Index: Alarms/Alarms.conf =================================================================== diff -u -raea1653335f854dc7931ac6506308bbc394cde6c -re98d63aeba011576977981439b291f7429d19f97 --- Alarms/Alarms.conf (.../Alarms.conf) (revision aea1653335f854dc7931ac6506308bbc394cde6c) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision e98d63aeba011576977981439b291f7429d19f97) @@ -1,1055 +1,1055 @@ [0] -Title = +Title = Indication for when no alarms is a possible situation ListTitle = Message = [1] -Title = Stuck Button +Title = Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up ListTitle = HD POST: Stuck Button -Message = A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. +Message = Stop or Off button are pressed for more than 1 sec during POST. [2] -Title = Service Required: Hemodialysis Device +Title = HD FPGA POST failure ListTitle = HD POST: FPGA Compatibility -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. [3] -Title = Service Required: Dialysate Device +Title = DG FPGA POST failure ListTitle = DG POST: FPGA Compatibility -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header SRSDG 602 SRSDG 814 SRSDG 1061 = [4] -Title = Service Required: Hemodialysis Device +Title = HD Watchdog POST failure ListTitle = HD POST: WD Safestate -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA [5] -Title = Service Required: Dialysate Device +Title = DG Watchdog POST failure ListTitle = DG POST: WD Safestate -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V SRSDG 53 = [6] -Title = Service Required: Hemodialysis Device +Title = UI communication POST failure. UI failed to communicate within a reasonable time after power up ListTitle = HD POST: Proc- UI Comm -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI failed to checkin via the CAN bus within 40 seconds [7] -Title = Service Required: Hemodialysis Device +Title = Real-time clock configuration error ListTitle = HD Alarm (Low): RTC Bat Low -Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. +Message = If the RTC is not functional as indicated by the bits in the first 3 control registers. [8] -Title = Service Required: Hemodialysis Device +Title = HD accelerometer failed POST ListTitle = HD POST: Acc Vector -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. [9] -Title = Service Required: Dialysate Device +Title = DG accelerometer failed POST ListTitle = DG POST: Acc Vector -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. SRSDG 323 = [10] -Title = Service Required: Hemodialysis Device +Title = HD blood leak sensor embedded mode communication failed ListTitle = HD POST: BLD Config Emb Mode -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the blood leak sensor embedded mode is unresponsive. [11] -Title = Low Dialysate Temperature +Title = Dialysate temperature is below the target temperature ListTitle = Dial. Gen: Dialysate Temp Low -Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. +Message = If the dialysate temperature is below the target temperature for a certain period of time. [12] -Title = Service Required: Hemodialysis Device +Title = HD firmware image integrity POST test failed ListTitle = HD POST: HD Image -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated CRC of the HD image does not match the expected CRC. [13] -Title = Service Required: Dialysate Device +Title = DG firmware image integrity POST test failed ListTitle = DG POST: DG Image -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated CRC of the DG firmware image does not matched the expected CRC SRSDG 813 = [14] -Title = Service Required: Dialysate Device +Title = DG non-volatile usage information group invalid record CRC ListTitle = DG POST: Usage Rec NVRAM CRC -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG usage information record's CRC does not match the calculated CRC in POST. SRSDG 957 = [15] -Title = High Dialysate Temperature +Title = Dialysate temperature is out of high safety range ListTitle = Dial. Gen: Dialysate Temp High -Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. +Message = If the dialysate temperature is > 46 C for a certain period of time. [16] -Title = Service Required: Hemodialysis Device +Title = HD alarm audio failed POST ListTitle = HD POST: Primary Alarm Current -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output [17] -Title = Service Required: Hemodialysis Device +Title = HD UI POST failed ListTitle = HD POST: UI POST Not Complete -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. [18] -Title = Communication Error +Title = HD to DG comm error. ListTitle = Loss of DG Comm (DG ACK) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = HD message to DG that requires acknowledge was not acknowledged within a certain time. [19] -Title = High Dialysate Temperature +Title = Dialysate temperature is above the target temperature ListTitle = Dial. Gen: Dialysate Temp High -Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. +Message = If the dialysate temperature is above the target temperature for a certain period of time. [20] -Title = Service Required: Dialysate Device +Title = DG conductivity sensors invalid calibration record ListTitle = DG POST: Conductivity Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 16 = [21] -Title = Service Required: Dialysate Device +Title = DG drain line volume invalid calibration record ListTitle = DG POST: Drain Line Vol Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 936 = [22] -Title = Service Required: Dialysate Device +Title = DG reservoirs invalid calibration record ListTitle = DG POST: Reservoir Vol Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 935 = [23] -Title = Service Required: Dialysate Device +Title = DG acid concentrate calibration record ListTitle = DG POST: Acid Conc Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the non-volatile data is valid and if not raise the alarm SRSDG 934 = [24] -Title = Service Required: Dialysate Device +Title = DG bicarb concentrate calibration record ListTitle = DG POST: Bicarb Conc Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the non-volatile data is valid and if not raise the alarm SRSDG 933 = [25] -Title = Service Required: Dialysate Device +Title = DG accelerometer invalid calibration record ListTitle = DG POST: Acc Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calibration data that was received from NVDataMgmt is NOT legitimate SRSDG 817 = [26] -Title = Service Required: Hemodialysis Device +Title = HD accelerometer invalid calibration record ListTitle = HD POST: Acc Cal -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the accel calibration data that was received from NVDataMgmt is NOT legitimate. [27] -Title = Self Test: Blood Leak +Title = HD BLD sensor failed zero sequence ListTitle = Process: Pre- Tx Wet Self- Test -Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. +Message = BLD fails to complete zero sequence. [28] -Title = Service Required: Dialysate Device +Title = DG two wire sensors FPGA fault ListTitle = DG Fault: Conductivity RTD Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 864 = [29] -Title = Service Required: Hemodialysis Device +Title = HD heparin force sensor invalid calibration record ListTitle = HD POST: SP DAC Cal -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. [30] -Title = Service Required: Hemodialysis Device +Title = HD Software fault. Software found itself in an unexpected state ListTitle = HD Fault: SW Invalid State -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the specific software faults listed in alarm management SW faults has been detected [31] -Title = Service Required: Hemodialysis Device +Title = Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running ListTitle = HD Fault: BP Current -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. [32] -Title = Service Required: Hemodialysis Device +Title = Blood pump failed motor off check. Measured speed while commanded off ListTitle = HD Fault: BP Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM [33] -Title = Service Required: Hemodialysis Device +Title = Blood pump failed motor direction check. Measured vs. commanded ListTitle = HD Fault: BP Direction -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. [34] -Title = Blood Pump Error +Title = Blood pump failed rotor speed check. Mismatch with rotor and motor speeds ListTitle = HD Fault: BP Rotor -Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. +Message = If the delta of the measured motor speed and the commanded motor speed is not within tolerance. [35] -Title = Service Required: Hemodialysis Device +Title = Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running ListTitle = HD Fault: DPi Current -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. [36] -Title = Service Required: Hemodialysis Device +Title = Dialysis inlet pump failed motor off check. Measured speed while commanded off ListTitle = HD Fault: DPi Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If during command off state, the DPi pump motor speed is measured > 100.0 RMP. [37] -Title = Service Required: Hemodialysis Device +Title = Dialysis inlet pump failed motor direction check. Measured vs. commanded ListTitle = HD Fault: DPi Direction -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. [38] -Title = Dialysate Pump Error +Title = Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds ListTitle = HD Fault: DPi Rotor -Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. +Message = If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. [39] -Title = Service Required: Hemodialysis Device +Title = Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running ListTitle = HD Fault: DPo Current -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. [40] -Title = Service Required: Hemodialysis Device +Title = Dialysis outlet pump failed motor off check. Measured speed while commanded off ListTitle = HD Fault: DPo Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. [41] -Title = Service Required: Hemodialysis Device +Title = Dialysis outlet pump failed motor direction check. Measured vs. commanded ListTitle = HD Fault: DPo Direction -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. [42] -Title = Dialysate Pump Error +Title = Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds ListTitle = HD Fault: DPo Rotor -Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. +Message = If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed [43] -Title = Service Required: Dialysate Device +Title = Main primary heater FPGA fault ListTitle = DG Fault: Iso ADC Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 954 = [44] -Title = Service Required: Hemodialysis Device +Title = User interface communication timeout ListTitle = HD Fault: UI Comm -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. [45] -Title = Service Required: Hemodialysis Device +Title = Too many bad CRCs detected on received system messages ListTitle = HD Fault: CAN CRC Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If too many bad messages CRCs are received within a set period of time [46] -Title = Service Required: Hemodialysis Device +Title = System message that required acknowledgment was not acknowledged ListTitle = HD Fault: CAN ACK UI Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If no retries left in the pending acknowlegement list. [47] Title = -ListTitle = +ListTitle = DG Fault: RO Pump High Message = [48] -Title = Service Required: Hemodialysis Device +Title = Ultrafiltration volume accuracy error during treatment ListTitle = HD Fault: UF Rate Tare Err -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UF volume (refUFvolume - measureVolume) > 100.0 mL. [49] -Title = Service Required: Hemodialysis Device +Title = HD FPGA communication down for too long ListTitle = HD Fault: FPGA Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If there are more than 3 comm errors in a 1 min period. [50] -Title = Service Required: Dialysate Device +Title = DG FPGA not accepting commanded valve states ListTitle = DG Fault: Valve Control -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. SRSDG 26 = [51] -Title = Service Required: Hemodialysis Device +Title = Blood pump failed motor speed check. Measured vs. commanded ListTitle = HD Fault: BP Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If measured motor speed versus commanded rotor speed exceeded limits. [52] -Title = Service Required: Hemodialysis Device +Title = Dialysate inlet pump failed motor speed check. Measured vs. commanded ListTitle = HD Fault: DPi Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. [53] -Title = Service Required: Hemodialysis Device +Title = Dialysate outlet pump failed motor speed check. Measured vs. commanded ListTitle = HD Fault: DPo Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. [54] -Title = Service Required: Hemodialysis Device +Title = HD critical data integrity check failed ListTitle = HD Fault: Critical Data Err -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the critical data failed range check. [55] -Title = Service Required: Dialysate Device +Title = DG critical data integrity check failed ListTitle = DG Fault: Critical Data Err -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the critical data failed range check. SRSDG 921 = [56] -Title = Service Required: Hemodialysis Device +Title = HD accelerometer error (no readings or FPGA reports error) ListTitle = HD Fault: Acc Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. [57] -Title = Service Required: Dialysate Device +Title = DG accelerometer error (no readings or FPGA reports error) ListTitle = DG Fault: Acc Com -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. SRSDG 869 SRSDG 870 = [58] -Title = Service Required: Hemodialysis Device +Title = HD valve homing failed ListTitle = HD Fault: Valve Home -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the number of failed homing > = 3 counts. [59] -Title = Service Required: Hemodialysis Device +Title = HD valve transition time out ListTitle = HD Fault: Valve Transition Err -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = if the valve's transition time has timed out (3000msec) [60] -Title = Service Required: Hemodialysis Device +Title = HD valve not functional ListTitle = HD Fault: Valve Control -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the valve bit masks are not as expected in different valve control mode. [61] -Title = Service Required: Hemodialysis Device +Title = HD valve current out of range ListTitle = HD Fault: Valve Current -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the valve current is over the threshold of 1.0 A for 1000 msec. [62] -Title = Service Required: Hemodialysis Device +Title = HD valve position out of target ListTitle = HD Fault: Valve Position -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the current position has deviated from the expected position for more than 1 second. [63] Title = ListTitle = Message = [64] -Title = Service Required: Dialysate Device +Title = Barometric pressure is out of range ListTitle = DG Fault: Baro Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. SRSDG 874 = [65] -Title = Communication Error +Title = HD requests DG command with invalid parameter fault ListTitle = Loss of DG Comm (Reject) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the latest DG command response indicated invalid parameter [66] -Title = Service Required: Hemodialysis Device +Title = HD blood leak sensor set point set failure ListTitle = HD POST: BLD Config Set Point -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the set point of the blood leak sensor does not match the set point in the calibration record. [67] -Title = Self Test: Cartridge +Title = HD blood pump occlusion self-test failure alarm ListTitle = Process: Pre- Tx Remove Cartridge -Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK +Message = If the blood pump pressure > 5000 [68] -Title = Dialysate Generation Error +Title = HD active reservoir recirculation out of range ListTitle = Dial. Gen: Dialysate not ready -Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. +Message = If the recirculation level has exceeded the max allowed [69] -Title = Service Required: Hemodialysis Device +Title = HD blood leak sensor invalid calibration record ListTitle = HD POST: BLD Cal -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the BLD data that was received from NVDataMgmt is NOT legitimate. [70] -Title = Self Test: Cartridge +Title = HD arterial pressure self-test failure alarm ListTitle = Process: Pre- Tx Remove Cartridge -Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK +Message = If arterial pressure < -300 or > 100 mmHg [71] -Title = Self Test: Cartridge +Title = HD venous pressure self-test failure alarm ListTitle = Process: Pre- Tx Remove Cartridge -Message = A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK +Message = If the venous pressure < -100 or > 600 mmHg [72] -Title = Communication Error +Title = HD No load cell data message receive at least once every 2 seconds ListTitle = Loss of DG Comm (LC) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = No new load cell data for 2 seconds. [73] -Title = Communication Error +Title = HD No dialysate temperature data message received at least once every 2 seconds ListTitle = Loss of DG Comm (TDI/TRO) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = No new dialysate temperature data for 2 seconds. [74] -Title = Service Required: Dialysate Device +Title = DG inlet UV reactor is on while there is no flow ListTitle = DG Fault: UVi On No Flow -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the inlet UV reactor is on while VPi is closed for a certain period of time. SRSDG 600 = [75] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump self-test failure alarm ListTitle = HD Fault: Syringe DAC -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. [76] -Title = Service Required: Hemodialysis Device +Title = HD monitored voltage is out of range ListTitle = HD Fault: Voltage Monitor -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the HD monitored voltages are out of range for a certain period of time. [77] -Title = Service Required: Dialysate Device +Title = DG monitored voltage is out of range ListTitle = DG Fault: Voltage Monitor -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG monitored voltages are out of range for a certain period of time. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump direction (from encoder) error ListTitle = HD Fault: Syringe Pump Enc Dir -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the measured encoder direction of the syringe pump does not agree with the expected direction [79] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump direction (from controller) error ListTitle = HD Fault: Syringe Pump Drv Dir -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the measured direction of the syringe pump motor controller does not agree with the expected direction [80] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump fault reported by FPGA ListTitle = HD Fault: Syringe Pump Driver -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA reported the syringe pump fault [81] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump over travel error ListTitle = HD Fault: Syringe Pump OT -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the syringe pump position is deemed to have overtraveled [82] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump DAC write failure ListTitle = HD Fault: Syringe Pump DAC Err -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the syringe pump DAC write error is reported by the FPGA [83] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump is running while the BP is off ListTitle = HD Fault: Syringe Pump On Err -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the syringe pump is on while the blood pump is off [84] -Title = Service Required: Dialysate Device +Title = DG RTC set year is invalid ListTitle = DG POST: DG RTC Time -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If DG RTC set year is < 2021. SRSDG 815 = [85] -Title = Cartridge Lever is Open +Title = HD pump track lever is not fully latched ListTitle = Blood Stop: Cartridge Lever Open -Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. +Message = If the pump track lever is detected as unlatched when it should be. [86] -Title = Service Required: Hemodialysis Device +Title = HD RTC set year is invalid ListTitle = HD POST: HD RTC Time -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If HD RTC set year is < 2021. [87] -Title = Service Required: Dialysate Device +Title = DG heating parameters invalid calibration record ListTitle = DG POST: Heat Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the heating paramters' non-volatile record's CRC does not match the calculated CRC. SRSDG 929 = [88] -Title = Service Required: Dialysate Device +Title = DG concentrate pump hall sensor out of range. ListTitle = DG Fault: CP1 CP2 Speed Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = Hall sensor pulse to pulse timing is < 1ms (speed too fast). [89] -Title = Service Required: Dialysate Device +Title = DG outlet UV reactor is on while there is no flow ListTitle = DG Fault: UVo On No Flow -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the outlet UV reactor is on while the RO flow is less than 200 mL/min. SRSDG 599 = [90] -Title = Service Required: Dialysate Device +Title = DG load cells A1/B1 FPGA fault ListTitle = DG Fault: LC A1/B1 Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time. SRSDG 604 = [91] -Title = Communication Error +Title = HD No dialysate flow data received in the last 3 seconds ListTitle = Loss of DG Comm (FMD) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If not receiving new dialysate flow readings in expected interval (3 seconds) [92] Title = ListTitle = Message = [93] -Title = Service Required: Dialysate Device +Title = DG outlet primary conductivity sensor out of range ListTitle = DG Fault: RO Permeate Cond High -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. SRSDG 430 = [94] -Title = Service Required: Dialysate Device +Title = DG pressure sensor out of range ListTitle = DG Fault: Pressure Sensor Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. SRSDG 629 = [95] -Title = Service Required: Dialysate Device +Title = DG watchdog expired ListTitle = DG Fault: WD Error -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the watchdog signal is low while not in POST. SRSDG 907 = [96] -Title = Incoming Water Temperature High +Title = DG inlet water temperature out of range ListTitle = Dial. Gen: Inlet Water Temp High -Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +Message = If the inlet water temperature is > 37 C. SRSDG 410 = [97] -Title = Conductivity Out of Range +Title = DG fill conductivity out of range ListTitle = Dial. Gen: Dialysate Cond. -Message = Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +Message = If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. SRSDG 397 = [98] -Title = Service Required: Hemodialysis Device +Title = HD battery or battery charger communication fault ListTitle = HD Alarm (Low): Battery Com -Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. +Message = If battery I2C communication fault persistence count > 5. [99] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump stall alarm ListTitle = HD Fault: Syringe Pump Stall -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. [100] Title = ListTitle = Message = [101] -Title = Service Required: Hemodialysis Device +Title = UI POST FW Communication failure ListTitle = HD POST: UI- Proc Comm -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the HD is not communicating with the UI during POST. [102] -Title = Disinfection Interrupted +Title = DG heat disinfection temperature gradient out of range ListTitle = Cleaning: Tpo THD Temp Delta -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If temperature gradient between Tpo and Thd exceeds maximum 15°C. SRSDG 856 SRSDG 930 = [103] -Title = Service Required: Hemodialysis Device +Title = HD non-volatile calibration group invalid record CRC ListTitle = HD POST: Cal NVRAM CRC -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the computed CRC of the calibration group record is not equal to the record CRC. [104] -Title = Service Required: Hemodialysis Device +Title = Air trap level sensors reporting illegal combination of air/fluid ListTitle = HD Fault: Air Trap Sensor -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the air trap level readings are not valid for more than 2 seconds. [105] -Title = Service Required: Dialysate Device +Title = DG non-volatile system group invalid record CRC ListTitle = DG POST: Cal NVRAM CRC -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG NVDATAMGMT CAL group calculated CRC does not matched the expected CRC. SRSDG 608 = [106] -Title = Service Required: Dialysate Device +Title = DG flow sensors calibration error ListTitle = DG POST: Flow Sensor Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] -Title = Service Required: Dialysate Device +Title = HD reports DG restarted fault ListTitle = DG Fault: DG Restarted -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. [108] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump ADC error ListTitle = HD Fault: Syringe Pump ADC Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the ADC read is not fresh after POST init and the ADC read counter has not changed. [109] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump volume check error ListTitle = HD Fault: Syringe Pump Vol -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the volume delivered is off by more than 5% or 0.1mL. [110] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump speed check error ListTitle = HD Fault: Syringe Pump Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the syringe measured rate is off by more than 5% or 0.1mL. [111] -Title = Service Required: Hemodialysis Device +Title = HD syringe pump not stopped in off state error ListTitle = HD Fault: Syringe Pump Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the syringe pump position is not equal to its last position. [112] -Title = Service Required: Dialysate Device +Title = DG load cells A2/B2 FPGA fault ListTitle = DG Fault: LC A2/B2 Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time. SRSDG 953 = [113] -Title = Service Required: Dialysate Device +Title = DG drain pump current out of range ListTitle = DG Fault: DRP Current -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. SRSDG 890 = [114] -Title = Self Test: Venous Air Detector +Title = HD venous air bubble detector self-test failure ListTitle = Process: Pre- Tx Wet Self- Test -Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. +Message = If the venous air bubble self test did not detect bubble for more than 500 msec. [115] -Title = Service Required: Dialysate Device +Title = DG temperature sensor out of range ListTitle = DG Fault: Temp Sensor Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. SRSDG 224 = [116] -Title = Service Required: Dialysate Device +Title = SD card is non-operational ListTitle = HD Alarm (Low): SD- Card Error -Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. +Message = SD card is non-operational [117] -Title = Communication Error +Title = System message that required acknowledgment was not acknowledged ListTitle = Loss of HD Comm (HD ACK) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If no retries left in the pending acknowlegement list. SRSDG 909 = [118] -Title = Service Required: Dialysate Device +Title = DG RTC configuration error ListTitle = DG Fault: RTC Config -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RTC configuration registers report an error. SRSDG 43 = [119] -Title = Service Required: Dialysate Device +Title = DG RTC battery low ListTitle = DG Alarm (Low): RTC Bat Low -Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. +Message = If the RTC battery bit report a low alarm. SRSDG 908 = [120] -Title = Self Test: Venous Air Detector +Title = HD pre-treatment wet prime test failure ListTitle = Process: Pre- Tx Wet Self- Test -Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. +Message = If the HD pre-treatment wet self test