: "Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off.", - 41: "Dialysis outlet pump failed motor direction check.\nMeasured vs commanded.", - 42: "Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", - 43: "Watchdog expired error.\nWatchdog was not \"pet\" within the appropriate time.", - 44: "User interface communication timeout.", - 45: "Too many bad CRCs detected on received system messages.", - 46: "System message that required acknowledgment was not acknowledged.", - 47: "Ultrafiltration rate is too high error during treatment.", - 48: "Ultrafiltration volume accuracy error during treatment.", - 49: "HD FPGA communication down for too long.", - 50: "FPGA not accepting commanded valve states.", - 51: "Blood pump failed motor speed check.\nMeasured vs commanded.", - 52: "Dialysate inlet pump failed motor speed check.\nMeasured vs commanded.", - 53: "Dialysate outlet pump failed motor speed check.\nMeasured vs commanded.", - 54: "HD critical data integrity check failed.", - 55: "DG critical data integrity check failed.", - 56: "HD accelerometer error (no readings or FPGA reports error).", - 57: "DG accelerometer error (no readings or FPGA reports error).", - 58: "HD valve homing failed.", - 59: "HD valve transition time out.", - 60: "HD valve not functional.", - 61: "HD valve current out of range.", - 62: "HD valve position out of target.", - 63: "FPGA reports failure reading arterial pressure sensor.", - 64: "FPGA reports failure reading venous pressure sensor.", - 65: "HD requests DG command with invalid parameter fault.", - 66: "HD blood leak sensor set point set failure.", - 67: "HD blood pump occlusion self-test failure alarm.", - 68: "HD active reservoir recirculation out of range.", - 69: "HD blood leak sensor invalid calibration record.", - 70: "HD arterial pressure self-test failure alarm." , - 71: "HD venous pressure self-test failure alarm.", - 72: "HD blood flow meter status check self-test failure alarm.", - 73: "HD dialysate flow meter status check self-test failure alarm.", - 74: "HD blood leak detector self-test failure alarm.", - 75: "HD syringe pump self-test failure alarm.", - 76: "HD monitored voltage is out of range.", - 77: "DG monitored voltage is out of range.", - 78: "HD syringe pump direction (from encoder) error.", - 79: "HD syringe pump direction (from controller) error.", - 80: "HD syringe pump fault reported by FPGA.", - 81: "HD syringe pump over travel error.", - 82: "HD syringe pump DAC write failure.", - 83: "HD syringe pump is running while the BP is off.", - 84: "HD venous pressure sensor not being read.", - 85: "HD pump track latch opened alarm.", - 86: "HD BP occlusion sensor not being read.", - 87: "DG heating invalid calibration record.", - 88: "DG concentrate pumps hall sensor out of range.", - 89: "HD BP occlusion sensor error.", - 90: "This alarm ID is available for use.", - 91: "HD No dialysate flow data receive in the last 3 seconds.", - 92: "DG conductivity sensor bad status.", - 93: "This alarm ID is available for use.", - 94: "This alarm ID is available for use.", - 95: "This alarm ID is available for use.", - 96: "Inlet water temperature in high range.", - 97: "Fill conductivity out of range.", - 98: "HD battery communication fault.", - 99: "HD syringe pump stall alarm.", - 100: "HD no cartridge self-test timeout.", - 