Index: Alarms/Alarms.conf =================================================================== diff -u -re20b995ef22c8516d2bf04ba8efb68406ad47804 -r7d5c934a4d7a28fc5f15832ec70c97f211be4074 --- Alarms/Alarms.conf (.../Alarms.conf) (revision e20b995ef22c8516d2bf04ba8efb68406ad47804) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 7d5c934a4d7a28fc5f15832ec70c97f211be4074) @@ -1,1077 +1,1077 @@ [0] -Title = Indication for when no alarms is a possible situation +Title = ` Message = [1] -Title = HD stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up -Message = HD +Title = HD POST: Stuck Button +Message = Stop or On button pressed during startup\n-Release button and power cycle\n-Call service to schedule a repair if problem persists [2] -Title = HD FPGA POST failure -Message = HD +Title = HD POST: FPGA Compatibility +Message = Call service to schedule a repair. [3] -Title = DG FPGA POST failure -Message = DG +Title = DG POST: FPGA Compatibility +Message = Call service to schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [4] -Title = HD Watchdog POST failure -Message = HD +Title = HD POST: WD Safestate +Message = Call service to schedule a repair. [5] -Title = DG Watchdog POST failure -Message = DG +Title = DG POST: WD Safestate +Message = Call service to schedule a repair. SRSDG 53 = [6] -Title = HD UI communication POST failure. UI failed to communicate within a reasonable time after power up -Message = HD +Title = HD POST: Proc-UI Comm +Message = Call service to schedule a repair. [7] -Title = HD RTC battery low -Message = HD +Title = HD Alarm (Low): RTC Bat Low +Message = Call service to schedule a repair. [8] -Title = HD accelerometer failed POST -Message = HD +Title = HD POST: Acc Vector +Message = Call service to schedule a repair. [9] -Title = DG accelerometer failed POST -Message = DG +Title = DG POST: Acc Vector +Message = Call service to schedule a repair. SRSDG 323 = [10] -Title = HD blood leak sensor setting embedded mode failure. -Message = HD +Title = HD POST: BLD Config Emb Mode +Message = Call service to schedule a repair. [11] -Title = HD dialysate temperature below target temperature or below safety temp -Message = HD +Title = Dial. Gen: Dialysate Temp Low +Message = Dialysate temperature is low\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [12] -Title = HD firmware image integrity POST test failed -Message = HD +Title = HD POST: HD Image +Message = Call service to schedule a repair. [13] -Title = DG firmware image integrity POST test failed -Message = DG +Title = DG POST: DG Image +Message = Call service to schedule a repair. SRSDG 813 = [14] -Title = DG invalid usage record CRC -Message = DG +Title = DG POST: Usage Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 957 = [15] -Title = HD dialysate temperature above high safety. -Message = HD +Title = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high\n-Treatment is paused and will be ready to resume when \ndialysate is within target range. [16] -Title = HD alarm audio failed POST -Message = HD +Title = HD POST: Primary Alarm Current +Message = Call service to schedule a repair. [17] -Title = HD UI POST failed -Message = HD +Title = HD POST: UI POST Not Complete +Message = Call service to schedule a repair. [18] -Title = HD didn't get ACK on message to DG that required acknowledgment -Message = HD +Title = Loss of DG Comm (DG ACK) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [19] -Title = HD dialysate temperature above target temperature. -Message = HD +Title = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [20] -Title = DG conductivity sensors invalid calibration record -Message = DG +Title = DG POST: Conductivity Cal +Message = Call service to schedule a repair. SRSDG 16 = [21] -Title = DG drain line volume invalid calibration record -Message = DG +Title = DG POST: Drain Line Vol Cal +Message = Call service to schedule a repair. SRSDG 936 = [22] -Title = DG reservoirs invalid calibration record -Message = DG +Title = DG POST: Reservoir Vol Cal +Message = Call service to schedule a repair. SRSDG 935 = [23] -Title = DG acid concentrate calibration record -Message = DG +Title = DG POST: Acid Conc Cal +Message = Call service to schedule a repair. SRSDG 934 = [24] -Title = DG bicarb concentrate calibration record -Message = DG +Title = DG POST: Bicarb Conc Cal +Message = Call service to schedule a repair. SRSDG 933 = [25] -Title = DG accelerometer invalid calibration record -Message = DG +Title = DG POST: Acc Cal +Message = Call service to schedule a repair. SRSDG 817 = [26] -Title = HD accelerometer invalid calibration record -Message = HD +Title = HD POST: Acc Cal +Message = Call service to schedule a repair. [27] -Title = HD blood leak sensor zero and self test sequence failed -Message = HD +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test detected has a problem with the blood leak sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [28] -Title = DG two wire sensors FPGA fault -Message = DG +Title = DG Fault: Conductivity RTD Comm +Message = Call service to schedule a repair. SRSDG 864 = [29] -Title = HD heparin force sensor invalid calibration record -Message = HD +Title = HD POST: SP DAC Cal +Message = Call service to schedule a repair. [30] -Title = HD Software fault. Software found itself in an unexpected state -Message = HD +Title = HD Fault: SW Invalid State +Message = Call service to schedule a repair. [31] -Title = HD blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = HD +Title = HD Fault: BP Current +Message = Call service to schedule a repair. [32] -Title = HD blood pump failed motor off check. Measured speed while commanded off -Message = HD +Title = HD Fault: BP Speed +Message = Call service to schedule a repair. [33] -Title = HD blood pump failed motor direction check. Measured vs. commanded -Message = HD +Title = HD Fault: BP Direction +Message = Call service to schedule a repair. [34] -Title = HD blood pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = HD +Title = HD Fault: BP Rotor +Message = Blood pump rotor failure.\n- Check the pump rotor for secure connection. [35] -Title = HD dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = HD +Title = HD Fault: DPi Current +Message = Call service to schedule a repair. [36] -Title = HD dialysis inlet pump failed motor off check. Measured speed while commanded off -Message = HD +Title = HD Fault: DPi Speed +Message = Call service to schedule a repair. [37] -Title = HD dialysis inlet pump failed motor direction check. Measured vs. commanded -Message = HD +Title = HD Fault: DPi Direction +Message = Call service to schedule a repair. [38] -Title = HD dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = HD +Title = HD Fault: DPi Rotor +Message = Dialysate in pump rotor failure.