Index: Alarms/Alarms.conf =================================================================== diff -u -r0cc88081de00352665176c2156dbc0e4264a7bdb -r59995bc9616e9ea9cbb29af043c7dcd71b570351 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 0cc88081de00352665176c2156dbc0e4264a7bdb) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 59995bc9616e9ea9cbb29af043c7dcd71b570351) @@ -684,7 +684,7 @@ [127] Title = Low Venous Pressure ListTitle = Blood Stop: Venous Pres. Low -Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap and venous pressure lines for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. +Message = Venous pressure is low.\n- Assess venous access and blood lines, correct issues if needed.\n- Check air trap and venous pressure lines for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if needed.\n- Inspect dialyzer for clots. If clots are present, follow facility\npolicy to end treatment.\n- Press Resume/OK to continue treatment. [128] Title = Venous Air Detected @@ -699,7 +699,7 @@ [130] Title = High Venous Pressure ListTitle = Blood Stop: Venous Pres. High -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap and venous pressure line for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. [131] Title = Low Arterial Pressure @@ -1140,8 +1140,8 @@ [208] Title = High Venous Pressure -ListTitle = Blood Stop: Venous Pres. High -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. +ListTitle = HD venous line occlusion fault +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap and venous pressure line for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. [209] Title = Disinfection Interrupted @@ -1177,7 +1177,7 @@ [215] Title = Air Trap Overfill ListTitle = Air Trap: Air Trap Overfill -Message = Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. +Message = Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. [216] Title = Treatment Complete @@ -1187,7 +1187,7 @@ [217] Title = Multiple Air Trap Events ListTitle = Air Trap: Air Trap Overfill -Message = Multiple Air Trap Events have been detected. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducer protectors are secured.\n- Inspect venous transducer protector for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. +Message = Multiple air trap events have been detected. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap and venous pressure line for foam and that\ntransducers are secured.\n- Inspect venous transducer for moisture; change if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. [218] Title = Syringe Detected