Fisheye: Tag 9a2fac022f74292e0b1518c9839a0f5749c4ac9c refers to a dead (removed) revision in file `Alarms/Alarms.conf'. Fisheye: No comparison available. Pass `N' to diff? Index: Alarms/Alarms_en.conf =================================================================== diff -u --- Alarms/Alarms_en.conf (revision 0) +++ Alarms/Alarms_en.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,1823 @@ +[0] +Title = +ListTitle = +Message = + +[1] +Title = Stuck Button +ListTitle = HD POST: Stuck Button +Message = A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. + +[2] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: FPGA Compatibility +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[3] +Title = Service Required: Dialysate Device +ListTitle = DG POST: FPGA Compatibility +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 602 +SRSDG 814 +SRSDG 1061 = + +[4] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: WD Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[5] +Title = Service Required: Dialysate Device +ListTitle = DG POST: WD Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 53 = + +[6] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Proc- UI Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[7] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[8] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Acc Vector +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[9] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acc Vector +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 323 = + +[10] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Config Emb Mode +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[11] +Title = Low Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp Low +Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. + +[12] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[13] +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG Image +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 813 = + +[14] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Usage Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 957 = + +[15] +Title = High Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. + +[16] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Primary Alarm Current +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[17] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI POST Not Complete +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[18] +Title = Communication Error +ListTitle = Loss of DG Comm (DG ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[19] +Title = High Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. + +[20] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Conductivity Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 16 = + +[21] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Drain Line Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 936 = + +[22] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Reservoir Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 935 = + +[23] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acid Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 934 = + +[24] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Bicarb Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 933 = + +[25] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 817 = + +[26] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Acc Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[27] +Title = Self Test: Blood Leak +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. + +[28] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Conductivity RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 864 = + +[29] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: SP DAC Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[30] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: SW Invalid State +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[31] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[32] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[33] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[34] +Title = Blood Pump Error +ListTitle = HD Fault: BP Rotor +Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. + +[35] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[36] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[37] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[38] +Title = Dialysate Pump Error +ListTitle = HD Fault: DPi Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. + +[39] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[40] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[41] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[42] +Title = Dialysate Pump Error +ListTitle = HD Fault: DPo Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. + +[43] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Iso ADC Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 954 = + +[44] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UI Comm +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[45] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: CAN CRC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[46] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: CAN ACK UI Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[47] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Pump High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[48] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UF Rate Tare Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[49] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: FPGA Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[50] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Valve Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 26 = + +[51] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[52] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[53] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[54] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Critical Data Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[55] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Critical Data Err +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 921 = + +[56] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Acc Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[57] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Acc Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 869 +SRSDG 870 = + +[58] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Home +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[59] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Transition Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[60] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Control +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[61] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[62] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Position +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[63] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Pre- Tx Hep. Pump Timeout. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[64] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 874 = + +[65] +Title = Communication Error +ListTitle = Loss of DG Comm (Reject) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[66] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Config Set Point +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[67] +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK + +[68] +Title = Dialysate Generation Error +ListTitle = Dial. Gen: Dialysate not ready +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. + +[69] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[70] +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK + +[71] +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK + +[72] +Title = Communication Error +ListTitle = Loss of DG Comm (LC) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[73] +Title = Communication Error +ListTitle = Loss of DG Comm (TDI/TRO) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[74] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVi On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 600 = + +[75] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe DAC +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[76] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Voltage Monitor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[77] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Voltage Monitor +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 46 +SRSDG 866 +SRSDG 1088 = + +[78] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Enc Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[79] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Drv Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[80] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Driver +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[81] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump OT +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[82] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump DAC Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[83] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump On Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[84] +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG RTC Time +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 815 = + +[85] +Title = Cartridge Lever is Open +ListTitle = Blood Stop: Cartridge Lever Open +Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. + +[86] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD RTC Time +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[87] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Heat Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 929 = + +[88] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1 CP2 Speed Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[89] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVo On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 599 = + +[90] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC A1/B1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 604 = + +[91] +Title = Communication Error +ListTitle = Loss of DG Comm (FMD) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[92] +Title = +ListTitle = +Message = + +[93] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Permeate Cond High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 430 = + +[94] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Pressure Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 629 = + +[95] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: WD Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 907 = + +[96] +Title = Incoming Water Temperature High +ListTitle = Dial. Gen: Inlet Water Temp High +Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 410 = + +[97] +Title = Conductivity Out of Range +ListTitle = Dial. Gen: Dialysate Cond. +Message = Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +SRSDG 397 = + +[98] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Battery Com +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[99] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Stall +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[100] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Conductivity Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[101] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI- Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[102] +Title = Disinfection Interrupted +ListTitle = Cleaning: Tpo THD Temp Delta +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 856 +SRSDG 930 = + +[103] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Cal NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[104] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Air Trap Sensor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[105] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Cal NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 608 = + +[106] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Flow Sensor Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 820 +SRSDG 821 +SRSDG 1087 = + +[107] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DG Restarted +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[108] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump ADC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[109] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Vol +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[110] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[111] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[112] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC A2/B2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 953 = + +[113] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Current +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 890 = + +[114] +Title = Self Test: Venous Air Detector +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. + +[115] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Temp Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 224 = + +[116] +Title = Service Required: Dialysate Device +ListTitle = HD Alarm (Low): SD- Card Error +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[117] +Title = Communication Error +ListTitle = Loss of HD Comm (HD ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 909 = + +[118] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RTC Config +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 43 = + +[119] +Title = Service Required: Dialysate Device +ListTitle = DG Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. +SRSDG 908 = + +[120] +Title = Self Test: Venous Air Detector +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. + +[121] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Prim Heater Lg Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 611 = + +[122] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Prim Heater Sm Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 612 = + +[123] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Trimmer Heater Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 613 = + +[124] +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. + +[125] +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. + +[126] +Title = Blood Leak +ListTitle = Blood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. + +[127] +Title = Low Venous Pressure +ListTitle = Blood Stop: Venous Pres. Low +Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. + +[128] +Title = Venous Air Detected +ListTitle = Blood Stop: Venous Air Bubble +Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. + +[129] +Title = Blood Leak: Bypass +ListTitle = Blood Stop: Dialyzer Blood Leak Bypass +Message = The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. + +[130] +Title = High Venous Pressure +ListTitle = Blood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. + +[131] +Title = Low Arterial Pressure +ListTitle = Blood Stop: Arterial Pres. Low +Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. + +[132] +Title = High Arterial Pressure +ListTitle = Blood Stop: Arterial Pres. High +Message = Arterial pressure is high.