\n- Check the pump rotor for secure connection. [39] -Title = HD dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = HD +Title = HD Fault: DPo Current +Message = Call service to schedule a repair. [40] -Title = HD dialysis outlet pump failed motor off check. Measured speed while commanded off -Message = HD +Title = HD Fault: DPo Speed +Message = Call service to schedule a repair. [41] -Title = HD dialysis outlet pump failed motor direction check. Measured vs. commanded -Message = HD +Title = HD Fault: DPo Direction +Message = Call service to schedule a repair. [42] -Title = HD dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = HD +Title = HD Fault: DPo Rotor +Message = Dialysate out pump rotor failure.\n- Check the pump rotor for secure connection. [43] -Title = DG main primary heater FPGA fault -Message = DG +Title = DG Fault: Iso ADC Comm +Message = Call service to schedule a repair. SRSDG 954 = [44] -Title = HD UI communication timeout -Message = HD +Title = HD Fault: UI Comm +Message = Call service to schedule a repair. [45] -Title = HD too many bad communications CRC -Message = HD +Title = HD Fault: CAN CRC Com +Message = Call service to schedule a repair. [46] -Title = HD didn't get ACK on message to UI that required acknowledgment -Message = HD +Title = HD Fault: CAN ACK UI Com +Message = Call service to schedule a repair. [47] -Title = Available for use +Title = Message = [48] -Title = HD ultrafiltration volume accuracy error during treatment -Message = HD +Title = HD Fault: UF Rate Tare Err +Message = Call service to schedule a repair. [49] -Title = HD FPGA communication down for too long -Message = HD +Title = HD Fault: FPGA Com +Message = Call service to schedule a repair. [50] -Title = DG FPGA not accepting commanded valve states -Message = DG +Title = DG Fault: Valve Control +Message = Call service to schedule a repair. SRSDG 26 = [51] -Title = HD blood pump failed motor speed check. Measured vs. commanded -Message = HD +Title = HD Fault: BP Speed +Message = Call service to schedule a repair. [52] -Title = HD dialysate inlet pump failed motor speed check. Measured vs. commanded -Message = HD +Title = HD Fault: DPi Speed +Message = Call service to schedule a repair. [53] -Title = HD dialysate outlet pump failed motor speed check. Measured vs. commanded -Message = HD +Title = HD Fault: DPo Speed +Message = Call service to schedule a repair. [54] -Title = HD critical data integrity check failed -Message = HD +Title = HD Fault: Critical Data Err +Message = Call service to schedule a repair. [55] -Title = DG critical data integrity check failed -Message = DG +Title = DG Fault: Critical Data Err +Message = Call service to schedule a repair. SRSDG 921 = [56] -Title = HD accelerometer error (no readings or FPGA reports error) -Message = HD +Title = HD Fault: Acc Com +Message = Call service to schedule a repair. [57] -Title = DG accelerometer error (no readings or FPGA reports error) -Message = DG +Title = DG Fault: Acc Com +Message = Call service to schedule a repair. SRSDG 869 SRSDG 870 = [58] -Title = HD valve homing failed -Message = HD +Title = HD Fault: Valve Home +Message = Call service to schedule a repair. [59] -Title = HD valve transition time out -Message = HD +Title = HD Fault: Valve Transition Err +Message = Call service to schedule a repair. [60] -Title = HD valve not functional -Message = HD +Title = HD Fault: Valve Control +Message = Call service to schedule a repair. [61] -Title = HD valve current out of range -Message = HD +Title = HD Fault: Valve Current +Message = Call service to schedule a repair. [62] -Title = HD valve position out of target -Message = HD +Title = HD Fault: Valve Position +Message = Call service to schedule a repair. [63] -Title = HD blood leak FPGA fault -Message = HD +Title = HD Fault: BLD Com +Message = Call service to schedule a repair. [64] -Title = DG barometric pressure sensor out of range -Message = DG +Title = DG Fault: Baro Range +Message = Call service to schedule a repair. SRSDG 874 = [65] -Title = HD DG requests DG command with invalid parameter fault -Message = HD +Title = Loss of DG Comm (Reject) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [66] -Title = HD blood leak sensor set point set failure -Message = HD +Title = HD POST: BLD Config Set Point +Message = Call service to schedule a repair. [67] -Title = HD blood pump occlusion self-test failure alarm -Message = HD +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [68] -Title = HD active reservoir recirculation out of range -Message = HD +Title = Dial. Gen: Dialysate not ready +Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. [69] -Title = HD blood leak sensor invalid calibration record -Message = HD +Title = HD POST: BLD Cal +Message = Call service to schedule a repair. [70] -Title = HD arterial pressure self-test failure alarm -Message = HD +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [71] -Title = HD venous pressure self-test failure alarm -Message = HD +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [72] -Title = HD No load cell data message receive at least once every 2 seconds -Message = HD +Title = Loss of DG Comm (LC) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [73] -Title = HD No dialysate temperature data message receive at least once every 2 seconds -Message = HD +Title = Loss of DG Comm (TDI/TRO) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [74] -Title = DG inlet UV reactor is on with no flow -Message = DG +Title = DG Fault: UVi On No Flow +Message = Call service to schedule a repair. SRSDG 600 = [75] -Title = HD syringe pump self-test failure alarm -Message = HD +Title = HD Fault: Syringe DAC +Message = Call service to schedule a repair. [76] -Title = HD monitored voltage is out of range -Message = HD +Title = HD Fault: Voltage Monitor +Message = Call service to schedule a repair. [77] -Title = DG monitored voltage is out of range -Message = DG +Title = DG Fault: Voltage Monitor +Message = Call service to schedule a repair. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] -Title = HD syringe pump direction (from encoder) error -Message = HD +Title = HD Fault: Syringe Pump Enc Dir +Message = Call service to schedule a repair. [79] -Title = HD syringe pump direction (from controller) error -Message = HD +Title = HD Fault: Syringe Pump Drv Dir +Message = Call service to schedule a repair. [80] -Title = HD syringe pump fault reported by FPGA -Message = HD +Title = HD Fault: Syringe Pump Driver +Message = Call service to schedule a repair. [81] -Title = HD syringe pump over travel error -Message = HD +Title = HD Fault: Syringe Pump OT +Message = Call service to schedule a repair. [82] -Title = HD syringe pump DAC write failure -Message = HD +Title = HD Fault: Syringe Pump DAC Err +Message = Call service to schedule a repair. [83] -Title = HD syringe pump is running while the BP is off -Message = HD +Title = HD Fault: Syringe Pump On Err +Message = Call service to schedule a repair. [84] -Title = DG set RTC year is invalid -Message = DG +Title = DG POST: HD RTC Time +Message = Call service to schedule a repair. SRSDG 815 = [85] -Title = HD pump track latch opened alarm -Message = HD +Title = Blood Stop: Cartridge Lever Open +Message = Cartridge Lock Lever is not closed.\n- Open front door and push lever all the way down.\n- Close door and press RESUME button to continue. [86] -Title = HD set RTC year is invalid -Message = HD +Title = HD POST: HD RTC Time +Message = Call service to schedule a repair. [87] -Title = DG heating invalid calibration record -Message = DG +Title = DG POST: Heat Cal +Message = Call service to schedule a repair. SRSDG 929 = [88] -Title = DG concentrate pumps hall sensor out of range -Message = DG +Title = DG Fault: CP1 CP2 Speed Range +Message = Call service to schedule a repair. [89] -Title = DG outlet UV reactor on with no flow -Message = DG +Title = DG Fault: UVo On No Flow +Message = Call service to schedule a repair. SRSDG 599 = [90] -Title = DG load cells A1/B1 FPGA fault -Message = DG +Title = DG Fault: LC A1/B1 Comm +Message = Call service to schedule a repair. SRSDG 604 = [91] -Title = HD No dialysate flow data receive in the last 3 seconds -Message = HD +Title = Loss of DG Comm (FMD) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [92] Title = Message = [93] -Title = DG outlet primary conductivity out of range -Message = DG +Title = DG Fault: RO Permeate Cond High +Message = Call service to schedule a repair. SRSDG 430 = [94] -Title = DG pressure out of range -Message = DG +Title = DG Fault: Pressure Sensor Range +Message = Call service to schedule a repair. SRSDG 629 = [95] -Title = DG watchdog expired -Message = DG +Title = DG Fault: WD Error +Message = Call service to schedule a repair. SRSDG 907 = [96] -Title = DG inlet water temperature in high range -Message = DG +Title = Dial. Gen: Inlet Water Temp High +Message = Inlet water temperature is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 410 = [97] -Title = DG fill conductivity out of range -Message = DG +Title = Dial. Gen: Dialysate Cond. +Message = Dialysate conductivity is out of range.\n- Check acid and bicarb concentrate is correct and not empty. SRSDG 397 = [98] -Title = HD battery communication fault -Message = HD +Title = HD Alarm (Low): Battery Com +Message = Call service to schedule a repair. [99] -Title = HD syringe pump stall alarm -Message = HD +Title = HD Fault: Syringe Pump Stall +Message = Call service to schedule a repair. [100] -Title = HD no cartridge self-test timeout -Message = HD +Title = HD Fault: Timeout No Cart ST +Message = Call service to schedule a repair. [101] -Title = UI POST HD communication failure -Message = UI +Title = HD POST: UI-Proc Comm +Message = Call service to schedule a repair. [102] -Title = DG heat disinfect temperature gradient out of range -Message = DG +Title = Cleaning: Tpo THD Temp Delta +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 856 SRSDG 930 = [103] -Title = HD invalid calibration CRC -Message = HD +Title = HD POST: Cal NVRAM CRC +Message = Call service to schedule a repair. [104] -Title = HD air trap level sensors reporting illegal combination of air/fluid -Message = HD +Title = HD Fault: Air Trap Sensor +Message = Call service to schedule a repair. [105] -Title = DG invalid calibration CRC -Message = DG +Title = DG POST: Cal NVRAM CRC +Message = Call service to schedule a repair. SRSDG 608 = [106] -Title = DG dialysate flow sensor invalid calibration record -Message = DG +Title = DG POST: Flow Sensor Cal +Message = Call service to schedule a repair. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] -Title = HD reports DG restarted fault -Message = HD +Title = DG Fault: DG Restarted +Message = Call service to schedule a repair. [108] -Title = HD syringe pump ADC FPGA fault -Message = HD +Title = HD Fault: Syringe Pump ADC Com +Message = Call service to schedule a repair. [109] -Title = HD syringe pump volume check error -Message = HD +Title = HD Fault: Syringe Pump Vol +Message = Call service to schedule a repair. [110] -Title = HD syringe pump speed check error -Message = HD +Title = HD Fault: Syringe Pump Speed +Message = Call service to schedule a repair. [111] -Title = HD syringe pump not stopped in off state error -Message = HD +Title = HD Fault: Syringe Pump Speed +Message = Call service to schedule a repair. [112] -Title = DG load cells A1/B1 FPGA fault -Message = DG +Title = DG Fault: LC A2/B2 Comm +Message = Call service to schedule a repair. SRSDG 953 = [113] -Title = DG Drain pump current out of range -Message = DG +Title = DG Fault: DRP Current +Message = Call service to schedule a repair. SRSDG 890 = [114] -Title = HD venous air bubble detector self-test failure -Message = HD +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem with the air bubble sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [115] -Title = DG temperature sensor out of range -Message = DG +Title = DG Fault: Temp Sensor Range +Message = Call service to schedule a repair. SRSDG 224 = [116] -Title = HD UI SD card failure -Message = UI HD +Title = HD Alarm (Low): SD-Card Error +Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. [117] -Title = DG CAN message not acked -Message = DG +Title = Loss of HD Comm (HD ACK) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. SRSDG 909 = [118] -Title = DG RTC config error -Message = DG +Title = DG Fault: RTC Config +Message = Call service to schedule a repair. SRSDG 43 = [119] -Title = DG RTC battery low error -Message = DG +Title = DG Alarm (Low): RTC Bat Low +Message = Call service to schedule a repair. SRSDG 908 = [120] -Title = HD pre-treatment mode wet self-test prime check failure -Message = HD +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a bubble in the venous bubble detector\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [121] -Title = DG main primary heater voltage out of range -Message = DG +Title = DG Fault: Prim Heater Lg Control +Message = Call service to schedule a repair. SRSDG 611 = [122] -Title = DG small primary heater voltage out of range -Message = DG +Title = DG Fault: Prim Heater Sm Control +Message = Call service to schedule a repair. SRSDG 612 = [123] -Title = DG trimmer heater voltage out of range -Message = DG +Title = DG Fault: Trimmer Heater Control +Message = Call