Index: Alarms/Alarms.conf =================================================================== diff -u -raaa0ce2a11504f70a62b0aa2c3057a04857216ae -rb7f38bd79b12800a9bdc75e65b15041980facf1e --- Alarms/Alarms.conf (.../Alarms.conf) (revision aaa0ce2a11504f70a62b0aa2c3057a04857216ae) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision b7f38bd79b12800a9bdc75e65b15041980facf1e) @@ -1,1077 +1,1077 @@ [0] -Title = -Message = \n\n +Title = ` +Message = [1] -Title = Stuck Button -Message = A button was pressed during start up and appears stuck. \n-Press and release the power and stop buttons on the front of the device. \n-Turn device off, then back on. \n-If problem continues, call service to schedule a repair. \n\nHD POST: Stuck Button +Title = HD POST: Stuck Button +Message = Stop or On button pressed during startup\n-Release button and power cycle\n-Call service to schedule a repair if problem persists [2] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: FPGA Compatibility +Title = HD POST: FPGA Compatibility +Message = Call service to schedule a repair. [3] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: FPGA Compatibility +Title = DG POST: FPGA Compatibility +Message = Call service to schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [4] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: WD Safestate +Title = HD POST: WD Safestate +Message = Call service to schedule a repair. [5] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: WD Safestate +Title = DG POST: WD Safestate +Message = Call service to schedule a repair. SRSDG 53 = [6] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Proc-UI Comm +Title = HD POST: Proc-UI Comm +Message = Call service to schedule a repair. [7] -Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): RTC Bat Low +Title = HD Alarm (Low): RTC Bat Low +Message = Call service to schedule a repair. [8] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Acc Vector +Title = HD POST: Acc Vector +Message = Call service to schedule a repair. [9] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acc Vector +Title = DG POST: Acc Vector +Message = Call service to schedule a repair. SRSDG 323 = [10] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Emb Mode +Title = HD POST: BLD Config Emb Mode +Message = Call service to schedule a repair. [11] -Title = Low Dialysate Temperature -Message = Dialysate temperature is low. \n-Treatment is currently paused.\n-When the temperature is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Dialysate Temp Low +Title = Dial. Gen: Dialysate Temp Low +Message = Dialysate temperature is low\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [12] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: HD Image +Title = HD POST: HD Image +Message = Call service to schedule a repair. [13] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: DG Image +Title = DG POST: DG Image +Message = Call service to schedule a repair. SRSDG 813 = [14] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Usage Rec NVRAM CRC +Title = DG POST: Usage Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 957 = [15] -Title = High Dialysate Temperature -Message = Dialysate temperature is high. \n-Treatment is currently paused.\n-When dialysate temperature is within target range,\n treatment will need to be resumed.\n\nDial. Gen: Dialysate Temp High +Title = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high\n-Treatment is paused and will be ready to resume when \ndialysate is within target range. [16] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Primary Alarm Current +Title = HD POST: Primary Alarm Current +Message = Call service to schedule a repair. [17] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI POST Not Complete +Title = HD POST: UI POST Not Complete +Message = Call service to schedule a repair. [18] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (DG ACK) +Title = Loss of DG Comm (DG ACK) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [19] -Title = High Dialysate Temperature -Message = Dialysate temperature is high. \n-Treatment is currently paused.\n-When dialysate temperature is within target range,\n treatment will need to be resumed. \n\nDial. Gen: Dialysate Temp High +Title = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [20] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Conductivity Cal +Title = DG POST: Conductivity Cal +Message = Call service to schedule a repair. SRSDG 16 = [21] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Drain Line Vol Cal +Title = DG POST: Drain Line Vol Cal +Message = Call service to schedule a repair. SRSDG 936 = [22] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Reservoir Vol Cal +Title = DG POST: Reservoir Vol Cal +Message = Call service to schedule a repair. SRSDG 935 = [23] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acid Conc Cal +Title = DG POST: Acid Conc Cal +Message = Call service to schedule a repair. SRSDG 934 = [24] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Bicarb Conc Cal +Title = DG POST: Bicarb Conc Cal +Message = Call service to schedule a repair. SRSDG 933 = [25] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acc Cal +Title = DG POST: Acc Cal +Message = Call service to schedule a repair. SRSDG 817 = [26] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Acc Cal +Title = HD POST: Acc Cal +Message = Call service to schedule a repair. [27] -Title = Self Test: Blood Leak -Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test detected has a problem with the blood leak sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [28] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Conductivity RTD Comm +Title = DG Fault: Conductivity RTD Comm +Message = Call service to schedule a repair. SRSDG 864 = [29] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: SP DAC Cal +Title = HD POST: SP DAC Cal +Message = Call service to schedule a repair. [30] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: SW Invalid State +Title = HD Fault: SW Invalid State +Message = Call service to schedule a repair. [31] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Current +Title = HD Fault: BP Current +Message = Call service to schedule a repair. [32] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed +Title = HD Fault: BP Speed +Message = Call service to schedule a repair. [33] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Direction +Title = HD Fault: BP Direction +Message = Call service to schedule a repair. [34] -Title = Blood Pump Error -Message = A problem has been detected with the blood pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: BP Rotor +Title = HD Fault: BP Rotor +Message = Blood pump rotor failure.\n- Check the pump rotor for secure connection. [35] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Current +Title = HD Fault: DPi Current +Message = Call service to schedule a repair. [36] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed +Title = HD Fault: DPi Speed +Message = Call service to schedule a repair. [37] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Direction +Title = HD Fault: DPi Direction +Message = Call service to schedule a repair. [38] -Title = Dialysate Pump Error -Message = A problem has been detected with the dialysate pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: DPi Rotor +Title = HD Fault: DPi Rotor +Message = Dialysate in pump rotor failure.\n- Check the pump rotor for secure connection. [39] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Current +Title = HD Fault: DPo Current +Message = Call service to schedule a repair. [40] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed +Title = HD Fault: DPo Speed +Message = Call service to schedule a repair. [41] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Direction +Title = HD Fault: DPo Direction +Message = Call service to schedule a repair. [42] -Title = Dialysate Pump Error -Message = A problem has been detected with the dialysate pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: DPo Rotor +Title = HD Fault: DPo Rotor +Message = Dialysate out pump rotor failure.