\n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. + +[133] +Title = Dialysate Device Fluid Leak +ListTitle = DG Fault: Fluid Leak in DG +Message = Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 328 = + +[134] +Title = Fluid Leak +ListTitle = Blood Stop: Fluid Leak in HD +Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. + +[135] +Title = Blood Pump Stopped: Excessive Movement +ListTitle = Blood Stop: HD Shock Detected +Message = Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. + +[136] +Title = Dialysate Device: Excessive Movement +ListTitle = Dial. Gen: Shock- Vibration +Message = Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. +SRSDG 327 = + +[137] +Title = Blood Pump Stopped: Excessive Tilt +ListTitle = Blood Stop: HD Tilted +Message = Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. + +[138] +Title = Dialysate Device: Excessive Tilt +ListTitle = Dial. Gen: Tilt +Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. +SRSDG 326 = + +[139] +Title = Power Loss +ListTitle = Loss of AC Power +Message = AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. + +[140] +Title = Communication Error +ListTitle = Loss of DG Comm (Timeout) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[141] +Title = Blood Pump Stopped: Venous Air Detector +ListTitle = Blood Stop: Air Trap Fill +Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. + +[142] +Title = Blood Pump Stopped: Occlusion +ListTitle = Blood Stop: Blood Pump Occlusion +Message = Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. + +[143] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TRo TDi Delta +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 912 = + +[144] +Title = Disinfection: Water Conductivity High +ListTitle = Cleaning: Inlet Water Cond High +Message = Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1066 +SRSDG 1054 +SRSDG 970 = + +[145] +Title = Conductivity Error During Prime +ListTitle = Dial. Gen: Acid Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +SRSDG 846 = + +[146] +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG RTC Timer +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 312 = + +[147] +Title = Dialysate Generation Error +ListTitle = Dial. Gen: Dialysate not ready +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. +SRSDG 398 = + +[148] +Title = Disinfection: Water Temperature High +ListTitle = Cleaning: Inlet Water Temp High +Message = Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1065 +SRSDG 1053 +SRSDG 969 = + +[149] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: WD Error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[150] +Title = Incoming Water Conductivity High +ListTitle = Dial. Gen: Inlet Water Cond High +Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 391 = + +[151] +Title = Incoming Water Conductivity Low +ListTitle = Dial. Gen: Inlet Water Cond Low +Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 826 = + +[152] +Title = Incoming Water Pressure Low +ListTitle = Dial. Gen: Inlet Water Pres Low +Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 417 = + +[153] +Title = Pre-Treatment Process Terminated +ListTitle = Process: Pre- Tx Terminated +Message = The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. + +[154] +Title = Disinfection: Water Conductivity Low +ListTitle = Cleaning: Inlet Water Cond Low +Message = Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1067 +SRSDG 1055 +SRSDG 971 = + +[155] +Title = Communication Error +ListTitle = Loss of DG Comm (Reservoir) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[156] +Title = Communication Error +ListTitle = Loss of DG Comm (Op Mode) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[157] +Title = Chemical Disinfectant Not Detected +ListTitle = Cleaning: Chem Disinf PAA Prime +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 988 = + +[158] +Title = Incoming Water Temperature Low +ListTitle = Dial. Gen: Inlet Water Temp Low +Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 408 = + +[159] +Title = Remove Chemical Disinfectant +ListTitle = Cleaning: Chem Disinf Connect Cap +Message = The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. +SRSDG 1044 = + +[160] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Rotor Max Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[161] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HD- DG Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 122 = + +[162] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Proc- FPGA Clock +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 867 = + +[163] +Title = Ultrafiltration Error +ListTitle = Dial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. + +[164] +Title = Ultrafiltration Error +ListTitle = Dial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. + +[165] +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. + +[166] +Title = Test Alarm - Alarm 166 +ListTitle = +Message = + +[167] +Title = Blood Pump Stopped: Door Open +ListTitle = Blood Stop: Door Open +Message = The device door is open.\n Close the device door. \n- Press Resume/OK to continue. + +[168] +Title = Test Alarm - Alarm 168 +ListTitle = +Message = + +[169] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Dialysate Flow Sens Rg +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 643 = + +[170] +Title = Heparin Syringe Empty +ListTitle = Heparin Syringe Empty +Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. + +[171] +Title = Heparin Pump Occlusion +ListTitle = Process: Pre- Tx Hep. Pump Occ. +Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. + +[172] +Title = Not Enough Heparin +ListTitle = Process: Pre- Tx Not Enough Hep. +Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. + +[173] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: RTC Config +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[174] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD RTC Timer +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[175] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Pump Commutation Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[176] +Title = Test Alarm - Alarm 176 +ListTitle = +Message = + +[177] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Software Invalid State +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 919 = + +[178] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UI- Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1066 + +[179] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: FPGA Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 605 = + +[180] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 883 = + +[181] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Tare Out of Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 849 +SRSDG 838 = + +[182] +Title = Service Required: Dialysate Device +ListTitle = DG POST: LC Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 162 = + +[183] +Title = Disinfection Interrupted +ListTitle = Cleaning: Invalid Resv Transfer +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 851 = + +[184] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVi Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 559 = + +[185] +Title = Service Required: Dialysate Device +ListTitle = DG Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. +SRSDG 628 = + +[186] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1/CP2 Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 891 = + +[187] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 905 = + +[188] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP2 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 906 = + +[189] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 448 = + +[190] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Speed Not Off +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[191] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 885 = + +[192] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow Low +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 945 = + +[193] +Title = Blood Pump Stopped: Partial Occlusion +ListTitle = Blood Stop: Blood Pump Partial Occlusion +Message = Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment. + +[194] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Driver Feedback +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 881 = + +[195] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Pump High Pressure +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 925 = + +[196] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPi/CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 955 = + +[197] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD1/CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 956 = + +[198] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HP On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 872 = + +[199] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HR On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 871 = + +[200] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Thermistor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 178 = + +[201] +Title = Self Test: Dialysate Flow Rate +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. + +[202] +Title = Self Test: Pressure +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. + +[203] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Fill Cond Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[204] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Rejection +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 429 = + +[205] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Cond Sensors Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 913 = + +[206] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Fill Timeout +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 437 = + +[207] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: FMP Flow Rate Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 240 = + +[208] +Title = High Venous Pressure +ListTitle = Blood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. + +[209] +Title = Disinfection Interrupted +ListTitle = Cleaning: Flush Circ Error +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 980 = + +[210] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Battery Fault +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[211] +Title = Approaching MAX Blood Pump Stop Time +ListTitle = Blood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. + +[212] +Title = Approaching MAX Blood Pump Stop Time +ListTitle = Blood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. + +[213] +Title = Prime Complete - 30 MIN WARNING +ListTitle = Process: Pre- Tx Prime Complete +Message = The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. + +[214] +Title = +ListTitle = +Message = + +[215] +Title = +ListTitle = +Message = + +[216] +Title = Treatment Complete +ListTitle = Process: End of Dialysis Tx +Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. + +[217] +Title = +ListTitle = +Message = + +[218] +Title = Syringe Detected +ListTitle = Process: Remove Syringe +Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. + +[219] +Title = Blood Stopped: Syringe Not Detected +ListTitle = Blood Stop: Syringe Not Detected +Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. + +[220] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: THD RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 947 = + +[221] +Title = Blood Pump Stopped: Saline Bag Empty +ListTitle = Blood Stop: Saline Bag Empty +Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. + +[222] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Occ Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[223] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Art. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[224] +Title = Blood Pump Stopped: Treatment Paused +ListTitle = Blood Stopped: Treatment Pause +Message = Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. + +[225] +Title = Treatment Complete +ListTitle = Process: End of Treatment +Message = Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. + +[226] +Title = Priming Complete +ListTitle = Process: Pre- Tx Priming Complete +Message = Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. + +[227] +Title = +ListTitle = +Message = + +[228] +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. + +[229] +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. + +[230] +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. + +[231] +Title = Self Test: Cartridge Check +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. + +[232] +Title = Self Test: Dialysate Flow Rate +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. + +[233] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Cal CRC +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 964 +SRSDG 949 +SRSDG 1089 = + +[234] +Title = Incoming Water Pressure High +ListTitle = Dial. Gen: Inlet Water Pres High +Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 1037 = + +[235] +Title = Rinseback Complete +ListTitle = Process: Rinseback Complete +Message = Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. + +[236] +Title = Self Test: Fluid Present +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. + +[237] +Title = +ListTitle = +Message = + +[238] +Title = Cartridge Not Detected +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. + +[239] +Title = Cartridge Still Detected +ListTitle = Process: Post- Tx Remove Cartridge +Message = A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. + +[240] +Title = Conductivity Error During Prime +ListTitle = Dial. Gen: Bicarb Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +SRSDG 482 = + +[241] +Title = Disinfection Interrupted +ListTitle = Cleaning: Resv Drain Timeout +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 848 +SRSDG 90 +SRSDG 561 +SRSDG 577 +SRSDG 590 +SRSDG 591 +SRSDG 336 +SRSDG 879 +SRSDG 340 +SRSDG 496 +SRSDG 501 +SRSDG 505 +SRSDG 532 +SRSDG 535 +SRSDG 556 +SRSDG 557 +SRSDG 1048 +SRSDG 1049 +SRSDG 651 +SRSDG 655 +SRSDG 744 +SRSDG 745 +SRSDG 1004 +SRSDG 1008 +SRSDG 1019 +SRSDG 1023 +SRSDG 1033 +SRSDG 1034 = + +[242] +Title = Disinfection Interrupted +ListTitle = Cleaning: Resv Fill Timeout +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 850 +SRSDG 573 +SRSDG 574 +SRSDG 579 +SRSDG 580 +SRSDG 492 +SRSDG 493 +SRSDG 498 +SRSDG 499 +SRSDG 508 +SRSDG 509 +SRSDG 519 +SRSDG 690 +SRSDG 691 +SRSDG 703 +SRSDG 711 +SRSDG 1020 +SRSDG 1024 = + +[243] +Title = Disinfection Interrupted +ListTitle = Cleaning: Fluid Loss Detected +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 853 = + +[244] +Title = Disinfection Interrupted +ListTitle = Cleaning: Temp Sensor Check +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 852 = + +[245] +Title = Disinfection Interrupted +ListTitle = Cleaning: Heat Disinfect Temp TO +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 854 = + +[246] +Title = Disinfection Interrupted +ListTitle = Cleaning: RO Water Cond Error +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. + +[247] +Title = +ListTitle = +Message = + +[248] +Title = Disinfection Interrupted +ListTitle = Cleaning: Chem Disinfect Proc +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. + +[249] +Title = Chemical Disinfectant Not Detected +ListTitle = Cleaning: Chem Disinf PAA Cond +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 992 = + +[250] +Title = Connect Chemical Disinfectant +ListTitle = Cleaning: Chem Disinf Connect PAA +Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. +SRSDG 1046 = + +[251] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Sys Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[252] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Serv Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[253] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Sys Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 609 = + +[254] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Serv Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 610 = + +[255] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Compatibility +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[256] +Title = Chemical Flush in Progress +ListTitle = Cleaning: Chem Flush In Progress +Message = Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + = SRSUI 497 + +[257] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Institutional Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[258] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: OS Version +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[259] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Temp Sensors Error +Message = A problem was detected with the hemodialysis device.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[260] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1067 + +[261] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI CAN Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1068 + +[262] +Title = +ListTitle = +Message = + +[263] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Touch Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1070 + +[264] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI SD- Card +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1071 + +[265] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI RTC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1072 + +[266] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI WiFi +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1073 + +[267] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Bluetooth +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1074 + +[268] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Ethernet +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1075 + +[269] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Sound +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1076 + +[270] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Proc Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[271] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Proc Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 822 = + +[272] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[273] +Title = +ListTitle = +Message = + +[274] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[275] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Arterial Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[276] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Venous Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[277] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Occlusion Sens Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[278] +Title = Dialysate Reservoir Error +ListTitle = Dial. Gen: Reservoir Overfill +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. + +[279] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Drain +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 752 = + +[280] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Direction Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 888 = + +[281] +Title = Acid Concentrate Low +ListTitle = Dial. Gen: Acid Conc. Low +Message = The acid concentrate is low.\n- Replace acid concentrate container. +SRSDG 836 = + +[282] +Title = Bicarb Concentrate Low +ListTitle = Dial. Gen: Bicarb Conc. Low +Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. +SRSDG 837 = + +[283] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 615 = + +[284] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Redundant LC Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 383 = + +[285] +Title = RO Permeate Sample in Progress +ListTitle = Cleaning: RO Permeate Sample In Progress +Message = RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[286] +Title = Concentrate Connection Cap Removed +ListTitle = Cleaning: Conc Cap Removed +Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. + +[287] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Proc- FPGA Clock +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[288] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC Drift Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[289] +Title = Dialysate Connection Cap Removed +ListTitle = Cleaning: Dialysate Cap Removed +Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. +SRSDG 928 = + +[290] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Ven. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[291] +Title = Flush In Progress +ListTitle = Cleaning: Flush In Progress +Message = System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[292] +Title = Heat Disinfection in Progress +ListTitle = Cleaning: Heat Disinf In Progress +Message = Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[293] +Title = Chemical Disinfection in Progress +ListTitle = Cleaning: Chem Disinfect In Progress +Message = Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[294] +Title = Active Cool in Progress +ListTitle = Cleaning: Active Cool In Progress +Message = Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[295] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI CloudSync +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1238 + +[296] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TDi RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 951 = + +[297] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TRo RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 950 = + +[298] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Pres Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 948 = + +[299] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Serial Number Check +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 952 = + +[300] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Serial Number Check +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[301] +Title = Remove Dialysate Cap +ListTitle = Cleaning: RO Permeate Sample Connect Cap +Message = Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. + +[302] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Processor RAM error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 126 = + +[303] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Hall Commutation +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 958 = + +[304] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Usage Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[305] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Processor RAM error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[306] +Title = Turn Off Inlet Water Valves +ListTitle = Cleaning: Turn Off Inlet Water Valves +Message = Heat disinfection is in progress.\n- Turn off inlet water valves. + +[307] +Title = Power Loss +ListTitle = Blood Stop: Loss of AC Power +Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. + +[308] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPi Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 959 = + +[309] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 960 = + +[310] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 962 = + +[311] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 961 = + +[312] +Title = Low Dialysate Flow +ListTitle = Dial. Gen: Dialysate In Flow +Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. + +[313] +Title = Disinfection: Water Temperature Low +ListTitle = Cleaning: Inlet Water Temp Low +Message = Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1064 +SRSDG 1052 +SRSDG 968 = + +[314] +Title = Disinfection: Water Pressure High +ListTitle = Cleaning: Inlet Water Press High +Message = Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1063 +SRSDG 1051 +SRSDG 967 = + +[315] +Title = Disinfection: Water Pressure Low +ListTitle = Cleaning: Inlet Water Press Low +Message = Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water pressure\n before initiating disinfection again. +SRSDG 1068 +SRSDG 1056 +SRSDG 966 = + +[316] +Title = +ListTitle = +Message = +SRSDG 983 = + +[317] +Title = +ListTitle = +Message = +SRSDG 984 = + +[318] +Title = +ListTitle = +Message = + +[319] +Title = Residual Chemical Sampling +ListTitle = Cleaning: Chem Residual Sampling +Message = A water sample is needed to confirm the \nabsence of residual chemical following disinfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. +SRSDG 1040 = + +[320] +Title = Chemical Residual Timeout +ListTitle = Cleaning: Chem Residual Timeout +Message = A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. +SRSDG 1043 = + +[321] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVo Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 1069 = + +[322] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST Alarm: UI Invalid Year +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[323] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST Alarm: UI Bad Settings +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + Index: Alarms/Alarms_es.conf =================================================================== diff -u --- Alarms/Alarms_es.conf (revision 0) +++ Alarms/Alarms_es.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,1823 @@ +[0] +Title = +ListTitle = +Message = + +[1] +Title = Ñtuck Button +ListTitle = ÑD POST: Stuck Button +Message = Ñ button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. + +[2] +Title = Ñervice Required: Hemodialysis Device +ListTitle = ÑD POST: FPGA Compatibility +Message = Ñ problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[3] +Title = Service Required: Dialysate Device +ListTitle = DG POST: FPGA Compatibility +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 602 +SRSDG 814 +SRSDG 1061 = + +[4] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: WD Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[5] +Title = Service Required: Dialysate Device +ListTitle = DG POST: WD Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 53 = + +[6] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Proc- UI Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[7] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[8] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Acc Vector +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[9] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acc Vector +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 323 = + +[10] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Config Emb Mode +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[11] +Title = Low Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp Low +Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. + +[12] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[13] +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG Image +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 813 = + +[14] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Usage Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 957 = + +[15] +Title = High Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. + +[16] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Primary Alarm Current +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[17] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI POST Not Complete +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[18] +Title = Communication Error +ListTitle = Loss of DG Comm (DG ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[19] +Title = High Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. + +[20] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Conductivity Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 16 = + +[21] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Drain