Index: Alarms/Alarms.conf =================================================================== diff -u -r04843b618095ef6dbe2aeb60cc9df4e70b12537a -r99267d15f547caea0dc88faaa8e8766123c9df69 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 04843b618095ef6dbe2aeb60cc9df4e70b12537a) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision 99267d15f547caea0dc88faaa8e8766123c9df69) @@ -424,9 +424,9 @@ HD venous pressure sensor not being read = ALARM_ID_HD_VENOUS_PRESSURE_READ_TIMEOUT_ERROR [85] -Title = Venous Pressure Error -Message = Call service to schedule a repair. -HD venous pressure sensor temperature out of range error = ALARM_ID_PUMP_TRACK_LATCH_OPENED +Title = Pump Track Lever Not Fully Latched +Message = Open front door and pull pump track lever all the way down. +HD pump track lever is not fully latched = ALARM_ID_PUMP_TRACK_LATCH_OPENED [86] Title = Occlusion Sensor Error @@ -1166,7 +1166,7 @@ [233] Title = Battery Error Message = Battery pack error. -RTC battery low = ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC +RTC battery low = ALARM_ID__AVAILABLE_8 [234] Title = Communication Failure