Index: Alarms/Alarms.conf =================================================================== diff -u -r0a0379316e7e6fc01cc5221376f9831c3578549b -r4e6489f8af98f00cf149db4602d7e65151356ee1 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 0a0379316e7e6fc01cc5221376f9831c3578549b) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 4e6489f8af98f00cf149db4602d7e65151356ee1) @@ -1175,9 +1175,9 @@ Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [215] -Title = -ListTitle = -Message = +Title = Air Trap Overfill +ListTitle = Air Trap: Air Trap Overfill +Message = Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. [216] Title = Treatment Complete