Line Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 936 = + +[22] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Reservoir Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 935 = + +[23] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acid Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 934 = + +[24] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Bicarb Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 933 = + +[25] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 817 = + +[26] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Acc Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[27] +Title = Self Test: Blood Leak +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. + +[28] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Conductivity RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 864 = + +[29] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: SP DAC Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[30] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: SW Invalid State +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[31] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[32] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[33] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[34] +Title = Blood Pump Error +ListTitle = HD Fault: BP Rotor +Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. + +[35] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[36] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[37] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[38] +Title = Dialysate Pump Error +ListTitle = HD Fault: DPi Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. + +[39] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[40] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[41] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[42] +Title = Dialysate Pump Error +ListTitle = HD Fault: DPo Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. + +[43] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Iso ADC Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 954 = + +[44] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UI Comm +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[45] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: CAN CRC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[46] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: CAN ACK UI Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[47] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Pump High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[48] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UF Rate Tare Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[49] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: FPGA Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[50] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Valve Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 26 = + +[51] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[52] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[53] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[54] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Critical Data Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[55] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Critical Data Err +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 921 = + +[56] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Acc Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[57] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Acc Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 869 +SRSDG 870 = + +[58] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Home +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[59] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Transition Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[60] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Control +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[61] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[62] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Position +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[63] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Pre- Tx Hep. Pump Timeout. +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[64] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 874 = + +[65] +Title = Communication Error +ListTitle = Loss of DG Comm (Reject) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[66] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Config Set Point +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[67] +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK + +[68] +Title = Dialysate Generation Error +ListTitle = Dial. Gen: Dialysate not ready +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. + +[69] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[70] +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK + +[71] +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK + +[72] +Title = Communication Error +ListTitle = Loss of DG Comm (LC) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[73] +Title = Communication Error +ListTitle = Loss of DG Comm (TDI/TRO) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[74] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVi On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 600 = + +[75] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe DAC +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[76] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Voltage Monitor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[77] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Voltage Monitor +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 46 +SRSDG 866 +SRSDG 1088 = + +[78] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Enc Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[79] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Drv Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[80] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Driver +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[81] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump OT +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[82] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump DAC Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[83] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump On Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[84] +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG RTC Time +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 815 = + +[85] +Title = Cartridge Lever is Open +ListTitle = Blood Stop: Cartridge Lever Open +Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. + +[86] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD RTC Time +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[87] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Heat Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 929 = + +[88] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1 CP2 Speed Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[89] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVo On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 599 = + +[90] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC A1/B1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 604 = + +[91] +Title = Communication Error +ListTitle = Loss of DG Comm (FMD) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[92] +Title = +ListTitle = +Message = + +[93] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Permeate Cond High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 430 = + +[94] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Pressure Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 629 = + +[95] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: WD Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 907 = + +[96] +Title = Incoming Water Temperature High +ListTitle = Dial. Gen: Inlet Water Temp High +Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 410 = + +[97] +Title = Conductivity Out of Range +ListTitle = Dial. Gen: Dialysate Cond. +Message = Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +SRSDG 397 = + +[98] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Battery Com +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[99] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Stall +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[100] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Conductivity Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[101] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI- Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[102] +Title = Disinfection Interrupted +ListTitle = Cleaning: Tpo THD Temp Delta +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 856 +SRSDG 930 = + +[103] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Cal NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[104] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Air Trap Sensor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[105] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Cal NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 608 = + +[106] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Flow Sensor Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 820 +SRSDG 821 +SRSDG 1087 = + +[107] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DG Restarted +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[108] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump ADC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[109] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Vol +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[110] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[111] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[112] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC A2/B2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 953 = + +[113] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Current +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 890 = + +[114] +Title = Self Test: Venous Air Detector +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. + +[115] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Temp Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 224 = + +[116] +Title = Service Required: Dialysate Device +ListTitle = HD Alarm (Low): SD- Card Error +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[117] +Title = Communication Error +ListTitle = Loss of HD Comm (HD ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 909 = + +[118] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RTC Config +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 43 = + +[119] +Title = Service Required: Dialysate Device +ListTitle = DG Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. +SRSDG 908 = + +[120] +Title = Self Test: Venous Air Detector +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. + +[121] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Prim Heater Lg Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 611 = + +[122] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Prim Heater Sm Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 612 = + +[123] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Trimmer Heater Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 613 = + +[124] +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. + +[125] +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. + +[126] +Title = Blood Leak +ListTitle = Blood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. + +[127] +Title = Low Venous Pressure +ListTitle = Blood Stop: Venous Pres. Low +Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- Check needle placement and resolve access issues, \nif applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment.\n- Press Resume/OK to continue treatment. + +[128] +Title = Venous Air Detected +ListTitle = Blood Stop: Venous Air Bubble +Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. + +[129] +Title = Blood Leak: Bypass +ListTitle = Blood Stop: Dialyzer Blood Leak Bypass +Message = The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. + +[130] +Title = High Venous Pressure +ListTitle = Blood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- Check air trap for foam and that transducers are secured.\n- Inspect transducer for moisture and change, if applicable.\n- If clots are present, end treatment without rinseback.\n- Press Resume/OK to continue treatment. + +[131] +Title = Low Arterial Pressure +ListTitle = Blood Stop: Arterial Pres. Low +Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. + +[132] +Title = High Arterial Pressure +ListTitle = Blood Stop: Arterial Pres. High +Message = Arterial pressure is high.\n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- Check to ensure transducers are securely connected.\n- Inspect transducer for moisture and change if applicable.