Index: Alarms/Alarms.conf =================================================================== diff -u -r383cbbc43bbca1b39dc7b6d5d68a163487163459 -r0d28971551d7704b41d77fff3d4f72c7e14b588c --- Alarms/Alarms.conf (.../Alarms.conf) (revision 383cbbc43bbca1b39dc7b6d5d68a163487163459) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 0d28971551d7704b41d77fff3d4f72c7e14b588c) @@ -1053,9 +1053,9 @@ SRSDG 945 = [193] -Title = -ListTitle = -Message = +Title = Blood Pump Stopped: Partial Occlusion +ListTitle = Blood Stop: Blood Pump Partial Occlusion +Message = Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [194] Title = Service Required: Dialysate Device @@ -1777,15 +1777,15 @@ SRSDG 966 = [316] -Title = Service Required: Dialysate Device -ListTitle = DG Fault: Acid CP Pump Park -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Title = +ListTitle = +Message = SRSDG 983 = [317] -Title = Service Required: Dialysate Device -ListTitle = DG Fault: Bicarb CP Pump Park -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Title = +ListTitle = +Message = SRSDG 984 = [318]