\n- Check the pump rotor for secure connection. [43] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Iso ADC Comm +Title = DG Fault: Iso ADC Comm +Message = Call service to schedule a repair. SRSDG 954 = [44] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UI Comm +Title = HD Fault: UI Comm +Message = Call service to schedule a repair. [45] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: CAN CRC Com +Title = HD Fault: CAN CRC Com +Message = Call service to schedule a repair. [46] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: CAN ACK UI Com +Title = HD Fault: CAN ACK UI Com +Message = Call service to schedule a repair. [47] Title = -Message = \n\n +Message = [48] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UF Rate Tare Err +Title = HD Fault: UF Rate Tare Err +Message = Call service to schedule a repair. [49] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: FPGA Com +Title = HD Fault: FPGA Com +Message = Call service to schedule a repair. [50] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Valve Control +Title = DG Fault: Valve Control +Message = Call service to schedule a repair. SRSDG 26 = [51] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed +Title = HD Fault: BP Speed +Message = Call service to schedule a repair. [52] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed +Title = HD Fault: DPi Speed +Message = Call service to schedule a repair. [53] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed +Title = HD Fault: DPo Speed +Message = Call service to schedule a repair. [54] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Critical Data Err +Title = HD Fault: Critical Data Err +Message = Call service to schedule a repair. [55] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Critical Data Err +Title = DG Fault: Critical Data Err +Message = Call service to schedule a repair. SRSDG 921 = [56] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Acc Com +Title = HD Fault: Acc Com +Message = Call service to schedule a repair. [57] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Acc Com +Title = DG Fault: Acc Com +Message = Call service to schedule a repair. SRSDG 869 SRSDG 870 = [58] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Home +Title = HD Fault: Valve Home +Message = Call service to schedule a repair. [59] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Transition Err +Title = HD Fault: Valve Transition Err +Message = Call service to schedule a repair. [60] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Control +Title = HD Fault: Valve Control +Message = Call service to schedule a repair. [61] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Current +Title = HD Fault: Valve Current +Message = Call service to schedule a repair. [62] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Position +Title = HD Fault: Valve Position +Message = Call service to schedule a repair. [63] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BLD Com +Title = HD Fault: BLD Com +Message = Call service to schedule a repair. [64] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Range +Title = DG Fault: Baro Range +Message = Call service to schedule a repair. SRSDG 874 = [65] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reject) +Title = Loss of DG Comm (Reject) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [66] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Set Point +Title = HD POST: BLD Config Set Point +Message = Call service to schedule a repair. [67] -Title = Self Test: Cartridge -Message = A cartridge has been detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [68] -Title = Dialysate Generation Error -Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. \n\nDial. Gen: Dialysate not ready +Title = Dial. Gen: Dialysate not ready +Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. [69] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: BLD Cal +Title = HD POST: BLD Cal +Message = Call service to schedule a repair. [70] -Title = Self Test: Cartridge -Message = A cartridge has been detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [71] -Title = Self Test: Cartridge -Message = A cartridge was detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [72] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (LC) +Title = Loss of DG Comm (LC) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [73] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (TDI/TRO) +Title = Loss of DG Comm (TDI/TRO) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [74] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVi On No Flow +Title = DG Fault: UVi On No Flow +Message = Call service to schedule a repair. SRSDG 600 = [75] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe DAC +Title = HD Fault: Syringe DAC +Message = Call service to schedule a repair. [76] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Voltage Monitor +Title = HD Fault: Voltage Monitor +Message = Call service to schedule a repair. [77] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Voltage Monitor +Title = DG Fault: Voltage Monitor +Message = Call service to schedule a repair. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Enc Dir +Title = HD Fault: Syringe Pump Enc Dir +Message = Call service to schedule a repair. [79] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Drv Dir +Title = HD Fault: Syringe Pump Drv Dir +Message = Call service to schedule a repair. [80] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Driver +Title = HD Fault: Syringe Pump Driver +Message = Call service to schedule a repair. [81] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump OT +Title = HD Fault: Syringe Pump OT +Message = Call service to schedule a repair. [82] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump DAC Err +Title = HD Fault: Syringe Pump DAC Err +Message = Call service to schedule a repair. [83] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump On Err +Title = HD Fault: Syringe Pump On Err +Message = Call service to schedule a repair. [84] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: HD RTC Time +Title = DG POST: HD RTC Time +Message = Call service to schedule a repair. SRSDG 815 = [85] -Title = Cartridge Lever is Open -Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press RESUME button to continue.\n\nBlood Stop: Cartridge Lever Open +Title = Blood Stop: Cartridge Lever Open +Message = Cartridge Lock Lever is not closed.\n- Open front door and push lever all the way down.\n- Close door and press RESUME button to continue. [86] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: HD RTC Time +Title = HD POST: HD RTC Time +Message = Call service to schedule a repair. [87] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Heat Cal +Title = DG POST: Heat Cal +Message = Call service to schedule a repair. SRSDG 929 = [88] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 CP2 Speed Range +Title = DG Fault: CP1 CP2 Speed Range +Message = Call service to schedule a repair. [89] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVo On No Flow +Title = DG Fault: UVo On No Flow +Message = Call service to schedule a repair. SRSDG 599 = [90] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC A1/B1 Comm +Title = DG Fault: LC A1/B1 Comm +Message = Call service to schedule a repair. SRSDG 604 = [91] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nLoss of DG Comm (FMD) +Title = Loss of DG Comm (FMD) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [92] Title = Message = [93] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Permeate Cond High +Title = DG Fault: RO Permeate Cond High +Message = Call service to schedule a repair. SRSDG 430 = [94] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Pressure Sensor Range +Title = DG Fault: Pressure Sensor Range +Message = Call service to schedule a repair. SRSDG 629 = [95] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: WD Error +Title = DG Fault: WD Error +Message = Call service to schedule a repair. SRSDG 907 = [96] -Title = Incoming Water Temperature High -Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the temperature is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp High +Title = Dial. Gen: Inlet Water Temp High +Message = Inlet water temperature is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 410 = [97] -Title = Conductivity Out of Range -Message = Treatment is currently paused.