Index: Configurations/DataList.conf =================================================================== diff -u -r34c495cade5fb00d29972b5051861263d1b3fccb -r9b3349c8a76d87e38996bf6fe093505e6f8df42d --- Configurations/DataList.conf (.../DataList.conf) (revision 34c495cade5fb00d29972b5051861263d1b3fccb) +++ Configurations/DataList.conf (.../DataList.conf) (revision 9b3349c8a76d87e38996bf6fe093505e6f8df42d) @@ -103,8 +103,8 @@ Blood_Pressure_Measure_Interval_Def = 0 # PRS 358, SRSUI 778 [Rinseback Flow Rate Ranges] # mL/min -Rinseback_Flow_Rate_Min = 50 # PRS 359, SRSUI 330, 770 -Rinseback_Flow_Rate_Max = 150 # PRS 359, SRSUI 330, 770 -Rinseback_Flow_Rate_Res = 25 # PRS 359, SRSUI 330, 770 -Rinseback_Flow_Rate_Def = 50 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Min = 100 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Max = 300 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Res = 0 # PRS 359, SRSUI 330, 770 # The adjustment parameter has been removed from the Create Treatment Parameter so this value is not used anymore. +Rinseback_Flow_Rate_Def = 200 # PRS 359, SRSUI 330, 770