101: "HD dry self-test timeout.", - 102: "Real-time clock communication error.", - 103: "HD non-volatile calibration group invalid record CRC.", - 104: "Air trap level sensors reporting illegal combination of air/fluid.", - 105: "DG non-volatile system group invalid record CRC.", - 106: "DG inlet water conductivity is greater than threshold.", - 107: "HD reports DG restarted fault.", - 108: "HD syringe pump ADC error.", - 109: "HD syringe pump volume check error.", - 110: "HD syringe pump speed check error.", - 111: "HD syringe pump not stopped in off state error.", - 112: "HD blood leak detector fault.", - 113: "HD arterial air bubble detector self-test failure.", - 114: "HD venous air bubble detector self-test failure.", - 115: "DG temperature sensor out of range.", - 116: "DG temperature sensor ADC out of range.", - 117: "DG primary heater internal temperature out of range.", - 118: "DG primary heater cold junction temperature out of range.", - 119: "DG trimmer heater internal temperature out of range.", - 120: "DG trimmer heater cold junction temperature out of range.", - 121: "DG main primary heater voltage out of range.", - 122: "DG small primary heater voltage out of range.", - 123: "DG trimmer heater voltage out of range.", - 124: "HD end of treatment alarm (high priority).", - 125: "Blood sitting too long after treatment stopped by user alarm (>5 min).", - 126: "HD blood leak detected alarm.", - 127: "Venous pressure too low during treatment." , - 128: "HD venous air bubble detected alarm.", - 129: "HD venous air bubble detected rinseback alarm.", - 130: "Venous pressure too high during treatment.", - 131: "Arterial pressure too low during treatment.", - 132: "Arterial pressure too high during treatment.", - 133: "DG fluid leak detected alarm.", - 134: "HD fluid leak detected alarm." , - 135: "HD experienced large acceleration.", - 136: "DG experienced large acceleration.", - 137: "HD tilt exceeds maximum.", - 138: "DG tilt exceeds maximum.", - 139: "HD AC power lost alarm.", - 140: "Dialysate generator communication timeout.", - 141: "Air trap fill timeout during treatment.", - 142: "Blood pump occlusion detected.", - 143: "DG dialysate temperature sensors drift timeout.", - 144: "DG software configuration record invalid CRC.", - 145: "Concentrate conductivity after adding acid out of range alarm.", - 146: "Dialysate conductivity out of range.", - 147: "Wait for the DG to produce dialysate.", - 148: "Inlet water temperature too high.", - 149: "Alarm ID available for use.", - 150: "Inlet water conductivity in high range.", - 151: "Inlet water conductivity in low range.", - 152: "Inlet water pressure in low range.", - 153: "HD prime completed high priority alarm.", - 154: "EEPROM operations (read, write, erase) failure.", - 155: "DG software configuration record invalid CRC.", - 156: "HW usage data (treatment time in HD and total consumed water in DG) failure.", - 157: "DG chemical disinfect prime acid line timeout.", - 158: "Inlet water temperature is in the low range.", - 159: "Dialysate inlet pump failed flow vs motor speed check.", - 160: "Blood pump rotor speed too high.", - 161: "This alarm ID is available for use.", - 162: "This alarm ID is available for use.", - 163: "HD sees primary load cell for reservoir 1 change too much too fast.", - 164: "HD sees