Index: shared/scripts/configuration/config.py =================================================================== diff -u -r7f5d92792f49814d405ae6266af82b19cd999d4b -ra4f1e264727015d92baa065dde52c33c0bc33ef3 --- shared/scripts/configuration/config.py (.../config.py) (revision 7f5d92792f49814d405ae6266af82b19cd999d4b) +++ shared/scripts/configuration/config.py (.../config.py) (revision a4f1e264727015d92baa065dde52c33c0bc33ef3) @@ -19,7 +19,6 @@ AUT_NAME = "denaliSquish" COMMON_PATH = os.environ['HOME']+"/Projects" - #create custom treatment HIGHER_VALUE_OF_BLOOD_FLOW_RATE = 500 @@ -37,6 +36,491 @@ POSITIVE_BUFFER = 10 NEGATIVE_BUFFER = -10 +ALARM_ID_AND_MESSAGES = { + 1: "Stuck button POST failure.\nStop or Off button detected to be pressed for at least 1 second during test shortly after power up.", + 2: "HD FPGA POST failure.", + 3: "DG FPGA POST failure.", + 4: "HD Watchdog POST failure.", + 5: "DG Watchdog POST failure.", + 6: "UI communication POST failure.\nUI failed to communicate within a reasonable time after power up.", + 7: "Real-time clock configuration error.", + 8: "HD accelerometer failed POST.", + 9: "DG accelerometer failed POST." , + 10: "RTC or MCU timer inaccurate.", + 11: "DG heaters failed POST.", + 12: "HD firmware image integrity POST test failed.", + 13: "DG firmware image integrity POST test failed.", + 14: "HD blood flow invalid calibration.", + 15: "HD dialysate flow invalid calibration.", + 16: "HD alarm audio failed POST.", + 17: "HD UI POST failed.", + 18: "DG load cells invalid calibration record.", + 19: "DG flow sensors invalid calibration record.", + 20: "DG conductivity sensors invalid calibration record.", + 21: "DG drain line volume invalid calibration record.", + 22: "DG reservoirs invalid calibration record.", + 23: "DG acid concentrate calibration record.", + 24: "DG bicarb concentrate calibration record.", + 25: "DG accelerometer invalid calibration record.", + 26: "HD accelerometer invalid calibration record.", + 27: "HD blood flow invalid calibration record.", + 28: "HD dialyzer flow invalid calibration record.", + 29: "HD heparin force sensor invalid calibration record.", + 30: "HD Software fault.\nSoftware found itself in an unexpected state.", + 31: "Blood pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 32: "Blood pump failed motor off check.\nMeasured speed while commanded off.", + 33: "Blood pump failed motor direction check.\nMeasured vs commanded.", + 34: "Blood pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 35: "Dialysis inlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 36: "Dialysis inlet pump failed motor off check.\nMeasured speed while commanded off.", + 37: "Dialysis inlet pump failed motor direction check.\nMeasured vs commanded.", + 38: "Dialysis inlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 39: "Dialysis outlet pump failed motor controller current check.\nToo high when pump should be off or out of range when pump should be running.", + 40: "Dialysis outlet pump failed motor off check.\nMeasured speed while commanded off.", + 41: "Dialysis outlet pump failed motor direction check.\nMeasured vs commanded.", + 42: "Dialysis outlet pump failed rotor speed check.\nMismatch with rotor and motor speeds.", + 43: "Watchdog expired error.