\n- If clots are present, end treatment without rinseback. + +[133] +Title = Dialysate Device Fluid Leak +ListTitle = DG Fault: Fluid Leak in DG +Message = Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 328 = + +[134] +Title = Fluid Leak +ListTitle = Blood Stop: Fluid Leak in HD +Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. + +[135] +Title = Blood Pump Stopped: Excessive Movement +ListTitle = Blood Stop: HD Shock Detected +Message = Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. + +[136] +Title = Dialysate Device: Excessive Movement +ListTitle = Dial. Gen: Shock- Vibration +Message = Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. +SRSDG 327 = + +[137] +Title = Blood Pump Stopped: Excessive Tilt +ListTitle = Blood Stop: HD Tilted +Message = Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. + +[138] +Title = Dialysate Device: Excessive Tilt +ListTitle = Dial. Gen: Tilt +Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. +SRSDG 326 = + +[139] +Title = Power Loss +ListTitle = Loss of AC Power +Message = AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. + +[140] +Title = Communication Error +ListTitle = Loss of DG Comm (Timeout) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[141] +Title = Blood Pump Stopped: Venous Air Detector +ListTitle = Blood Stop: Air Trap Fill +Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. + +[142] +Title = Blood Pump Stopped: Occlusion +ListTitle = Blood Stop: Blood Pump Occlusion +Message = Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. + +[143] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TRo TDi Delta +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 912 = + +[144] +Title = Disinfection: Water Conductivity High +ListTitle = Cleaning: Inlet Water Cond High +Message = Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1066 +SRSDG 1054 +SRSDG 970 = + +[145] +Title = Conductivity Error During Prime +ListTitle = Dial. Gen: Acid Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +SRSDG 846 = + +[146] +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG RTC Timer +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 312 = + +[147] +Title = Dialysate Generation Error +ListTitle = Dial. Gen: Dialysate not ready +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. +SRSDG 398 = + +[148] +Title = Disinfection: Water Temperature High +ListTitle = Cleaning: Inlet Water Temp High +Message = Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1065 +SRSDG 1053 +SRSDG 969 = + +[149] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: WD Error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[150] +Title = Incoming Water Conductivity High +ListTitle = Dial. Gen: Inlet Water Cond High +Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 391 = + +[151] +Title = Incoming Water Conductivity Low +ListTitle = Dial. Gen: Inlet Water Cond Low +Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 826 = + +[152] +Title = Incoming Water Pressure Low +ListTitle = Dial. Gen: Inlet Water Pres Low +Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 417 = + +[153] +Title = Pre-Treatment Process Terminated +ListTitle = Process: Pre- Tx Terminated +Message = The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. + +[154] +Title = Disinfection: Water Conductivity Low +ListTitle = Cleaning: Inlet Water Cond Low +Message = Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1067 +SRSDG 1055 +SRSDG 971 = + +[155] +Title = Communication Error +ListTitle = Loss of DG Comm (Reservoir) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[156] +Title = Communication Error +ListTitle = Loss of DG Comm (Op Mode) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[157] +Title = Chemical Disinfectant Not Detected +ListTitle = Cleaning: Chem Disinf PAA Prime +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 988 = + +[158] +Title = Incoming Water Temperature Low +ListTitle = Dial. Gen: Inlet Water Temp Low +Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 408 = + +[159] +Title = Remove Chemical Disinfectant +ListTitle = Cleaning: Chem Disinf Connect Cap +Message = The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. +SRSDG 1044 = + +[160] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Rotor Max Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[161] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HD- DG Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 122 = + +[162] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Proc- FPGA Clock +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 867 = + +[163] +Title = Ultrafiltration Error +ListTitle = Dial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. + +[164] +Title = Ultrafiltration Error +ListTitle = Dial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. + +[165] +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. + +[166] +Title = Test Alarm - Alarm 166 +ListTitle = +Message = + +[167] +Title = Blood Pump Stopped: Door Open +ListTitle = Blood Stop: Door Open +Message = The device door is open.\n Close the device door. \n- Press Resume/OK to continue. + +[168] +Title = Test Alarm - Alarm 168 +ListTitle = +Message = + +[169] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Dialysate Flow Sens Rg +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 643 = + +[170] +Title = Heparin Syringe Empty +ListTitle = Heparin Syringe Empty +Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. + +[171] +Title = Heparin Pump Occlusion +ListTitle = Process: Pre- Tx Hep. Pump Occ. +Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. + +[172] +Title = Not Enough Heparin +ListTitle = Process: Pre- Tx Not Enough Hep. +Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. + +[173] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: RTC Config +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[174] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD RTC Timer +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[175] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Pump Commutation Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[176] +Title = Test Alarm - Alarm 176 +ListTitle = +Message = + +[177] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Software Invalid State +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 919 = + +[178] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UI- Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1066 + +[179] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: FPGA Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 605 = + +[180] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 883 = + +[181] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Tare Out of Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 849 +SRSDG 838 = + +[182] +Title = Service Required: Dialysate Device +ListTitle = DG POST: LC Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 162 = + +[183] +Title = Disinfection Interrupted +ListTitle = Cleaning: Invalid Resv Transfer +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 851 = + +[184] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVi Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 559 = + +[185] +Title = Service Required: Dialysate Device +ListTitle = DG Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. +SRSDG 628 = + +[186] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1/CP2 Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 891 = + +[187] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 905 = + +[188] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP2 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 906 = + +[189] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 448 = + +[190] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Speed Not Off +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[191] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 885 = + +[192] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow Low +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 945 = + +[193] +Title = Blood Pump Stopped: Partial Occlusion +ListTitle = Blood Stop: Blood Pump Partial Occlusion +Message = Partial occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- Inspect dialyzer for clots. If clots are present, follow\n facility policy to end treatment. + +[194] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Driver Feedback +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 881 = + +[195] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Pump High Pressure +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 925 = + +[196] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPi/CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 955 = + +[197] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD1/CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 956 = + +[198] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HP On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 872 = + +[199] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HR On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 871 = + +[200] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Thermistor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 178 = + +[201] +Title = Self Test: Dialysate Flow Rate +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. + +[202] +Title = Self Test: Pressure +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. + +[203] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Fill Cond Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[204] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Rejection +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 429 = + +[205] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Cond Sensors Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 913 = + +[206] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Fill Timeout +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 437 = + +[207] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: FMP Flow Rate Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 240 = + +[208] +Title = High Venous Pressure +ListTitle = Blood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. + +[209] +Title = Disinfection Interrupted +ListTitle = Cleaning: Flush Circ Error +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 980 = + +[210] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Battery Fault +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[211] +Title = Approaching MAX Blood Pump Stop Time +ListTitle = Blood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. + +[212] +Title = Approaching MAX Blood Pump Stop Time +ListTitle = Blood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. + +[213] +Title = Prime Complete - 30 MIN WARNING +ListTitle = Process: Pre- Tx Prime Complete +Message = The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. + +[214] +Title = +ListTitle = +Message = + +[215] +Title = +ListTitle = +Message = + +[216] +Title = Treatment Complete +ListTitle = Process: End of Dialysis Tx +Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. + +[217] +Title = +ListTitle = +Message = + +[218] +Title = Syringe Detected +ListTitle = Process: Remove Syringe +Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. + +[219] +Title = Blood Stopped: Syringe Not Detected +ListTitle = Blood Stop: Syringe Not Detected +Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. + +[220] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: THD RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 947 = + +[221] +Title = Blood Pump Stopped: Saline Bag Empty +ListTitle = Blood Stop: Saline Bag Empty +Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. + +[222] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Occ Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[223] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Art. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[224] +Title = Blood Pump Stopped: Treatment Paused +ListTitle = Blood Stopped: Treatment Pause +Message = Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. + +[225] +Title = Treatment Complete +ListTitle = Process: End of Treatment +Message = Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. + +[226] +Title = Priming Complete +ListTitle = Process: Pre- Tx Priming Complete +Message = Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. + +[227] +Title = +ListTitle = +Message = + +[228] +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. + +[229] +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. + +[230] +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. + +[231] +Title = Self Test: Cartridge Check +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. + +[232] +Title = Self Test: Dialysate Flow Rate +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. + +[233] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Cal CRC +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 964 +SRSDG 949 +SRSDG 1089 = + +[234] +Title = Incoming Water Pressure High +ListTitle = Dial. Gen: Inlet Water Pres High +Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. +SRSDG 1037 = + +[235] +Title = Rinseback Complete +ListTitle = Process: Rinseback Complete +Message = Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. + +[236] +Title = Self Test: Fluid Present +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. + +[237] +Title = +ListTitle = +Message = + +[238] +Title = Cartridge Not Detected +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. + +[239] +Title = Cartridge Still Detected +ListTitle = Process: Post- Tx Remove Cartridge +Message = A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. + +[240] +Title = Conductivity Error During Prime +ListTitle = Dial. Gen: Bicarb Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. +SRSDG 482 = + +[241] +Title = Disinfection Interrupted +ListTitle = Cleaning: Resv Drain Timeout +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 848 +SRSDG 90 +SRSDG 561 +SRSDG 577 +SRSDG 590 +SRSDG 591 +SRSDG 336 +SRSDG 879 +SRSDG 340 +SRSDG 496 +SRSDG 501 +SRSDG 505 +SRSDG 532 +SRSDG 535 +SRSDG 556 +SRSDG 557 +SRSDG 1048 +SRSDG 1049 +SRSDG 651 +SRSDG 655 +SRSDG 744 +SRSDG 745 +SRSDG 1004 +SRSDG 1008 +SRSDG 1019 +SRSDG 1023 +SRSDG 1033 +SRSDG 1034 = + +[242] +Title = Disinfection Interrupted +ListTitle = Cleaning: Resv Fill Timeout +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 850 +SRSDG 573 +SRSDG 574 +SRSDG 579 +SRSDG 580 +SRSDG 492 +SRSDG 493 +SRSDG 498 +SRSDG 499 +SRSDG 508 +SRSDG 509 +SRSDG 519 +SRSDG 690 +SRSDG 691 +SRSDG 703 +SRSDG 711 +SRSDG 1020 +SRSDG 1024 = + +[243] +Title = Disinfection Interrupted +ListTitle = Cleaning: Fluid Loss Detected +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 853 = + +[244] +Title = Disinfection Interrupted +ListTitle = Cleaning: Temp Sensor Check +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 852 = + +[245] +Title = Disinfection Interrupted +ListTitle = Cleaning: Heat Disinfect Temp TO +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 854 = + +[246] +Title = Disinfection Interrupted +ListTitle = Cleaning: RO Water Cond Error +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. + +[247] +Title = +ListTitle = +Message = + +[248] +Title = Disinfection Interrupted +ListTitle = Cleaning: Chem Disinfect Proc +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. + +[249] +Title = Chemical Disinfectant Not Detected +ListTitle = Cleaning: Chem Disinf PAA Cond +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. +SRSDG 992 = + +[250] +Title = Connect Chemical Disinfectant +ListTitle = Cleaning: Chem Disinf Connect PAA +Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. +SRSDG 1046 = + +[251] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Sys Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[252] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Serv Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[253] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Sys Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 609 = + +[254] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Serv Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 610 = + +[255] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Compatibility +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[256] +Title = Chemical Flush in Progress +ListTitle = Cleaning: Chem Flush In Progress +Message = Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + = SRSUI 497 + +[257] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Institutional Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[258] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: OS Version +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[259] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Temp Sensors Error +Message = A problem was detected with the hemodialysis device.