Index: Alarms/Alarms.conf =================================================================== diff -u -rcf601fa4517faee2868c7677e4308d736ba137f9 -r6a9f67a29509a880ed3d6062f66313f352049eda --- Alarms/Alarms.conf (.../Alarms.conf) (revision cf601fa4517faee2868c7677e4308d736ba137f9) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 6a9f67a29509a880ed3d6062f66313f352049eda) @@ -1110,9 +1110,9 @@ Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [203] -Title = -ListTitle = -Message = +Title = Service Required: Dialysate Device +ListTitle = DG POST: Fill Cond Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [204] Title = Service Required: Dialysate Device @@ -1714,9 +1714,9 @@ Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [306] -Title = Service Required: Dialysate Device -ListTitle = DG Fault: CD1 CD2 Check -Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +Title = +ListTitle = +Message = [307] Title = Power Loss