\nWatchdog was not \"pet\" within the appropriate time.", + 44: "User interface communication timeout.", + 45: "Too many bad CRCs detected on received system messages.", + 46: "System message that required acknowledgment was not acknowledged.", + 47: "Ultrafiltration rate is too high error during treatment.", + 48: "Ultrafiltration volume accuracy error during treatment.", + 49: "HD FPGA communication down for too long.", + 50: "FPGA not accepting commanded valve states.", + 51: "Blood pump failed motor speed check.\nMeasured vs commanded.", + 52: "Dialysate inlet pump failed motor speed check.\nMeasured vs commanded.", + 53: "Dialysate outlet pump failed motor speed check.\nMeasured vs commanded.", + 54: "HD critical data integrity check failed.", + 55: "DG critical data integrity check failed.", + 56: "HD accelerometer error (no readings or FPGA reports error).", + 57: "DG accelerometer error (no readings or FPGA reports error).", + 58: "HD valve homing failed.", + 59: "HD valve transition time out.", + 60: "HD valve not functional.", + 61: "HD valve current out of range.", + 62: "HD valve position out of target.", + 63: "FPGA reports failure reading arterial pressure sensor.", + 64: "FPGA reports failure reading venous pressure sensor.", + 65: "HD requests DG command with invalid parameter fault.", + 66: "HD blood leak sensor set point set failure.", + 67: "HD blood pump occlusion self-test failure alarm.", + 68: "HD active reservoir recirculation out of range.", + 69: "HD blood leak sensor invalid calibration record.", + 70: "HD arterial pressure self-test failure alarm." , + 71: "HD venous pressure self-test failure alarm.", + 72: "HD blood flow meter status check self-test failure alarm.", + 73: "HD dialysate flow meter status check self-test failure alarm.", + 74: "HD blood leak detector self-test failure alarm.", + 75: "HD syringe pump self-test failure alarm.", + 76: "HD monitored voltage is out of range.", + 77: "DG monitored voltage is out of range.", + 78: "HD syringe pump direction (from encoder) error.", + 79: "HD syringe pump direction (from controller) error.", + 80: "HD syringe pump fault reported by FPGA.", + 81: "HD syringe pump over travel error.", + 82: "HD syringe pump DAC write failure.", + 83: "HD syringe pump is running while the BP is off.", + 84: "HD venous pressure sensor not being read.", + 85: "HD pump track latch opened alarm.", + 86: "HD BP occlusion sensor not being read.", + 87: "DG heating invalid calibration record.", + 88: "DG concentrate pumps hall sensor out of range.", + 89: "HD BP occlusion sensor error.", + 90: "This alarm ID is available for use.", + 91: "HD No dialysate flow data receive in the last 3 seconds.", + 92: "DG conductivity sensor bad status.", + 93: "This alarm ID is available for use.", + 94: "This alarm ID is available for use.", + 95: "This alarm ID is available for use.", + 96: "This alarm ID is available for use.", + 97: "This alarm ID is available for use.", + 98: "HD battery communication fault.", + 99: "HD syringe pump stall alarm.", + 100: "HD no cartridge self-test timeout.", + 101: "HD dry self-test timeout.", + 102: "Real-time clock communication error.", + 103: "HD non-volatile calibration group invalid record