service to schedule a repair. SRSDG 613 = [124] -Title = HD end of treatment alarm (high priority - user still not acting to end treatment) -Message = HD +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [125] -Title = HD blood sitting too long after treatment stopped by user alarm (>5 min) -Message = HD +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [126] -Title = HD blood leak detected alarm -Message = HD +Title = Blood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. Treatment must end. [127] -Title = HD venous pressure too low during treatment -Message = HD +Title = Blood Stop: Venous Pres. Low +Message = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured \nproperly. [128] -Title = HD venous air bubble detected alarm -Message = HD +Title = Blood Stop: Venous Air Bubble +Message = Venous air bubble detected indicating failure of air trap \ncontrol system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [129] -Title = Available for use +Title = Message = [130] -Title = HD venous pressure too high during treatment -Message = HD +Title = Blood Stop: Venous Pres. High +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [131] -Title = HD arterial pressure too low during treatment -Message = HD +Title = Blood Stop: Arterial Pres. Low +Message = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [132] -Title = HD arterial pressure too high during treatment -Message = HD +Title = Blood Stop: Arterial Pres. High +Message = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [133] -Title = DG fluid leak detected alarm -Message = DG +Title = DG Fault: Fluid Leak in DG +Message = Fluid leak detected in Dialysate Generation Delivery Device.\n-Treatment must be terminated.\n- Shut off water source and disconnect water and \nelectrical connections from system.\n- Call service to schedule a repair. SRSDG 328 = [134] -Title = HD fluid leak detected alarm -Message = HD +Title = Blood Stop: Fluid Leak in HD +Message = Fluid leak detected in Hemodialysis Delivery device.\n- If in treatment, BLOOD STOPPED.\n-Identify and correct the cause of the spill.\n- Clean up the spill and dry leak sensor.\n- If in treatment, resume treatment. [135] -Title = HD experienced large acceleration -Message = HD +Title = Blood Stop: HD Shock Detected +Message = Excessive movement detected with Hemodialysis Delivery \ndevice.\n- Confirm the device is stationary and level before continuing. [136] -Title = DG experienced large acceleration -Message = DG +Title = Dial. Gen: Shock-Vibration +Message = Excessive movement detected with Dialysis Generation \ndevice.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. SRSDG 327 = [137] -Title = HD tilt exceeds maximum -Message = HD +Title = Blood Stop: HD Tilted +Message = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. [138] -Title = DG tilt exceeds maximum -Message = DG +Title = Dial. Gen: Tilt +Message = Excessive tilt detected with Dialysis Generation device.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing. SRSDG 326 = [139] -Title = HD AC power lost alarm -Message = HD +Title = Loss of AC Power +Message = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns. [140] -Title = HD DG communication timeout -Message = HD +Title = Loss of DG Comm (Timeout) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [141] -Title = HD air trap fill timeout during treatment -Message = HD +Title = Blood Stop: Air Trap Fill +Message = Failure to fill the Air Trap.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [142] -Title = HD blood pump occlusion detected -Message = HD +Title = Blood Stop: Blood Pump Occlusion +Message = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the \nvenous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. [143] -Title = DG dialysate temperature sensors out of range -Message = DG +Title = DG Fault: TRo TDi Delta +Message = Call service to schedule a repair. SRSDG 912 = [144] -Title = DG cleaning mode inlet water conductivity out of high range. -Message = DG +Title = Cleaning: Inlet Water Cond High +Message = Cleaning Mode Inlet water conductivity too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = DG concentrate conductivity after adding acid out of range alarm -Message = DG +Title = Dial. Gen: Acid Conc. Cond. +Message = Acid concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 846 = [146] -Title = DG RTC or timer accuracy failure. -Message = DG +Title = DG POST: DG RTC Timer +Message = Call service to schedule a repair. SRSDG 312 = [147] -Title = DG wait for the DG to produce dialysate -Message = DG +Title = Dial. Gen: Dialysate not ready +Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. SRSDG 398 = [148] -Title = DG Cleaning mode inlet water temperature too high -Message = DG +Title = Cleaning: Inlet Water Temp High +Message = Cleaning Mode Inlet water temperature too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] -Title = HD watchdog expired -Message = HD +Title = HD Fault: WD Error +Message = Call service to schedule a repair. [150] -Title = DG inlet water conductivity in high range -Message = DG +Title = Dial. Gen: Inlet Water Cond High +Message = Inlet water conductivity is high.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 391 = [151] -Title = DG inlet water conductivity in low range -Message = DG +Title = Dial. Gen: Inlet Water Cond Low +Message = Inlet water conductivity is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 826 = [152] -Title = DG inlet water pressure in low range -Message = DG +Title = Dial. Gen: Inlet Water Pres Low +Message = Inlet water pressure is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 417 = [153] -Title = HD prime completed high priority alarm -Message = HD +Title = Process: Pre-Tx Terminated +Message = Pre-Treatment timeout.