\n-Confirm that acid and bicarbonate concentrates are securely connected. \n- If empty, replace. \n-Treatment will need to be resumed when conductivity is within target range.\n\nDial. Gen: Dialysate Cond. +Title = Dial. Gen: Dialysate Cond. +Message = Dialysate conductivity is out of range.\n- Check acid and bicarb concentrate is correct and not empty. SRSDG 397 = [98] -Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Com +Title = HD Alarm (Low): Battery Com +Message = Call service to schedule a repair. [99] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Stall +Title = HD Fault: Syringe Pump Stall +Message = Call service to schedule a repair. [100] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Timeout No Cart ST +Title = HD Fault: Timeout No Cart ST +Message = Call service to schedule a repair. [101] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI-Proc Comm +Title = HD POST: UI-Proc Comm +Message = Call service to schedule a repair. [102] -Title = Disinfection Interrupted -Message = Disinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Tpo THD Temp Delta +Title = Cleaning: Tpo THD Temp Delta +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 856 SRSDG 930 = [103] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: Cal NVRAM CRC +Title = HD POST: Cal NVRAM CRC +Message = Call service to schedule a repair. [104] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Air Trap Sensor +Title = HD Fault: Air Trap Sensor +Message = Call service to schedule a repair. [105] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Cal NVRAM CRC +Title = DG POST: Cal NVRAM CRC +Message = Call service to schedule a repair. SRSDG 608 = [106] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Flow Sensor Cal +Title = DG POST: Flow Sensor Cal +Message = Call service to schedule a repair. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DG Restarted +Title = DG Fault: DG Restarted +Message = Call service to schedule a repair. [108] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump ADC Com +Title = HD Fault: Syringe Pump ADC Com +Message = Call service to schedule a repair. [109] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Vol +Title = HD Fault: Syringe Pump Vol +Message = Call service to schedule a repair. [110] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed +Title = HD Fault: Syringe Pump Speed +Message = Call service to schedule a repair. [111] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed +Title = HD Fault: Syringe Pump Speed +Message = Call service to schedule a repair. [112] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC A2/B2 Comm +Title = DG Fault: LC A2/B2 Comm +Message = Call service to schedule a repair. SRSDG 953 = [113] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Current +Title = DG Fault: DRP Current +Message = Call service to schedule a repair. SRSDG 890 = [114] -Title = Self Test: Venous Air Detector -Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem with the air bubble sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [115] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Temp Sensor Range +Title = DG Fault: Temp Sensor Range +Message = Call service to schedule a repair. SRSDG 224 = [116] -Title = Service Required: Dialysate Device -Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): SD-Card Error +Title = HD Alarm (Low): SD-Card Error +Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. [117] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of HD Comm (HD ACK) +Title = Loss of HD Comm (HD ACK) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. SRSDG 909 = [118] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RTC Config +Title = DG Fault: RTC Config +Message = Call service to schedule a repair. SRSDG 43 = [119] -Title = Service Required: Dialysate Device -Message = A non-critical problem was detected with the dialysate device.\n-Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): RTC Bat Low +Title = DG Alarm (Low): RTC Bat Low +Message = Call service to schedule a repair. SRSDG 908 = [120] -Title = Self Test: Venous Air Detector -Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a bubble in the venous bubble detector\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [121] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Lg Control +Title = DG Fault: Prim Heater Lg Control +Message = Call service to schedule a repair. SRSDG 611 = [122] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Sm Control +Title = DG Fault: Prim Heater Sm Control +Message = Call service to schedule a repair. SRSDG 612 = [123] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Trimmer Heater Control +Title = DG Fault: Trimmer Heater Control +Message = Call service to schedule a repair. SRSDG 613 = [124] -Title = Blood Pump Stop Time Exceeded -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [125] -Title = Blood Pump Stop Time Exceeded -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [126] -Title = Blood Leak -Message = A blood leak has been detected. \n-Treatment must be terminated without rinseback.\n\nBlood Stop: Dialyzer Blood Leak +Title = Blood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. Treatment must end. [127] -Title = Low Venous Pressure -Message = Venous pressure is low.\n- Ensure venous access connections are properly secured. \n-If there has been an access disconnection or needle dislodgement, \nfollow facility policy to address and terminate treatment without rinseback. \n-If access connections are properly secured, \npress RESUME to continue treatment.\n\nBlood Stop: Venous Pres. Low +Title = Blood Stop: Venous Pres. Low +Message = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured \nproperly. [128] -Title = Venous Air Detected -Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm connections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Venous Air Bubble +Title = Blood Stop: Venous Air Bubble +Message = Venous air bubble detected indicating failure of air trap \ncontrol system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [129] Title = -Message = \n\n +Message = [130] -Title = High Venous Pressure -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n-Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Venous Pres. High +Title = Blood Stop: Venous Pres. High +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [131] -Title = Low Arterial Pressure -Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. Low +Title = Blood Stop: Arterial Pres. Low +Message = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [132] -Title = High Arterial Pressure -Message = Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n-Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. High +Title = Blood Stop: Arterial Pres. High +Message = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [133] -Title = Dialysate Device Fluid Leak -Message = Fluid leak detected in the dialysate device.\n-Treatment must be terminated.\n- Turn off water source. \n-Disconnect water and electrical connections from the device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Fluid Leak in DG +Title = DG Fault: Fluid Leak in DG +Message = Fluid leak detected in Dialysate Generation Delivery Device.\n-Treatment must be terminated.\n- Shut off water source and disconnect water and \nelectrical connections from system.\n- Call service to schedule a repair. SRSDG 328 = [134] -Title = Fluid Leak -Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n-Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, resume treatment.