primary load cell for reservoir 2 change too much too fast.", - 165: "HD in treatment rinseback operation timeout.", - 166: "HD in treatment re-circ sub-mode for too long.", - 167: "HD cartridge door opened alarm.", - 168: "HD active reservoir depletion timeout.", - 169: "Dialysate flow rate out of range.", - 170: "HD syringe empty alarm.", - 171: "HD syringe pump occlusion alarm.", - 172: "HD syringe pump not enough Heparin alarm.", - 173: "HD arterial air bubble detected alarm.", - 174: "HD arterial air bubble detected rinseback alarm.", - 175: "HD pump direction status error.", - 176: "HD active reservoir is full during treatment and DG not ready to switch reservoirs.", - 177: "DG software fault.\nSoftware found itself in an unexpected state.", - 178: "HD communication timeout.", - 179: "DG FPGA communication down for too long.", - 180: "DG load cell ADC error.", - 181: "DG load cells weight out of range for tare.", - 182: "DG load cells invalid calibration.", - 183: "DG invalid load cell value.", - 184: "Inlet or Outlet UV reactor not healthy.", - 185: "DG fan RPM out of range.", - 186: "DG concentrate pump fault.", - 187: "DG concentrate pump CP1 speed control error.", - 188: "DG concentrate pump CP2 speed control error.", - 189: "DG drain pump RPM out of range.", - 190: "DG drain pump off fault.", - 191: "DG flow rate out of upper range.", - 192: "DG flow rate out of lower range.", - 193: "RO pump flow rate out of range.", - 194: "DG RO pump duty cycle out of range.", - 195: "DG RO pump pressure out of range.", - 196: "DG temperature sensor error flag fault.", - 197: "DG temperature sensors ADC fault.", - 198: "DG heaters cold junction temperature out of range.", - 199: "DG heaters fault.", - 200: "DG thermistors/sensors temperature out of range.", - 201: "Inlet water quality is bad (temperature, conductivity, pressure).", - 202: "Inlet water pressure fault.", - 203: "Pressure sensor fault.", - 204: "RO rejection ratio out of range.", - 205: "Conductivity sensor fault.", - 206: "DG dialysate fill runs out of time.", - 207: "DG flow meter check failure alarm.", - 208: "DG primary heater on with no flow time out.", - 209: "DG drain circulation line timeout.", - 210: "HD battery pack detected an error.", - 211: "Blood sitting too long warning (>4 min).", - 212: "HD end of treatment alarm (user not acting to end treatment).", - 213: "HD prime completed medium priority alarm.", - 214: "HD Saline bolus volume check failure.", - 215: "HD rinseback volume check failure.", - 216: "HD end treatment sub-mode timeout alarm.", - 217: "HD blood prime volume check failure.", - 218: "HD syringe detected alarm.", - 219: "HD syringe pump syringe removed alarm.", - 220: "HD battery pack SOC is too low.", - 221: "HD Saline bag is empty.", - 222: "HD dialysate temperature too high alarm.", - 223: "HD dialysate temperature too low alarm.", - 224: "Treatment stopped by user action.", - 225: "HD end of treatment warning.", - 226: "HD prime completed low priority alarm.", - 227: "HD prime out of time alarm.", - 228: "HD time out on prime sub-mode purge air state.", - 229: "HD prime dialysate dialyzer time out alarm.", - 230: "HD prime dialysate bypass time out alarm.", - 231: "HD pre-treatment mode dry self-tests failure.", - 232: "HD pre-treatment mode wet self-tests failure.", - 233: "RTC