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[260] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1067 + +[261] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI CAN Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1068 + +[262] +Title = +ListTitle = +Message = + +[263] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Touch Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1070 + +[264] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI SD- Card +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1071 + +[265] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI RTC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1072 + +[266] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI WiFi +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1073 + +[267] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Bluetooth +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1074 + +[268] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Ethernet +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1075 + +[269] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Sound +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1076 + +[270] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Proc Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[271] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Proc Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 822 = + +[272] +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. + +[273] +Title = +ListTitle = +Message = + +[274] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[275] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Arterial Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[276] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Venous Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[277] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Occlusion Sens Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[278] +Title = Dialysate Reservoir Error +ListTitle = Dial. Gen: Reservoir Overfill +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \ndialysate is ready for use. + +[279] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Drain +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 752 = + +[280] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Direction Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 888 = + +[281] +Title = Acid Concentrate Low +ListTitle = Dial. Gen: Acid Conc. Low +Message = The acid concentrate is low.\n- Replace acid concentrate container. +SRSDG 836 = + +[282] +Title = Bicarb Concentrate Low +ListTitle = Dial. Gen: Bicarb Conc. Low +Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. +SRSDG 837 = + +[283] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 615 = + +[284] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Redundant LC Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 383 = + +[285] +Title = RO Permeate Sample in Progress +ListTitle = Cleaning: RO Permeate Sample In Progress +Message = RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[286] +Title = Concentrate Connection Cap Removed +ListTitle = Cleaning: Conc Cap Removed +Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. + +[287] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Proc- FPGA Clock +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[288] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC Drift Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[289] +Title = Dialysate Connection Cap Removed +ListTitle = Cleaning: Dialysate Cap Removed +Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. +SRSDG 928 = + +[290] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Ven. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[291] +Title = Flush In Progress +ListTitle = Cleaning: Flush In Progress +Message = System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[292] +Title = Heat Disinfection in Progress +ListTitle = Cleaning: Heat Disinf In Progress +Message = Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[293] +Title = Chemical Disinfection in Progress +ListTitle = Cleaning: Chem Disinfect In Progress +Message = Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[294] +Title = Active Cool in Progress +ListTitle = Cleaning: Active Cool In Progress +Message = Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. + +[295] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI CloudSync +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + = SRSUI 1238 + +[296] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TDi RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 951 = + +[297] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TRo RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 950 = + +[298] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Pres Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 948 = + +[299] +Title = Service Required: Dialysate Device +ListTitle = DG POST: Serial Number Check +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 952 = + +[300] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Serial Number Check +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[301] +Title = Remove Dialysate Cap +ListTitle = Cleaning: RO Permeate Sample Connect Cap +Message = Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. + +[302] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Processor RAM error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 126 = + +[303] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Hall Commutation +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 958 = + +[304] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Usage Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[305] +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Processor RAM error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[306] +Title = Turn Off Inlet Water Valves +ListTitle = Cleaning: Turn Off Inlet Water Valves +Message = Heat disinfection is in progress.\n- Turn off inlet water valves. + +[307] +Title = Power Loss +ListTitle = Blood Stop: Loss of AC Power +Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. + +[308] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPi Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 959 = + +[309] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 960 = + +[310] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 962 = + +[311] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 961 = + +[312] +Title = Low Dialysate Flow +ListTitle = Dial. Gen: Dialysate In Flow +Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. + +[313] +Title = Disinfection: Water Temperature Low +ListTitle = Cleaning: Inlet Water Temp Low +Message = Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1064 +SRSDG 1052 +SRSDG 968 = + +[314] +Title = Disinfection: Water Pressure High +ListTitle = Cleaning: Inlet Water Press High +Message = Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. +SRSDG 1063 +SRSDG 1051 +SRSDG 967 = + +[315] +Title = Disinfection: Water Pressure Low +ListTitle = Cleaning: Inlet Water Press Low +Message = Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water pressure\n before initiating disinfection again. +SRSDG 1068 +SRSDG 1056 +SRSDG 966 = + +[316] +Title = +ListTitle = +Message = +SRSDG 983 = + +[317] +Title = +ListTitle = +Message = +SRSDG 984 = + +[318] +Title = +ListTitle = +Message = + +[319] +Title = Residual Chemical Sampling +ListTitle = Cleaning: Chem Residual Sampling +Message = A water sample is needed to confirm the \nabsence of residual chemical following disinfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. +SRSDG 1040 = + +[320] +Title = Chemical Residual Timeout +ListTitle = Cleaning: Chem Residual Timeout +Message = A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. +SRSDG 1043 = + +[321] +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVo Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. +SRSDG 1069 = + +[322] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST Alarm: UI Invalid Year +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + +[323] +Title = Service Required: Hemodialysis Device +ListTitle = HD POST Alarm: UI Bad Settings +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. + Fisheye: Tag 9a2fac022f74292e0b1518c9839a0f5749c4ac9c refers to a dead (removed) revision in file `Alarms/Events.conf'. Fisheye: No comparison available. Pass `N' to diff? Index: Alarms/Events_en.conf =================================================================== diff -u --- Alarms/Events_en.conf (revision 0) +++ Alarms/Events_en.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,66 @@ +[0] +Title=Blood flow rate change event + +[1] +Title=Dialysate flow rate change event + +[2] +Title=Treatment duration change event + +[3] +Title=Ultrafiltration volume change event + +[4] +Title=Ultrafiltration rate change event + +[5] +Title=Ultrafiltration start or resume event + +[6] +Title=Ultrafiltration pause event + +[7] +Title=Arterial pressure limit window change event + +[8] +Title=Venous pressure limit window change event + +[9] +Title=Venous pressure limit asymmetric change event + +[10] +Title=Heparin bolus start event + +[11] +Title=Heparin bolus end event + +[12] +Title=Heparin start or resume event + +[13] +Title=Heparin stop or pause event + +[14] +Title=Saline boluses change event + +[15] +Title=Mid treatment rinse-back event + +[16] +Title=Alarm audio volume changed event + +[17] +Title=Front door switch changed event + +[18] +Title=Pump track switch changed event + +[19] +Title=Syringe detection switch changed event + +[20] +Title=Off button pressed event + +[21] +Title=Stop button pressed event + Index: Alarms/Events_es.conf =================================================================== diff -u --- Alarms/Events_es.conf (revision 0) +++ Alarms/Events_es.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,66 @@ +[0] +Title=Ñlood flow rate change event + +[1] +Title=Ñialysate flow rate change event + +[2] +Title=Ñreatment duration change event + +[3] +Title=Ultrafiltration volume change event + +[4] +Title=Ultrafiltration rate change event + +[5] +Title=Ultrafiltration start or resume event + +[6] +Title=Ultrafiltration pause event + +[7] +Title=Arterial pressure limit window change event + +[8] +Title=Venous pressure limit window change event + +[9] +Title=Venous pressure limit asymmetric change event + +[10] +Title=Heparin bolus start event + +[11] +Title=Heparin bolus end event + +[12] +Title=Heparin start or resume event + +[13] +Title=Heparin stop or pause event + +[14] +Title=Saline boluses change event + +[15] +Title=Mid treatment rinse-back event + +[16] +Title=Alarm audio volume changed event + +[17] +Title=Front door switch changed event + +[18] +Title=Pump track switch changed event + +[19] +Title=Syringe detection switch changed event + +[20] +Title=Off button pressed event + +[21] +Title=Stop button pressed event + Fisheye: Tag 9a2fac022f74292e0b1518c9839a0f5749c4ac9c refers to a dead (removed) revision in file `Configurations/DataList.conf'. Fisheye: No comparison available. Pass `N' to diff? Index: Configurations/DataList_en.conf =================================================================== diff -u --- Configurations/DataList_en.conf (revision 0) +++ Configurations/DataList_en.