Index: Alarms/Alarms.conf =================================================================== diff -u -rffd732f4855260631d8606ab887c3f7da61fd521 -r0567f0d0212467a2b1060f85dfcdea3fe88b323d --- Alarms/Alarms.conf (.../Alarms.conf) (revision ffd732f4855260631d8606ab887c3f7da61fd521) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 0567f0d0212467a2b1060f85dfcdea3fe88b323d) @@ -1175,9 +1175,9 @@ Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [215] -Title = -ListTitle = -Message = +Title = Air Trap Overfill +ListTitle = Air Trap: Air Trap Overfill +Message = Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. [216] Title = Treatment Complete