battery low.", - 234: "RTC (i.e read time) or RAM operations failure (read or write).", - 235: "HD in treatment stopped sub-mode after rinseback completed (no escalation).", - 236: "HD needs new cartridge to be installed.", - 237: "HD prime saline dialyzer time out alarm.", - 238: "HD no cartridge loaded or installed improperly alarm.", - 239: "HD fail to remove cartridge alarm.", - 240: "Bicarb conductivity out of range during bicarb pump check alarm.", - 241: "DG reservoir drain time out.", - 242: "DG reservoir fill time out.", - 243: "DG reservoir leak time out.", - 244: "DG temperature sensors difference out of range.", - 245: "DG heat disinfect target temperature time out (could not reach to temperature).", - 246: "DG heat disinfect inlet pressure and temperature sensors out of range.", - 247: "DG heat disinfect inlet conductivity and temperature out of range.", - 248: "DG chemical disinfect could not reach to target temperature.", - 249: "DG chemical disinfect inlet pressure and temperature sensors out of range.", - 250: "DG chemical disinfect inlet conductivity and temperature out of range.", - 251: "HD invalid system record.", - 252: "HD invalid service record.", - 253: "DG invalid system record.", - 254: "DG invalid service record.", - 255: "HD and UI software builds are not compatible.", - 256: "HD and DG software builds are not compatible.", - 257: "DG FPGA power out timeout.", - 258: "HD FPGA power out timeout.", - 259: "HD temperatures out of range.", - 260: "UI POST Application Integrity (Sha256Sum) failure.", - 261: "UI POST CANBus failure.", - 262: "UI POST Display failure.", - 263: "UI POST Touch failure.", - 264: "UI POST SD-Card failure.", - 265: "UI POST RTC failure.", - 266: "UI POST WiFi failure.", - 267: "UI POST Bluetooth failure.", - 268: "UI POST Ethernet failure.", - 269: "UI POST Sound failure.", - 270: "HD POST Safety Shutdown failure.", - 271: "DG POST Safety Shutdown failure.", - 272: "HD Fan RPM out of range.", - 273: "HD measured blood pump flow rate is out of range.", - 274: "HD measured dialysate inlet pump flow rate is out of range.", - 275: "HD arterial pressure sensor is reading out of range.", - 276: "HD venous pressure sensor is reading out of range.", - 277: "HD BP occlusion sensor is reading out of range.", - 278: "HD active reservoir weight out of range.", - 279: "DG dialysate drain time out.", - 280: "HD arterial pressure sensor read timeout error.", - 281: "Acid concentration bottle low volume alarm.", - 282: "Bicarbonate concentration bottle low volume alarm.", - 283: "DG load cells weight out of range.", - 284: "DG load cells primary/back up drift out of range.", - 285: "HD treatment recirculate timeout warning.", - 286: "HD treatment rinseback complete timeout warning.", - 287: "HD processor clock speed checks against FPGA clock failure.", - 288: "DG trimmer heater on with flow timeout.", - 289: "DG dialysate or concentrate caps not closed." - } - -ALARM_PRIORITIES = {1: "LOW", 2: "MEDIUM", 3: "HIGH"} -NUM_OF_ALARM_PRIORITIES = 4 # Total number of alarm priorities -ACCEPTED = True -REJECTED = False -ALARM_LIST_TITLE = "Alarm list" - - -NUM_OF_REQUEST_REJECT_REASONS = 43 -NUM_OF_MSG_ID = 289 -UNDEFINED_ALARM_ID_MSG = "ALARM_ID_UNDEFINED" -ALARM_TITLE = "Alarm" - - -ALARM_ID_NO_ALARM = 0 -ALARM_ID_OUT_OF_RANGE = 290 - -ALARM_PRIORITY_OPTIONS = { - 0 : 'ALARM_PRIORITY_NONE' , - 1 : 'ALARM_PRIORITY_LOW' , - 2 : 'ALARM_PRIORITY_MEDIUM', - 3 : 'ALARM_PRIORITY_HIGH', -} - -# alarm colors -ALARMS_COLORS_HEADER = { - "ALARM_PRIORITY_NONE" : "#438feb", - "ALARM_PRIORITY_LOW" : "#db8f00", - "ALARM_PRIORITY_MEDIUM" : "#db8f00", - "ALARM_PRIORITY_HIGH" : "#831913" -} - -ALARMS_COLORS_BODY = { - "ALARM_PRIORITY_NONE" : "#254670", - "ALARM_PRIORITY_LOW" : "#f5a623", - "ALARM_PRIORITY_MEDIUM" : "#f5a623", - "ALARM_PRIORITY_HIGH" : "#c53b33" -} - -ALARM_MUTE_FLAG = 0B0000001000000000 -ALARM_UNMUTE_FLAG = 0B0000000000000000 - -ALARM_BUTTONS_OK = 0B0000000010111000 - -ALARM_BUTTONS_RESUME = 0B0000000000110000 -ALARM_BUTTONS_RESUME_RINSEBACK = 0B0000000000100000 -ALARM_BUTTONS_RESUME_END = 0B0000000000010000 - -ALARM_BUTTONS_RINSEBACK = 0B0000000000101000 -ALARM_BUTTONS_RINSEBACK_RESUME = ALARM_BUTTONS_RESUME_RINSEBACK -ALARM_BUTTONS_RINSEBACK_END = 0B0000000000001000 - -ALARM_BUTTONS_END = 0B0000000000011000 -ALARM_BUTTONS_END_RESUME = ALARM_BUTTONS_RESUME_END -ALARM_BUTTONS_END_RINSEBACK = ALARM_BUTTONS_RINSEBACK_END - -ALARM_BUTTONS_RESUME_RINSEBACK_END = 0B0000000000000000 -ALARM_BUTTONS_RINSEBACK_RESUME_END = ALARM_BUTTONS_RESUME_RINSEBACK_END -ALARM_BUTTONS_END_RESUME_RINSEBACK = ALARM_BUTTONS_RESUME_RINSEBACK_END - -CURRENT_COLOR = '#000000' -COMPLETE_COLOR= '#4290ec' -ENABLED_COLOR = '#fcfcfc' -INCOMPLETE_COLOR = '#607a91' -OUT_OF_RANGE_COLOR = "#c53b33" - -#standby mode -GOODMORNING_START_TIME_SEC = 0 -GOODEVENING_START_TIME_SEC = 43200 -BLOOD_PRIMING_TEXT = "Blood Priming" -LIQUID_UNIT = "mL" -BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" -SALINE_BOLUS_CUMULATIVE = 300 -SALINE_UNIT = "mL" - -UF_MINIMUM_VALUE = 0 -UF_MAXIMUM_VALUE = 600 -ULTRAFILTRATION_PAUSED_TEXT = "Ultrafiltration Paused" -RESUME_TEXT = "RESUME ULTRAFILTRATION" -EDIT_TEXT = "EDIT ULTRAFILTRATION VOLUME" -TITLE_TEXT = "ULTRAFILTRATION VOLUME (L)" - -#main treatment pressure -PRESSURE_TITLE = "PRESSURES" -ARTERIAL_TITLE = "Arterial" -VENOUS_TITLE = "Venous" -VENOUS_UOM = "mmHg" -ARTERIAL_UOM = "mmHg" -LOW_TEXT = "LOW" -HIGH_TEXT = "HIGH" -ACCEPTED = True -REJECTED = False -NUM_OF_REQUEST_REJECT_REASONS = 43 - -#pressure -PRESSURE_STEPS = 10 -ARTERIAL_PRESSURE_MINIMUM = -400 -ARTERIAL_PRESSURE_MAXIMUM = +600 -ARTERIAL_PRESSURE_LOW_MIN = -300 -ARTERIAL_PRESSURE_LOW_DEF = -300 -ARTERIAL_PRESSURE_LOW_MAX = +200 -ARTERIAL_PRESSURE_HIGH_MIN = -300 -ARTERIAL_PRESSURE_HIGH_DEF = +100 -ARTERIAL_PRESSURE_HIGH_MAX = +200 - -VENOUS_PRESSURE_MINIMUM = -100 -VENOUS_PRESSURE_MAXIMUM = +700 -VENOUS_PRESSURE_LOW_MIN = -100 -VENOUS_PRESSURE_LOW_DEF = -100 -VENOUS_PRESSURE_LOW_MAX = +600 -VENOUS_PRESSURE_HIGH_MIN = +100 -VENOUS_PRESSURE_HIGH_DEF = +400 -VENOUS_PRESSURE_HIGH_MAX = +600 - -#main_treatment_flows -BLOOD_FLOW_MINIMUM_VALUE = 100 -BLOOD_FLOW_BUFFER = 25 -FLOW_SLIDER_WIDTH_BUFFER = 37 -BLOOD_FLOW_MAXIMUM_VALUE = 500 -DIALYSATE_FLOW_MINIMUM_VALUE = 0 -DIALYSATE_FLOW_MAXIMUM_VALUE = 0 # Seems to be bug. need change -FLOW_UNIT = "mL/min" -BLOOD_FLOW_TEXT = "blood" -DIALYSATE_FLOW_TEXT = "dialysate" - -#main treatment screen -VITALS_TITLE = "VITALS" -OUT_OF_RANGE_COLOR = "#c53b33" -IN_RANGE_COLOR = "#fcfcfc" -SYSTOLIC_LOWER_LIMIT = 60 -SYSTOLIC_UPPER_LIMIT = 250 -DIASTOLIC_LOWER_LIMIT = 40 -DIASTOLIC_UPPER_LIMIT = 200 -HEART_RATE_LOWER_LIMIT = 40 -HEART_RATE_UPPER_LIMIT = 180 -HEART_RATE_TITLE = "Heart Rate" -BLOOD_PRESSURE_TITLE = "Blood Pressure" -BLOOD_PRESSURE_UNIT = "mmHg" -HEART_RATE_UNIT = "BPM" -BLOOD_PRIMING_RANGE = 300 -BLOOD_PRIMING_TARGET_MAXIMUM = 300 - - -#tst_treatment_monitoring -FLOWS_TEXT = "FLOWS" -BLOOD_UNIT = "mL/min" -DIALYSATE_TEXT = "Dialysate" -BLOOD_TEXT = "Blood" -VITALS_TEXT = "VITALS" -BLOOD_PRESSURE_TEXT = "mmHg" -HEARTRATE_UNIT = "BPM" -PRESSURE_TEXT = "PRESSURE (mmHg)" -ARTEREAL_TEXT = "Arterial" -VENOUS_TEXT = "Venous" -TIME_DURATION_TEXT = "Time Remaining" -ULTRAFILTERATION_TEXT = "ULTRAFILTRATION VOLUME (L)" -SALINE_BOLUS = "SALINE BOLUS" -VOLUME_DELIVERED = "Volume Delivered" -CUMULATIVE_DELIVERED = "Cumulative Delivered" -HEPARIN_TEXT = "HEPARIN" -VOLUME_DELIVERED = "Volume Delivered" - -#patient id -PATIENT_ID = { - "qwerty" : "qwerty", - "sampleid" : "sampleid", - "abcdefghijklmnop" : "abcdefghijklmnop", - "patientid" : "patientid", - "username" : "username", - "sA@MpLE#123" : "sAMpLE123", - "A@BH&321" : "ABH321", - "5432@qwerty" : "5432qwerty" -} - -KEYBOARD_SWITCH = ["ABC", "&123"] - - -#main-treatment heparin -HEPARIN_UNIT = "mL" -HEPARIN_DELIVERY_TEXT = "HEPARIN DELIVERY" -HEPARIN_DELIVERY_OFF_TEXT = "OFF" -HEPARIN_STOP_TEXT = "Delivery Stopped" -HEPARIN_RESUME_DELIVERY_TEXT = "RESUME DELIVERY" -HEPARIN_BOLUS_ACTIVE_TEXT = "Bolus Active" -HEPARIN_PAUSE_DELIVERY_TEXT = "PAUSE DELIVERY" -HEPARIN_MAXIMUM_VOLUME_DELIVERED_TEXT = "Maximum Cumulative Heparin Volume Delivered" -HEPARIN_SYRINGE_EMPTY_TEXT = "Syringe Empty" - #Time_Duration CREATE_TREATMENT_PARAMETERS = ["Blood Flow Rate", "Dialysate Flow Rate", "Duration","Heparin Dispensing Rate", "Heparin Bolus Volume","Heparin Stop Time", "Saline Bolus","Dialysate Temperature","Blood Pressure Measurement Interval", @@ -589,159 +110,4 @@ "arterial_pressure_limit_high" : [20, 10, 11.931], "blood_pressure_measurement_interval" : [0, 5, 62], "rinseback_flow_rate" : [50, 25, 187], -} - -#ultrafilteration volume based on flow parameters. Python list contains Rotational speed, Motor speed, -# MC speed, MC current and PWM. -TREATMENT_ULTRAFILTRATION_FLOW_PARAMETERS = { - 0 : [100, 1, 1, 1, 1, 1, 1], - 1 : [550, 24, 32, 425, 85, 18, 36], - 2 : [80, 125, 187, 257, 530, 458, 985], - 3 : [100, 0, 0, 0, 0, 0, 0], - 4 : [220, 1.45, 1.45, 1.45, 1.45, 1.45, 1.45], - } - -#dictionary contains uf maximum and minimum value's to set. -TREATMENT_ULTRAFILTRATION_TESTING_OPTION = { - 0 : {"uf_maximum" : 600, "uf_minimum" : 0}, - 1 : {"uf_maximum" : 500, "uf_minimum" : 0}, - 2 : {"uf_maximum" : 550, "uf_minimum" : 0}, - 3 : {"uf_maximum" : 300, "uf_minimum" : 0}, - 4 : {"uf_maximum" : 280, "uf_minimum" : 0}, - 5 : {"uf_maximum" : 765, "uf_minimum" : 500}, - 6 : {"uf_maximum" : 680, "uf_minimum" : 450}, - 7 : {"uf_maximum" : 480, "uf_minimum" : 380}, - 8 : {"uf_maximum" : 390, "uf_minimum" : 675}, - 9 : {"uf_maximum" : 290, "uf_minimum" : 125}, - } - -#tst_pretreatment_screens -PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] -NUM_OF_PRETREATMENT_BULLETS = len(PRE_TREATMENT_SCREENS) - - -#dictionary contains saline cumulative and saline delivered value -#key : saline cumulative value. -#value: saline delivered value -SALINE_BOLUS_VALUES = { - 0 : 0, - 1.0 : 1.0, - 125.56 : 100.56, - 165.87 : 145.87, - 220.65 : 220.65, - 325.25 : 300.25, - 