Index: Alarms/Alarms.conf =================================================================== diff -u -r079ccbf302493b44bf66c218d1d03d696dbfb090 -r5d0b098f15d890e6ae97bb3590344d1c12a55eb8 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 079ccbf302493b44bf66c218d1d03d696dbfb090) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 5d0b098f15d890e6ae97bb3590344d1c12a55eb8) @@ -1177,7 +1177,7 @@ [215] Title = Air Trap Overfill ListTitle = Air Trap: Air Trap Upper Fluid -Message = Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots, address as clinically indicated per facility policy. \n- Press Resume/OK to continue treatment. +Message = Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. [216] Title = Treatment Complete