\n\nBlood Stop: Fluid Leak in HD +Title = Blood Stop: Fluid Leak in HD +Message = Fluid leak detected in Hemodialysis Delivery device.\n- If in treatment, BLOOD STOPPED.\n-Identify and correct the cause of the spill.\n- Clean up the spill and dry leak sensor.\n- If in treatment, resume treatment. [135] -Title = Blood Pump Stopped: Excessive Movement -Message = Excessive movement of the hemodialysis device has been detected. \n-Treatment is currently paused. \n-When system movement has stablized, \ntreatment will need to be resumed.\n-To resume treatment, press OK.\n\nBlood Stop: HD Shock Detected +Title = Blood Stop: HD Shock Detected +Message = Excessive movement detected with Hemodialysis Delivery \ndevice.\n- Confirm the device is stationary and level before continuing. [136] -Title = Dialysate Device: Excessive Movement -Message = Excessive movement of the dialysate device has been detected.\n-Treatment is currently paused. \n-Treatment can be resumed when system movement has stablized.\n\nDial. Gen: Shock-Vibration +Title = Dial. Gen: Shock-Vibration +Message = Excessive movement detected with Dialysis Generation \ndevice.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. SRSDG 327 = [137] -Title = Blood Pump Stopped: Excessive Tilt -Message = Excessive tilt of the hemodialysis device has been detected. \n-Treatment is currently paused. \n-When the device is stationary and level, \ntreatment will need to be resumed.\n-To resume treatment, press OK.\n\nBlood Stop: HD Tilted +Title = Blood Stop: HD Tilted +Message = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. [138] -Title = Dialysate Device: Excessive Tilt -Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing.\n\nDial. Gen: Tilt +Title = Dial. Gen: Tilt +Message = Excessive tilt detected with Dialysis Generation device.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing. SRSDG 326 = [139] -Title = Power Loss -Message = AC Power has been lost.\n-Check all electrical connections.\n- Press OK to return to welcome screen. \n\nLoss of AC Power +Title = Loss of AC Power +Message = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns. [140] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Timeout) +Title = Loss of DG Comm (Timeout) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [141] -Title = Blood Pump Stopped: Venous Air Detector -Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure connections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Air Trap Fill +Title = Blood Stop: Air Trap Fill +Message = Failure to fill the Air Trap.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [142] -Title = Blood Pump Stopped: Occulusion -Message = Occlusion detected after the blood pump.\n-Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Blood Pump Occlusion +Title = Blood Stop: Blood Pump Occlusion +Message = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the \nvenous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. [143] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TRo TDi Delta +Title = DG Fault: TRo TDi Delta +Message = Call service to schedule a repair. SRSDG 912 = [144] -Title = Disinfection: Water Conductivity High -Message = Incoming water conductivity is too high.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond High +Title = Cleaning: Inlet Water Cond High +Message = Cleaning Mode Inlet water conductivity too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = Conductivity Error During Prime -Message = Conductivity out of range during prime.\n-Confirm that acid and bicarbonate concentrates are securely connected. \n- If empty, replace. \n-Treatment will need to be resumed when conductivity is within target range.\n\nDial. Gen: Acid Conc. Cond. +Title = Dial. Gen: Acid Conc. Cond. +Message = Acid concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 846 = [146] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: DG RTC Timer +Title = DG POST: DG RTC Timer +Message = Call service to schedule a repair. SRSDG 312 = [147] -Title = Dialysate Generation Error -Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \nthe dialysate is ready for use. \n\nDial. Gen: Dialysate not ready +Title = Dial. Gen: Dialysate not ready +Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. SRSDG 398 = [148] -Title = Disinfection: Water Temperature High -Message = Incoming water temperature is too high.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp High +Title = Cleaning: Inlet Water Temp High +Message = Cleaning Mode Inlet water temperature too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: WD Error +Title = HD Fault: WD Error +Message = Call service to schedule a repair. [150] -Title = Incoming Water Conductivity High -Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the conductivity is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond High +Title = Dial. Gen: Inlet Water Cond High +Message = Inlet water conductivity is high.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 391 = [151] -Title = Incoming Water Conductivity Low -Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the conductivity is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond Low +Title = Dial. Gen: Inlet Water Cond Low +Message = Inlet water conductivity is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 826 = [152] -Title = Incoming Water Pressure Low -Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the pressure is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres Low +Title = Dial. Gen: Inlet Water Pres Low +Message = Inlet water pressure is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 417 = [153] -Title = Pre-Treatment Process Terminated -Message = The pre-treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n-Press OK for continue with next steps to remove cartridge from the device.\n\nProcess: Pre-Tx Terminated +Title = Process: Pre-Tx Terminated +Message = Pre-Treatment timeout.\n- Patient was not connected to device\n- Pre-Treatment has been terminated\n- Proceed to end treatment to remove cartridge from system. [154] -Title = Disinfection: Water Conductivity Low -Message = Incoming water conductivity is too low.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond Low +Title = Cleaning: Inlet Water Cond Low +Message = Cleaning Mode Inlet water conductivity too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reservoir) +Title = Loss of DG Comm (Reservoir) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [156] -Title = Communication Error -Message = A communication problem has been detected within the Diality Hemodialysis System.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Op Mode) +Title = Loss of DG Comm (Op Mode) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [157] -Title = Chemical Disinfectant Not Detected -Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the concentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Prime +Title = Cleaning: Chem Disinf PAA Prime +Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 988 = [158] -Title = Incoming Water Temperature Low -Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the temperature is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp Low +Title = Dial. Gen: Inlet Water Temp Low +Message = Inlet water temperature is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 408 = [159] -Title = Remove Chemical Disinfectant -Message = The chemical disinfectant is no longer needed. \n-Disconnect chemical disinfectant from the device. \n-Place the concentrate connection cap on the device.\n-Keep dialysate cap on the device.