CRC.", + 104: "Air trap level sensors reporting illegal combination of air/fluid.", + 105: "DG non-volatile system group invalid record CRC.", + 106: "This alarm ID is available for use.", + 107: "HD reports DG restarted fault.", + 108: "HD syringe pump ADC error.", + 109: "HD syringe pump volume check error.", + 110: "HD syringe pump speed check error.", + 111: "HD syringe pump not stopped in off state error.", + 112: "HD blood leak detector fault.", + 113: "HD arterial air bubble detector self-test failure.", + 114: "HD venous air bubble detector self-test failure.", + 115: "DG temperature sensor out of range.", + 116: "DG temperature sensor ADC out of range.", + 117: "This alarm ID is available for use.", + 118: "This alarm ID is available for use.", + 119: "This alarm ID is available for use.", + 120: "This alarm ID is available for use.", + 121: "DG main primary heater voltage out of range.", + 122: "DG small primary heater voltage out of range.", + 123: "DG trimmer heater voltage out of range.", + 124: "HD end of treatment alarm (high priority - user still not acting to end treatment).", + 125: "Blood sitting too long after treatment stopped by user alarm (>5 min).", + 126: "HD blood leak detected alarm.", + 127: "Venous pressure too low during treatment." , + 128: "HD venous air bubble detected alarm.", + 129: "HD venous air bubble detected rinseback alarm.", + 130: "Venous pressure too high during treatment.", + 131: "Arterial pressure too low during treatment.", + 132: "Arterial pressure too high during treatment.", + 133: "DG fluid leak detected alarm.", + 134: "HD fluid leak detected alarm." , + 135: "HD experienced large acceleration.", + 136: "DG experienced large acceleration.", + 137: "HD tilt exceeds maximum.", + 138: "DG tilt exceeds maximum.", + 139: "HD AC power lost alarm.", + 140: "Dialysate generator communication timeout.", + 141: "Air trap fill timeout during treatment.", + 142: "Blood pump occlusion detected.", + 143: "DG dialysate temperature sensors drift timeout.", + 144: "DG software configuration record invalid CRC.", + 145: "Concentrate conductivity after adding acid out of range alarm.", + 146: "Dialysate conductivity out of range.", + 147: "DG dialysate generation conductivity fault alarm.", + 148: "Inlet water temperature too high.", + 149: "Inlet water temperature too low.", + 150: "Inlet water conductivity too high.", + 151: "Inlet water conductivity too low.", + 152: "Inlet water pressure too low.", + 153: "HD prime completed high priority alarm.", + 154: "EEPROM operations (read, write, erase) failure.", + 155: "DG software configuration record invalid CRC.", + 156: "HW usage data (treatment time in HD and total consumed water in DG) failure.", + 157: "DG chemical disinfect prime acid line timeout.", + 158: "This alarm ID is available for use.", + 159: "Dialysate inlet pump failed flow vs motor speed check.", + 160: "Blood pump rotor speed too high.", + 161: "This alarm ID is available for use.", + 162: "This alarm ID is available for use.", + 163: "HD sees primary load cell for reservoir 1 change too much too fast.", + 164: "HD sees primary load cell for reservoir 2 change too much too fast.", + 165: "HD in treatment rinseback operation timeout.", + 166: "HD in