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,110 @@ +[Water Sample Result] +Fail +Pass + +[Blood Flow Rate Ranges] # U32 - mL/min +Blood_Flow_Rate_Min = +100 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Max = +500 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Res = 25 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Def = 100 # PRS 30, SRSUI 619, 639, 765 + +[Dialysate Flow Rate Ranges] # U32 - mL/min +Dialysate_Flow_Rate_Min = +100 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Max = +600 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Res = 50 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Def = 100 # PRS 345, SRSUI 620, 640, 766 + +[Treatment Duration Ranges] # U32 - min +Treatment_Duration_Min = +60 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Max = +480 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Res = 15 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Def = 60 # PRS 346, SRSUI 569, 767 + +[Heparin Dispensing Rate Ranges] # F32 - mL/hr (0 = OFF) +Heparin_Dispensing_Rate_Min = 0.2 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Max = 1.0 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Res = 0.1 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Def = 0.0 # PRS 347, SRSUI 768 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. + +[Heparin Bolus Volume Ranges] # F32 - mL (0 = OFF) +Heparin_Bolus_Volume_Min = 0.2 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Max = 2.0 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Res = 0.1 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Def = 0.0 # PRS 348, SRSUI 769 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. + +[Heparin Stop Time Ranges] #U32 - min +Heparin_Stop_Time_Min = 0 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Max = 480 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Res = 10 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Def = 0 # PRS 464. SRSUI 771 + +[Saline Bolus Volume Ranges] # U32 - mL +Saline_Bolus_Volume_Min = 100 # PRS 350 +Saline_Bolus_Volume_Max = 300 # PRS 350 +Saline_Bolus_Volume_Res = 100 # PRS 350 +Saline_Bolus_Volume_Def = 100 # PRS 350 + +[Heparin Type Options] # U32 - PRS ???, SRSUI ??? +UFH 1,000 IU/mL + +[Acid Concentrate Options] # U32 - PRS 35, SRSUI 772 +1.0 K 2.5 Ca # (08-1251-1) +2.0 K 2.5 Ca # (08-2251-0) +3.0 K 2.5 Ca # (08-3251-9) + +[Bicarbonate Concentrate Options] # U32 - PRS 351, SRSUI 773 +Fres. Centrisol + +[Dialyzer Type Options] # U32 - PRS 352, SRSUI 774 +BB Diacap Pro 13H +BB Diacap Pro 16H +BB Diacap Pro 19H +F Optiflux F160NRe +F Optiflux F180NRe +F Optiflux F200NRe + +[Dialysate Temperature] # Celsius +Dialysate_Temperature_Min = 35 # PRS 353, SRSUI 775 +Dialysate_Temperature_Max = 37 # PRS 353, SRSUI 775 +Dialysate_Temperature_Res = 0.5 # PRS 353, SRSUI 775 +Dialysate_Temperature_Def = 37 # PRS 353, SRSUI 775 + +[Arterial Blood Pressure Ranges] +# Adjustment, Arterial Window +Arterial_Adjust_Window_Min = +120 # PRS 354, SRSUI 776, SRSUI 726 +Arterial_Adjust_Window_Max = +200 # PRS 354, SRSUI 776, SRSUI 726 +Arterial_Adjust_Window_Res = +40 # PRS 354, SRSUI 776, SRSUI 726 +Arterial_Adjust_Window_Def = +120 # PRS 354, SRSUI 776 +#Monitoring +Arterial_Monitor_Min = -400 # PRS 362, SRSUI 617 +Arterial_Monitor_Max = +100 # PRS 362, SRSUI 617 +Arterial_Monitor_Res = 1 # PRS 362, SRSUI 617 + +[Venous Blood Pressure Ranges] +# Adjustment, Venous Window +Venous_Adjust_Window_Min = +100 # PRS 356, SRSUI 777, SRSUI 731 +Venous_Adjust_Window_Max = +200 # PRS 356, SRSUI 777, SRSUI 731 +Venous_Adjust_Window_Res = +20 # PRS 356, SRSUI 777, SRSUI 731 +Venous_Adjust_Window_Def = +100 # PRS 356, SRSUI 777, SRSUI 731 +# Adjustment, Venous Asymmetric +Venous_Adjust_Asymmetric_Min = +20 # PRS 357, SRSUI 781, SRUSUI 732 +Venous_Adjust_Asymmetric_Max = +35 # PRS 357, SRSUI 781, SRUSUI 732 +Venous_Adjust_Asymmetric_Res = +5 # PRS 357, SRSUI 781, SRUSUI 732 +Venous_Adjust_Asymmetric_Def = +20 # PRS 357, SRSUI 781, SRUSUI 732 +##Monitoring +Venous_Monitor_Min = -100 # PRS 361, SRSUI 616 +Venous_Monitor_Max = +500 # PRS 361, SRSUI 616 +Venous_Monitor_Res = 1 # PRS 361, SRSUI 616 + +[Blood Pressure Measure Interval Ranges] # min (0 = OFF) +Blood_Pressure_Measure_Interval_Min = 5 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Max = 60 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Res = 5 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Def = 0 # PRS 358, SRSUI 778 + +[Rinseback Flow Rate Ranges] # mL/min +Rinseback_Flow_Rate_Min = 100 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Max = 300 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Res = 0 # PRS 359, SRSUI 330, 770 # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. +Rinseback_Flow_Rate_Def = 200 # PRS 359, SRSUI 330, 770 + Index: Configurations/DataList_es.conf =================================================================== diff -u --- Configurations/DataList_es.conf (revision 0) +++ Configurations/DataList_es.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,110 @@ +[Water Sample Result] +Ñail +Ñass + +[Blood Flow Rate Ranges] # U32 - mL/min +Blood_Flow_Rate_Min = +100 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Max = +500 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Res = 25 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Def = 100 # PRS 30, SRSUI 619, 639, 765 + +[Dialysate Flow Rate Ranges] # U32 - mL/min +Dialysate_Flow_Rate_Min = +100 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Max = +600 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Res = 50 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Def = 100 # PRS 345, SRSUI 620, 640, 766 + +[Treatment Duration Ranges] # U32 - min +Treatment_Duration_Min = +60 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Max = +480 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Res = 15 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Def = 60 # PRS 346, SRSUI 569, 767 + +[Heparin Dispensing Rate Ranges] # F32 - mL/hr (0 = OFF) +Heparin_Dispensing_Rate_Min = 0.2 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Max = 1.0 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Res = 0.1 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Def = 0.0 # PRS 347, SRSUI 768 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. + +[Heparin Bolus Volume Ranges] # F32 - mL (0 = OFF) +Heparin_Bolus_Volume_Min = 0.2 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Max = 2.0 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Res = 0.1 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Def = 0.0 # PRS 348, SRSUI 769 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. + +[Heparin Stop Time Ranges] #U32 - min +Heparin_Stop_Time_Min = 0 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Max = 480 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Res = 10 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Def = 0 # PRS 464. SRSUI 771 + +[Saline Bolus Volume Ranges] # U32 - mL +Saline_Bolus_Volume_Min = 100 # PRS 350 +Saline_Bolus_Volume_Max = 300 # PRS 350 +Saline_Bolus_Volume_Res = 100 # PRS 350 +Saline_Bolus_Volume_Def = 100 # PRS 350 + +[Heparin Type Options] # U32 - PRS ???, SRSUI ??? +ÑFH 1,000 IU/mL + +[Acid Concentrate Options] # U32 - PRS 35, SRSUI 772 +Ñ1.0 K 2.5 Ca # (08-1251-1) +Ñ2.0 K 2.5 Ca # (08-2251-0) +Ñ3.0 K 2.5 Ca # (08-3251-9) + +[Bicarbonate Concentrate Options] # U32 - PRS 351, SRSUI 773 +Ñres. Centrisol + +[Dialyzer Type Options] # U32 - PRS 352, SRSUI 774 +ÑB Diacap Pro 13H +ÑB Diacap Pro 16H +ÑB Diacap Pro 19H +Ñ Optiflux F160NRe +Ñ Optiflux F180NRe +Ñ Optiflux F200NRe + +[Dialysate Temperature] # Celsius +Dialysate_Temperature_Min = 35 # PRS 353, SRSUI 775 +Dialysate_Temperature_Max = 37 # PRS 353, SRSUI 775 +Dialysate_Temperature_Res = 0.5 # PRS 353, SRSUI 775 +Dialysate_Temperature_Def = 37 # PRS 353, SRSUI 775 + +[Arterial Blood Pressure Ranges] +# Adjustment, Arterial Window +Arterial_Adjust_Window_Min = +120 # PRS 354, SRSUI 776, SRSUI 726 +Arterial_Adjust_Window_Max = +200 # PRS 354, SRSUI 776, SRSUI 726 +Arterial_Adjust_Window_Res = +40 # PRS 354, SRSUI 776, SRSUI 726 +Arterial_Adjust_Window_Def = +120 # PRS 354, SRSUI 776 +#Monitoring +Arterial_Monitor_Min = -400 # PRS 362, SRSUI 617 +Arterial_Monitor_Max = +100 # PRS 362, SRSUI 617 +Arterial_Monitor_Res = 1 # PRS 362, SRSUI 617 + +[Venous Blood Pressure Ranges] +# Adjustment, Venous Window +Venous_Adjust_Window_Min = +100 # PRS 356, SRSUI 777, SRSUI 731 +Venous_Adjust_Window_Max = +200 # PRS 356, SRSUI 777, SRSUI 731 +Venous_Adjust_Window_Res = +20 # PRS 356, SRSUI 777, SRSUI 731 +Venous_Adjust_Window_Def = +100 # PRS 356, SRSUI 777, SRSUI 731 +# Adjustment, Venous Asymmetric +Venous_Adjust_Asymmetric_Min = +20 # PRS 357, SRSUI 781, SRUSUI 732 +Venous_Adjust_Asymmetric_Max = +35 # PRS 357, SRSUI 781, SRUSUI 732 +Venous_Adjust_Asymmetric_Res = +5 # PRS 357, SRSUI 781, SRUSUI 732 +Venous_Adjust_Asymmetric_Def = +20 # PRS 357, SRSUI 781, SRUSUI 732 +##Monitoring +Venous_Monitor_Min = -100 # PRS 361, SRSUI 616 +Venous_Monitor_Max = +500 # PRS 361, SRSUI 616 +Venous_Monitor_Res = 1 # PRS 361, SRSUI 616 + +[Blood Pressure Measure Interval Ranges] # min (0 = OFF) +Blood_Pressure_Measure_Interval_Min = 5 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Max = 60 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Res = 5 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Def = 0 # PRS 358, SRSUI 778 + +[Rinseback Flow Rate Ranges] # mL/min +Rinseback_Flow_Rate_Min = 100 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Max = 300 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Res = 0 # PRS 359, SRSUI 330, 770 # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. +Rinseback_Flow_Rate_Def = 200 # PRS 359, SRSUI 330, 770 + Fisheye: Tag 9a2fac022f74292e0b1518c9839a0f5749c4ac9c refers to a dead (removed) revision in file `Confirm/Confirm.conf'. Fisheye: No comparison available. Pass `N' to diff? Index: Confirm/Confirm_en.conf =================================================================== diff -u --- Confirm/Confirm_en.conf (revision 0) +++ Confirm/Confirm_en.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,55 @@ +[1] +Title = Water Flush +Message = If stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = CONFIRM +Cancel = CANCEL + +[2] +Title = Heat Disinfect +Message = If stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = CONFIRM +Cancel = CANCEL + +[3] +Title = Chemical Disinfect +Message = If stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = CONFIRM +Cancel = CANCEL + +[4] +Title = Chemical Flush +Message = If stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = CONFIRM +Cancel = CANCEL + +[5] +Title = Chemical Flush Sample Pass/Fail +Message = Please select sample pass/fail status +Confirm = PASS +Cancel = FAIL + +[6] +Title = Treatment End +Message = Are you sure you want to end the treatment? +Confirm = CONFIRM +Cancel = CANCEL + +[7] +Title = Active Cool +Message = Are you sure you want to stop heat disinfect active cool? +Confirm = CONFIRM +Cancel = CANCEL + +[8] +Title = RO Permeate Sample +Message = Are you sure you want to stop RO permeate sample? +Confirm = CONFIRM +Cancel = CANCEL + +[9] +Title = RO Permeate Sample +Message = Please select dispense or cancel option +Confirm = DISPENSE +Cancel = CANCEL + + Index: Confirm/Confirm_es.conf =================================================================== diff -u --- Confirm/Confirm_es.conf (revision 0) +++ Confirm/Confirm_es.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,55 @@ +[1] +Title = Ñater Flush +Message = Ñf stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = ÑONFIRM +Cancel = ÑANCEL + +[2] +Title = Ñeat Disinfect +Message = Ñf stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = ÑONFIRM +Cancel = ÑANCEL + +[3] +Title = Ñhemical Disinfect +Message = Ñf stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = ÑONFIRM +Cancel = ÑANCEL + +[4] +Title = Ñhemical Flush +Message = Ñf stopped, the device will have to have a complete cycle\nbefore you can create a new treatment.\n\nAre you sure you want to stop the process?\n +Confirm = ÑONFIRM +Cancel = ÑANCEL + +[5] +Title = Ñhemical Flush Sample Pass/Fail +Message = Ñlease select sample pass/fail status +Confirm = ÑASS +Cancel = ÑAIL + +[6] +Title = Ñreatment End +Message = Ñre you sure you want to end the treatment? +Confirm = ÑONFIRM +Cancel = ÑANCEL + +[7] +Title = Ñctive Cool +Message = Ñre you sure you want to stop heat disinfect active cool? +Confirm = ÑONFIRM +Cancel = ÑANCEL + +[8] +Title = ÑO Permeate Sample +Message = Ñre you sure you want to stop RO permeate sample? +Confirm = ÑONFIRM +Cancel = ÑANCEL + +[9] +Title = ÑO Permeate Sample +Message = Ñlease select dispense or cancel option +Confirm = ÑISPENSE +Cancel = ÑANCEL + + Fisheye: Tag 9a2fac022f74292e0b1518c9839a0f5749c4ac9c refers to a dead (removed) revision in file `Instructions/Instructions.conf'. Fisheye: No comparison available. Pass `N' to diff? Index: Instructions/Instructions_en.conf =================================================================== diff -u --- Instructions/Instructions_en.conf (revision 0) +++ Instructions/Instructions_en.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,134 @@ +# ###################################################################################################### +# Do not use data message with change interval in here, it affects the instructions update performance. # +# ###################################################################################################### +# multiple value reserved blocks can be used even in each line. +# format is as the following (look at the Messages_List.xlsx for the id and parameters): +# {:::} +# extra string can be used for unit(mL) which also accepts space +# In case extra string or precision is not being used leave it empty eg. {77:1::} +# the format shall always have three ':' and message id and parameter number shall not be ignored. +# otherwise will NOT be identified as a value reserved block and will be used as part of the string. +# ###################################################################################################### + +[Sample^Water Sample] +Place cup in the water sampling pocket. Press and hold the "Water Sample" button. = 9_water sampling.png + +[Sample^Water Sample Result] +Test the water sample per your clinic's instructions and enter the result. = 42_water test pass fail.png + +[Sample^Water Sample Result Failed] +Water sample result failed. = 40_water test fail.png + +[Consumables^Consumables Installation] +Remove the concentrate and dialysate caps. = 47_remove DG caps.png +Connect the acid concentrate to the acid port. = 10_install acid line.png +Connect the bicarbonate concentrate to the bicarbonate port. = 11_install bicarb line.png + +[Disposables^Cartridge Installation] +Open the front panel door. = 1_HD door open-01.png +Unlock the pump track by moving the lever to its full upward position. = 69_pump track open without cart.png +Install the cartridge using the alignment pins. = 3_insert casette-01.png +Insert tubing into blood and dialysate pumps. = 4_loop casette tubes-01.png +Lock the pump track by moving the lever to its downward position. = 6_Pump track lock-01.png + +[Disposables^Cartridge Connection] +Connect arterial and venous bloodlines using the recirculation connector. = 51_shunt patient lines.png +Place the arterial pressure line behind the arterial and venous bloodlines\nand connect to the arterial pressure transducer port. = 70_PBa connection Rev02.png +Connect the venous pressure line to the venous pressure transducer port. = 23_PBo connection.png +Connect the bloodlines to the dialyzer. = 20_connect blood lines to dialyzer.png +Connect the dialysate lines to the dialyzer. = 21_connect dialysate lines to dialyzer.png +Insert dialyzer into the dialyzer holder. = 19_insert dialyzer.png +Clamp the heparin line. = 50_clamp_heparin.png +Clamp venous chamber line. = 49_clamp venous chamber line.png +Connect the dialysate lines to the dialysate fluid ports. = 16_connect dialysate lines to DG.png + +[Disposables^Heparin Syringe] +Fill a syringe with {77:2:mL:1} heparin and connect to the heparin tubing line. = 25_fill syringe with heparin.png +Insert heparin filled syringe into the syringe pump. Unclamp heparin line. = 26_load syringe.png + +[Disposables^Saline Bag] +Clamp saline line and y-connector. = 53_pinch saline line and y connector.png +Hang the saline bag on the hook on the left side of the system.