failed. [121] -Title = Service Required: Dialysate Device +Title = DG main primary heater voltage out of range ListTitle = DG Fault: Prim Heater Lg Control -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the main primary heater's voltage is out of range. SRSDG 611 = [122] -Title = Service Required: Dialysate Device +Title = DG small primary heater voltage out of range ListTitle = DG Fault: Prim Heater Sm Control -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the small primary heater's voltage is out of range. SRSDG 612 = [123] -Title = Service Required: Dialysate Device +Title = DG trimmer heater voltage out of range ListTitle = DG Fault: Trimmer Heater Control -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the trimmer voltage is out of range. SRSDG 613 = [124] -Title = Blood Pump Stop Time Exceeded +Title = HD end of treatment alarm (high priority - user still not acting to end treatment) ListTitle = Blood Stop: Blood Clot Risk -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. +Message = In treatment end paused state for > 5 minutes. [125] -Title = Blood Pump Stop Time Exceeded +Title = Blood sitting too long after treatment stopped by user alarm (> 5 min) ListTitle = Blood Stop: Blood Clot Risk -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. +Message = If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. [126] -Title = Blood Leak +Title = HD blood leak detected alarm ListTitle = Blood Stop: Dialyzer Blood Leak -Message = A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. +Message = If the blood leak is detected based on the blood leak status reported by the sensor. [127] -Title = Low Venous Pressure +Title = Venous pressure too low during treatment ListTitle = Blood Stop: Venous Pres. Low -Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and\nterminate treatment without rinseback.\n- If access connections are properly secured, \npress Resume/OK to continue treatment. +Message = If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. [128] -Title = Venous Air Detected +Title = HD venous air bubble detected alarm ListTitle = Blood Stop: Venous Air Bubble -Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. +Message = If bubbles are detected during the treatment. [129] -Title = Blood Leak: Bypass +Title = HD blood leak detected and it is in the process of attempting to remove the bubbles/foam ListTitle = Blood Stop: Dialyzer Blood Leak Bypass -Message = The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. +Message = If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. [130] -Title = High Venous Pressure +Title = Venous pressure too high during treatment ListTitle = Blood Stop: Venous Pres. High -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. +Message = If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. [131] -Title = Low Arterial Pressure +Title = Arterial pressure too low during treatment ListTitle = Blood Stop: Arterial Pres. Low -Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. +Message = If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. [132] -Title = High Arterial Pressure +Title = Arterial pressure too high during treatment ListTitle = Blood Stop: Arterial Pres. High -Message = Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback. +Message = If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. [133] -Title = Dialysate Device Fluid Leak +Title = DG fluid leak detected alarm ListTitle = DG Fault: Fluid Leak in DG -Message = Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the fluid leak is detected as reported by the fluid leak detector. SRSDG 328 = [134] -Title = Fluid Leak +Title = HD fluid leak detected alarm ListTitle = Blood Stop: Fluid Leak in HD -Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. +Message = If the fluid leak is detected as reported by the fluid leak detector. [135] -Title = Blood Pump Stopped: Excessive Movement +Title = HD experienced large acceleration ListTitle = Blood Stop: HD Shock Detected -Message = Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. +Message = If the system experienced excessive shock > 2.5 g. [136] -Title = Dialysate Device: Excessive Movement +Title = DG experienced large acceleration ListTitle = Dial. Gen: Shock- Vibration -Message = Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. +Message = If the system experienced excessive shock > 2.5 g. SRSDG 327 = [137] -Title = Blood Pump Stopped: Excessive Tilt +Title = HD tilt exceeds maximum ListTitle = Blood Stop: HD Tilted -Message = Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. +Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. [138] -Title = Dialysate Device: Excessive Tilt +Title = DG tilt exceeds maximum ListTitle = Dial. Gen: Tilt -Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. +Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. SRSDG 326 = [139] -Title = Power Loss +Title = HD AC power lost alarm ListTitle = Loss of AC Power -Message = AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. +Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. [140] -Title = Communication Error +Title = Dialysate generator communication timeout ListTitle = Loss of DG Comm (Timeout) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If DG communication timeout is detected for more than 2 seconds. [141] -Title = Blood Pump Stopped: Venous Air Detector +Title = Air trap fill timeout during treatment ListTitle = Blood Stop: Air Trap Fill -Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. +Message = If the air trap fill timeout during treatment. [142] -Title = Blood Pump Stopped: Occlusion +Title = Blood pump occlusion detected ListTitle = Blood Stop: Blood Pump Occlusion -Message = Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. +Message = If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall [143] -Title = Service Required: Dialysate Device +Title = DG dialysate temperature sensors drift timeout. ListTitle = DG Fault: TRo TDi Delta -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If TDi and TRo have drifted for more 1 C for a certain period of time. SRSDG 912 = [144] -Title = Disinfection: Water Conductivity High +Title = DG cleaning mode inlet water conductivity out of high range ListTitle = Cleaning: Inlet Water Cond High -Message = Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = Conductivity Error During Prime +Title = DG Concentrate conductivity after adding acid out of range alarm ListTitle = Dial. Gen: Acid Conc. Cond. -Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +Message = If the conductivity value for the acid concentrate is out of range for more than 45 secs. SRSDG 846 = [146] -Title = Service Required: Dialysate Device +Title = DG RTC vs. Timer accuracy error ListTitle = DG POST: DG RTC Timer -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If RTC and timer are not aligned on 1 second elapsed time. SRSDG 312 = [147] -Title = Dialysate Generation Error +Title = DG dialysate generation conductivity fault alarm ListTitle = Dial. Gen: Dialysate not ready -Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. +Message = If dialysate conductivity is not with 5% of the TCD SRSDG 398 = [148] -Title = Disinfection: Water Temperature High +Title = Cleaning Mode Inlet water temperature too high ListTitle = Cleaning: Inlet Water Temp High -Message = Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +Message = Inlet temperature > 37 C. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] -Title = Service Required: Hemodialysis Device +Title = HD watchdog expired ListTitle = HD Fault: WD Error -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the watchdog signal is low while not in POST. [150] -Title = Incoming Water Conductivity High +Title = Inlet water conductivity too high ListTitle = Dial. Gen: Inlet Water Cond High -Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). SRSDG 391 = [151] -Title = Incoming Water Conductivity Low +Title = Inlet water conductivity too low ListTitle = Dial. Gen: Inlet Water Cond Low -Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). SRSDG 826 = [152] -Title = Incoming Water Pressure Low +Title = Inlet water pressure too low ListTitle = Dial. Gen: Inlet Water Pres Low -Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +Message = Inlet water pressure < 25 PSI. SRSDG 417 = [153] -Title = Pre-Treatment Process Terminated +Title = HD prime completed high priority alarm ListTitle = Process: Pre- Tx Terminated -Message = The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. +Message = In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) [154] -Title = Disinfection: Water Conductivity Low +Title = Cleaning mode inlet water conductivity too low ListTitle = Cleaning: Inlet Water Cond Low -Message = Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = Communication Error +Title = HD No reservoir data message receive at least once every 2 seconds ListTitle = Loss of DG Comm (Reservoir) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = No new reservoir data for 2 seconds. [156] -Title = Communication Error +Title = HD No DG op mode message receive at least once every 2 seconds ListTitle = Loss of DG Comm (Op Mode) -Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = No DG op mode for 2 seconds. [157] -Title = Chemical Disinfectant Not Detected +Title = Chemical disinfect priming the acid line timeout ListTitle = Cleaning: Chem Disinf PAA Prime -Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If the disinfectant acid line is not primed within a certain period of time. SRSDG 988 = [158] -Title = Incoming Water Temperature Low +Title = Inlet water temperature low ListTitle = Dial. Gen: Inlet Water Temp Low -Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +Message = Inlet temperature < 24 °C for 5 seconds. SRSDG 408 = [159] -Title = Remove Chemical Disinfectant +Title = Close concentrate line be able to flush it ListTitle = Cleaning: Chem Disinf Connect Cap -Message = The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. +Message = Start of chemical disinfect flush. SRSDG 1044 = [160] -Title = Service Required: Hemodialysis Device +Title = Blood pump rotor speed too high ListTitle = HD Fault: BP Rotor Max Speed -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If blood pump motor speed exceeded 100.0 RPM. [161] -Title = Service Required: Dialysate Device +Title = Too many bad CRCs detected on received system messages ListTitle = DG Fault: HD- DG Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If >= 5 bad CRCs detected on received CAN messages within 10 minutes. SRSDG 122 = [162] -Title = Service Required: Dialysate Device +Title = FPGA clock speed failure ListTitle = DG Fault: Proc- FPGA Clock -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA clock speed is relatively close to the processor clock for a certain period of time. SRSDG 867 = [163] -Title = Ultrafiltration Error +Title = HD sees primary load cell for reservoir 1 change too much too fast ListTitle = Dial. Gen: UF Rate Check -Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. +Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). [164] -Title = Ultrafiltration Error +Title = HD sees primary load cell for reservoir 2 change too much too fast ListTitle = Dial. Gen: UF Rate Check -Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. +Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). [165] -Title = Blood Pump Stop Time Exceeded +Title = HD in treatment rinseback operation timeout ListTitle = Blood Stop: Blood Clot Risk -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. +Message = If the maximum time inrinseback operation has been exceeded. [166] -Title = Test Alarm - Alarm 166 +Title = ListTitle = Message = [167] -Title = Blood Pump Stopped: Door Open +Title = HD cartridge door opened alarm ListTitle = Blood Stop: Door Open -Message = The device door is open.\n Close the device door. \n- Press Resume/OK to continue. +Message = If the cartridge door is opened as reported by the FPGA. [168] -Title = Test Alarm - Alarm 168 +Title = ListTitle = Message = [169] -Title = Service Required: Dialysate Device +Title = Dialysate flow rate out of range ListTitle = DG Fault: Dialysate Flow Sens Rg -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the dialysate flow rate > 2.00 L/min. SRSDG 643 = [170] -Title = Heparin Syringe Empty +Title = HD syringe empty alarm ListTitle = Heparin Syringe Empty -Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. +Message = If the syringe pump position is near empty position. [171] -Title = Heparin Pump Occlusion +Title = HD syringe pump occlusion alarm ListTitle = Process: Pre- Tx Hep. Pump Occ. -Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. +Message = If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. [172] -Title = Not Enough Heparin +Title = HD syringe pump not enough Heparin alarm ListTitle = Process: Pre- Tx Not Enough Hep. -Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. +Message = If the estimated syringe volume needed for treatment is not sufficient [173] -Title = Service Required: Hemodialysis Device +Title = HD RTC configuration error ListTitle = HD Fault: RTC Config -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RTC configuration registers report an error. [174] -Title = Service Required: Hemodialysis Device +Title = HD RTC vs. Timer accuracy error ListTitle = HD POST: HD RTC Timer -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If RTC and timer are not aligned on 1 second elapsed time. [175] -Title = Service Required: Hemodialysis Device +Title = HD pump commutation error ListTitle = HD Fault: Pump Commutation Err -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the set pump direction does not agree with the hall sensor direction. [176] -Title = Test Alarm - Alarm 176 +Title = ListTitle = Message = [177] -Title = Service Required: Dialysate Device +Title = DG software fault. Software found itself in an unexpected state ListTitle = DG Fault: Software Invalid State -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the specific software faults listed in alarm management SW faults has been detected SRSDG 919 = [178] -Title = Service Required: Hemodialysis Device +Title = HD communication timeout ListTitle = HD Fault: UI- Proc Comm -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the HD stops communicating with the UI. = SRSUI 1066 [179] -Title = Service Required: Dialysate Device +Title = DG FPGA communication down for too long ListTitle = DG Fault: FPGA Com -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If retries for commands exceeds limit or FPGA reports comm error. SRSDG 605 = [180] -Title = Service Required: Dialysate Device +Title = RO flow rate out of range ListTitle = DG Fault: RO Flow Sensor Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RO flow rate > 2.00 L/min. SRSDG 883 = [181] -Title = Service Required: Dialysate Device +Title = DG load cells weight out of range for tare ListTitle = DG Fault: Tare Out of Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the load cell weight is out of range for tare. SRSDG 849 SRSDG 838 = [182] -Title = Service Required: Dialysate Device +Title = DG load cells invalid calibration ListTitle = DG POST: LC Cal -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the CRC of load cells calibration data does not match the calculated CRC. SRSDG 162 = [183] -Title = Disinfection Interrupted +Title = DG invalid load cell value ListTitle = Cleaning: Invalid Resv Transfer -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. SRSDG 851 = [184] -Title = Service Required: Dialysate Device +Title = Inlet reactor not healthy ListTitle = DG Fault: UVi Error -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the inlet reactor are not healthy. SRSDG 559 = [185] -Title = Service Required: Dialysate Device +Title = DG fan RPM out of range ListTitle = DG Alarm (Low): Fan Speed -Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. +Message = If the fan RPM is out of Min/Max range. SRSDG 628 = [186] -Title = Service Required: Dialysate Device +Title = DG concentrate pump fault ListTitle = DG Fault: CP1/CP2 Fault -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the concentrate pumps fault is reported by the FPGA. SRSDG 891 = [187] -Title = Service Required: Dialysate Device +Title = DG concentrate pump CP1 speed control error ListTitle = DG Fault: CP1 Speed Fault -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. SRSDG 905 = [188] -Title = Service Required: Dialysate Device +Title = DG concentrate pump CP2 speed control error ListTitle = DG Fault: CP2 Speed Fault -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. SRSDG 906 = [189] -Title = Service Required: Dialysate Device +Title = DG drain pump RPM out of range ListTitle = DG Fault: DRP Speed Fault -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the drain pump RPM > 100.0 of the target RPM. SRSDG 448 = [190] -Title = Service Required: Dialysate Device +Title = DG drain pump off fault ListTitle = DG Fault: DRP Speed Not Off -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. [191] -Title = Service Required: Dialysate Device +Title = DG flow rate out of upper range ListTitle = DG Fault: RO Flow High -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RO flow rate > 10% from the target flow rate. SRSDG 885 = [192] -Title = Service Required: Dialysate Device +Title = DG flow rate out of lower range ListTitle = DG Fault: RO Flow Low -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RO flow rate < 10% from the target flow rate. SRSDG 945 = [193] @@ -1058,116 +1058,116 @@ Message = [194] -Title = Service Required: Dialysate Device +Title = DG RO pump duty cycle out of range ListTitle = DG Fault: RO Driver Feedback -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. SRSDG 881 = [195] -Title = Service Required: Dialysate Device +Title = DG RO pump pressure out of range ListTitle = DG Fault: RO Pump High Pressure -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG RO pump pressure > 135 psi for a certain period of time. SRSDG 925 = [196] -Title = Service Required: Dialysate Device +Title = DG CPi/CPo conductivity sensors FPGA fault ListTitle = DG Fault: CPi/CPo Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA reports error for a certain period of time. SRSDG 955 = [197] -Title = Service Required: Dialysate Device +Title = DG CD1/CD2 conductivity sensors FPGA fault ListTitle = DG Fault: CD1/CD2 Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA reports error for a certain period of time. SRSDG 956 = [198] -Title = Service Required: Dialysate Device +Title = DG primary heater while RO flow is low ListTitle = DG Fault: HP On With No Flow -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RO flow is < 0.2 L/min for a certain period of time. SRSDG 872 = [199] -Title = Service Required: Dialysate