\n\nCleaning: Chem Disinf Connect Cap +Title = Cleaning: Chem Disinf Connect Cap +Message = Remove chemical disinfectant connection and close cap. SRSDG 1044 = [160] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Rotor Max Speed +Title = HD Fault: BP Rotor Max Speed +Message = Call service to schedule a repair. [161] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HD-DG Comm +Title = DG Fault: HD-DG Comm +Message = Call service to schedule a repair. SRSDG 122 = [162] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Proc-FPGA Clock +Title = DG Fault: Proc-FPGA Clock +Message = Call service to schedule a repair. SRSDG 867 = [163] -Title = Ultrafiltration Error -Message = A problem was detected with the dialysate device. \n-Confirm the device is level with minimal movement. \n-Resume treatment. \n-If problem persists, call service to report the issue and schedule a repair. \n\nDial. Gen: UF Rate Check +Title = Dial. Gen: UF Rate Check +Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system\n(shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a\nrepair. [164] -Title = Ultrafiltration Error -Message = A problem was detected with the dialysate device. \n-Confirm the device is level with minimal movement. \n-Resume treatment. \n-If problem persists, call service to report the issue and schedule a repair. \n\nDial. Gen: UF Rate Check +Title = Dial. Gen: UF Rate Check +Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system \n(shock/vibration).\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a \nrepair. [165] -Title = Blood Pump Stop Time Exceeded -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [166] Title = -Message = \n\n +Message = [167] -Title = Blood Pump Stopped: Door Open -Message = The device door is open.\n Close the device door. \n-Press RESUME to continue.\n\nBlood Stop: Door Open +Title = Blood Stop: Door Open +Message = Device door is open.\n - Close door and press RESUME button to continue. [168] Title = -Message = \n\n +Message = [169] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Dialysate Flow Sens Rg +Title = DG Fault: Dialysate Flow Sens Rg +Message = Call service to schedule a repair. SRSDG 643 = [170] Title = Heparin Syringe Empty -Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to complete treatment.\n\nHeparin Syringe Empty +Message = Heparin syringe empty.\n- Treatment will continue\n- If additional heparin is required to complete treatment, \nadminister manually. [171] -Title = Heparin Pump Occlusion -Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped.\n\nProcess: Pre-Tx Hep. Pump Occ. +Title = Process: Pre-Tx Hep. Pump Occ. +Message = Heparin pump occlusion detected.\n- Check heparin line is unclamped.\n- Check there are no kinks in the heparin line. [172] -Title = Not Enough Heparin -Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger, end treatment.\n- If syringe pump is not touching the syringe plunger, \n- ensure the correct amount of heparin has been drawn and reload the syringe. \n\nProcess: Pre-Tx Not Enough Hep. +Title = Process: Pre-Tx Not Enough Hep. +Message = Not enough heparin is available for treatment.\n- Check if syringe pump is touching syringe plunger.\n- If syringe pump is touching syringe plunger, end treatment.\n- If syringe pump is not touching syringe plunger, reload \nsyringe with enough heparin. [173] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: RTC Config +Title = HD Fault: RTC Config +Message = Call service to schedule a repair. [174] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: HD RTC Timer +Title = HD POST: HD RTC Timer +Message = Call service to schedule a repair. [175] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Pump Commutation Err +Title = HD Fault: Pump Commutation Err +Message = Call service to schedule a repair. [176] Title = -Message = \n\n +Message = [177] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Software Invalid State +Title = DG Fault: Software Invalid State +Message = Call service to schedule a repair. SRSDG 919 = [178] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UI-Proc Comm +Title = HD Fault: UI-Proc Comm +Message = Call service to schedule a repair. = SRSUI 1066 [179] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: FPGA Com +Title = DG Fault: FPGA Com +Message = Call service to schedule a repair. SRSDG 605 = [180] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Sensor Range +Title = DG Fault: RO Flow Sensor Range +Message = Call service to schedule a repair. SRSDG 883 = [181] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Tare Out of Range +Title = DG Fault: Tare Out of Range +Message = Call service to schedule a repair. SRSDG 849 SRSDG 838 = [182] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: LC Cal +Title = DG POST: LC Cal +Message = Call service to schedule a repair. SRSDG 162 = [183] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Invalid Resv Transfer +Title = Cleaning: Invalid Resv Transfer +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 851 = [184] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVi Error +Title = DG Fault: UVi Error +Message = Call service to schedule a repair. SRSDG 559 = [185] -Title = Service Required: Dialysate Device -Message = A non-critical problem was detected with the dialysate device.\n-Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): Fan Speed +Title = DG Alarm (Low): Fan Speed +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. SRSDG 628 = [186] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1/CP2 Fault +Title = DG Fault: CP1/CP2 Fault +Message = Call service to schedule a repair. SRSDG 891 = [187] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 Speed Fault +Title = DG Fault: CP1 Speed Fault +Message = Call service to schedule a repair. SRSDG 905 = [188] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP2 Speed Fault +Title = DG Fault: CP2 Speed Fault +Message = Call service to schedule a repair. SRSDG 906 = [189] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Fault +Title = DG Fault: DRP Speed Fault +Message = Call service to schedule a repair. SRSDG 448 = [190] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Not Off +Title = DG Fault: DRP Speed Not Off +Message = Call service to schedule a repair. [191] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow High +Title = DG Fault: RO Flow High +Message = Call service to schedule a repair. SRSDG 885 = [192] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Low +Title = DG Fault: RO Flow Low +Message = Call service to schedule a repair. SRSDG 945 = [193] Title = -Message = \n\n +Message = [194] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Driver Feedback +Title = DG Fault: RO Driver Feedback +Message = Call service to schedule a repair. SRSDG 881 = [195] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Pump High Pressure +Title = DG Fault: RO Pump High Pressure +Message = Call service to schedule a repair. SRSDG 925 = [196] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPi/CPo Comm +Title = DG Fault: CPi/CPo Comm +Message = Call service to schedule a repair. SRSDG 955 = [197] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1/CD2 Comm +Title = DG Fault: CD1/CD2 Comm +Message = Call service to schedule a repair. SRSDG 956 = [198] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HP On With No Flow +Title = DG Fault: HP On With No Flow +Message = Call service to schedule a repair. SRSDG 872 = [199] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HR On With No Flow +Title = DG Fault: HR On With No Flow +Message = Call service to schedule a repair. SRSDG 871 = [200] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Thermistor Range +Title = DG Fault: Thermistor Range +Message = Call service to schedule a repair. SRSDG 178 = [201] -Title = Self Test: Dialysate Flow Rate -Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem in the dialysate flow \nmeasurement or there is a leak in the dialysate in/out lines.\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [202] -Title = Self Test: Pressure -Message = A problem has been detected during self-test. \n-Confirm the cartridge is properly installed. \n-Ensure arterial and venous pressure sensor lines are not kinked. \n-Press OK to resume. \n-If problem persists, install a new cartridge.\n\nProcess: Pre-Tx Dry Self-Test +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected blood circuit pressure cannot be \nrelieved.