400.86 : 380.86, - 500.00 : 500.25, - 634.76 : 634.76, - 823.65 : 800.65, - 999.99 : 999.99, - 1000 : 1000 -} - - -#tst_pretreatment_screens color palettes -CURRENT_COLOR = '#000000' -COMPLETE_COLOR= '#4290ec' -ENABLED_COLOR = '#fcfcfc' -INCOMPLETE_COLOR = '#607a91' - -CONSUMABLE_SELF_TEST_TEXT = "Consumables Self Test" -BICARB_PUMP_CHECK_TEXT = "BiCarb Pump Check" -ACID_PUMP_CHECK_TEXT = "Acid Pump Check" -SELF_TEST_COMPLETE_TEXT = "Self Test Complete!" - - -#pre-treatment_priming -NOT_VISIBLE = False -VISIBLE = True -FOCUS = True -NOT_IN_FOCUS = False -BUSY = "Busy" -CHECK = "Check" -BEGIN_PRIME_TITLE = "Begin Prime" -PRIMING_TITLE = "Priming" -SYSTEM_SELF_TEST_TITLE = "System Self Test" -FILTER_FLUSH_TITLE = "Filter Flush" -TIME_OUT_DURATION_300 = 300 -TIME_OUT_DURATION_200 = 200 -TIME_OUT_DURATION_100 = 100 -COUNT_DOWN_TIME_100 = 100 -MINIMUM_COUNTDOWN_TIME = 0 -ENABLED = True -DISABLED = False -CONSUMABLE_SELF_TEST_BICARB_PUMP_CHECK_STATE = 3 -LOOD_PRIMING_TEXT = "Blood Priming" -SALINE_UNIT = "mL" -BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" - -#Pre-Treatment Disposables -NEXT_BUTTON_TEXT = "NEXT" -BACK_BUTTON_TEXT = "BACK" -ENABLED = True -DISABLED = False -CONFIRM_BUTTON_TEXT = "CONFIRM" - -#Filter Flush - -#Pre_treatment_water_sample -SAMPLE_TEXT = "Sample" -CREATE_TEXT = "Create" -NEXT_BUTTON_TEXT = "NEXT" -WATER_SAMPLE_BUTTON_TEXT = "WATER SAMPLE" -BACK_BUTTON_TEXT = "BACK" -PASS_TEXT = "PASS" -FAIL_TEXT = "FAIL" -OK_TEXT = "OK" - -# pretreatment_patient_connection -SYSTOLIC_PRESSSURE_120 = 120 -SYSTOLIC_PRESSSURE_113 = 113 -SYSTOLIC_PRESSSURE_150 = 150 -SYSTOLIC_PRESSSURE_175 = 175 -SYSTOLIC_PRESSSURE_200 = 200 -SYSTOLIC_PRESSSURE_200 = 200 -SYSTOLIC_PRESSSURE_251 = 251 -SYSTOLIC_PRESSSURE_59 = 59 -SYSTOLIC_PRESSSURE_15 = 15 -DIASTOLIC_PRESSSURE_25 = 25 -DIASTOLIC_PRESSSURE_39 = 39 -DIASTOLIC_PRESSSURE_74 = 74 -DIASTOLIC_PRESSSURE_80 = 80 -DIASTOLIC_PRESSSURE_85 = 85 -DIASTOLIC_PRESSSURE_60 = 60 -DIASTOLIC_PRESSSURE_150 = 150 -DIASTOLIC_PRESSSURE_200 = 200 -DIASTOLIC_PRESSSURE_201 = 201 -HEART_RATE_VAL_39 = 39 -HEART_RATE_VAL_60 = 60 -HEART_RATE_VAL_70 = 70 -HEART_RATE_VAL_85 = 85 -HEART_RATE_VAL_100 = 100 -HEART_RATE_VAL_101 = 101 -HEART_RATE_VAL_110 = 110 -HEART_RATE_VAL_181 = 181 -HEART_RATE_VAL_200 = 200 -SYSTOLIC_LOWER_LIMIT = 60 -SYSTOLIC_UPPER_LIMIT = 250 -DIASTOLIC_LOWER_LIMIT = 40 -DIASTOLIC_UPPER_LIMIT = 200 -HEART_RATE_LOWER_LIMIT = 40 -HEART_RATE_UPPER_LIMIT = 180 -UF_MINIMUM_SLIDER_WIDTH = -2 -UF_MAXIMUM_SLIDER_WIDTH = 638.00 -UF_VALID_RANGE = 700 -TRAINING_INDICATOR_SCREEN = 4 -TREATMENT_ULTRAFILTERATION_TESTING_OPTION = { - "OPTION_1" : {"uf_maximum" : 800, "uf_minimum" : 0}, - "OPTION_2" : {"uf_maximum" : 500, "uf_minimum" : 0}, - "OPTION_3" : {"uf_maximum" : 550, "uf_minimum" : 0}, - "OPTION_4" : {"uf_maximum" : 300, "uf_minimum" : 0}, - "OPTION_5" : {"uf_maximum" : 280, "uf_minimum" : 0}, - "OPTION_6" : {"uf_maximum" : 765, "uf_minimum" : 0}, - "OPTION_7" : {"uf_maximum" : 680, "uf_minimum" : 0}, - "OPTION_8" : {"uf_maximum" : 1500, "uf_minimum" : 0}, - "OPTION_9" : {"uf_maximum" : 390, "uf_minimum" : 0}, - "OPTION_10" : {"uf_maximum" : 800, "uf_minimum" : 0}, - "OPTION_11" : {"uf_maximum" : 1000, "uf_minimum" : 0}, - "OPTION_12" : {"uf_maximum" : 1250, "uf_minimum" : 0}, - } \ No newline at end of file +} \ No newline at end of file