Device +Title = DG trimmer heater while dialysate flow is low ListTitle = DG Fault: HR On With No Flow -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. SRSDG 871 = [200] -Title = Service Required: Dialysate Device +Title = DG thermistor or temperature sensor out of range ListTitle = DG Fault: Thermistor Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If thermistor > 80.0 C or < 0.0 C. SRSDG 178 = [201] -Title = Self Test: Dialysate Flow Rate +Title = HD pre-treatment wet flow test failure ListTitle = Process: Pre- Tx Wet Self- Test -Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. +Message = If the pre-treatment wet flow test has failed. [202] -Title = Self Test: Pressure +Title = HD pre-treatment dry pressure normal test failure ListTitle = Process: Pre- Tx Dry Self- Test -Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. +Message = If the pre-treatment dry pressure normal test failed. [203] Title = ListTitle = Message = [204] -Title = Service Required: Dialysate Device +Title = RO rejection ratio out of range ListTitle = DG Fault: RO Rejection -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If RO rejection ratio > 0.1 SRSDG 429 = [205] -Title = Service Required: Dialysate Device +Title = Conductivity sensor fault ListTitle = DG Fault: Cond Sensors Error -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If conductivity sensor reports 0x11 for more than 2 consecutive reads. SRSDG 913 = [206] -Title = Service Required: Dialysate Device +Title = DG dialysate fill runs out of time ListTitle = DG Fault: Reservoir Fill Timeout -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If DG dialysate fill > 5 minutes. SRSDG 437 = [207] -Title = Service Required: Dialysate Device +Title = DG flow meter check failure alarm ListTitle = DG Fault: FMP Flow Rate Error -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If integrated Volume To Load Cell Reading Percent > 0.1. SRSDG 240 = [208] -Title = High Venous Pressure +Title = HD venous line occlusion fault ListTitle = Blood Stop: Venous Pres. High -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. +Message = If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. [209] -Title = Disinfection Interrupted +Title = DG drain circulation line timeout ListTitle = Cleaning: Flush Circ Error -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If the integrated volume < 0.1 for more than 60 secs. SRSDG 980 = [210] -Title = Service Required: Hemodialysis Device +Title = HD battery pack detected an error ListTitle = HD Alarm (Low): Battery Fault -Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. +Message = If the battery pack status error is detected from the slave device I2C address 0x0B. [211] -Title = Approaching MAX Blood Pump Stop Time +Title = Blood sitting too long warning (> 4 min) ListTitle = Blood Stopped: Warning -Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. +Message = During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. [212] -Title = Approaching MAX Blood Pump Stop Time +Title = Blood sitting too long warning (> 4 min) ListTitle = Blood Stopped: Warning -Message = The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. +Message = During treatment end paused state, if the blood is in the line for more than 4 minutes. [213] -Title = Prime Complete - 30 MIN WARNING +Title = HD prime completed medium priority alarm ListTitle = Process: Pre- Tx Prime Complete -Message = The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. +Message = In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) [214] Title = @@ -1180,140 +1180,140 @@ Message = [216] -Title = Treatment Complete +Title = HD end treatment sub-mode timeout alarm ListTitle = Process: End of Dialysis Tx -Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. +Message = If the treatment ended successfully. [217] Title = ListTitle = Message = [218] -Title = Syringe Detected +Title = HD syringe detected alarm ListTitle = Process: Remove Syringe -Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. +Message = If the syringe detection switch > 2.0 volts. [219] -Title = Blood Stopped: Syringe Not Detected +Title = HD syringe pump syringe removed alarm ListTitle = Blood Stop: Syringe Not Detected -Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. +Message = If the syringe detection switch < 2.0 volts. [220] -Title = Service Required: Dialysate Device +Title = DG THd sensor FPGA fault ListTitle = DG Fault: THD RTD Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 947 = [221] -Title = Blood Pump Stopped: Saline Bag Empty +Title = HD Saline bag is empty ListTitle = Blood Stop: Saline Bag Empty -Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. +Message = If the measured arterial pressure < -300.0 mmHg for more than 250 msec. [222] -Title = Service Required: Hemodialysis Device +Title = HD occlusion sensor FPGA fault ListTitle = HD Fault: BP Occ Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data or indicates an error for a certain period of time. [223] -Title = Service Required: Hemodialysis Device +Title = HD arterial sensor FPGA fault ListTitle = HD Fault: Art. Pres. Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time [224] -Title = Blood Pump Stopped: Treatment Paused +Title = Treatment stopped by user action - pressed stop button ListTitle = Blood Stopped: Treatment Pause -Message = Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. +Message = If the user has pressed the stop button. [225] -Title = Treatment Complete +Title = HD end of treatment warning ListTitle = Process: End of Treatment -Message = Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. +Message = If the treatment duration has exceeded the prescribed treatment time entered by the user. [226] -Title = Priming Complete +Title = HD prime completed low priority alarm ListTitle = Process: Pre- Tx Priming Complete -Message = Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. +Message = Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection [227] Title = ListTitle = Message = [228] -Title = Priming Error +Title = HD time out on prime sub-mode purge air state. ListTitle = Process: Pre- Tx Prime Timeout -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. +Message = During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. [229] -Title = Priming Error +Title = HD prime dialysate dialyzer time out alarm ListTitle = Process: Pre- Tx Prime Timeout -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. +Message = If the priming of the dialyzer dialysate flow path has exceeded the time limit. [230] -Title = Priming Error +Title = HD prime dialysate bypass time out alarm ListTitle = Process: Pre- Tx Prime Timeout -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. +Message = If the priming of the dialyzer bypass flow path has exceeded the time limit. [231] -Title = Self Test: Cartridge Check +Title = HD pre-treatment mode dry self-tests failure ListTitle = Process: Pre- Tx Dry Self- Test -Message = A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. +Message = If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState [232] -Title = Self Test: Dialysate Flow Rate +Title = HD pre-treatment mode wet self-tests failure ListTitle = Process: Pre- Tx Wet Self- Test -Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. +Message = If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState [233] -Title = Service Required: Dialysate Device +Title = DG barometric pressure sensor bad coefficients CRC ListTitle = DG Fault: Baro Cal CRC -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = Incoming Water Pressure High +Title = DG inlet water pressure is too high ListTitle = Dial. Gen: Inlet Water Pres High -Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +Message = If the inlet water pressure is > 80 psig. SRSDG 1037 = [235] -Title = Rinseback Complete +Title = HD in treatment stopped sub-mode after rinseback completed (no escalation) ListTitle = Process: Rinseback Complete -Message = Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. +Message = If there is no active alarm and rinseback has completed. [236] -Title = Self Test: Fluid Present +Title = HD needs new cartridge to be installed ListTitle = Process: Pre- Tx Dry Self- Test -Message = A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. +Message = If the air trap sensor reported AIR and bubble sensor detected bubbles. [237] Title = ListTitle = Message = [238] -Title = Cartridge Not Detected +Title = HD no cartridge loaded or installed improperly alarm ListTitle = Process: Pre- Tx Dry Self- Test -Message = A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. +Message = If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. [239] -Title = Cartridge Still Detected +Title = HD fail to remove cartridge alarm ListTitle = Process: Post- Tx Remove Cartridge -Message = A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. +Message = In the post treatment state, if the cartridge has not been removed. [240] -Title = Conductivity Error During Prime +Title = Bicarb conductivity out of range during bicarb pump check alarm ListTitle = Dial. Gen: Bicarb Conc. Cond. -Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +Message = If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. SRSDG 482 = [241] -Title = Disinfection Interrupted +Title = DG reservoir drain time out ListTitle = Cleaning: Resv Drain Timeout -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If