\n- Check PBo connection to HD for kinks\n -Check PBo transducer protector for occlusion\n- If problem persists, install a new cartridge and \nPBo transducer protector. [203] Title = -Message = \n\n +Message = [204] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Rejection +Title = DG Fault: RO Rejection +Message = Call service to schedule a repair. SRSDG 429 = [205] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Cond Sensors Error +Title = DG Fault: Cond Sensors Error +Message = Call service to schedule a repair. SRSDG 913 = [206] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Fill Timeout +Title = DG Fault: Reservoir Fill Timeout +Message = Call service to schedule a repair. SRSDG 437 = [207] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: FMP Flow Rate Error +Title = DG Fault: FMP Flow Rate Error +Message = Call service to schedule a repair. SRSDG 240 = [208] -Title = High Venous Pressure -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n-Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. \n\nBlood Stop: Venous Pres. High +Title = Blood Stop: Venous Pres. High +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [209] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Flush Circ Error +Title = Cleaning: Flush Circ Error +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 980 = [210] -Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Fault +Title = HD Alarm (Low): Battery Fault +Message = Call service to schedule a repair. [211] -Title = Approaching MAX Blood Pump Stop Time -Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time timer expires. \n- If treatment is not resumed before the timer expires, treatment will be terminated without rinseback. \n\nBlood Stopped: Warning +Title = Blood Stopped: Warning +Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [212] -Title = Approaching MAX Blood Pump Stop Time -Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time timer expires. \n- If treatment is not resumed before the timer expires, treatment will be terminated without rinseback.\n\nBlood Stopped: Warning +Title = Blood Stopped: Warning +Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [213] -Title = Prime Complete - 30 MIN WARNING -Message = The priming process completed 90 minutes ago. \n-Treatment must begin within 30 minutes or the pre-treatment process will be terminated. \n-Press OK to continue. \n\nProcess: Pre-Tx Prime Complete +Title = Process: Pre-Tx Prime Complete +Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [214] Title = -Message = \n\n +Message = [215] Title = -Message = \n\n +Message = [216] -Title = Treatment Complete -Message = Dialysis treatment is complete.\n-Press OK to end treatment and rinseback.\n\nProcess: End of Dialysis Tx +Title = Process: End of Dialysis Tx +Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [217] Title = -Message = \n\n +Message = [218] -Title = Syringe Detected -Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n-Press OK to continue.\n\nProcess: Remove Syringe +Title = Process: Remove Syringe +Message = Syringe not removed from machine.\n- Remove heparin syringe. [219] -Title = Blood Stopped: Syringe Not Detected -Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n-Press Ok to resume treatment.\n\nBlood Stop: Syringe Not Detected +Title = Blood Stop: Syringe Not Detected +Message = Heparin is needed for treatment but syringe has been removed \nfrom syringe pump.\n- Install Heparin syringe. [220] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: THD RTD Comm +Title = DG Fault: THD RTD Comm +Message = Call service to schedule a repair. SRSDG 947 = [221] -Title = Blood Pump Stopped: Saline Bag Empty -Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. \n\nBlood Stop: Saline Bag Empty +Title = Blood Stop: Saline Bag Empty +Message = Saline bag appears empty.\n- Check saline bag and line and replace if saline bag is \nempty. [222] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Occ Com +Title = HD Fault: BP Occ Com +Message = Call service to schedule a repair. [223] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Art. Pres. Com +Title = HD Fault: Art. Pres. Com +Message = Call service to schedule a repair. [224] -Title = Blood Pump Stopped: Treatment Paused -Message = Treatment has been stopped by used. \n- Press OK to resume treatment. \n-If treatment is not resumed within five minutes, treatment will be terminated without rinseback. \n\nBlood Stopped: Treatment Pause +Title = Blood Stopped: Treatment Pause +Message = Treatment paused by user. [225] -Title = Treatment Complete -Message = Dialysis treatment is complete.\n-Press OK to end treatment and rinseback.\n\nProcess: End of Treatment +Title = Process: End of Treatment +Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [226] -Title = Priming Complete -Message = Priming is complete. \n-Treatment must begin within 120 minutes \nor the pre-treatment process will be terminated.\n-Press OK to continue. \n\nProcess: Pre-Tx Priming Complete +Title = Process: Pre-Tx Priming Complete +Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [227] Title = -Message = \n\n +Message = [228] -Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Title = Process: Pre-Tx Prime Timeout +Message = Blood line prime with saline timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [229] -Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [230] -Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [231] -Title = Self Test: Pressure -Message = A problem has been detected during self-test. \n- Confirm cartridge is properly installed without kinks or clamps. \n-Ensure all connections are secure. \n-Press OK to resume. \n- If problem persists, install a new cartridge.\n\nProcess: Pre-Tx Dry Self-Test +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected a leak in the circuit\n- Check blood line connections\n- Check dialysate line connections\n- Check PBo and PBa transducer protectors for occlusion.\n- Check clamps.\n- If problem persists, install a new cartridge. [232] -Title = Self Test: Dialysate Flow Rate -Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem in the load cell \nmeasurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation onto machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [233] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Cal CRC +Title = DG Fault: Baro Cal CRC +Message = Call service to schedule a repair. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = Incoming Water Pressure High -Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the pressure is within target range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres High +Title = Dial. Gen: Inlet Water Pres High +Message = Inlet water pressure is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 1037 = [235] -Title = Rinseback Complete -Message = Rinseback is complete. \n-Follow facililty policy prior to disconnecting patient. \n-Press OK to continue. \n\nProcess: Rinseback Complete +Title = Process: Rinseback Complete +Message = Rinseback has been completed. [236] -Title = Self Test: Fluid Present -Message = A problem has been detected during self-test. \n-Cartridges are single-use only and fluid is present.\n- Install new cartridge to resume.\n\nProcess: Pre-Tx Dry Self-Test +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected fluid in the cartridge.\n- Cartridges are single-use only\n- Install new cartridge to resume. [237] -Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [238] -Title = Cartridge Not Dectected -Message = A problem has been detected during self-test. \n-Cartridge is not detected.\n- Confirm cartridge is properly installed.\n\nProcess: Pre-Tx Dry Self-Test +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected cartridge is not installed.\n- Check that a cartridge is properly installed. [239] -Title = Cartridge Still Detected -Message = A problem has been detected during self-test. \n-Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed.\n\nProcess: Post-Tx Remove Cartridge +Title = Process: Post-Tx Remove Cartridge +Message = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. [240] -Title = Conductivity Error During Prime -Message = Conductivity out of range during prime.\n-Confirm that acid and bicarbonate concentrates are securely connected. \n- If empty, replace. \n-Treatment will need to be resumed when conductivity is within target range.\n\nDial. Gen: Bicarb Conc. Cond. +Title = Dial. Gen: Bicarb Conc. Cond. +Message = Bicarbonate conductivity during prime is out of range.\n- Check correct concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 482 = [241] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Resv Drain Timeout +Title = Cleaning: Resv Drain Timeout +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 848 SRSDG 90 SRSDG 561 @@ -1102,8 +1102,8 @@ SRSDG 1034 = [242] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Resv Fill Timeout +Title = Cleaning: Resv Fill Timeout +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 850 SRSDG 573 SRSDG 574 @@ -1124,373 +1124,373 @@ SRSDG 1024 = [243] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Fluid Loss Detected +Title = Cleaning: Fluid Loss Detected +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 853 = [244] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Temp Sensor Check +Title = Cleaning: Temp Sensor Check +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 852 = [245] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Heat Disinfect Temp TO +Title = Cleaning: Heat Disinfect Temp TO +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 854 = [246] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: RO Water Cond Error +Title = Cleaning: RO Water Cond Error +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [247] Title = -Message = \n\n +Message = [248] -Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinfect Proc +Title = Cleaning: Chem Disinfect Proc +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [249] -Title = Chemical Disinfectant Not Detected -Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the concentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Cond +Title = Cleaning: Chem Disinf PAA Cond +Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 992 = [250] -Title = Connect Chemical Disinfectant -Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n-Keep dialysate cap on the device.\n-Place disinfectant straw in the disinfectant bottle. \n-Attach straw to concentrate connection and ensure connection is secure. \n-Press OK to continue.\n\nCleaning: Chem Disinf Connect PAA +Title = Cleaning: Chem Disinf Connect PAA +Message = Connect chemical disinfectant bottle to bicarb port. SRSDG 1046 = [251] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Sys Rec NVRAM CRC +Title = HD POST: Sys Rec NVRAM CRC +Message = Call service to schedule a repair. [252] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Serv Rec NVRAM CRC +Title = HD POST: Serv Rec NVRAM CRC +Message = Call service to schedule a repair. [253] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Sys Rec NVRAM CRC +Title = DG POST: Sys Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 609 = [254] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Serv Rec NVRAM CRC +Title = DG POST: Serv Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 610 = [255] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Compatibility +Title = HD POST: UI Compatibility +Message = Call service to schedule a repair. [256] -Title = Chemical Flush in Progress -Message = Chemical flush in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Flush In Progress +Title = Cleaning: Chem Flush In Progress +Message = Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. = SRSUI 497 [257] Title = -Message = \n\n +Message = [258] Title = -Message = \n\n +Message = [259] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Temp Sensors Error +Title = HD Fault: Temp Sensors Error +Message = Call service to schedule a repair. [260] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Image +Title = HD POST: UI Image +Message = Call service to schedule a repair. = SRSUI 1067 [261] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI CAN Driver +Title = HD POST: UI CAN Driver +Message = Call service to schedule a repair. = SRSUI 1068 [262] -Title = -Message = \n\n +Title = HD POST: UI Disp Driver +Message = Call service to schedule a repair. = SRSUI 1069 [263] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Touch Driver +Title = HD POST: UI Touch Driver +Message = Call service to schedule a repair. = SRSUI 1070 [264] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI SD-Card +Title = HD POST: UI SD-Card +Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. = SRSUI 1071 [265] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI RTC +Title = HD POST: UI RTC +Message = Call service to schedule a repair. = SRSUI 1072 [266] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI WiFi +Title = HD POST: UI WiFi +Message = Call service to schedule a repair. = SRSUI 1073 [267] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Bluetooth +Title = HD POST: UI Bluetooth +Message = Call service to schedule a repair. = SRSUI 1074 [268] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Ethernet +Title = HD POST: UI Ethernet +Message = Call service to schedule a repair. = SRSUI 1075 [269] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Sound +Title = HD POST: UI Sound +Message = Call service to schedule a repair. = SRSUI 1076 [270] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Proc Safestate +Title = HD POST: Proc Safestate +Message = Call service to schedule a repair. [271] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Proc Safestate +Title = DG POST: Proc Safestate +Message = Call service to schedule a repair. SRSDG 822 = [272] -Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Fan Speed +Title = HD Alarm (Low): Fan Speed +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. [273] Title = -Message = \n\n +Message = [274] Title = -Message = \n\n +Message = [275] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Arterial Press Range +Title = HD Fault: Arterial Press Range +Message = Call service to schedule a repair. [276] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Venous Press Range +Title = HD Fault: Venous Press Range +Message = Call service to schedule a repair. [277] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Occlusion Sens Range +Title = HD Fault: Occlusion Sens Range +Message = Call service to schedule a repair. [278] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Overfill +Title = DG Fault: Reservoir Overfill +Message = Call service to schedule a repair. [279] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Drain +Title = DG Fault: Reservoir Drain +Message = Call service to schedule a repair. SRSDG 752 = [280] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Direction Fault +Title = DG Fault: DRP Direction Fault +Message = Call service to schedule a repair. SRSDG 888 = [281] -Title = Acid Concentrate Low -Message = The acid concentrate is low.\n- Replace acid concentrate container. \n\nDial. Gen: Acid Conc. Low +Title = Dial. Gen: Acid Conc. Low +Message = Acid concentrate low.\n- Replace acid concentrate bottle. SRSDG 836 = [282] -Title = Bicarb Concentrate Low -Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container.\n\nDial. Gen: Bicarb Conc. Low +Title = Dial. Gen: Bicarb Conc. Low +Message = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. SRSDG 837 = [283] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC Range +Title = DG Fault: LC Range +Message = Call service to schedule a repair. SRSDG 615 = [284] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Redundant LC Check +Title = DG Fault: Redundant LC Check +Message = Call service to schedule a repair. SRSDG 383 = [285] Title = -Message = \n\n +Message = [286] -Title = Concentrate Connection Cap Removed -Message = The concentrate connection cap was removed from the dialysate device during disinfection.