\n- Patient was not connected to device\n- Pre-Treatment has been terminated\n- Proceed to end treatment to remove cartridge from system. [154] -Title = DG cleaning mode inlet water conductivity out of low range. -Message = DG +Title = Cleaning: Inlet Water Cond Low +Message = Cleaning Mode Inlet water conductivity too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = HD No reservoirs data message receive at least once every 2 seconds -Message = HD +Title = Loss of DG Comm (Reservoir) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [156] -Title = HD No DG operation mode message receive at least once every 2 seconds -Message = HD +Title = Loss of DG Comm (Op Mode) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [157] -Title = DG chemical disinfect prime acid line timeout. -Message = DG +Title = Cleaning: Chem Disinf PAA Prime +Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 988 = [158] -Title = DG inlet water temperature is in the low range -Message = DG +Title = Dial. Gen: Inlet Water Temp Low +Message = Inlet water temperature is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 408 = [159] -Title = DG chemical disinfect flush remove acid and close the concentrate cap. -Message = DG +Title = Cleaning: Chem Disinf Connect Cap +Message = Remove chemical disinfectant connection and close cap. SRSDG 1044 = [160] -Title = HD blood pump rotor speed too high -Message = HD +Title = HD Fault: BP Rotor Max Speed +Message = Call service to schedule a repair. [161] -Title = DG comm too many bad CRCs -Message = DG +Title = DG Fault: HD-DG Comm +Message = Call service to schedule a repair. SRSDG 122 = [162] -Title = DG FPGA clock speed check failure. -Message = DG +Title = DG Fault: Proc-FPGA Clock +Message = Call service to schedule a repair. SRSDG 867 = [163] -Title = HD sees primary load cell for reservoir 1 change too much too fast -Message = HD +Title = Dial. Gen: UF Rate Check +Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system\n(shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a\nrepair. [164] -Title = HD sees primary load cell for reservoir 2 change too much too fast -Message = HD +Title = Dial. Gen: UF Rate Check +Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system \n(shock/vibration).\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a \nrepair. [165] -Title = HD in treatment rinseback operation timeout -Message = HD +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [166] -Title = Available for use +Title = Message = [167] -Title = HD cartridge door opened alarm -Message = HD +Title = Blood Stop: Door Open +Message = Device door is open.\n - Close door and press RESUME button to continue. [168] -Title = Available for use +Title = Message = [169] -Title = DG Dialysate flow rate out of maximum range -Message = DG +Title = DG Fault: Dialysate Flow Sens Rg +Message = Call service to schedule a repair. SRSDG 643 = [170] -Title = HD syringe empty alarm -Message = HD +Title = Heparin Syringe Empty +Message = Heparin syringe empty.\n- Treatment will continue\n- If additional heparin is required to complete treatment, \nadminister manually. [171] -Title = HD syringe pump occlusion alarm -Message = HD +Title = Process: Pre-Tx Hep. Pump Occ. +Message = Heparin pump occlusion detected.\n- Check heparin line is unclamped.\n- Check there are no kinks in the heparin line. [172] -Title = HD syringe pump not enough Heparin alarm -Message = HD +Title = Process: Pre-Tx Not Enough Hep. +Message = Not enough heparin is available for treatment.\n- Check if syringe pump is touching syringe plunger.\n- If syringe pump is touching syringe plunger, end treatment.\n- If syringe pump is not touching syringe plunger, reload \nsyringe with enough heparin. [173] -Title = HD RTC configuration error -Message = HD +Title = HD Fault: RTC Config +Message = Call service to schedule a repair. [174] -Title = HD RTC or timer accuracy failure -Message = HD +Title = HD POST: HD RTC Timer +Message = Call service to schedule a repair. [175] -Title = HD pump direction status error -Message = HD +Title = HD Fault: Pump Commutation Err +Message = Call service to schedule a repair. [176] -Title = HD dialysate temperature above high safety range +Title = Message = [177] -Title = DG software fault. Software found itself in an unexpected state -Message = DG +Title = DG Fault: Software Invalid State +Message = Call service to schedule a repair. SRSDG 919 = [178] -Title = HD communication timeout -Message = UI +Title = HD Fault: UI-Proc Comm +Message = Call service to schedule a repair. = SRSUI 1066 [179] -Title = DG FPGA communication down for too long -Message = DG +Title = DG Fault: FPGA Com +Message = Call service to schedule a repair. SRSDG 605 = [180] -Title = DG RO flow out of maximum range -Message = DG +Title = DG Fault: RO Flow Sensor Range +Message = Call service to schedule a repair. SRSDG 883 = [181] -Title = DG load cells weight out of range for tare -Message = DG +Title = DG Fault: Tare Out of Range +Message = Call service to schedule a repair. SRSDG 849 SRSDG 838 = [182] -Title = DG load cells invalid calibration -Message = DG +Title = DG POST: LC Cal +Message = Call service to schedule a repair. SRSDG 162 = [183] -Title = DG invalid load cell value -Message = DG +Title = Cleaning: Invalid Resv Transfer +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 851 = [184] -Title = DG inlet UV reactor not healthy -Message = DG +Title = DG Fault: UVi Error +Message = Call service to schedule a repair. SRSDG 559 = [185] -Title = DG fan RPM out of range -Message = DG +Title = DG Alarm (Low): Fan Speed +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. SRSDG 628 = [186] -Title = DG concentrate pump fault -Message = DG +Title = DG Fault: CP1/CP2 Fault +Message = Call service to schedule a repair. SRSDG 891 = [187] -Title = DG concentrate pump CP1 speed control error -Message = DG +Title = DG Fault: CP1 Speed Fault +Message = Call service to schedule a repair. SRSDG 905 = [188] -Title = DG concentrate pump CP2 speed control error -Message = DG +Title = DG Fault: CP2 Speed Fault +Message = Call service to schedule a repair. SRSDG 906 = [189] -Title = DG drain pump RPM out of range -Message = DG +Title = DG Fault: DRP Speed Fault +Message = Call service to schedule a repair. SRSDG 448 = [190] -Title = DG drain pump off fault -Message = DG +Title = DG Fault: DRP Speed Not Off +Message = Call service to schedule a repair. [191] -Title = DG flow rate out of upper range -Message = DG +Title = DG Fault: RO Flow High +Message = Call service to schedule a repair. SRSDG 885 = [192] -Title = DG flow rate out of lower range -Message = DG +Title = DG Fault: RO Flow Low +Message = Call service to schedule a repair. SRSDG 945 = [193] -Title = HD dialysate temperature out of low safety range. +Title = Message = [194] -Title = DG RO pump duty cycle out of range -Message = DG +Title = DG Fault: RO Driver Feedback +Message = Call service to schedule a repair. SRSDG 881 = [195] -Title = DG RO pump pressure out of range -Message = DG +Title = DG Fault: RO Pump High Pressure +Message = Call service to schedule a repair. SRSDG 925 = [196] -Title = DG CPi/CPo sensors FPGA fault -Message = DG +Title = DG Fault: CPi/CPo Comm +Message = Call service to schedule a repair. SRSDG 955 = [197] -Title = DG CD1/CD2 sensors FPGA fault -Message = DG +Title = DG Fault: CD1/CD2 Comm +Message = Call service to schedule a repair. SRSDG 956 = [198] -Title = DG RO flow too low while primary heater is on -Message = DG +Title = DG Fault: HP On With No Flow +Message = Call service to schedule a repair. SRSDG 872 = [199] -Title = DG Dialysate flow too low while trimmer heater is on -Message = DG +Title = DG Fault: HR On With No Flow +Message = Call service to schedule a repair. SRSDG 871 = [200] -Title = DG thermistors/sensors temperature out of range -Message = DG +Title = DG Fault: Thermistor Range +Message = Call service to schedule a repair. SRSDG 178 = [201] -Title = HD pre-treatment mode wet self-test lc vs FMD failure -Message = HD +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem in the dialysate flow \nmeasurement or there is a leak in the dialysate in/out lines.