Index: Alarms/Alarms.conf =================================================================== diff -u -rc5862e4d4760c1ac231b0b95c6fa7ec9615ebf88 -r7787c6f24e8b93f0f17d54c03edfaf0c144139e1 --- Alarms/Alarms.conf (.../Alarms.conf) (revision c5862e4d4760c1ac231b0b95c6fa7ec9615ebf88) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 7787c6f24e8b93f0f17d54c03edfaf0c144139e1) @@ -552,7 +552,7 @@ [126] Title = Blood Leak -Message = A blood leak has been detected. \n- Treatment must be terminated without rinseback.\n\nBlood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press OK to restart blood pump. Dialysate will\n remain in bypass.\n\nBlood Stop: Dialyzer Blood Leak [127] Title = Low Venous Pressure @@ -563,8 +563,8 @@ Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Venous Air Bubble [129] -Title = -Message = \n\n +Title = Blood Leak +Message = The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to RESUME or END TREATMENT.\n\nBlood Stop: Dialyzer Blood Leak Bypass [130] Title = High Venous Pressure @@ -965,7 +965,7 @@ [216] Title = Treatment Complete -Message = Dialysis treatment is complete.\n- Press OK to end treatment and rinseback.\n\nProcess: End of Dialysis Tx +Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback.\n\nProcess: End of Dialysis Tx [217] Title =