Index: Alarms/Alarms.conf =================================================================== diff -u -re98d63aeba011576977981439b291f7429d19f97 -r310cf32ee8ff2a58a0cbfd6b568c7dc7485aa84e --- Alarms/Alarms.conf (.../Alarms.conf) (revision e98d63aeba011576977981439b291f7429d19f97) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 310cf32ee8ff2a58a0cbfd6b568c7dc7485aa84e) @@ -1,1055 +1,1055 @@ [0] -Title = Indication for when no alarms is a possible situation +Title = ListTitle = Message = [1] -Title = Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up +Title = Stuck Button ListTitle = HD POST: Stuck Button -Message = Stop or Off button are pressed for more than 1 sec during POST. +Message = A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. [2] -Title = HD FPGA POST failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: FPGA Compatibility -Message = Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [3] -Title = DG FPGA POST failure +Title = Service Required: Dialysate Device ListTitle = DG POST: FPGA Compatibility -Message = Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [4] -Title = HD Watchdog POST failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: WD Safestate -Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [5] -Title = DG Watchdog POST failure +Title = Service Required: Dialysate Device ListTitle = DG POST: WD Safestate -Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 53 = [6] -Title = UI communication POST failure. UI failed to communicate within a reasonable time after power up +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Proc- UI Comm -Message = If the UI failed to checkin via the CAN bus within 40 seconds +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [7] -Title = Real-time clock configuration error +Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): RTC Bat Low -Message = If the RTC is not functional as indicated by the bits in the first 3 control registers. +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [8] -Title = HD accelerometer failed POST +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Acc Vector -Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [9] -Title = DG accelerometer failed POST +Title = Service Required: Dialysate Device ListTitle = DG POST: Acc Vector -Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 323 = [10] -Title = HD blood leak sensor embedded mode communication failed +Title = Service Required: Hemodialysis Device ListTitle = HD POST: BLD Config Emb Mode -Message = If the blood leak sensor embedded mode is unresponsive. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [11] -Title = Dialysate temperature is below the target temperature +Title = Low Dialysate Temperature ListTitle = Dial. Gen: Dialysate Temp Low -Message = If the dialysate temperature is below the target temperature for a certain period of time. +Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [12] -Title = HD firmware image integrity POST test failed +Title = Service Required: Hemodialysis Device ListTitle = HD POST: HD Image -Message = If the calculated CRC of the HD image does not match the expected CRC. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [13] -Title = DG firmware image integrity POST test failed +Title = Service Required: Dialysate Device ListTitle = DG POST: DG Image -Message = If the calculated CRC of the DG firmware image does not matched the expected CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 813 = [14] -Title = DG non-volatile usage information group invalid record CRC +Title = Service Required: Dialysate Device ListTitle = DG POST: Usage Rec NVRAM CRC -Message = If the DG usage information record's CRC does not match the calculated CRC in POST. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 957 = [15] -Title = Dialysate temperature is out of high safety range +Title = High Dialysate Temperature ListTitle = Dial. Gen: Dialysate Temp High -Message = If the dialysate temperature is > 46 C for a certain period of time. +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [16] -Title = HD alarm audio failed POST +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Primary Alarm Current -Message = If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [17] -Title = HD UI POST failed +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI POST Not Complete -Message = If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [18] -Title = HD to DG comm error. +Title = Communication Error ListTitle = Loss of DG Comm (DG ACK) -Message = HD message to DG that requires acknowledge was not acknowledged within a certain time. +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [19] -Title = Dialysate temperature is above the target temperature +Title = High Dialysate Temperature ListTitle = Dial. Gen: Dialysate Temp High -Message = If the dialysate temperature is above the target temperature for a certain period of time. +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [20] -Title = DG conductivity sensors invalid calibration record +Title = Service Required: Dialysate Device ListTitle = DG POST: Conductivity Cal -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 16 = [21] -Title = DG drain line volume invalid calibration record +Title = Service Required: Dialysate Device ListTitle = DG POST: Drain Line Vol Cal -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 936 = [22] -Title = DG reservoirs invalid calibration record +Title = Service Required: Dialysate Device ListTitle = DG POST: Reservoir Vol Cal -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 935 = [23] -Title = DG acid concentrate calibration record +Title = Service Required: Dialysate Device ListTitle = DG POST: Acid Conc Cal -Message = If the non-volatile data is valid and if not raise the alarm +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 934 = [24] -Title = DG bicarb concentrate calibration record +Title = Service Required: Dialysate Device ListTitle = DG POST: Bicarb Conc Cal -Message = If the non-volatile data is valid and if not raise the alarm +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 933 = [25] -Title = DG accelerometer invalid calibration record +Title = Service Required: Dialysate Device ListTitle = DG POST: Acc Cal -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 817 = [26] -Title = HD accelerometer invalid calibration record +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Acc Cal -Message = If the accel calibration data that was received from NVDataMgmt is NOT legitimate. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [27] -Title = HD BLD sensor failed zero sequence +Title = Self Test: Blood Leak ListTitle = Process: Pre- Tx Wet Self- Test -Message = BLD fails to complete zero sequence. +Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. [28] -Title = DG two wire sensors FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: Conductivity RTD Comm -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 864 = [29] -Title = HD heparin force sensor invalid calibration record +Title = Service Required: Hemodialysis Device ListTitle = HD POST: SP DAC Cal -Message = If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [30] -Title = HD Software fault. Software found itself in an unexpected state +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: SW Invalid State -Message = If the specific software faults listed in alarm management SW faults has been detected +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [31] -Title = Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Current -Message = If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [32] -Title = Blood pump failed motor off check. Measured speed while commanded off +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Speed -Message = If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [33] -Title = Blood pump failed motor direction check. Measured vs. commanded +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Direction -Message = If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [34] -Title = Blood pump failed rotor speed check. Mismatch with rotor and motor speeds +Title = Blood Pump Error ListTitle = HD Fault: BP Rotor -Message = If the delta of the measured motor speed and the commanded motor speed is not within tolerance. +Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [35] -Title = Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Current -Message = If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [36] -Title = Dialysis inlet pump failed motor off check. Measured speed while commanded off +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Speed -Message = If during command off state, the DPi pump motor speed is measured > 100.0 RMP. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [37] -Title = Dialysis inlet pump failed motor direction check. Measured vs. commanded +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Direction -Message = If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [38] -Title = Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds +Title = Dialysate Pump Error ListTitle = HD Fault: DPi Rotor -Message = If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [39] -Title = Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Current -Message = If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [40] -Title = Dialysis outlet pump failed motor off check. Measured speed while commanded off +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Speed -Message = If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [41] -Title = Dialysis outlet pump failed motor direction check. Measured vs. commanded +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Direction -Message = If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [42] -Title = Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds +Title = Dialysate Pump Error ListTitle = HD Fault: DPo Rotor -Message = If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [43] -Title = Main primary heater FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: Iso ADC Comm -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 954 = [44] -Title = User interface communication timeout +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: UI Comm -Message = If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [45] -Title = Too many bad CRCs detected on received system messages +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: CAN CRC Com -Message = If too many bad messages CRCs are received within a set period of time +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [46] -Title = System message that required acknowledgment was not acknowledged +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: CAN ACK UI Com -Message = If no retries left in the pending acknowlegement list. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [47] Title = -ListTitle = DG Fault: RO Pump High +ListTitle = Message = [48] -Title = Ultrafiltration volume accuracy error during treatment +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: UF Rate Tare Err -Message = If the UF volume (refUFvolume - measureVolume) > 100.0 mL. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [49] -Title = HD FPGA communication down for too long +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: FPGA Com -Message = If there are more than 3 comm errors in a 1 min period. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [50] -Title = DG FPGA not accepting commanded valve states +Title = Service Required: Dialysate Device ListTitle = DG Fault: Valve Control -Message = If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 26 = [51] -Title = Blood pump failed motor speed check. Measured vs. commanded +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Speed -Message = If measured motor speed versus commanded rotor speed exceeded limits. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [52] -Title = Dialysate inlet pump failed motor speed check. Measured vs. commanded +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPi Speed -Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [53] -Title = Dialysate outlet pump failed motor speed check. Measured vs. commanded +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: DPo Speed -Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [54] -Title = HD critical data integrity check failed +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Critical Data Err -Message = If the critical data failed range check. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [55] -Title = DG critical data integrity check failed +Title = Service Required: Dialysate Device ListTitle = DG Fault: Critical Data Err -Message = If the critical data failed range check. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 921 = [56] -Title = HD accelerometer error (no readings or FPGA reports error) +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Acc Com -Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [57] -Title = DG accelerometer error (no readings or FPGA reports error) +Title = Service Required: Dialysate Device ListTitle = DG Fault: Acc Com -Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 869 SRSDG 870 = [58] -Title = HD valve homing failed +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Home -Message = If the number of failed homing > = 3 counts. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [59] -Title = HD valve transition time out +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Transition Err -Message = if the valve's transition time has timed out (3000msec) +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [60] -Title = HD valve not functional +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Control -Message = If the valve bit masks are not as expected in different valve control mode. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [61] -Title = HD valve current out of range +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Current -Message = If the valve current is over the threshold of 1.0 A for 1000 msec. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [62] -Title = HD valve position out of target +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Valve Position -Message = If the current position has deviated from the expected position for more than 1 second. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [63] Title = ListTitle = Message = [64] -Title = Barometric pressure is out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Baro Range -Message = If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 874 = [65] -Title = HD requests DG command with invalid parameter fault +Title = Communication Error ListTitle = Loss of DG Comm (Reject) -Message = If the latest DG command response indicated invalid parameter +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [66] -Title = HD blood leak sensor set point set failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: BLD Config Set Point -Message = If the set point of the blood leak sensor does not match the set point in the calibration record. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [67] -Title = HD blood pump occlusion self-test failure alarm +Title = Self Test: Cartridge ListTitle = Process: Pre- Tx Remove Cartridge -Message = If the blood pump pressure > 5000 +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [68] -Title = HD active reservoir recirculation out of range +Title = Dialysate Generation Error ListTitle = Dial. Gen: Dialysate not ready -Message = If the recirculation level has exceeded the max allowed +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. [69] -Title = HD blood leak sensor invalid calibration record +Title = Service Required: Hemodialysis Device ListTitle = HD POST: BLD Cal -Message = If the BLD data that was received from NVDataMgmt is NOT legitimate. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [70] -Title = HD arterial pressure self-test failure alarm +Title = Self Test: Cartridge ListTitle = Process: Pre- Tx Remove Cartridge -Message = If arterial pressure < -300 or > 100 mmHg +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [71] -Title = HD venous pressure self-test failure alarm +Title = Self Test: Cartridge ListTitle = Process: Pre- Tx Remove Cartridge -Message = If the venous pressure < -100 or > 600 mmHg +Message = A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [72] -Title = HD No load cell data message receive at least once every 2 seconds +Title = Communication Error ListTitle = Loss of DG Comm (LC) -Message = No new load cell data for 2 seconds. +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [73] -Title = HD No dialysate temperature data message received at least once every 2 seconds +Title = Communication Error ListTitle = Loss of DG Comm (TDI/TRO) -Message = No new dialysate temperature data for 2 seconds. +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [74] -Title = DG inlet UV reactor is on while there is no flow +Title = Service Required: Dialysate Device ListTitle = DG Fault: UVi On No Flow -Message = If the inlet UV reactor is on while VPi is closed for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 600 = [75] -Title = HD syringe pump self-test failure alarm +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe DAC -Message = If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [76] -Title = HD monitored voltage is out of range +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Voltage Monitor -Message = If the HD monitored voltages are out of range for a certain period of time. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [77] -Title = DG monitored voltage is out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Voltage Monitor -Message = If the DG monitored voltages are out of range for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] -Title = HD syringe pump direction (from encoder) error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Enc Dir -Message = If the measured encoder direction of the syringe pump does not agree with the expected direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [79] -Title = HD syringe pump direction (from controller) error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Drv Dir -Message = If the measured direction of the syringe pump motor controller does not agree with the expected direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [80] -Title = HD syringe pump fault reported by FPGA +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Driver -Message = If the FPGA reported the syringe pump fault +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [81] -Title = HD syringe pump over travel error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump OT -Message = If the syringe pump position is deemed to have overtraveled +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [82] -Title = HD syringe pump DAC write failure +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump DAC Err -Message = If the syringe pump DAC write error is reported by the FPGA +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [83] -Title = HD syringe pump is running while the BP is off +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump On Err -Message = If the syringe pump is on while the blood pump is off +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [84] -Title = DG RTC set year is invalid +Title = Service Required: Dialysate Device ListTitle = DG POST: DG RTC Time -Message = If DG RTC set year is < 2021. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 815 = [85] -Title = HD pump track lever is not fully latched +Title = Cartridge Lever is Open ListTitle = Blood Stop: Cartridge Lever Open -Message = If the pump track lever is detected as unlatched when it should be. +Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. [86] -Title = HD RTC set year is invalid +Title = Service Required: Hemodialysis Device ListTitle = HD POST: HD RTC Time -Message = If HD RTC set year is < 2021. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [87] -Title = DG heating parameters invalid calibration record +Title = Service Required: Dialysate Device ListTitle = DG POST: Heat Cal -Message = If the heating paramters' non-volatile record's CRC does not match the calculated CRC. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 929 = [88] -Title = DG concentrate pump hall sensor out of range. +Title = Service Required: Dialysate Device ListTitle = DG Fault: CP1 CP2 Speed Range -Message = Hall sensor pulse to pulse timing is < 1ms (speed too fast). +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [89] -Title = DG outlet UV reactor is on while there is no flow +Title = Service Required: Dialysate Device ListTitle = DG Fault: UVo On No Flow -Message = If the outlet UV reactor is on while the RO flow is less than 200 mL/min. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 599 = [90] -Title = DG load cells A1/B1 FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: LC A1/B1 Comm -Message = If the FPGA does not report fresh data for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 604 = [91] -Title = HD No dialysate flow data received in the last 3 seconds +Title = Communication Error ListTitle = Loss of DG Comm (FMD) -Message = If not receiving new dialysate flow readings in expected interval (3 seconds) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [92] Title = ListTitle = Message = [93] -Title = DG outlet primary conductivity sensor out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Permeate Cond High -Message = If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 430 = [94] -Title = DG pressure sensor out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Pressure Sensor Range -Message = If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 629 = [95] -Title = DG watchdog expired +Title = Service Required: Dialysate Device ListTitle = DG Fault: WD Error -Message = If the watchdog signal is low while not in POST. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 907 = [96] -Title = DG inlet water temperature out of range +Title = Incoming Water Temperature High ListTitle = Dial. Gen: Inlet Water Temp High -Message = If the inlet water temperature is > 37 C. +Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 410 = [97] -Title = DG fill conductivity out of range +Title = Conductivity Out of Range ListTitle = Dial. Gen: Dialysate Cond. -Message = If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. +Message = Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 397 = [98] -Title = HD battery or battery charger communication fault +Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): Battery Com -Message = If battery I2C communication fault persistence count > 5. +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [99] -Title = HD syringe pump stall alarm +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Stall -Message = If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [100] Title = ListTitle = Message = [101] -Title = UI POST FW Communication failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI- Proc Comm -Message = If the HD is not communicating with the UI during POST. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [102] -Title = DG heat disinfection temperature gradient out of range +Title = Disinfection Interrupted ListTitle = Cleaning: Tpo THD Temp Delta -Message = If temperature gradient between Tpo and Thd exceeds maximum 15°C. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 856 SRSDG 930 = [103] -Title = HD non-volatile calibration group invalid record CRC +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Cal NVRAM CRC -Message = If the computed CRC of the calibration group record is not equal to the record CRC. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [104] -Title = Air trap level sensors reporting illegal combination of air/fluid +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Air Trap Sensor -Message = If the air trap level readings are not valid for more than 2 seconds. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [105] -Title = DG non-volatile system group invalid record CRC +Title = Service Required: Dialysate Device ListTitle = DG POST: Cal NVRAM CRC -Message = If the DG NVDATAMGMT CAL group calculated CRC does not matched the expected CRC. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 608 = [106] -Title = DG flow sensors calibration error +Title = Service Required: Dialysate Device ListTitle = DG POST: Flow Sensor Cal -Message = If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] -Title = HD reports DG restarted fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: DG Restarted -Message = If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [108] -Title = HD syringe pump ADC error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump ADC Com -Message = If the ADC read is not fresh after POST init and the ADC read counter has not changed. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [109] -Title = HD syringe pump volume check error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Vol -Message = If the volume delivered is off by more than 5% or 0.1mL. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [110] -Title = HD syringe pump speed check error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Speed -Message = If the syringe measured rate is off by more than 5% or 0.1mL. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [111] -Title = HD syringe pump not stopped in off state error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Syringe Pump Speed -Message = If the syringe pump position is not equal to its last position. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [112] -Title = DG load cells A2/B2 FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: LC A2/B2 Comm -Message = If the FPGA does not report fresh data for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 953 = [113] -Title = DG drain pump current out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Current -Message = If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 890 = [114] -Title = HD venous air bubble detector self-test failure +Title = Self Test: Venous Air Detector ListTitle = Process: Pre- Tx Wet Self- Test -Message = If the venous air bubble self test did not detect bubble for more than 500 msec. +Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. [115] -Title = DG temperature sensor out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Temp Sensor Range -Message = If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 224 = [116] -Title = SD card is non-operational +Title = Service Required: Dialysate Device ListTitle = HD Alarm (Low): SD- Card Error -Message = SD card is non-operational +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [117] -Title = System message that required acknowledgment was not acknowledged +Title = Communication Error ListTitle = Loss of HD Comm (HD ACK) -Message = If no retries left in the pending acknowlegement list. +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 909 = [118] -Title = DG RTC configuration error +Title = Service Required: Dialysate Device ListTitle = DG Fault: RTC Config -Message = If the RTC configuration registers report an error. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 43 = [119] -Title = DG RTC battery low +Title = Service Required: Dialysate Device ListTitle = DG Alarm (Low): RTC Bat Low -Message = If the RTC battery bit report a low alarm. +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. SRSDG 908 = [120] -Title = HD pre-treatment wet prime test failure +Title = Self Test: Venous Air Detector ListTitle = Process: Pre- Tx Wet Self- Test -Message = If the HD pre-treatment wet self test failed. +Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [121] -Title = DG main primary heater voltage out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Prim Heater Lg Control -Message = If the main primary heater's voltage is out of range. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 611 = [122] -Title = DG small primary heater voltage out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Prim Heater Sm Control -Message = If the small primary heater's voltage is out of range. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 612 = [123] -Title = DG trimmer heater voltage out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Trimmer Heater Control -Message = If the trimmer voltage is out of range. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 613 = [124] -Title = HD end of treatment alarm (high priority - user still not acting to end treatment) +Title = Blood Pump Stop Time Exceeded ListTitle = Blood Stop: Blood Clot Risk -Message = In treatment end paused state for > 5 minutes. +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [125] -Title = Blood sitting too long after treatment stopped by user alarm (> 5 min) +Title = Blood Pump Stop Time Exceeded ListTitle = Blood Stop: Blood Clot Risk -Message = If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [126] -Title = HD blood leak detected alarm +Title = Blood Leak ListTitle = Blood Stop: Dialyzer Blood Leak -Message = If the blood leak is detected based on the blood leak status reported by the sensor. +Message = A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. [127] -Title = Venous pressure too low during treatment +Title = Low Venous Pressure ListTitle = Blood Stop: Venous Pres. Low -Message = If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. +Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and\nterminate treatment without rinseback.\n- If access connections are properly secured, \npress Resume/OK to continue treatment. [128] -Title = HD venous air bubble detected alarm +Title = Venous Air Detected ListTitle = Blood Stop: Venous Air Bubble -Message = If bubbles are detected during the treatment. +Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. [129] -Title = HD blood leak detected and it is in the process of attempting to remove the bubbles/foam +Title = Blood Leak: Bypass ListTitle = Blood Stop: Dialyzer Blood Leak Bypass -Message = If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. +Message = The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. [130] -Title = Venous pressure too high during treatment +Title = High Venous Pressure ListTitle = Blood Stop: Venous Pres. High -Message = If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [131] -Title = Arterial pressure too low during treatment +Title = Low Arterial Pressure ListTitle = Blood Stop: Arterial Pres. Low -Message = If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. +Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [132] -Title = Arterial pressure too high during treatment +Title = High Arterial Pressure ListTitle = Blood Stop: Arterial Pres. High -Message = If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. +Message = Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback. [133] -Title = DG fluid leak detected alarm +Title = Dialysate Device Fluid Leak ListTitle = DG Fault: Fluid Leak in DG -Message = If the fluid leak is detected as reported by the fluid leak detector. +Message = Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 328 = [134] -Title = HD fluid leak detected alarm +Title = Fluid Leak ListTitle = Blood Stop: Fluid Leak in HD -Message = If the fluid leak is detected as reported by the fluid leak detector. +Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. [135] -Title = HD experienced large acceleration +Title = Blood Pump Stopped: Excessive Movement ListTitle = Blood Stop: HD Shock Detected -Message = If the system experienced excessive shock > 2.5 g. +Message = Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. [136] -Title = DG experienced large acceleration +Title = Dialysate Device: Excessive Movement ListTitle = Dial. Gen: Shock- Vibration -Message = If the system experienced excessive shock > 2.5 g. +Message = Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. SRSDG 327 = [137] -Title = HD tilt exceeds maximum +Title = Blood Pump Stopped: Excessive Tilt ListTitle = Blood Stop: HD Tilted -Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. +Message = Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. [138] -Title = DG tilt exceeds maximum +Title = Dialysate Device: Excessive Tilt ListTitle = Dial. Gen: Tilt -Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. +Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. SRSDG 326 = [139] -Title = HD AC power lost alarm +Title = Power Loss ListTitle = Loss of AC Power -Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. +Message = AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. [140] -Title = Dialysate generator communication timeout +Title = Communication Error ListTitle = Loss of DG Comm (Timeout) -Message = If DG communication timeout is detected for more than 2 seconds. +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [141] -Title = Air trap fill timeout during treatment +Title = Blood Pump Stopped: Venous Air Detector ListTitle = Blood Stop: Air Trap Fill -Message = If the air trap fill timeout during treatment. +Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. [142] -Title = Blood pump occlusion detected +Title = Blood Pump Stopped: Occlusion ListTitle = Blood Stop: Blood Pump Occlusion -Message = If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall +Message = Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [143] -Title = DG dialysate temperature sensors drift timeout. +Title = Service Required: Dialysate Device ListTitle = DG Fault: TRo TDi Delta -Message = If TDi and TRo have drifted for more 1 C for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 912 = [144] -Title = DG cleaning mode inlet water conductivity out of high range +Title = Disinfection: Water Conductivity High ListTitle = Cleaning: Inlet Water Cond High -Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode +Message = Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = DG Concentrate conductivity after adding acid out of range alarm +Title = Conductivity Error During Prime ListTitle = Dial. Gen: Acid Conc. Cond. -Message = If the conductivity value for the acid concentrate is out of range for more than 45 secs. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 846 = [146] -Title = DG RTC vs. Timer accuracy error +Title = Service Required: Dialysate Device ListTitle = DG POST: DG RTC Timer -Message = If RTC and timer are not aligned on 1 second elapsed time. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 312 = [147] -Title = DG dialysate generation conductivity fault alarm +Title = Dialysate Generation Error ListTitle = Dial. Gen: Dialysate not ready -Message = If dialysate conductivity is not with 5% of the TCD +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. SRSDG 398 = [148] -Title = Cleaning Mode Inlet water temperature too high +Title = Disinfection: Water Temperature High ListTitle = Cleaning: Inlet Water Temp High -Message = Inlet temperature > 37 C. +Message = Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] -Title = HD watchdog expired +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: WD Error -Message = If the watchdog signal is low while not in POST. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [150] -Title = Inlet water conductivity too high +Title = Incoming Water Conductivity High ListTitle = Dial. Gen: Inlet Water Cond High -Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). +Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 391 = [151] -Title = Inlet water conductivity too low +Title = Incoming Water Conductivity Low ListTitle = Dial. Gen: Inlet Water Cond Low -Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). +Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 826 = [152] -Title = Inlet water pressure too low +Title = Incoming Water Pressure Low ListTitle = Dial. Gen: Inlet Water Pres Low -Message = Inlet water pressure < 25 PSI. +Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 417 = [153] -Title = HD prime completed high priority alarm +Title = Pre-Treatment Process Terminated ListTitle = Process: Pre- Tx Terminated -Message = In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) +Message = The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. [154] -Title = Cleaning mode inlet water conductivity too low +Title = Disinfection: Water Conductivity Low ListTitle = Cleaning: Inlet Water Cond Low -Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode +Message = Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = HD No reservoir data message receive at least once every 2 seconds +Title = Communication Error ListTitle = Loss of DG Comm (Reservoir) -Message = No new reservoir data for 2 seconds. +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [156] -Title = HD No DG op mode message receive at least once every 2 seconds +Title = Communication Error ListTitle = Loss of DG Comm (Op Mode) -Message = No DG op mode for 2 seconds. +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [157] -Title = Chemical disinfect priming the acid line timeout +Title = Chemical Disinfectant Not Detected ListTitle = Cleaning: Chem Disinf PAA Prime -Message = If the disinfectant acid line is not primed within a certain period of time. +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 988 = [158] -Title = Inlet water temperature low +Title = Incoming Water Temperature Low ListTitle = Dial. Gen: Inlet Water Temp Low -Message = Inlet temperature < 24 °C for 5 seconds. +Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 408 = [159] -Title = Close concentrate line be able to flush it +Title = Remove Chemical Disinfectant ListTitle = Cleaning: Chem Disinf Connect Cap -Message = Start of chemical disinfect flush. +Message = The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. SRSDG 1044 = [160] -Title = Blood pump rotor speed too high +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Rotor Max Speed -Message = If blood pump motor speed exceeded 100.0 RPM. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [161] -Title = Too many bad CRCs detected on received system messages +Title = Service Required: Dialysate Device ListTitle = DG Fault: HD- DG Comm -Message = If >= 5 bad CRCs detected on received CAN messages within 10 minutes. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 122 = [162] -Title = FPGA clock speed failure +Title = Service Required: Dialysate Device ListTitle = DG Fault: Proc- FPGA Clock -Message = If the FPGA clock speed is relatively close to the processor clock for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 867 = [163] -Title = HD sees primary load cell for reservoir 1 change too much too fast +Title = Ultrafiltration Error ListTitle = Dial. Gen: UF Rate Check -Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. [164] -Title = HD sees primary load cell for reservoir 2 change too much too fast +Title = Ultrafiltration Error ListTitle = Dial. Gen: UF Rate Check -Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. [165] -Title = HD in treatment rinseback operation timeout +Title = Blood Pump Stop Time Exceeded ListTitle = Blood Stop: Blood Clot Risk -Message = If the maximum time inrinseback operation has been exceeded. +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [166] -Title = +Title = Test Alarm - Alarm 166 ListTitle = Message = [167] -Title = HD cartridge door opened alarm +Title = Blood Pump Stopped: Door Open ListTitle = Blood Stop: Door Open -Message = If the cartridge door is opened as reported by the FPGA. +Message = The device door is open.\n Close the device door. \n- Press Resume/OK to continue. [168] -Title = +Title = Test Alarm - Alarm 168 ListTitle = Message = [169] -Title = Dialysate flow rate out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Dialysate Flow Sens Rg -Message = If the dialysate flow rate > 2.00 L/min. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 643 = [170] -Title = HD syringe empty alarm +Title = Heparin Syringe Empty ListTitle = Heparin Syringe Empty -Message = If the syringe pump position is near empty position. +Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. [171] -Title = HD syringe pump occlusion alarm +Title = Heparin Pump Occlusion ListTitle = Process: Pre- Tx Hep. Pump Occ. -Message = If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. +Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. [172] -Title = HD syringe pump not enough Heparin alarm +Title = Not Enough Heparin ListTitle = Process: Pre- Tx Not Enough Hep. -Message = If the estimated syringe volume needed for treatment is not sufficient +Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. [173] -Title = HD RTC configuration error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: RTC Config -Message = If the RTC configuration registers report an error. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [174] -Title = HD RTC vs. Timer accuracy error +Title = Service Required: Hemodialysis Device ListTitle = HD POST: HD RTC Timer -Message = If RTC and timer are not aligned on 1 second elapsed time. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [175] -Title = HD pump commutation error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Pump Commutation Err -Message = If the set pump direction does not agree with the hall sensor direction. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [176] -Title = +Title = Test Alarm - Alarm 176 ListTitle = Message = [177] -Title = DG software fault. Software found itself in an unexpected state +Title = Service Required: Dialysate Device ListTitle = DG Fault: Software Invalid State -Message = If the specific software faults listed in alarm management SW faults has been detected +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 919 = [178] -Title = HD communication timeout +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: UI- Proc Comm -Message = If the HD stops communicating with the UI. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1066 [179] -Title = DG FPGA communication down for too long +Title = Service Required: Dialysate Device ListTitle = DG Fault: FPGA Com -Message = If retries for commands exceeds limit or FPGA reports comm error. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 605 = [180] -Title = RO flow rate out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Flow Sensor Range -Message = If the RO flow rate > 2.00 L/min. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 883 = [181] -Title = DG load cells weight out of range for tare +Title = Service Required: Dialysate Device ListTitle = DG Fault: Tare Out of Range -Message = If the load cell weight is out of range for tare. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 849 SRSDG 838 = [182] -Title = DG load cells invalid calibration +Title = Service Required: Dialysate Device ListTitle = DG POST: LC Cal -Message = If the CRC of load cells calibration data does not match the calculated CRC. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 162 = [183] -Title = DG invalid load cell value +Title = Disinfection Interrupted ListTitle = Cleaning: Invalid Resv Transfer -Message = If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 851 = [184] -Title = Inlet reactor not healthy +Title = Service Required: Dialysate Device ListTitle = DG Fault: UVi Error -Message = If the inlet reactor are not healthy. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 559 = [185] -Title = DG fan RPM out of range +Title = Service Required: Dialysate Device ListTitle = DG Alarm (Low): Fan Speed -Message = If the fan RPM is out of Min/Max range. +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. SRSDG 628 = [186] -Title = DG concentrate pump fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: CP1/CP2 Fault -Message = If the concentrate pumps fault is reported by the FPGA. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 891 = [187] -Title = DG concentrate pump CP1 speed control error +Title = Service Required: Dialysate Device ListTitle = DG Fault: CP1 Speed Fault -Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 905 = [188] -Title = DG concentrate pump CP2 speed control error +Title = Service Required: Dialysate Device ListTitle = DG Fault: CP2 Speed Fault -Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 906 = [189] -Title = DG drain pump RPM out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Speed Fault -Message = If the drain pump RPM > 100.0 of the target RPM. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 448 = [190] -Title = DG drain pump off fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Speed Not Off -Message = If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [191] -Title = DG flow rate out of upper range +Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Flow High -Message = If the RO flow rate > 10% from the target flow rate. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 885 = [192] -Title = DG flow rate out of lower range +Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Flow Low -Message = If the RO flow rate < 10% from the target flow rate. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 945 = [193] @@ -1058,116 +1058,116 @@ Message = [194] -Title = DG RO pump duty cycle out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Driver Feedback -Message = If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 881 = [195] -Title = DG RO pump pressure out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Pump High Pressure -Message = If the DG RO pump pressure > 135 psi for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 925 = [196] -Title = DG CPi/CPo conductivity sensors FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: CPi/CPo Comm -Message = If the FPGA reports error for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 955 = [197] -Title = DG CD1/CD2 conductivity sensors FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: CD1/CD2 Comm -Message = If the FPGA reports error for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 956 = [198] -Title = DG primary heater while RO flow is low +Title = Service Required: Dialysate Device ListTitle = DG Fault: HP On With No Flow -Message = If the RO flow is < 0.2 L/min for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 872 = [199] -Title = DG trimmer heater while dialysate flow is low +Title = Service Required: Dialysate Device ListTitle = DG Fault: HR On With No Flow -Message = If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 871 = [200] -Title = DG thermistor or temperature sensor out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Thermistor Range -Message = If thermistor > 80.0 C or < 0.0 C. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 178 = [201] -Title = HD pre-treatment wet flow test failure +Title = Self Test: Dialysate Flow Rate ListTitle = Process: Pre- Tx Wet Self- Test -Message = If the pre-treatment wet flow test has failed. +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [202] -Title = HD pre-treatment dry pressure normal test failure +Title = Self Test: Pressure ListTitle = Process: Pre- Tx Dry Self- Test -Message = If the pre-treatment dry pressure normal test failed. +Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [203] Title = ListTitle = Message = [204] -Title = RO rejection ratio out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: RO Rejection -Message = If RO rejection ratio > 0.1 +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 429 = [205] -Title = Conductivity sensor fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: Cond Sensors Error -Message = If conductivity sensor reports 0x11 for more than 2 consecutive reads. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 913 = [206] -Title = DG dialysate fill runs out of time +Title = Service Required: Dialysate Device ListTitle = DG Fault: Reservoir Fill Timeout -Message = If DG dialysate fill > 5 minutes. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 437 = [207] -Title = DG flow meter check failure alarm +Title = Service Required: Dialysate Device ListTitle = DG Fault: FMP Flow Rate Error -Message = If integrated Volume To Load Cell Reading Percent > 0.1. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 240 = [208] -Title = HD venous line occlusion fault +Title = High Venous Pressure ListTitle = Blood Stop: Venous Pres. High -Message = If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [209] -Title = DG drain circulation line timeout +Title = Disinfection Interrupted ListTitle = Cleaning: Flush Circ Error -Message = If the integrated volume < 0.1 for more than 60 secs. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 980 = [210] -Title = HD battery pack detected an error +Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): Battery Fault -Message = If the battery pack status error is detected from the slave device I2C address 0x0B. +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [211] -Title = Blood sitting too long warning (> 4 min) +Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood Stopped: Warning -Message = During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. [212] -Title = Blood sitting too long warning (> 4 min) +Title = Approaching MAX Blood Pump Stop Time ListTitle = Blood Stopped: Warning -Message = During treatment end paused state, if the blood is in the line for more than 4 minutes. +Message = The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. [213] -Title = HD prime completed medium priority alarm +Title = Prime Complete - 30 MIN WARNING ListTitle = Process: Pre- Tx Prime Complete -Message = In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) +Message = The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. [214] Title = @@ -1180,140 +1180,140 @@ Message = [216] -Title = HD end treatment sub-mode timeout alarm +Title = Treatment Complete ListTitle = Process: End of Dialysis Tx -Message = If the treatment ended successfully. +Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. [217] Title = ListTitle = Message = [218] -Title = HD syringe detected alarm +Title = Syringe Detected ListTitle = Process: Remove Syringe -Message = If the syringe detection switch > 2.0 volts. +Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. [219] -Title = HD syringe pump syringe removed alarm +Title = Blood Stopped: Syringe Not Detected ListTitle = Blood Stop: Syringe Not Detected -Message = If the syringe detection switch < 2.0 volts. +Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. [220] -Title = DG THd sensor FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: THD RTD Comm -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 947 = [221] -Title = HD Saline bag is empty +Title = Blood Pump Stopped: Saline Bag Empty ListTitle = Blood Stop: Saline Bag Empty -Message = If the measured arterial pressure < -300.0 mmHg for more than 250 msec. +Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. [222] -Title = HD occlusion sensor FPGA fault +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: BP Occ Com -Message = If the FPGA does not report fresh data or indicates an error for a certain period of time. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [223] -Title = HD arterial sensor FPGA fault +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Art. Pres. Com -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [224] -Title = Treatment stopped by user action - pressed stop button +Title = Blood Pump Stopped: Treatment Paused ListTitle = Blood Stopped: Treatment Pause -Message = If the user has pressed the stop button. +Message = Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. [225] -Title = HD end of treatment warning +Title = Treatment Complete ListTitle = Process: End of Treatment -Message = If the treatment duration has exceeded the prescribed treatment time entered by the user. +Message = Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. [226] -Title = HD prime completed low priority alarm +Title = Priming Complete ListTitle = Process: Pre- Tx Priming Complete -Message = Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection +Message = Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. [227] Title = ListTitle = Message = [228] -Title = HD time out on prime sub-mode purge air state. +Title = Priming Error ListTitle = Process: Pre- Tx Prime Timeout -Message = During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [229] -Title = HD prime dialysate dialyzer time out alarm +Title = Priming Error ListTitle = Process: Pre- Tx Prime Timeout -Message = If the priming of the dialyzer dialysate flow path has exceeded the time limit. +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [230] -Title = HD prime dialysate bypass time out alarm +Title = Priming Error ListTitle = Process: Pre- Tx Prime Timeout -Message = If the priming of the dialyzer bypass flow path has exceeded the time limit. +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [231] -Title = HD pre-treatment mode dry self-tests failure +Title = Self Test: Cartridge Check ListTitle = Process: Pre- Tx Dry Self- Test -Message = If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState +Message = A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [232] -Title = HD pre-treatment mode wet self-tests failure +Title = Self Test: Dialysate Flow Rate ListTitle = Process: Pre- Tx Wet Self- Test -Message = If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [233] -Title = DG barometric pressure sensor bad coefficients CRC +Title = Service Required: Dialysate Device ListTitle = DG Fault: Baro Cal CRC -Message = If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = DG inlet water pressure is too high +Title = Incoming Water Pressure High ListTitle = Dial. Gen: Inlet Water Pres High -Message = If the inlet water pressure is > 80 psig. +Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 1037 = [235] -Title = HD in treatment stopped sub-mode after rinseback completed (no escalation) +Title = Rinseback Complete ListTitle = Process: Rinseback Complete -Message = If there is no active alarm and rinseback has completed. +Message = Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. [236] -Title = HD needs new cartridge to be installed +Title = Self Test: Fluid Present ListTitle = Process: Pre- Tx Dry Self- Test -Message = If the air trap sensor reported AIR and bubble sensor detected bubbles. +Message = A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. [237] Title = ListTitle = Message = [238] -Title = HD no cartridge loaded or installed improperly alarm +Title = Cartridge Not Detected ListTitle = Process: Pre- Tx Dry Self- Test -Message = If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. +Message = A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. [239] -Title = HD fail to remove cartridge alarm +Title = Cartridge Still Detected ListTitle = Process: Post- Tx Remove Cartridge -Message = In the post treatment state, if the cartridge has not been removed. +Message = A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. [240] -Title = Bicarb conductivity out of range during bicarb pump check alarm +Title = Conductivity Error During Prime ListTitle = Dial. Gen: Bicarb Conc. Cond. -Message = If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 482 = [241] -Title = DG reservoir drain time out +Title = Disinfection Interrupted ListTitle = Cleaning: Resv Drain Timeout -Message = If the DG reservoir drain is not complete in 120 secs. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 848 SRSDG 90 SRSDG 561 @@ -1344,9 +1344,9 @@ SRSDG 1034 = [242] -Title = DG reservoir fill time out +Title = Disinfection Interrupted ListTitle = Cleaning: Resv Fill Timeout -Message = If the DG reservoir fill is not complete in 120 secs. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 850 SRSDG 573 SRSDG 574 @@ -1367,81 +1367,81 @@ SRSDG 1024 = [243] -Title = DG reservoir leak time out +Title = Disinfection Interrupted ListTitle = Cleaning: Fluid Loss Detected -Message = If the reservoir level is drifted by 100 mL for more than 30 seconds. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 853 = [244] -Title = DG temperature sensors difference out of range +Title = Disinfection Interrupted ListTitle = Cleaning: Temp Sensor Check -Message = If any of the temperature sensors are out of the average of the sensors by more than 3 C. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 852 = [245] -Title = DG heat disinfect target temperature time out (could not reach to temperature) +Title = Disinfection Interrupted ListTitle = Cleaning: Heat Disinfect Temp TO -Message = If heating up to minimum temperature for heat disinfect failed after 4 hours. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 854 = [246] -Title = DG chemical disinfect inlet conducitivty out of range +Title = Disinfection Interrupted ListTitle = Cleaning: RO Water Cond Error -Message = If CPo and CD2 conductivity values are above 100 uS/cm. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. [247] Title = ListTitle = Message = [248] -Title = DG chemical disinfect out of target temperature +Title = Disinfection Interrupted ListTitle = Cleaning: Chem Disinfect Proc -Message = If TPo > 50 C after 1 second. +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. [249] -Title = DG chemical disinfect out of target conductivity +Title = Chemical Disinfectant Not Detected ListTitle = Cleaning: Chem Disinf PAA Cond -Message = If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 992 = [250] -Title = Insert the acid to start chemical disinfect +Title = Connect Chemical Disinfectant ListTitle = Cleaning: Chem Disinf Connect PAA -Message = Start of chemical disinfect +Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. SRSDG 1046 = [251] -Title = HD invalid system record +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Sys Rec NVRAM CRC -Message = If the calculated CRC of the HD system record image does not match the expected CRC. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [252] -Title = HD invalid service record +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Serv Rec NVRAM CRC -Message = If the calculated CRC of the HD service record image does not match the expected CRC. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [253] -Title = DG invalid system record +Title = Service Required: Dialysate Device ListTitle = DG POST: Sys Rec NVRAM CRC -Message = If the calculated CRC of the DG system record image does not match the expected CRC. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 609 = [254] -Title = DG invalid service record +Title = Service Required: Dialysate Device ListTitle = DG POST: Serv Rec NVRAM CRC -Message = If the calculated CRC of the DG service record image does not match the expected CRC. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 610 = [255] -Title = HD and UI software builds are not compatible +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Compatibility -Message = If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [256] -Title = DG flush started +Title = Chemical Flush in Progress ListTitle = Cleaning: Chem Flush In Progress -Message = If DG flush successfully started from the UI. +Message = Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. = SRSUI 497 [257] @@ -1455,84 +1455,84 @@ Message = [259] -Title = HD temperatures out of range +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Temp Sensors Error -Message = If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [260] -Title = UI POST Application Integrity (Sha256Sum) failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Image -Message = If the UI POST reports a shasum faillure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1067 [261] -Title = UI POST CANBus failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI CAN Driver -Message = If the UI POST reports a can bus failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1068 [262] -Title = UI POST Display failure +Title = ListTitle = -Message = If the UI POST reports a display failure. +Message = [263] -Title = UI POST Touch failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Touch Driver -Message = If the UI POST reports a touch failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1070 [264] -Title = UI POST SD-Card failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI SD- Card -Message = If the UI POST reports SD card failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1071 [265] -Title = UI POST RTC failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI RTC -Message = If the UI POST reports RTC failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1072 [266] -Title = UI POST WiFi failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI WiFi -Message = If the UI POST reports WiFi failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1073 [267] -Title = UI POST Bluetooth failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Bluetooth -Message = If the UI POST reports bluetooth failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1074 [268] -Title = UI POST Ethernet failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Ethernet -Message = If the UI POST reports ethernet failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1075 [269] -Title = UI POST Sound failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI Sound -Message = If the UI POST reports sound failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1076 [270] -Title = HD POST Safety Shutdown failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Proc Safestate -Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [271] -Title = DG POST Safety Shutdown failure +Title = Service Required: Dialysate Device ListTitle = DG POST: Proc Safestate -Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 822 = [272] -Title = HD Fan RPM out of range +Title = Service Required: Hemodialysis Device ListTitle = HD Alarm (Low): Fan Speed -Message = If the fan RPM is out of Min/Max range. +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [273] Title = @@ -1545,276 +1545,276 @@ Message = [275] -Title = HD arterial pressure sensor is reading out of range +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Arterial Press Range -Message = If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [276] -Title = HD venous pressure sensor is reading out of range +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Venous Press Range -Message = If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [277] -Title = HD BP occlusion sensor is reading out of range +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Occlusion Sens Range -Message = If the blood pump occlusion count is < 2000. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [278] -Title = HD active reservoir weight out of range. +Title = Service Required: Dialysate Device ListTitle = DG Fault: Reservoir Overfill -Message = If the active reservoir weight is > 1900.0 g. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [279] -Title = DG dialysate drain time out +Title = Service Required: Dialysate Device ListTitle = DG Fault: Reservoir Drain -Message = If the DG dialysate drain exceeded 120 seconds. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 752 = [280] -Title = DG drain pump invalid direction +Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Direction Fault -Message = If the drain pump's direction is not 1 when draining. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 888 = [281] -Title = DG Acid concentration bottle low volume alarm +Title = Acid Concentrate Low ListTitle = Dial. Gen: Acid Conc. Low -Message = If the acid concentration bottle volume > 3240 (3340 - 100). +Message = The acid concentrate is low.\n- Replace acid concentrate container. SRSDG 836 = [282] -Title = DG Bicarbonate concentration bottle low volume alarm +Title = Bicarb Concentrate Low ListTitle = Dial. Gen: Bicarb Conc. Low -Message = If the bicarbonate concentration bottle volume > 3680 (3780 - 100). +Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. SRSDG 837 = [283] -Title = DG load cells weight out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: LC Range -Message = If the weight of the DG load cells > 4500.0 g or < 0.0 g. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 615 = [284] -Title = DG load cells primary/back up drift out of range +Title = Service Required: Dialysate Device ListTitle = DG Fault: Redundant LC Check -Message = If redundant load cells deviate by more than 60g after tare for 5 seconds. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 383 = [285] -Title = DG RO permeate sample started +Title = RO Permeate Sample in Progress ListTitle = Cleaning: RO Permeate Sample In Progress -Message = If DG RO permeate sample successfully started from UI. +Message = RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [286] -Title = DG cap in wrong state +Title = Concentrate Connection Cap Removed ListTitle = Cleaning: Conc Cap Removed -Message = If the concentrate cap is not in the right position in the cleaning modes. +Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. [287] -Title = HD processor clock speed checks against FPGA clock failure +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Proc- FPGA Clock -Message = If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [288] -Title = HD load cell inconsistency +Title = Service Required: Dialysate Device ListTitle = DG Fault: LC Drift Check -Message = If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [289] -Title = DG cap in wrong state +Title = Dialysate Connection Cap Removed ListTitle = Cleaning: Dialysate Cap Removed -Message = If the dialysate cap is not in the right position in the cleaning modes. +Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. SRSDG 928 = [290] -Title = HD venous sensor FPGA fault +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Ven. Pres. Com -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [291] -Title = DG flush started +Title = Flush In Progress ListTitle = Cleaning: Flush In Progress -Message = If DG flush successfully started from the UI. +Message = System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [292] -Title = DG heat disinfect started +Title = Heat Disinfection in Progress ListTitle = Cleaning: Heat Disinf In Progress -Message = If DG heat disinfect successfully started from the UI. +Message = Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [293] -Title = DG chem disinfect started +Title = Chemical Disinfection in Progress ListTitle = Cleaning: Chem Disinfect In Progress -Message = If DG flush successfully started from the UI. +Message = Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [294] -Title = DG active cool started +Title = Active Cool in Progress ListTitle = Cleaning: Active Cool In Progress -Message = If DG active cool successfully started. +Message = Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [295] -Title = UI cloud sync POST failure +Title = Service Required: Hemodialysis Device ListTitle = HD POST: UI CloudSync -Message = If the UI POST reports cloud sync failure. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1238 [296] -Title = DG TDi sensor FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: TDi RTD Comm -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 951 = [297] -Title = DG TRo sensor FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: TRo RTD Comm -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 950 = [298] -Title = DG barometric sensor FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: Baro Pres Comm -Message = If the FPGA does not report fresh data for a certain period of time +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 948 = [299] -Title = DG invalid serial number +Title = Service Required: Dialysate Device ListTitle = DG POST: Serial Number Check -Message = If the DG serial number is empty and does not have any characters. +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 952 = [300] -Title = HD invalid serial number +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Serial Number Check -Message = If the HD serial number is empty and does not have any characters. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [301] -Title = Open dialysate cap to be able to collecte the dispensing fluid +Title = Remove Dialysate Cap ListTitle = Cleaning: RO Permeate Sample Connect Cap -Message = RO permeate sample collection state +Message = Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. [302] -Title = DG_CPU RAM error +Title = Service Required: Dialysate Device ListTitle = DG Fault: Processor RAM error -Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 126 = [303] -Title = DG drain pump direction FPGA fault +Title = Service Required: Dialysate Device ListTitle = DG Fault: DRP Hall Commutation -Message = If the FPGA indicates an error for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 958 = [304] -Title = HD invalid usage reocord +Title = Service Required: Hemodialysis Device ListTitle = HD POST: Usage Rec NVRAM CRC -Message = If the CRC of the HD usage record does not match the calculated CRC. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [305] -Title = HD_CPU RAM error +Title = Service Required: Hemodialysis Device ListTitle = HD Fault: Processor RAM error -Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [306] -Title = +Title = Service Required: Dialysate Device ListTitle = DG Fault: CD1 CD2 Check -Message = CD1 and CD2 conductivity measurement during acid check is not within 5% of each other +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [307] -Title = HD AC power lost alarm +Title = Power Loss ListTitle = Blood Stop: Loss of AC Power -Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. +Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. [308] -Title = DG CPi conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device ListTitle = DG Fault: CPi Comm -Message = If the CPi conductivity sensor transmits invalid characters for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 959 = [309] -Title = DG CPo conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device ListTitle = DG Fault: CPo Comm -Message = If the CPo conductivity sensor transmits invalid characters for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 960 = [310] -Title = DG CD1 conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device ListTitle = DG Fault: CD1 Comm -Message = If the CD1 conductivity sensor transmits invalid characters for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 962 = [311] -Title = DG CD2 conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device ListTitle = DG Fault: CD2 Comm -Message = If the CD2 conductivity sensor transmits invalid characters for a certain period of time. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 961 = [312] -Title = The dialysate inlet flow is too low. +Title = Low Dialysate Flow ListTitle = Dial. Gen: Dialysate In Flow -Message = If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. +Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. [313] -Title = DG cleaning mode inlet water temperature too low +Title = Disinfection: Water Temperature Low ListTitle = Cleaning: Inlet Water Temp Low -Message = If inlet water temperature is < 24 C. +Message = Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = DG cleaning mode inlet water pressure too high +Title = Disinfection: Water Pressure High ListTitle = Cleaning: Inlet Water Press High -Message = If inlet water temperature is > 80 psig. +Message = Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = DG cleaning mode inlet water pressure too low +Title = Disinfection: Water Pressure Low ListTitle = Cleaning: Inlet Water Press Low -Message = If inlet water pressure is < 14 psig. +Message = Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water initiating disinfection again. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = Acid concentrate pump did not park successfully +Title = Service Required: Dialysate Device ListTitle = DG Fault: Acid CP Pump Park -Message = If the park took more than 1 second. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 983 = [317] -Title = Bicarb concentrate pump did not park successfully +Title = Service Required: Dialysate Device ListTitle = DG Fault: Bicarb CP Pump Park -Message = If the park took more than 1 second. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 984 = [318] -Title = HD Air Pump operation timeout +Title = Air Trap Control Error ListTitle = HD Fault: Air Pump Timeout -Message = If air pump has been turned on for greater than 10s. +Message = A problem has been detected with the air trap level\ncontrol system.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [319] -Title = Collect sample to check the chemical disinfectant acid +Title = Residual Chemical Sampling ListTitle = Cleaning: Chem Residual Sampling -Message = Chemical disinfect flush sample alarm is prompted. +Message = A water sample is needed to confirm the \nabsence of residual chemical following disininfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. SRSDG 1040 = [320] -Title = The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes +Title = Chemical Residual Timeout ListTitle = Cleaning: Chem Residual Timeout -Message = Chemical disinfect flush sample was not collected form more than 30 minutes. +Message = A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. SRSDG 1043 = [321] -Title = Outlet reactor not healthy +Title = Service Required: Dialysate Device ListTitle = DG Fault: UVo Error -Message = If the outlet reactor are not healthy. +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 1069 = [322] -Title = UI POST Invalid Year +Title = Service Required: Hemodialysis Device ListTitle = HD POST Alarm: UI Invalid Year -Message = If the UI POST reports a invalid year. +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [323] Title = Service Required: Hemodialysis Device