\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [202] -Title = HD pre-treatment mode dry pressure normal self-test failure -Message = HD +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected blood circuit pressure cannot be \nrelieved.\n- Check PBo connection to HD for kinks\n -Check PBo transducer protector for occlusion\n- If problem persists, install a new cartridge and \nPBo transducer protector. [203] -Title = Available for use +Title = Message = [204] -Title = DG RO rejection ratio out of range -Message = DG +Title = DG Fault: RO Rejection +Message = Call service to schedule a repair. SRSDG 429 = [205] -Title = DG conductivity sensor fault -Message = DG +Title = DG Fault: Cond Sensors Error +Message = Call service to schedule a repair. SRSDG 913 = [206] -Title = DG dialysate fill runs out of time -Message = DG +Title = DG Fault: Reservoir Fill Timeout +Message = Call service to schedule a repair. SRSDG 437 = [207] -Title = DG flow meter check failure alarm -Message = DG +Title = DG Fault: FMP Flow Rate Error +Message = Call service to schedule a repair. SRSDG 240 = [208] -Title = HD venous pressure occlusion alarm -Message = HD +Title = Blood Stop: Venous Pres. High +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [209] -Title = DG drain circulation line timeout -Message = DG +Title = Cleaning: Flush Circ Error +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 980 = [210] -Title = HD battery pack detected an error -Message = HD +Title = HD Alarm (Low): Battery Fault +Message = Call service to schedule a repair. [211] -Title = HD blood sitting too long warning (>4 min) -Message = HD +Title = Blood Stopped: Warning +Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [212] -Title = HD end of treatment alarm (user not acting to end treatment) -Message = HD +Title = Blood Stopped: Warning +Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [213] -Title = HD prime completed medium priority alarm -Message = HD +Title = Process: Pre-Tx Prime Complete +Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [214] -Title = Available for use +Title = Message = [215] -Title = Available for use +Title = Message = [216] -Title = HD end treatment sub-mode timeout alarm -Message = HD +Title = Process: End of Dialysis Tx +Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [217] -Title = Available for use +Title = Message = [218] -Title = HD syringe detected alarm -Message = HD +Title = Process: Remove Syringe +Message = Syringe not removed from machine.\n- Remove heparin syringe. [219] -Title = HD syringe pump syringe removed alarm -Message = HD +Title = Blood Stop: Syringe Not Detected +Message = Heparin is needed for treatment but syringe has been removed \nfrom syringe pump.\n- Install Heparin syringe. [220] -Title = DG THd sensors FPGA fault -Message = DG +Title = DG Fault: THD RTD Comm +Message = Call service to schedule a repair. SRSDG 947 = [221] -Title = HD Saline bag is empty -Message = HD +Title = Blood Stop: Saline Bag Empty +Message = Saline bag appears empty.\n- Check saline bag and line and replace if saline bag is \nempty. [222] -Title = HD occlusion sensor FPGA fault -Message = HD +Title = HD Fault: BP Occ Com +Message = Call service to schedule a repair. [223] -Title = HD arterial sensor FPGA fault -Message = HD +Title = HD Fault: Art. Pres. Com +Message = Call service to schedule a repair. [224] -Title = HD treatment stopped by user action - pressed stop button -Message = HD +Title = Blood Stopped: Treatment Pause +Message = Treatment paused by user. [225] -Title = HD end of treatment warning -Message = HD +Title = Process: End of Treatment +Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [226] -Title = HD prime completed low priority alarm -Message = HD +Title = Process: Pre-Tx Priming Complete +Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [227] -Title = Available for use +Title = Message = [228] -Title = HD time out on prime saline purge air state. -Message = HD +Title = Process: Pre-Tx Prime Timeout +Message = Blood line prime with saline timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [229] -Title = HD prime dialysate dialyzer time out alarm -Message = HD +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [230] -Title = HD prime dialysate bypass time out alarm -Message = HD +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [231] -Title = HD pre-treatment mode dry pressure self-test failure -Message = HD +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected a leak in the circuit\n- Check blood line connections\n- Check dialysate line connections\n- Check PBo and PBa transducer protectors for occlusion.\n- Check clamps.\n- If problem persists, install a new cartridge. [232] -Title = HD pre-treatment mode wet self-test LC vs. LC failure -Message = HD +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem in the load cell \nmeasurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation onto machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [233] -Title = DG barometric sensor coefficients bad CRC -Message = DG +Title = DG Fault: Baro Cal CRC +Message = Call service to schedule a repair. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = DG inlet water pressure in high range. -Message = DG +Title = Dial. Gen: Inlet Water Pres High +Message = Inlet water pressure is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 1037 = [235] -Title = HD in treatment stopped sub-mode after rinseback completed (no escalation) -Message = HD +Title = Process: Rinseback Complete +Message = Rinseback has been completed. [236] -Title = HD needs new cartridge to be installed -Message = HD +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected fluid in the cartridge.\n- Cartridges are single-use only\n- Install new cartridge to resume. [237] -Title = HD prime saline dialyzer time out alarm -Message = HD +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [238] -Title = HD no cartridge loaded or installed improperly alarm -Message = HD +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected cartridge is not installed.