Index: Alarms/Alarms.conf =================================================================== diff -u -rde1c5d548f0121f78c48484d439fe4dd4e526101 -r1fff96e1a3b712aafecb75e138b479db91502c02 --- Alarms/Alarms.conf (.../Alarms.conf) (revision de1c5d548f0121f78c48484d439fe4dd4e526101) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 1fff96e1a3b712aafecb75e138b479db91502c02) @@ -1185,9 +1185,9 @@ Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. [217] -Title = -ListTitle = -Message = +Title = Air Trap Overfill +ListTitle = Air Trap: Air Trap Overfill +Message = Air trap level high for too long. \n- Assess venous access and resolve access issues, if applicable.\n- Ensure venous chamber line is securely clamped.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect blood lines for clots, address as clinically indicated\n per facility policy. \n- Press Resume/OK to continue treatment. [218] Title = Syringe Detected