Index: Alarms/Alarms.conf =================================================================== diff -u -rac1f72effe1248f2c5ffa111aefad93fa463308f -r079ccbf302493b44bf66c218d1d03d696dbfb090 --- Alarms/Alarms.conf (.../Alarms.conf) (revision ac1f72effe1248f2c5ffa111aefad93fa463308f) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 079ccbf302493b44bf66c218d1d03d696dbfb090) @@ -1175,9 +1175,9 @@ Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [215] -Title = Air Trap Upper Level Fluid +Title = Air Trap Overfill ListTitle = Air Trap: Air Trap Upper Fluid -Message = Air trap upper level is detecting fluid for too long \n- Visually inspect air trap line.\n- Press Resume/OK to proceed. +Message = Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots, address as clinically indicated per facility policy. \n- Press Resume/OK to continue treatment. [216] Title = Treatment Complete