Index: Alarms/Alarms.conf =================================================================== diff -u -r1db04161e368c14fe70a7c6e017ed4befe56debe -r3cfb50d5dd4cb017bbbc2bfb2224c125c2f47b6e --- Alarms/Alarms.conf (.../Alarms.conf) (revision 1db04161e368c14fe70a7c6e017ed4befe56debe) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 3cfb50d5dd4cb017bbbc2bfb2224c125c2f47b6e) @@ -1053,9 +1053,9 @@ SRSDG 945 = [193] -Title = -ListTitle = -Message = +Title = Blood Pump Stopped: Partial Occlusion +ListTitle = Blood Stop: Blood Pump Partial Occlusion +Message = Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [194] Title = Service Required: Dialysate Device