\nUse the saline line to spike a saline bag. = 29_spike and hang saline bag.png +Close the front panel door. Ensure lines are not kinked. = 66_hd door close with cart.png +Unclamp the saline line. Leave the y-connector clamped. = 54_unpinch saline and clamped y connector.png + +[Connection^Patient Connection] +Open the front panel door. = 67_hd door open with cart.png +Remove dialyzer from dialyzer holder and invert dialyzer. Reinstall dialyzer into dialyzer holder. = 52_invert_dialyzer.png +Close the front panel door. Ensure lines are not kinked. = 66_hd door close with cart.png +Draw a dialysate sample from the luer lock connection port.\nTest the dialysate sample per your facility's policy = 17_dialysate sample.png +Wash your hands and establish vascular access per your facility's policy. = 43_wash hands.png +Clamp arterial and venous bloodlines. = 56_clamp shunted patient lines.png +Remove arterial and venous line recirculation connector. = 34_remove recirc connector.png +Connect arterial and venous bloodlines to your patient access lines. = 31_connect arterial and venous lines to patient.png + +[Connection^Start Treatment] +Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png + +[^Rinseback Setup] +Clamp arterial bloodlines and disconnect. = 33_disconnect arterial lines.png +Connect arterial bloodline to the saline y-connector. = 35_connect saline y.png +Unclamp arterial bloodline and saline y-connector. = 38_unclamp saline lines.png + +[^Rinseback Complete] +If rinseback is complete, clamp arterial bloodline and saline y-connector.\nFollow your facility's policy prior to disconnecting patient. = 64_rinseback clamp arterial and saline lines.png + +[^Rinseback Reconnect] +Clamp arterial bloodline and saline y-connector. Disconnect the arterial bloodline. = 48_disconnect saline y.png +Connect arterial bloodline to your patient access line. = 71_rinseback arterial re-connect.png +Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png + +[Disconnection^Patient Disconnection] +Clamp venous bloodlines and disconnect. = 63_disconnect venous lines.png +Connect the venous bloodline to the luer lock connection port on the venous line. = 58_end treatment connect venous to sampling port.png + +[Disposables^Disposables Removal] +Open the front panel door. = 67_hd door open with cart.png +Disconnect the arterial and venous pressure lines. = 24_PBo and PBA removal.png +If heparin was used during treatment, remove syringe. = 28_remove syringe.png +Clamp dialysate lines and disconnect from the dialysate ports. = 18_disconect dialysate lines from DG.png +Replace dialysate cap. = 12_replace dialysate line cap.png +Disconnect acid and bicarbonate concentrates lines from the concentrate ports.= 13_remove acid and bicarb connect.png +Replace concentrate cap. = 59_install concentrate caps.png +Unlock the pump track by moving the lever to its upward position. = 8_pump track unlock.png +Remove cartridge, dialyzer, and saline bag from the HD device. = 5_remove casette-01.png +Lock the pump track by moving the lever to its downward position and close front panel door. = 2_HD door closed-01.png + + +[^Recirculate Disconnect] +Clamp arterial bloodline and saline y-connector. Disconnect the arterial bloodline. = 48_disconnect saline y.png +Clamp venous bloodlines and disconnect. = 63_disconnect venous lines.png +Connect arterial and venous bloodlines using the recirculation connector. = 60_recirc shunt patient lines.png +Unclamp arterial and venous bloodlines. = 62_recirc unclamp recirc lines.png + +[^Recirculate] +Treatment is paused. Blood and dialysate pumps are recirculating = 61_recirc recirculating pumps.png + +[^Recirculate Stopped] +Recirculation is stopped. Blood pump is stopped. = 65_recirc stop TEMP.png + +[^Recirculate Reconnect] +Clamp arterial and venous bloodlines. = 56_clamp shunted patient lines.png +Remove arterial and venous line recirculation connector. = 34_remove recirc connector.png +Connect arterial and venous bloodlines to your patient access lines. = 31_connect arterial and venous lines to patient.png +Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png + +[Water Flush^Water Flush] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[Heat Disinfect^Heat Disinfect] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[Heat Disinfect with Active Cool^Heat Disinfect with Active Cool] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[Chemical Disinfect^Chemical Disinfect] +Remove the concentrate cap and connect the chemical disinfectant = 14_connect PAA.png + +[Chemical Flush^Chemical Flush] +Disconnect the chemical disinfectant and replace concentrate cap. = 15_disconnect PAA.png +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[RO Permeate Sample^RO Permeate Sample] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + Index: Instructions/Instructions_es.conf =================================================================== diff -u --- Instructions/Instructions_es.conf (revision 0) +++ Instructions/Instructions_es.conf (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -0,0 +1,134 @@ +# ###################################################################################################### +# Do not use data message with change interval in here, it affects the instructions update performance. # +# ###################################################################################################### +# multiple value reserved blocks can be used even in each line. +# format is as the following (look at the Messages_List.xlsx for the id and parameters): +# {:::} +# extra string can be used for unit(mL) which also accepts space +# In case extra string or precision is not being used leave it empty eg. {77:1::} +# the format shall always have three ':' and message id and parameter number shall not be ignored. +# otherwise will NOT be identified as a value reserved block and will be used as part of the string. +# ###################################################################################################### + +[Sample^Water Sample] +Ñlace cup in the water sampling pocket. Press and hold the "Water Sample" button. = 9_water sampling.png + +[Sample^Water Sample Result] +Ñest the water sample per your clinic's instructions and enter the result. = 42_water test pass fail.png + +[Cample^Cater Sample Result Failed] +Ñater sample result failed. = 40_water test fail.png + +[Consumables^Consumables Installation] +Ñemove the concentrate and dialysate caps. = 47_remove DG caps.png +Ñonnect the acid concentrate to the acid port. = 10_install acid line.png +Ñonnect the bicarbonate concentrate to the bicarbonate port. = 11_install bicarb line.png + +[Disposables^Cartridge Installation] +Ñpen the front panel door. = 1_HD door open-01.png +Ñnlock the pump track by moving the lever to its full upward position. = 69_pump track open without cart.png +Ñnstall the cartridge using the alignment pins. = 3_insert casette-01.png +Ñnsert tubing into blood and dialysate pumps. = 4_loop casette tubes-01.png +Ñock the pump track by moving the lever to its downward position. = 6_Pump track lock-01.png + +[Disposables^Cartridge Connection] +Ñonnect arterial and venous bloodlines using the recirculation connector. = 51_shunt patient lines.png +Ñlace the arterial pressure line behind the arterial and venous bloodlines\nand connect to the arterial pressure transducer port. = 70_PBa connection Rev02.png +Ñonnect the venous pressure line to the venous pressure transducer port. = 23_PBo connection.png +Ñonnect the bloodlines to the dialyzer. = 20_connect blood lines to dialyzer.png +Ñonnect the dialysate lines to the dialyzer. = 21_connect dialysate lines to dialyzer.png +Ñnsert dialyzer into the dialyzer holder. = 19_insert dialyzer.png +Ñlamp the heparin line. = 50_clamp_heparin.png +Ñlamp venous chamber line. = 49_clamp venous chamber line.png +Ñonnect the dialysate lines to the dialysate fluid ports. = 16_connect dialysate lines to DG.png + +[Disposables^Heparin Syringe] +Fill a syringe with {77:2:mL:1} heparin and connect to the heparin tubing line. = 25_fill syringe with heparin.png +Insert heparin filled syringe into the syringe pump. Unclamp heparin line. = 26_load syringe.png + +[Disposables^Saline Bag] +Clamp saline line and y-connector. = 53_pinch saline line and y connector.png +Hang the saline bag on the hook on the left side of the system.\nUse the saline line to spike a saline bag. = 29_spike and hang saline bag.png +Close the front panel door. Ensure lines are not kinked. = 66_hd door close with cart.png +Unclamp the saline line. Leave the y-connector clamped. = 54_unpinch saline and clamped y connector.png + +[Connection^Patient Connection] +Open the front panel door. = 67_hd door open with cart.png +Remove dialyzer from dialyzer holder and invert dialyzer. Reinstall dialyzer into dialyzer holder. = 52_invert_dialyzer.png +Close the front panel door. Ensure lines are not kinked. = 66_hd door close with cart.png +Draw a dialysate sample from the luer lock connection port.\nTest the dialysate sample per your facility's policy = 17_dialysate sample.png +Wash your hands and establish vascular access per your facility's policy. = 43_wash hands.png +Clamp arterial and venous bloodlines. = 56_clamp shunted patient lines.png +Remove arterial and venous line recirculation connector. = 34_remove recirc connector.png +Connect arterial and venous bloodlines to your patient access lines. = 31_connect arterial and venous lines to patient.png + +[Connection^Start Treatment] +Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png + +[^Rinseback Setup] +Clamp arterial bloodlines and disconnect. = 33_disconnect arterial lines.png +Connect arterial bloodline to the saline y-connector. = 35_connect saline y.png +Unclamp arterial bloodline and saline y-connector. = 38_unclamp saline lines.png + +[^Rinseback Complete] +If rinseback is complete, clamp arterial bloodline and saline y-connector.\nFollow your facility's policy prior to disconnecting patient. = 64_rinseback clamp arterial and saline lines.png + +[^Rinseback Reconnect] +Clamp arterial bloodline and saline y-connector. Disconnect the arterial bloodline. = 48_disconnect saline y.png +Connect arterial bloodline to your patient access line. = 71_rinseback arterial re-connect.png +Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png + +[Disconnection^Patient Disconnection] +Clamp venous bloodlines and disconnect. = 63_disconnect venous lines.png +Connect the venous bloodline to the luer lock connection port on the venous line. = 58_end treatment connect venous to sampling port.png + +[Disposables^Disposables Removal] +Open the front panel door. = 67_hd door open with cart.png +Disconnect the arterial and venous pressure lines. = 24_PBo and PBA removal.png +If heparin was used during treatment, remove syringe. = 28_remove syringe.png +Clamp dialysate lines and disconnect from the dialysate ports. = 18_disconect dialysate lines from DG.png +Replace dialysate cap. = 12_replace dialysate line cap.png +Disconnect acid and bicarbonate concentrates lines from the concentrate ports.= 13_remove acid and bicarb connect.png +Replace concentrate cap. = 59_install concentrate caps.png +Unlock the pump track by moving the lever to its upward position. = 8_pump track unlock.png +Remove cartridge, dialyzer, and saline bag from the HD device. = 5_remove casette-01.png +Lock the pump track by moving the lever to its downward position and close front panel door. = 2_HD door closed-01.png + + +[^Recirculate Disconnect] +Clamp arterial bloodline and saline y-connector. Disconnect the arterial bloodline. = 48_disconnect saline y.png +Clamp venous bloodlines and disconnect. = 63_disconnect venous lines.png +Connect arterial and venous bloodlines using the recirculation connector. = 60_recirc shunt patient lines.png +Unclamp arterial and venous bloodlines. = 62_recirc unclamp recirc lines.png + +[^Recirculate] +Treatment is paused. Blood and dialysate pumps are recirculating = 61_recirc recirculating pumps.png + +[^Recirculate Stopped] +Recirculation is stopped. Blood pump is stopped. = 65_recirc stop TEMP.png + +[^Recirculate Reconnect] +Clamp arterial and venous bloodlines. = 56_clamp shunted patient lines.png +Remove arterial and venous line recirculation connector. = 34_remove recirc connector.png +Connect arterial and venous bloodlines to your patient access lines. = 31_connect arterial and venous lines to patient.png +Unclamp arterial and venous bloodlines. = 32_unpinch arterial and venous for blood prime.png + +[Water Flush^Water Flush] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[Heat Disinfect^Heat Disinfect] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[Heat Disinfect with Active Cool^Heat Disinfect with Active Cool] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[Chemical Disinfect^Chemical Disinfect] +Remove the concentrate cap and connect the chemical disinfectant = 14_connect PAA.png + +[Chemical Flush^Chemical Flush] +Disconnect the chemical disinfectant and replace concentrate cap. = 15_disconnect PAA.png +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + +[RO Permeate Sample^RO Permeate Sample] +Ensure dialysate and concentrate caps are fully closed. = 46_install caps.png + Index: Settings/System.conf =================================================================== diff -u -rb72556f863ba789b0a66a262d1fb00ad70f02685 -r9a2fac022f74292e0b1518c9839a0f5749c4ac9c --- Settings/System.conf (.../System.conf) (revision b72556f863ba789b0a66a262d1fb00ad70f02685) +++ Settings/System.conf (.../System.conf) (revision 9a2fac022f74292e0b1518c9839a0f5749c4ac9c) @@ -5,6 +5,10 @@ [Date] Format = MM/dd/yyyy +[Localization] +Locale = es +Language = Spanish + [RoWaterMode] RoWaterMode = 0