\n- Check that a cartridge is properly installed. [239] -Title = HD fail to remove cartridge alarm -Message = HD +Title = Process: Post-Tx Remove Cartridge +Message = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. [240] -Title = DG bicarb conductivity out of range during bicarb pump check alarm -Message = DG +Title = Dial. Gen: Bicarb Conc. Cond. +Message = Bicarbonate conductivity during prime is out of range.\n- Check correct concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 482 = [241] -Title = DG reservoir drain time out -Message = DG +Title = Cleaning: Resv Drain Timeout +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 848 SRSDG 90 SRSDG 561 @@ -1102,8 +1102,8 @@ SRSDG 1034 = [242] -Title = DG reservoir fill time out -Message = DG +Title = Cleaning: Resv Fill Timeout +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 850 SRSDG 573 SRSDG 574 @@ -1124,373 +1124,373 @@ SRSDG 1024 = [243] -Title = DG reservoir leak time out -Message = DG +Title = Cleaning: Fluid Loss Detected +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 853 = [244] -Title = DG cleaning mode temperature sensors difference out of range -Message = DG +Title = Cleaning: Temp Sensor Check +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 852 = [245] -Title = DG heat disinfect target temperature time out (could not reach to temperature) -Message = DG +Title = Cleaning: Heat Disinfect Temp TO +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 854 = [246] -Title = DG cleaning mode conductivity sensors out of range -Message = DG +Title = Cleaning: RO Water Cond Error +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [247] -Title = Alarm Id available +Title = Message = [248] -Title = DG chemical disinfect could not reach to target temperature out of range -Message = DG +Title = Cleaning: Chem Disinfect Proc +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [249] -Title = DG chemical disinfect target conductivity out of range -Message = DG +Title = Cleaning: Chem Disinf PAA Cond +Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 992 = [250] -Title = DG chemical disinfect insert acid and remove the concentrate cap -Message = DG +Title = Cleaning: Chem Disinf Connect PAA +Message = Connect chemical disinfectant bottle to bicarb port. SRSDG 1046 = [251] -Title = HD invalid system record -Message = HD +Title = HD POST: Sys Rec NVRAM CRC +Message = Call service to schedule a repair. [252] -Title = HD invalid service record -Message = HD +Title = HD POST: Serv Rec NVRAM CRC +Message = Call service to schedule a repair. [253] -Title = DG invalid system record -Message = DG +Title = DG POST: Sys Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 609 = [254] -Title = DG invalid service record -Message = DG +Title = DG POST: Serv Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 610 = [255] -Title = HD and UI software builds are not compatible -Message = HD +Title = HD POST: UI Compatibility +Message = Call service to schedule a repair. [256] -Title = HD processor is in Disinfect Chemical Flush mode -Message = HD +Title = Cleaning: Chem Flush In Progress +Message = Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. = SRSUI 497 [257] -Title = Alarm ID available for use +Title = Message = [258] -Title = Alarm Id available +Title = Message = [259] -Title = HD temperatures out of range -Message = HD +Title = HD Fault: Temp Sensors Error +Message = Call service to schedule a repair. [260] -Title = HD UI POST Application Integrity (Sha256Sum) failure -Message = UI HD +Title = HD POST: UI Image +Message = Call service to schedule a repair. = SRSUI 1067 [261] -Title = HD UI POST CANBus failure -Message = UI HD +Title = HD POST: UI CAN Driver +Message = Call service to schedule a repair. = SRSUI 1068 [262] -Title = HD UI POST Display failure -Message = UI HD +Title = HD POST: UI Disp Driver +Message = Call service to schedule a repair. = SRSUI 1069 [263] -Title = HD UI POST Touch failure -Message = UI HD +Title = HD POST: UI Touch Driver +Message = Call service to schedule a repair. = SRSUI 1070 [264] -Title = HD UI POST SD-Card failure -Message = UI HD +Title = HD POST: UI SD-Card +Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. = SRSUI 1071 [265] -Title = HD UI POST RTC failure -Message = UI HD +Title = HD POST: UI RTC +Message = Call service to schedule a repair. = SRSUI 1072 [266] -Title = HD UI POST WiFi failure -Message = UI HD +Title = HD POST: UI WiFi +Message = Call service to schedule a repair. = SRSUI 1073 [267] -Title = HD UI POST Bluetooth failure -Message = UI HD +Title = HD POST: UI Bluetooth +Message = Call service to schedule a repair. = SRSUI 1074 [268] -Title = HD UI POST Ethernet failure -Message = UI HD +Title = HD POST: UI Ethernet +Message = Call service to schedule a repair. = SRSUI 1075 [269] -Title = HD UI POST Sound failure -Message = UI HD +Title = HD POST: UI Sound +Message = Call service to schedule a repair. = SRSUI 1076 [270] -Title = HD POST Safety Shutdown failure -Message = HD +Title = HD POST: Proc Safestate +Message = Call service to schedule a repair. [271] -Title = DG POST Safety Shutdown failure -Message = DG +Title = DG POST: Proc Safestate +Message = Call service to schedule a repair. SRSDG 822 = [272] -Title = HD Fan RPM out of range -Message = HD +Title = HD Alarm (Low): Fan Speed +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. [273] -Title = Available for use +Title = Message = [274] -Title = Available for use +Title = Message = [275] -Title = HD arterial pressure sensor is reading out of range -Message = HD +Title = HD Fault: Arterial Press Range +Message = Call service to schedule a repair. [276] -Title = HD venous pressure sensor is reading out of range -Message = HD +Title = HD Fault: Venous Press Range +Message = Call service to schedule a repair. [277] -Title = HD BP occlusion sensor is reading out of range -Message = HD +Title = HD Fault: Occlusion Sens Range +Message = Call service to schedule a repair. [278] -Title = HD active reservoir weight out of range. -Message = HD +Title = DG Fault: Reservoir Overfill +Message = Call service to schedule a repair. [279] -Title = DG dialysate drain time out -Message = DG +Title = DG Fault: Reservoir Drain +Message = Call service to schedule a repair. SRSDG 752 = [280] -Title = DG drain pump direction invalid -Message = DG +Title = DG Fault: DRP Direction Fault +Message = Call service to schedule a repair. SRSDG 888 = [281] -Title = Acid concentration bottle low volume alarm -Message = DG +Title = Dial. Gen: Acid Conc. Low +Message = Acid concentrate low.