treatment re-circ sub-mode for too long.", + 167: "HD cartridge door opened alarm.", + 168: "HD active reservoir depletion timeout.", + 169: "Dialysate flow rate out of range.", + 170: "HD syringe empty alarm.", + 171: "HD syringe pump occlusion alarm.", + 172: "HD syringe pump not enough Heparin alarm.", + 173: "HD arterial air bubble detected alarm.", + 174: "HD arterial air bubble detected rinseback alarm.", + 175: "HD pump direction status error.", + 176: "HD active reservoir is full during treatment and DG not ready to switch reservoirs.", + 177: "DG software fault.\nSoftware found itself in an unexpected state.", + 178: "HD communication timeout.", + 179: "DG FPGA communication down for too long.", + 180: "DG load cell ADC error.", + 181: "DG load cells weight out of range for tare.", + 182: "DG load cells invalid calibration.", + 183: "DG invalid load cell value.", + 184: "Inlet or Outlet UV reactor not healthy.", + 185: "DG fan RPM out of range.", + 186: "DG concentrate pump fault.", + 187: "DG concentrate pump CP1 speed control error.", + 188: "DG concentrate pump CP2 speed control error.", + 189: "DG drain pump RPM out of range.", + 190: "DG drain pump off fault.", + 191: "DG flow rate out of upper range.", + 192: "DG flow rate out of lower range.", + 193: "RO pump flow rate out of range.", + 194: "DG RO pump duty cycle out of range.", + 195: "DG RO pump pressure out of range.", + 196: "DG temperature sensor error flag fault.", + 197: "DG temperature sensors ADC fault.", + 198: "DG heaters cold junction temperature out of range.", + 199: "DG heaters fault.", + 200: "DG thermistors/sensors temperature out of range.", + 201: "Inlet water quality is bad (temperature, conductivity, pressure).", + 202: "Inlet water pressure fault.", + 203: "Pressure sensor fault.", + 204: "RO rejection ratio out of range.", + 205: "Conductivity sensor fault.", + 206: "DG dialysate fill runs out of time.", + 207: "DG flow meter check failure alarm.", + 208: "DG primary heater on with no flow time out.", + 209: "DG drain circulation line timeout.", + 210: "HD battery pack detected an error.", + 211: "Blood sitting too long warning (>4 min).", + 212: "HD end of treatment alarm (user not acting to end treatment).", + 213: "HD prime completed medium priority alarm.", + 214: "HD Saline bolus volume check failure.", + 215: "HD rinseback volume check failure.", + 216: "HD end treatment sub-mode timeout alarm.", + 217: "HD blood prime volume check failure.", + 218: "HD syringe detected alarm.", + 219: "HD syringe pump syringe removed alarm.", + 220: "HD battery pack SOC is too low.", + 221: "HD Saline bag is empty.", + 222: "HD dialysate temperature too high alarm.", + 223: "HD dialysate temperature too low alarm.", + 224: "Treatment stopped by user action - pressed stop button.", + 225: "HD end of treatment warning.", + 226: "HD prime completed low priority alarm.", + 227: "HD prime out of time alarm.", + 228: "HD time out on prime sub-mode purge air state.", + 229: "HD prime dialysate dialyzer time out alarm.", + 230: "HD prime dialysate bypass time out alarm.", + 231: "HD pre-treatment mode dry self-tests failure.", + 232: "HD pre-treatment mode wet self-tests failure.", + 233: "RTC battery low.", + 234: "RTC (i.e read