\n- Disinfection terminated.\n-Replace cap before attempting disinfection again. \n\nCleaning: Conc Cap Removed +Title = Cleaning: Conc Cap Removed +Message = Concentrate cap removed from DG during cleaning.\n- Cleaning Process Terminated. [287] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Proc-FPGA Clock +Title = HD Fault: Proc-FPGA Clock +Message = Call service to schedule a repair. [288] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC Drift Check +Title = DG Fault: LC Drift Check +Message = Call service to schedule a repair. [289] -Title = Dialysate Connection Cap Removed -Message = The dialysate connection cap was removed from the dialysate device during disinfection.\n- Disinfection terminated.\n-Replace cap before attempting disinfection again. \n\nCleaning: Dialysate Cap Removed +Title = Cleaning: Dialysate Cap Removed +Message = Dialysate cap removed from DG during cleaning.\n- Cleaning Process Terminated. SRSDG 928 = [290] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Ven. Pres. Com +Title = HD Fault: Ven. Pres. Com +Message = Call service to schedule a repair. [291] -Title = Flush In Progress -Message = System flush is in progress\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Flush In Progress +Title = Cleaning: Flush In Progress +Message = System flush in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [292] -Title = Heat Disinfection in Progress -Message = Heat disinfection is in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Heat Disinf In Progress +Title = Cleaning: Heat Disinf In Progress +Message = Heat disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [293] -Title = Chemical Disinfection in Progress -Message = Chemical disinfection in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process is finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Disinfect In Progress +Title = Cleaning: Chem Disinf In Progress +Message = Chemical disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [294] Title = -Message = \n\n +Message = [295] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: UI CloudSync +Title = HD POST: UI CloudSync +Message = Call service to schedule a repair. = SRSUI 1238 [296] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TDi RTD Comm +Title = DG Fault: TDi RTD Comm +Message = Call service to schedule a repair. SRSDG 951 = [297] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TRo RTD Comm +Title = DG Fault: TRo RTD Comm +Message = Call service to schedule a repair. SRSDG 950 = [298] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Pres Comm +Title = DG Fault: Baro Pres Comm +Message = Call service to schedule a repair. SRSDG 948 = [299] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Serial Number Check +Title = DG POST: Serial Number Check +Message = Call service to schedule a repair. SRSDG 952 = [300] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Serial Number Check +Title = HD POST: Serial Number Check +Message = Call service to schedule a repair. [301] Title = -Message = \n\n +Message = [302] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Processor RAM error +Title = DG Fault: Processor RAM error +Message = Call service to schedule a repair. SRSDG 126 = [303] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Hall Commutation +Title = DG Fault: DRP Hall Commutation +Message = Call service to schedule a repair. SRSDG 958 = [304] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: Usage Rec NVRAM CRC +Title = HD POST: Usage Rec NVRAM CRC +Message = Call service to schedule a repair. [305] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Processor RAM error +Title = HD Fault: Processor RAM error +Message = Call service to schedule a repair. [306] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 CD2 Check +Title = DG Fault: CD1 CD2 Check +Message = Call service to schedule a repair. [307] -Title = Power Loss -Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n-Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed.\n\nBlood Stop: Loss of AC Power +Title = Blood Stop: Loss of AC Power +Message = HD power has been lost. \n- Treatment must be terminated.\n-Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns.\n-If power is returned prior to blood stopped countdown elapses, \nrinseback will be allowed. [308] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPi Comm +Title = DG Fault: CPi Comm +Message = Call service to schedule a repair. SRSDG 959 = [309] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPo Comm +Title = DG Fault: CPo Comm +Message = Call service to schedule a repair. SRSDG 960 = [310] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 Comm +Title = DG Fault: CD1 Comm +Message = Call service to schedule a repair. SRSDG 962 = [311] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD2 Comm +Title = DG Fault: CD2 Comm +Message = Call service to schedule a repair. SRSDG 961 = [312] -Title = Low Dialysate Flow -Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nRESUME button will be available.\n-Press RESUME to continue.\n\nDial. Gen: Dialysate In Flow +Title = Dial. Gen: Dialysate In Flow +Message = Dialysate in flow rate is low\n- Check dialysate line connection to the DG. [313] -Title = Disinfection: Water Temperature Low -Message = Incoming water temperature is too low.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp Low +Title = Cleaning: Inlet Water Temp Low +Message = Cleaning Mode Inlet water temperature too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = Disinfection: Water Pressure High -Message = Incoming water pressure is too high. \n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Press High +Title = Cleaning: Inlet Water Press High +Message = Cleaning Mode Inlet water pressure too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = Disinfection: Water Pressure Low -Message = Incoming water pressure is too low. \n-Disinfection has been terminated.\n- Check incoming water initiating disinfection again. \n\nCleaning: Inlet Water Press Low +Title = Cleaning: Inlet Water Press Low +Message = Cleaning Mode Inlet water pressure too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Acid CP Pump Park +Title = DG Fault: Acid CP Pump Park +Message = Call service to schedule a repair. SRSDG 983 = [317] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Bicarb CP Pump Park +Title = DG Fault: Bicarb CP Pump Park +Message = Call service to schedule a repair. SRSDG 984 = [318] -Title = Air Trap Control Error -Message = A problem has been detected with the air trap level control system. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue.\n\nHD Fault: Air Pump Timeout +Title = HD Fault: Air Pump Timeout +Message = Unable to reduce fluid level.\n- Air trap may be overfilled.\n- Pressure transducer protector may be contaminated.\n- Treatment must be terminated without option for rinseback.\n- Call service to schedule a repair. [319] -Title = Residual Chemical Sampling -Message = A water sample is needed to confirm the \nabsence of residual chemical following disininfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant.\n\nCleaning: Chem Residual Sampling +Title = Cleaning: Chem Residual Sampling +Message = Collect chemical residuals water sample from drain line. SRSDG 1040 = [320] -Title = Chemical Residual Timeout -Message = A chemical residual sample was not collected \n-within 30min of chemical flush.\n-To collect a sample for chemical residual, you must repeat the chemical flush.\n\nCleaning: Chem Residual Timeout +Title = Cleaning: Chem Residual Timeout +Message = During Chemical Disinfection Flush\n-user residual sampling timed out. SRSDG 1043 = [321] -Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVo Error +Title = DG Fault: UVo Error +Message = Call service to schedule a repair. SRSDG 1069 = [322] -Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner of the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST Alarm: UI Invalid Year +Title = HD POST Alarm: UI Invalid Year +Message = Call service to schedule a repair.