the DG reservoir drain is not complete in 120 secs. SRSDG 848 SRSDG 90 SRSDG 561 @@ -1344,9 +1344,9 @@ SRSDG 1034 = [242] -Title = Disinfection Interrupted +Title = DG reservoir fill time out ListTitle = Cleaning: Resv Fill Timeout -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If the DG reservoir fill is not complete in 120 secs. SRSDG 850 SRSDG 573 SRSDG 574 @@ -1367,81 +1367,81 @@ SRSDG 1024 = [243] -Title = Disinfection Interrupted +Title = DG reservoir leak time out ListTitle = Cleaning: Fluid Loss Detected -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If the reservoir level is drifted by 100 mL for more than 30 seconds. SRSDG 853 = [244] -Title = Disinfection Interrupted +Title = DG temperature sensors difference out of range ListTitle = Cleaning: Temp Sensor Check -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If any of the temperature sensors are out of the average of the sensors by more than 3 C. SRSDG 852 = [245] -Title = Disinfection Interrupted +Title = DG heat disinfect target temperature time out (could not reach to temperature) ListTitle = Cleaning: Heat Disinfect Temp TO -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If heating up to minimum temperature for heat disinfect failed after 4 hours. SRSDG 854 = [246] -Title = Disinfection Interrupted +Title = DG chemical disinfect inlet conducitivty out of range ListTitle = Cleaning: RO Water Cond Error -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If CPo and CD2 conductivity values are above 100 uS/cm. [247] Title = ListTitle = Message = [248] -Title = Disinfection Interrupted +Title = DG chemical disinfect out of target temperature ListTitle = Cleaning: Chem Disinfect Proc -Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If TPo > 50 C after 1 second. [249] -Title = Chemical Disinfectant Not Detected +Title = DG chemical disinfect out of target conductivity ListTitle = Cleaning: Chem Disinf PAA Cond -Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +Message = If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. SRSDG 992 = [250] -Title = Connect Chemical Disinfectant +Title = Insert the acid to start chemical disinfect ListTitle = Cleaning: Chem Disinf Connect PAA -Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. +Message = Start of chemical disinfect SRSDG 1046 = [251] -Title = Service Required: Hemodialysis Device +Title = HD invalid system record ListTitle = HD POST: Sys Rec NVRAM CRC -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated CRC of the HD system record image does not match the expected CRC. [252] -Title = Service Required: Hemodialysis Device +Title = HD invalid service record ListTitle = HD POST: Serv Rec NVRAM CRC -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated CRC of the HD service record image does not match the expected CRC. [253] -Title = Service Required: Dialysate Device +Title = DG invalid system record ListTitle = DG POST: Sys Rec NVRAM CRC -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated CRC of the DG system record image does not match the expected CRC. SRSDG 609 = [254] -Title = Service Required: Dialysate Device +Title = DG invalid service record ListTitle = DG POST: Serv Rec NVRAM CRC -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the calculated CRC of the DG service record image does not match the expected CRC. SRSDG 610 = [255] -Title = Service Required: Hemodialysis Device +Title = HD and UI software builds are not compatible ListTitle = HD POST: UI Compatibility -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. [256] -Title = Chemical Flush in Progress +Title = DG flush started ListTitle = Cleaning: Chem Flush In Progress -Message = Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. +Message = If DG flush successfully started from the UI. = SRSUI 497 [257] @@ -1455,84 +1455,84 @@ Message = [259] -Title = Service Required: Hemodialysis Device +Title = HD temperatures out of range ListTitle = HD Fault: Temp Sensors Error -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. [260] -Title = Service Required: Hemodialysis Device +Title = UI POST Application Integrity (Sha256Sum) failure ListTitle = HD POST: UI Image -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports a shasum faillure. = SRSUI 1067 [261] -Title = Service Required: Hemodialysis Device +Title = UI POST CANBus failure ListTitle = HD POST: UI CAN Driver -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports a can bus failure. = SRSUI 1068 [262] -Title = +Title = UI POST Display failure ListTitle = -Message = +Message = If the UI POST reports a display failure. [263] -Title = Service Required: Hemodialysis Device +Title = UI POST Touch failure ListTitle = HD POST: UI Touch Driver -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports a touch failure. = SRSUI 1070 [264] -Title = Service Required: Hemodialysis Device +Title = UI POST SD-Card failure ListTitle = HD POST: UI SD- Card -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports SD card failure. = SRSUI 1071 [265] -Title = Service Required: Hemodialysis Device +Title = UI POST RTC failure ListTitle = HD POST: UI RTC -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports RTC failure. = SRSUI 1072 [266] -Title = Service Required: Hemodialysis Device +Title = UI POST WiFi failure ListTitle = HD POST: UI WiFi -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports WiFi failure. = SRSUI 1073 [267] -Title = Service Required: Hemodialysis Device +Title = UI POST Bluetooth failure ListTitle = HD POST: UI Bluetooth -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports bluetooth failure. = SRSUI 1074 [268] -Title = Service Required: Hemodialysis Device +Title = UI POST Ethernet failure ListTitle = HD POST: UI Ethernet -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports ethernet failure. = SRSUI 1075 [269] -Title = Service Required: Hemodialysis Device +Title = UI POST Sound failure ListTitle = HD POST: UI Sound -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports sound failure. = SRSUI 1076 [270] -Title = Service Required: Hemodialysis Device +Title = HD POST Safety Shutdown failure ListTitle = HD POST: Proc Safestate -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. [271] -Title = Service Required: Dialysate Device +Title = DG POST Safety Shutdown failure ListTitle = DG POST: Proc Safestate -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. SRSDG 822 = [272] -Title = Service Required: Hemodialysis Device +Title = HD Fan RPM out of range ListTitle = HD Alarm (Low): Fan Speed -Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. +Message = If the fan RPM is out of Min/Max range. [273] Title = @@ -1545,276 +1545,276 @@ Message = [275] -Title = Service Required: Hemodialysis Device +Title = HD arterial pressure sensor is reading out of range ListTitle = HD Fault: Arterial Press Range -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. [276] -Title = Service Required: Hemodialysis Device +Title = HD venous pressure sensor is reading out of range ListTitle = HD Fault: Venous Press Range -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. [277] -Title = Service Required: Hemodialysis Device +Title = HD BP occlusion sensor is reading out of range ListTitle = HD Fault: Occlusion Sens Range -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the blood pump occlusion count is < 2000. [278] -Title = Service Required: Dialysate Device +Title = HD active reservoir weight out of range. ListTitle = DG Fault: Reservoir Overfill -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the active reservoir weight is > 1900.0 g. [279] -Title = Service Required: Dialysate Device +Title = DG dialysate drain time out ListTitle = DG Fault: Reservoir Drain -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG dialysate drain exceeded 120 seconds. SRSDG 752 = [280] -Title = Service Required: Dialysate Device +Title = DG drain pump invalid direction ListTitle = DG Fault: DRP Direction Fault -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the drain pump's direction is not 1 when draining. SRSDG 888 = [281] -Title = Acid Concentrate Low +Title = DG Acid concentration bottle low volume alarm ListTitle = Dial. Gen: Acid Conc. Low -Message = The acid concentrate is low.