\n- Replace acid concentrate bottle. SRSDG 836 = [282] -Title = Bicarbonate concentration bottle low volume alarm -Message = DG +Title = Dial. Gen: Bicarb Conc. Low +Message = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. SRSDG 837 = [283] -Title = DG load cells weight out of range -Message = DG +Title = DG Fault: LC Range +Message = Call service to schedule a repair. SRSDG 615 = [284] -Title = DG load cells primary/back up drift out of range -Message = DG +Title = DG Fault: Redundant LC Check +Message = Call service to schedule a repair. SRSDG 383 = [285] -Title = Available for use +Title = Message = [286] -Title = DG concentrate cap not in proper position -Message = DG +Title = Cleaning: Conc Cap Removed +Message = Concentrate cap removed from DG during cleaning.\n- Cleaning Process Terminated. [287] -Title = HD processor clock speed checks against FPGA clock failure -Message = HD +Title = HD Fault: Proc-FPGA Clock +Message = Call service to schedule a repair. [288] -Title = HD load cells primary/back up drift out of range -Message = HD +Title = DG Fault: LC Drift Check +Message = Call service to schedule a repair. [289] -Title = DG dialysate cap not closed -Message = DG +Title = Cleaning: Dialysate Cap Removed +Message = Dialysate cap removed from DG during cleaning.\n- Cleaning Process Terminated. SRSDG 928 = [290] -Title = HD venous pressure sensor FPGA fault -Message = HD +Title = HD Fault: Ven. Pres. Com +Message = Call service to schedule a repair. [291] -Title = HD processor is in Disinfect Flush mode -Message = HD +Title = Cleaning: Flush In Progress +Message = System flush in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [292] -Title = HD processor is in Disinfect Heat mode -Message = HD +Title = Cleaning: Heat Disinf In Progress +Message = Heat disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [293] -Title = HD processor is in Disinfect Chemical mode -Message = HD +Title = Cleaning: Chem Flush In Progress +Message = Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [294] -Title = Available for use +Title = Message = [295] -Title = HD UI POST CloudSync failure -Message = UI HD +Title = HD POST: UI CloudSync +Message = Call service to schedule a repair. = SRSUI 1238 [296] -Title = DG TDi sensors FPGA fault -Message = DG +Title = DG Fault: TDi RTD Comm +Message = Call service to schedule a repair. SRSDG 951 = [297] -Title = DG TRo sensors FPGA fault -Message = DG +Title = DG Fault: TRo RTD Comm +Message = Call service to schedule a repair. SRSDG 950 = [298] -Title = DG baro sensor FPGA fault -Message = DG +Title = DG Fault: Baro Pres Comm +Message = Call service to schedule a repair. SRSDG 948 = [299] -Title = DG invalid serial number -Message = DG +Title = DG POST: Serial Number Check +Message = Call service to schedule a repair. SRSDG 952 = [300] -Title = HD invalid serial number -Message = HD +Title = HD POST: Serial Number Check +Message = Call service to schedule a repair. [301] -Title = Alarm ID available +Title = Message = [302] -Title = DG processor RAM error -Message = DG +Title = DG Fault: Processor RAM error +Message = Call service to schedule a repair. SRSDG 126 = [303] -Title = DG drain pump direction FPGA fault -Message = DG +Title = DG Fault: DRP Hall Commutation +Message = Call service to schedule a repair. SRSDG 958 = [304] -Title = HD invalid usage record -Message = HD +Title = HD POST: Usage Rec NVRAM CRC +Message = Call service to schedule a repair. [305] -Title = HD processor RAM error -Message = HD +Title = HD Fault: Processor RAM error +Message = Call service to schedule a repair. [306] -Title = DG CD1 CD2 check fault -Message = DG +Title = DG Fault: CD1 CD2 Check +Message = Call service to schedule a repair. [307] -Title = HD AC power lost second alarm -Message = HD +Title = Blood Stop: Loss of AC Power +Message = HD power has been lost. \n- Treatment must be terminated.\n-Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns.\n-If power is returned prior to blood stopped countdown elapses, \nrinseback will be allowed. [308] -Title = DG CPi conductivity sensor invalid character received -Message = DG +Title = DG Fault: CPi Comm +Message = Call service to schedule a repair. SRSDG 959 = [309] -Title = DG CPo conductivity sensor invalid character received -Message = DG +Title = DG Fault: CPo Comm +Message = Call service to schedule a repair. SRSDG 960 = [310] -Title = DG CD1 conductivity sensor invalid character received -Message = DG +Title = DG Fault: CD1 Comm +Message = Call service to schedule a repair. SRSDG 962 = [311] -Title = DG CD2 conductivity sensor invalid character received -Message = DG +Title = DG Fault: CD2 Comm +Message = Call service to schedule a repair. SRSDG 961 = [312] -Title = HD Dialysate In flow rate to too low. -Message = HD +Title = Dial. Gen: Dialysate In Flow +Message = Dialysate in flow rate is low\n- Check dialysate line connection to the DG. [313] -Title = DG cleaning mode inlet water temperature too low. -Message = DG +Title = Cleaning: Inlet Water Temp Low +Message = Cleaning Mode Inlet water temperature too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = DG cleaning mode inlet water pressure too high. -Message = DG +Title = Cleaning: Inlet Water Press High +Message = Cleaning Mode Inlet water pressure too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = DG cleaning mode inlet water pressure too low. -Message = DG +Title = Cleaning: Inlet Water Press Low +Message = Cleaning Mode Inlet water pressure too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = DG acid concentrate pump park fault. -Message = DG +Title = DG Fault: Acid CP Pump Park +Message = Call service to schedule a repair. SRSDG 983 = [317] -Title = DG bicarb concentrate pump part fault. -Message = DG +Title = DG Fault: Bicarb CP Pump Park +Message = Call service to schedule a repair. SRSDG 984 = [318] -Title = HD Air Pump timeout alarm -Message = HD +Title = HD Fault: Air Pump Timeout +Message = Unable to reduce fluid level.\n- Air trap may be overfilled.\n- Pressure transducer protector may be contaminated.\n- Treatment must be terminated without option for rinseback.\n- Call service to schedule a repair. [319] -Title = DG chemical disinfect flush flush sample. -Message = DG +Title = Cleaning: Chem Residual Sampling +Message = Collect chemical residuals water sample from drain line. SRSDG 1040 = [320] -Title = DG chemical disinfect flush sample timeout. -Message = DG +Title = Cleaning: Chem Residual Timeout +Message = During Chemical Disinfection Flush\n-user residual sampling timed out. SRSDG 1043 = [321] -Title = DG outlet UV reactor not healthy -Message = DG +Title = DG Fault: UVo Error +Message = Call service to schedule a repair. SRSDG 1069 = [322] -Title = HD UI POST Year check failure -Message = UI HD +Title = HD POST Alarm: UI Invalid Year +Message = Call service to schedule a repair.