time) or RAM operations failure (read or write).", + 235: "HD in treatment stopped sub-mode after rinseback completed (no escalation).", + 236: "HD needs new cartridge to be installed.", + 237: "HD prime saline dialyzer time out alarm.", + 238: "HD no cartridge loaded or installed improperly alarm.", + 239: "HD fail to remove cartridge alarm.", + 240: "Bicarb conductivity out of range during bicarb pump check alarm.", + 241: "DG reservoir drain time out.", + 242: "DG reservoir fill time out.", + 243: "DG reservoir leak time out.", + 244: "DG temperature sensors difference out of range.", + 245: "DG heat disinfect target temperature time out (could not reach to temperature).", + 246: "DG heat disinfect inlet pressure and temperature sensors out of range.", + 247: "DG heat disinfect inlet conductivity and temperature out of range.", + 248: "DG chemical disinfect could not reach to target temperature.", + 249: "DG chemical disinfect inlet pressure and temperature sensors out of range.", + 250: "DG chemical disinfect inlet conductivity and temperature out of range.", + 251: "HD invalid system record.", + 252: "HD invalid service record.", + 253: "DG invalid system record.", + 254: "DG invalid service record.", + 255: "HD and UI software builds are not compatible.", + 256: "HD and DG software builds are not compatible.", + 257: "DG FPGA power out timeout.", + 258: "HD FPGA power out timeout.", + 259: "HD temperatures out of range.", + 260: "UI POST Application Integrity (Sha256Sum) failure.", + 261: "UI POST CANBus failure.", + 262: "UI POST Display failure.", + 263: "UI POST Touch failure.", + 264: "UI POST SD-Card failure.", + 265: "UI POST RTC failure.", + 266: "UI POST WiFi failure.", + 267: "UI POST Bluetooth failure.", + 268: "UI POST Ethernet failure.", + 269: "UI POST Sound failure.", + 270: "HD POST Safety Shutdown failure.", + 271: "DG POST Safety Shutdown failure.", + 272: "HD Fan RPM out of range.", + 273: "HD measured blood pump flow rate is out of range.", + 274: "HD measured dialysate inlet pump flow rate is out of range.", + 275: "HD arterial pressure sensor is reading out of range.", + 276: "HD venous pressure sensor is reading out of range.", + 277: "HD BP occlusion sensor is reading out of range.", + 278: "HD active reservoir weight out of range.", + 279: "DG dialysate drain time out.", + 280: "HD arterial pressure sensor read timeout error.", + 281: "Acid concentration bottle low volume alarm.", + 282: "Bicarbonate concentration bottle low volume alarm.", + 283: "DG load cells weight out of range.", + 284: "DG load cells primary/back up drift out of range.", + 285: "HD treatment recirculate timeout warning.", + 286: "HD treatment rinseback complete timeout warning.", + 287: "HD processor clock speed checks against FPGA clock failure.", + 288: "DG trimmer heater on with flow timeout.", + 289: "DG dialysate or concentrate caps not closed." + } + +ALARM_PRIORITIES = {1: "LOW", 2: "MEDIUM", 3: "HIGH"} +NUM_OF_ALARM_PRIORITIES = 4 # Total number of alarm priorities +ACCEPTED = True +REJECTED = False +ALARM_LIST_TITLE = "Alarm list" + + +NUM_OF_REQUEST_REJECT_REASONS = 43 +NUM_OF_MSG_ID = 289 +UNDEFINED_ALARM_ID_MSG = "ALARM_ID_UNDEFINED" +ALARM_TITLE = "Alarm" + + +ALARM_ID_NO_ALARM = 0 +ALARM_ID_OUT_OF_RANGE = 290 + +ALARM_PRIORITY_OPTIONS = { + 0 : 'ALARM_PRIORITY_NONE' , + 1 : 'ALARM_PRIORITY_LOW' , + 2 : 'ALARM_PRIORITY_MEDIUM', + 3 : 'ALARM_PRIORITY_HIGH', +} + +# alarm colors +ALARMS_COLORS_HEADER = { + "ALARM_PRIORITY_NONE" : "#438feb", + "ALARM_PRIORITY_LOW" : "#db8f00", + "ALARM_PRIORITY_MEDIUM" : "#db8f00", + "ALARM_PRIORITY_HIGH" : "#831913" +} + +ALARMS_COLORS_BODY = { + "ALARM_PRIORITY_NONE" : "#254670", + "ALARM_PRIORITY_LOW" : "#f5a623", + "ALARM_PRIORITY_MEDIUM" : "#f5a623", + "ALARM_PRIORITY_HIGH" : "#c53b33" +} + +ALARM_MUTE_FLAG = 0B0000001000000000 +ALARM_UNMUTE_FLAG = 0B0000000000000000 + +ALARM_BUTTONS_OK = 0B0000000010111000 + +ALARM_BUTTONS_RESUME = 0B0000000000110000 +ALARM_BUTTONS_RESUME_RINSEBACK = 0B0000000000100000 +ALARM_BUTTONS_RESUME_END = 0B0000000000010000 + +ALARM_BUTTONS_RINSEBACK = 0B0000000000101000 +ALARM_BUTTONS_RINSEBACK_RESUME = ALARM_BUTTONS_RESUME_RINSEBACK +ALARM_BUTTONS_RINSEBACK_END = 0B0000000000001000 + +ALARM_BUTTONS_END = 0B0000000000011000 +ALARM_BUTTONS_END_RESUME = ALARM_BUTTONS_RESUME_END +ALARM_BUTTONS_END_RINSEBACK = ALARM_BUTTONS_RINSEBACK_END + +ALARM_BUTTONS_RESUME_RINSEBACK_END = 0B0000000000000000 +ALARM_BUTTONS_RINSEBACK_RESUME_END = ALARM_BUTTONS_RESUME_RINSEBACK_END +ALARM_BUTTONS_END_RESUME_RINSEBACK = ALARM_BUTTONS_RESUME_RINSEBACK_END + +CURRENT_COLOR = '#000000' +COMPLETE_COLOR= '#4290ec' +ENABLED_COLOR = '#fcfcfc' +INCOMPLETE_COLOR = '#607a91' +OUT_OF_RANGE_COLOR = "#c53b33" + +#standby mode +GOODMORNING_START_TIME_SEC = 0 +GOODEVENING_START_TIME_SEC = 43200 +BLOOD_PRIMING_TEXT = "Blood Priming" +LIQUID_UNIT = "mL" +BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" +SALINE_BOLUS_CUMULATIVE = 300 +SALINE_UNIT = "mL" + +UF_MINIMUM_VALUE = 0 +UF_MAXIMUM_VALUE = 600 +ULTRAFILTRATION_PAUSED_TEXT = "Ultrafiltration Paused" +RESUME_TEXT = "RESUME ULTRAFILTRATION" +EDIT_TEXT = "EDIT ULTRAFILTRATION VOLUME" +TITLE_TEXT = "ULTRAFILTRATION VOLUME (L)" + +#main treatment pressure +PRESSURE_TITLE = "PRESSURES" +ARTERIAL_TITLE = "Arterial" +VENOUS_TITLE = "Venous" +VENOUS_UOM = "mmHg" +ARTERIAL_UOM = "mmHg" +LOW_TEXT = "LOW" +HIGH_TEXT = "HIGH" +ACCEPTED = True +REJECTED = False +NUM_OF_REQUEST_REJECT_REASONS = 43 + +#pressure +PRESSURE_STEPS = 10 +ARTERIAL_PRESSURE_MINIMUM = -400 +ARTERIAL_PRESSURE_MAXIMUM = +600 +ARTERIAL_PRESSURE_LOW_MIN = -300 +ARTERIAL_PRESSURE_LOW_DEF = -300 +ARTERIAL_PRESSURE_LOW_MAX = +200 +ARTERIAL_PRESSURE_HIGH_MIN = -300 +ARTERIAL_PRESSURE_HIGH_DEF = +100 +ARTERIAL_PRESSURE_HIGH_MAX = +200 + +VENOUS_PRESSURE_MINIMUM = -100 +VENOUS_PRESSURE_MAXIMUM = +700 +VENOUS_PRESSURE_LOW_MIN = -100 +VENOUS_PRESSURE_LOW_DEF = -100 +VENOUS_PRESSURE_LOW_MAX = +600 +VENOUS_PRESSURE_HIGH_MIN = +100 +VENOUS_PRESSURE_HIGH_DEF = +400 +VENOUS_PRESSURE_HIGH_MAX = +600 + +#main_treatment_flows +BLOOD_FLOW_MINIMUM_VALUE = 100 +BLOOD_FLOW_BUFFER = 25 +FLOW_SLIDER_WIDTH_BUFFER = 37 +BLOOD_FLOW_MAXIMUM_VALUE = 500 +DIALYSATE_FLOW_MINIMUM_VALUE = 0 +DIALYSATE_FLOW_MAXIMUM_VALUE = 0 # Seems to be bug. need change +FLOW_UNIT = "mL/min" +BLOOD_FLOW_TEXT = "blood" +DIALYSATE_FLOW_TEXT = "dialysate" + +#main treatment screen +VITALS_TITLE = "VITALS" +OUT_OF_RANGE_COLOR = "#c53b33" +IN_RANGE_COLOR = "#fcfcfc" +SYSTOLIC_LOWER_LIMIT = 60 +SYSTOLIC_UPPER_LIMIT = 250 +DIASTOLIC_LOWER_LIMIT = 40 +DIASTOLIC_UPPER_LIMIT = 200 +HEART_RATE_LOWER_LIMIT = 40 +HEART_RATE_UPPER_LIMIT = 180 +HEART_RATE_TITLE = "Heart Rate" +BLOOD_PRESSURE_TITLE = "Blood Pressure" +BLOOD_PRESSURE_UNIT = "mmHg" +HEART_RATE_UNIT = "BPM" +BLOOD_PRIMING_RANGE = 300 +BLOOD_PRIMING_TARGET_MAXIMUM = 300 + + +#tst_treatment_monitoring +FLOWS_TEXT = "FLOWS" +BLOOD_UNIT = "mL/min" +DIALYSATE_TEXT = "Dialysate" +BLOOD_TEXT = "Blood" +VITALS_TEXT = "VITALS" +BLOOD_PRESSURE_TEXT = "mmHg" +HEARTRATE_UNIT = "BPM" +PRESSURE_TEXT = "PRESSURE (mmHg)" +ARTEREAL_TEXT = "Arterial" +VENOUS_TEXT = "Venous" +TIME_DURATION_TEXT = "Time Remaining" +ULTRAFILTERATION_TEXT = "ULTRAFILTRATION VOLUME (L)" +SALINE_BOLUS = "SALINE BOLUS" +VOLUME_DELIVERED = "Volume Delivered" +CUMULATIVE_DELIVERED = "Cumulative Delivered" +HEPARIN_TEXT = "HEPARIN" +VOLUME_DELIVERED = "Volume Delivered" + +#patient id +PATIENT_ID = { + "qwerty" : "qwerty", + "sampleid" : "sampleid", + "abcdefghijklmnop" : "abcdefghijklmnop", + "patientid" : "patientid", + "username" : "username", + "sA@MpLE#123" : "sAMpLE123", + "A@BH&321" : "ABH321", + "5432@qwerty" : "5432qwerty" +} + +KEYBOARD_SWITCH = ["ABC", "&123"] + + +#main-treatment heparin +HEPARIN_UNIT = "mL" +HEPARIN_DELIVERY_TEXT = "HEPARIN DELIVERY" +HEPARIN_DELIVERY_OFF_TEXT = "OFF" +HEPARIN_STOP_TEXT = "Delivery Stopped" +HEPARIN_RESUME_DELIVERY_TEXT = "RESUME DELIVERY" +HEPARIN_BOLUS_ACTIVE_TEXT = "Bolus Active" +HEPARIN_PAUSE_DELIVERY_TEXT = "PAUSE DELIVERY" +HEPARIN_MAXIMUM_VOLUME_DELIVERED_TEXT = "Maximum Cumulative Heparin Volume Delivered" +HEPARIN_SYRINGE_EMPTY_TEXT = "Syringe Empty" + +#Time_Duration +CREATE_TREATMENT_PARAMETERS = ["Blood Flow Rate", "Dialysate Flow Rate", "Duration","Heparin Dispensing Rate", + "Heparin Bolus Volume","Heparin Stop Time", "Saline Bolus","Dialysate Temperature","Blood Pressure Measurement Interval", + "Rinseback Flow Rate"] +TREATMENT_PAUSED_TEXT = "Treatment Paused" + #dictionary consist of custom treatment values CREATE_CUSTOM_TREATMENT = { "Blood Flow Rate": 250, @@ -88,4 +572,140 @@ "rinseback_flow_rate" : [50, 25, 187], } +#ultrafilteration volume based on flow parameters. Python list contains Rotational speed, Motor speed, +# MC speed, MC current and PWM. +TREATMENT_ULTRAFILTRATION_FLOW_PARAMETERS = { + 0 : [100, 1, 1, 1, 1, 1, 1], + 1 : [550, 24, 32, 425, 85, 18, 36], + 2 : [80, 125, 187, 257, 530, 458, 985], + 3 : [100, 0, 0, 0, 0, 0, 0], + 4 : [220, 1.45, 1.45, 1.45, 1.45, 1.45, 1.45], + } +#dictionary contains uf maximum and minimum value's to set. +TREATMENT_ULTRAFILTRATION_TESTING_OPTION = { + 0 : {"uf_maximum" : 600, "uf_minimum" : 0}, + 1 : {"uf_maximum" : 500, "uf_minimum" : 0}, + 2 : {"uf_maximum" : 550, "uf_minimum" : 0}, + 3 : {"uf_maximum" : 300, "uf_minimum" : 0}, + 4 : {"uf_maximum" : 280, "uf_minimum" : 0}, + 5 : {"uf_maximum" : 765, "uf_minimum" : 500}, + 6 : {"uf_maximum" : 680, "uf_minimum" : 450}, + 7 : {"uf_maximum" : 480, "uf_minimum" : 380}, + 8 : {"uf_maximum" : 390, "uf_minimum" : 675}, + 9 : {"uf_maximum" : 290, "uf_minimum" : 125}, + } + +#tst_pretreatment_screens +PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] +NUM_OF_PRETREATMENT_BULLETS = len(PRE_TREATMENT_SCREENS) + +#tst_pretreatment_screens +PRE_TREATMENT_SCREENS = ["Create" , "Sample" , "Consumables" , "Disposables" , "Prime" , "Ultrafiltration" , "BP/HR" , "Connection" , "Start"] + + +#tst_pretreatment_screens color palettes +CURRENT_COLOR = '#000000' +COMPLETE_COLOR= '#4290ec' +ENABLED_COLOR = '#fcfcfc' +INCOMPLETE_COLOR = '#607a91' + +CONSUMABLE_SELF_TEST_TEXT = "Consumables Self Test" +BICARB_PUMP_CHECK_TEXT = "BiCarb Pump Check" +ACID_PUMP_CHECK_TEXT = "Acid Pump Check" +SELF_TEST_COMPLETE_TEXT = "Self Test Complete!" + + +#pre-treatment_priming +NOT_VISIBLE = False +VISIBLE = True +FOCUS = True +NOT_IN_FOCUS = False +BUSY = "Busy" +CHECK = "Check" +BEGIN_PRIME_TITLE = "Begin Prime" +PRIMING_TITLE = "Priming" +SYSTEM_SELF_TEST_TITLE = "System Self Test" +FILTER_FLUSH_TITLE = "Filter Flush" +MAXIMUM_COUNTDOWN_TIME = 300 +COUNT_DOWN_TIME_100 = 100 +MINIMUM_COUNTDOWN_TIME = 0 +ENABLED = True +DISABLED = False +CONSUMABLE_SELF_TEST_BICARB_PUMP_CHECK_STATE = 3 + +LOOD_PRIMING_TEXT = "Blood Priming" +SALINE_UNIT = "mL" +BLOOD_PRIMING_DEFAULT_VALUE = "0 mL" + +#Pre-Treatment Disposables +NEXT_BUTTON_TEXT = "NEXT" +BACK_BUTTON_TEXT = "BACK" +ENABLED = True +DISABLED = False +CONFIRM_BUTTON_TEXT = "CONFIRM" + +#Filter Flush + +#Pre_treatment_water_sample +SAMPLE_TEXT = "Sample" +CREATE_TEXT = "Create" +NEXT_BUTTON_TEXT = "NEXT" +WATER_SAMPLE_BUTTON_TEXT = "WATER SAMPLE" +BACK_BUTTON_TEXT = "BACK" +PASS_TEXT = "PASS" +FAIL_TEXT = "FAIL" +OK_TEXT = "OK" + +# pretreatment_patient_connection +SYSTOLIC_PRESSSURE_120 = 120 +SYSTOLIC_PRESSSURE_113 = 113 +SYSTOLIC_PRESSSURE_150 = 150 +SYSTOLIC_PRESSSURE_175 = 175 +SYSTOLIC_PRESSSURE_200 = 200 +SYSTOLIC_PRESSSURE_200 = 200 +SYSTOLIC_PRESSSURE_251 = 251 +SYSTOLIC_PRESSSURE_59 = 59 +SYSTOLIC_PRESSSURE_15 = 15 +DIASTOLIC_PRESSSURE_25 = 25 +DIASTOLIC_PRESSSURE_39 = 39 +DIASTOLIC_PRESSSURE_74 = 74 +DIASTOLIC_PRESSSURE_80 = 80 +DIASTOLIC_PRESSSURE_85 = 85 +DIASTOLIC_PRESSSURE_60 = 60 +DIASTOLIC_PRESSSURE_150 = 150 +DIASTOLIC_PRESSSURE_200 = 200 +DIASTOLIC_PRESSSURE_201 = 201 +HEART_RATE_VAL_39 = 39 +HEART_RATE_VAL_60 = 60 +HEART_RATE_VAL_70 = 70 +HEART_RATE_VAL_85 = 85 +HEART_RATE_VAL_100 = 100 +HEART_RATE_VAL_101 = 101 +HEART_RATE_VAL_110 = 110 +HEART_RATE_VAL_181 = 181 +HEART_RATE_VAL_200 = 200 +SYSTOLIC_LOWER_LIMIT = 60 +SYSTOLIC_UPPER_LIMIT = 250 +DIASTOLIC_LOWER_LIMIT = 40 +DIASTOLIC_UPPER_LIMIT = 200 +HEART_RATE_LOWER_LIMIT = 40 +HEART_RATE_UPPER_LIMIT = 180 +UF_MINIMUM_SLIDER_WIDTH = -2 +UF_MAXIMUM_SLIDER_WIDTH = 638.00 +UF_VALID_RANGE = 700 +TRAINING_INDICATOR_SCREEN = 4 +TREATMENT_ULTRAFILTERATION_TESTING_OPTION = { + "OPTION_1" : {"uf_maximum" : 800, "uf_minimum" : 0}, + "OPTION_2" : {"uf_maximum" : 500, "uf_minimum" : 0}, + "OPTION_3" : {"uf_maximum" : 550, "uf_minimum" : 0}, + "OPTION_4" : {"uf_maximum" : 300, "uf_minimum" : 0}, + "OPTION_5" : {"uf_maximum" : 280, "uf_minimum" : 0}, + "OPTION_6" : {"uf_maximum" : 765, "uf_minimum" : 0}, + "OPTION_7" : {"uf_maximum" : 680, "uf_minimum" : 0}, + "OPTION_8" : {"uf_maximum" : 1500, "uf_minimum" : 0}, + "OPTION_9" : {"uf_maximum" : 390, "uf_minimum" : 0}, + "OPTION_10" : {"uf_maximum" : 800, "uf_minimum" : 0}, + "OPTION_11" : {"uf_maximum" : 1000, "uf_minimum" : 0}, + "OPTION_12" : {"uf_maximum" : 1250, "uf_minimum" : 0}, + }