\n- Replace acid concentrate container. +Message = If the acid concentration bottle volume > 3240 (3340 - 100). SRSDG 836 = [282] -Title = Bicarb Concentrate Low +Title = DG Bicarbonate concentration bottle low volume alarm ListTitle = Dial. Gen: Bicarb Conc. Low -Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. +Message = If the bicarbonate concentration bottle volume > 3680 (3780 - 100). SRSDG 837 = [283] -Title = Service Required: Dialysate Device +Title = DG load cells weight out of range ListTitle = DG Fault: LC Range -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the weight of the DG load cells > 4500.0 g or < 0.0 g. SRSDG 615 = [284] -Title = Service Required: Dialysate Device +Title = DG load cells primary/back up drift out of range ListTitle = DG Fault: Redundant LC Check -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If redundant load cells deviate by more than 60g after tare for 5 seconds. SRSDG 383 = [285] -Title = RO Permeate Sample in Progress +Title = DG RO permeate sample started ListTitle = Cleaning: RO Permeate Sample In Progress -Message = RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. +Message = If DG RO permeate sample successfully started from UI. [286] -Title = Concentrate Connection Cap Removed +Title = DG cap in wrong state ListTitle = Cleaning: Conc Cap Removed -Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. +Message = If the concentrate cap is not in the right position in the cleaning modes. [287] -Title = Service Required: Hemodialysis Device +Title = HD processor clock speed checks against FPGA clock failure ListTitle = HD Fault: Proc- FPGA Clock -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. [288] -Title = Service Required: Dialysate Device +Title = HD load cell inconsistency ListTitle = DG Fault: LC Drift Check -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. [289] -Title = Dialysate Connection Cap Removed +Title = DG cap in wrong state ListTitle = Cleaning: Dialysate Cap Removed -Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. +Message = If the dialysate cap is not in the right position in the cleaning modes. SRSDG 928 = [290] -Title = Service Required: Hemodialysis Device +Title = HD venous sensor FPGA fault ListTitle = HD Fault: Ven. Pres. Com -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time [291] -Title = Flush In Progress +Title = DG flush started ListTitle = Cleaning: Flush In Progress -Message = System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. +Message = If DG flush successfully started from the UI. [292] -Title = Heat Disinfection in Progress +Title = DG heat disinfect started ListTitle = Cleaning: Heat Disinf In Progress -Message = Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. +Message = If DG heat disinfect successfully started from the UI. [293] -Title = Chemical Disinfection in Progress +Title = DG chem disinfect started ListTitle = Cleaning: Chem Disinfect In Progress -Message = Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. +Message = If DG flush successfully started from the UI. [294] -Title = Active Cool in Progress +Title = DG active cool started ListTitle = Cleaning: Active Cool In Progress -Message = Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. +Message = If DG active cool successfully started. [295] -Title = Service Required: Hemodialysis Device +Title = UI cloud sync POST failure ListTitle = HD POST: UI CloudSync -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports cloud sync failure. = SRSUI 1238 [296] -Title = Service Required: Dialysate Device +Title = DG TDi sensor FPGA fault ListTitle = DG Fault: TDi RTD Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 951 = [297] -Title = Service Required: Dialysate Device +Title = DG TRo sensor FPGA fault ListTitle = DG Fault: TRo RTD Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 950 = [298] -Title = Service Required: Dialysate Device +Title = DG barometric sensor FPGA fault ListTitle = DG Fault: Baro Pres Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA does not report fresh data for a certain period of time SRSDG 948 = [299] -Title = Service Required: Dialysate Device +Title = DG invalid serial number ListTitle = DG POST: Serial Number Check -Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the DG serial number is empty and does not have any characters. SRSDG 952 = [300] -Title = Service Required: Hemodialysis Device +Title = HD invalid serial number ListTitle = HD POST: Serial Number Check -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the HD serial number is empty and does not have any characters. [301] -Title = Remove Dialysate Cap +Title = Open dialysate cap to be able to collecte the dispensing fluid ListTitle = Cleaning: RO Permeate Sample Connect Cap -Message = Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. +Message = RO permeate sample collection state [302] -Title = Service Required: Dialysate Device +Title = DG_CPU RAM error ListTitle = DG Fault: Processor RAM error -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. SRSDG 126 = [303] -Title = Service Required: Dialysate Device +Title = DG drain pump direction FPGA fault ListTitle = DG Fault: DRP Hall Commutation -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the FPGA indicates an error for a certain period of time. SRSDG 958 = [304] -Title = Service Required: Hemodialysis Device +Title = HD invalid usage reocord ListTitle = HD POST: Usage Rec NVRAM CRC -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the CRC of the HD usage record does not match the calculated CRC. [305] -Title = Service Required: Hemodialysis Device +Title = HD_CPU RAM error ListTitle = HD Fault: Processor RAM error -Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. [306] -Title = Service Required: Dialysate Device +Title = ListTitle = DG Fault: CD1 CD2 Check -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = CD1 and CD2 conductivity measurement during acid check is not within 5% of each other [307] -Title = Power Loss +Title = HD AC power lost alarm ListTitle = Blood Stop: Loss of AC Power -Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. +Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. [308] -Title = Service Required: Dialysate Device +Title = DG CPi conductivity sensor transmits invalid characters ListTitle = DG Fault: CPi Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the CPi conductivity sensor transmits invalid characters for a certain period of time. SRSDG 959 = [309] -Title = Service Required: Dialysate Device +Title = DG CPo conductivity sensor transmits invalid characters ListTitle = DG Fault: CPo Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the CPo conductivity sensor transmits invalid characters for a certain period of time. SRSDG 960 = [310] -Title = Service Required: Dialysate Device +Title = DG CD1 conductivity sensor transmits invalid characters ListTitle = DG Fault: CD1 Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the CD1 conductivity sensor transmits invalid characters for a certain period of time. SRSDG 962 = [311] -Title = Service Required: Dialysate Device +Title = DG CD2 conductivity sensor transmits invalid characters ListTitle = DG Fault: CD2 Comm -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the CD2 conductivity sensor transmits invalid characters for a certain period of time. SRSDG 961 = [312] -Title = Low Dialysate Flow +Title = The dialysate inlet flow is too low. ListTitle = Dial. Gen: Dialysate In Flow -Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. +Message = If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. [313] -Title = Disinfection: Water Temperature Low +Title = DG cleaning mode inlet water temperature too low ListTitle = Cleaning: Inlet Water Temp Low -Message = Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +Message = If inlet water temperature is < 24 C. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = Disinfection: Water Pressure High +Title = DG cleaning mode inlet water pressure too high ListTitle = Cleaning: Inlet Water Press High -Message = Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +Message = If inlet water temperature is > 80 psig. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = Disinfection: Water Pressure Low +Title = DG cleaning mode inlet water pressure too low ListTitle = Cleaning: Inlet Water Press Low -Message = Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water initiating disinfection again. +Message = If inlet water pressure is < 14 psig. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = Service Required: Dialysate Device +Title = Acid concentrate pump did not park successfully ListTitle = DG Fault: Acid CP Pump Park -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the park took more than 1 second. SRSDG 983 = [317] -Title = Service Required: Dialysate Device +Title = Bicarb concentrate pump did not park successfully ListTitle = DG Fault: Bicarb CP Pump Park -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the park took more than 1 second. SRSDG 984 = [318] -Title = Air Trap Control Error +Title = HD Air Pump operation timeout ListTitle = HD Fault: Air Pump Timeout -Message = A problem has been detected with the air trap level\ncontrol system.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. +Message = If air pump has been turned on for greater than 10s. [319] -Title = Residual Chemical Sampling +Title = Collect sample to check the chemical disinfectant acid ListTitle = Cleaning: Chem Residual Sampling -Message = A water sample is needed to confirm the \nabsence of residual chemical following disininfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. +Message = Chemical disinfect flush sample alarm is prompted. SRSDG 1040 = [320] -Title = Chemical Residual Timeout +Title = The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes ListTitle = Cleaning: Chem Residual Timeout -Message = A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. +Message = Chemical disinfect flush sample was not collected form more than 30 minutes. SRSDG 1043 = [321] -Title = Service Required: Dialysate Device +Title = Outlet reactor not healthy ListTitle = DG Fault: UVo Error -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the outlet reactor are not healthy. SRSDG 1069 = [322] -Title = Service Required: Hemodialysis Device +Title = UI POST Invalid Year ListTitle = HD POST Alarm: UI Invalid Year -Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Message = If the UI POST reports a invalid year. [323] Title = Service Required: Hemodialysis Device