Index: Alarms/Alarms.conf =================================================================== diff -u -r43ace154f10ce6a8589ebe1326407aa66a049f51 -r682a2e558ef4f7448f2d744d20aab352928a1dcf --- Alarms/Alarms.conf (.../Alarms.conf) (revision 43ace154f10ce6a8589ebe1326407aa66a049f51) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 682a2e558ef4f7448f2d744d20aab352928a1dcf) @@ -249,9 +249,9 @@ Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [47] -Title = -ListTitle = -Message = +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Pump High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [48] Title = Service Required: Hemodialysis Device @@ -333,9 +333,9 @@ Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [63] -Title = Heparin Syringe Pump Prime Timeout -ListTitle = Process: Pre- Tx Hep. Pump Timeout. -Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Pre- Tx Hep. Pump Timeout. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [64] Title = Service Required: Dialysate Device