Index: Alarms/Alarms.conf =================================================================== diff -u -rc87ec62a8eca9e9eb322c277dfc5abd1b98a17e7 -rce2ae4430ca7adf5a8e63d5d6528cab6e295a8da --- Alarms/Alarms.conf (.../Alarms.conf) (revision c87ec62a8eca9e9eb322c277dfc5abd1b98a17e7) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision ce2ae4430ca7adf5a8e63d5d6528cab6e295a8da) @@ -1,1055 +1,1055 @@ [0] Title = ListTitle = -Message = Indication for when no alarms is a possible situation +Message = [1] -Title = HD POST: Stuck Button -ListTitle = Stop or Off button are pressed for more than 1 sec during POST. -Message = Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up +Title = Stuck Button +ListTitle = HD POST: Stuck Button +Message = A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair. [2] -Title = HD POST: FPGA Compatibility -ListTitle = Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. -Message = HD FPGA POST failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: FPGA Compatibility +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [3] -Title = DG POST: FPGA Compatibility -ListTitle = Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header -Message = DG FPGA POST failure +Title = Service Required: Dialysate Device +ListTitle = DG POST: FPGA Compatibility +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [4] -Title = HD POST: WD Safestate -ListTitle = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA -Message = HD Watchdog POST failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: WD Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [5] -Title = DG POST: WD Safestate -ListTitle = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V -Message = DG Watchdog POST failure +Title = Service Required: Dialysate Device +ListTitle = DG POST: WD Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 53 = [6] -Title = HD POST: Proc- UI Comm -ListTitle = If the UI failed to checkin via the CAN bus within 40 seconds -Message = UI communication POST failure. UI failed to communicate within a reasonable time after power up +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Proc- UI Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [7] -Title = HD Alarm (Low): RTC Bat Low -ListTitle = If the RTC is not functional as indicated by the bits in the first 3 control registers. -Message = Real-time clock configuration error +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [8] -Title = HD POST: Acc Vector -ListTitle = If the calculated vector length of the accelerometer x, y, z is not the nominal length. -Message = HD accelerometer failed POST +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Acc Vector +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [9] -Title = DG POST: Acc Vector -ListTitle = If the calculated vector length of the accelerometer x, y, z is not the nominal length. -Message = DG accelerometer failed POST +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acc Vector +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 323 = [10] -Title = HD POST: BLD Config Emb Mode -ListTitle = If the blood leak sensor embedded mode is unresponsive. -Message = HD blood leak sensor embedded mode communication failed +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Config Emb Mode +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [11] -Title = Dial. Gen: Dialysate Temp Low -ListTitle = If the dialysate temperature is below the target temperature for a certain period of time. -Message = Dialysate temperature is below the target temperature +Title = Low Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp Low +Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [12] -Title = HD POST: HD Image -ListTitle = If the calculated CRC of the HD image does not match the expected CRC. -Message = HD firmware image integrity POST test failed +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [13] -Title = DG POST: DG Image -ListTitle = If the calculated CRC of the DG firmware image does not matched the expected CRC -Message = DG firmware image integrity POST test failed +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG Image +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 813 = [14] -Title = DG POST: Usage Rec NVRAM CRC -ListTitle = If the DG usage information record's CRC does not match the calculated CRC in POST. -Message = DG non-volatile usage information group invalid record CRC +Title = Service Required: Dialysate Device +ListTitle = DG POST: Usage Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 957 = [15] -Title = Dial. Gen: Dialysate Temp High -ListTitle = If the dialysate temperature is > 46 C for a certain period of time. -Message = Dialysate temperature is out of high safety range +Title = High Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [16] -Title = HD POST: Primary Alarm Current -ListTitle = If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output -Message = HD alarm audio failed POST +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Primary Alarm Current +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [17] -Title = HD POST: UI POST Not Complete -ListTitle = If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. -Message = HD UI POST failed +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI POST Not Complete +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [18] -Title = Loss of DG Comm (DG ACK) -ListTitle = HD message to DG that requires acknowledge was not acknowledged within a certain time. -Message = HD to DG comm error. +Title = Communication Error +ListTitle = Loss of DG Comm (DG ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [19] -Title = Dial. Gen: Dialysate Temp High -ListTitle = If the dialysate temperature is above the target temperature for a certain period of time. -Message = Dialysate temperature is above the target temperature +Title = High Dialysate Temperature +ListTitle = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. [20] -Title = DG POST: Conductivity Cal -ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate -Message = DG conductivity sensors invalid calibration record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Conductivity Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 16 = [21] -Title = DG POST: Drain Line Vol Cal -ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate -Message = DG drain line volume invalid calibration record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Drain Line Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 936 = [22] -Title = DG POST: Reservoir Vol Cal -ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate -Message = DG reservoirs invalid calibration record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Reservoir Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 935 = [23] -Title = DG POST: Acid Conc Cal -ListTitle = If the non-volatile data is valid and if not raise the alarm -Message = DG acid concentrate calibration record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acid Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 934 = [24] -Title = DG POST: Bicarb Conc Cal -ListTitle = If the non-volatile data is valid and if not raise the alarm -Message = DG bicarb concentrate calibration record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Bicarb Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 933 = [25] -Title = DG POST: Acc Cal -ListTitle = If the calibration data that was received from NVDataMgmt is NOT legitimate -Message = DG accelerometer invalid calibration record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Acc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 817 = [26] -Title = HD POST: Acc Cal -ListTitle = If the accel calibration data that was received from NVDataMgmt is NOT legitimate. -Message = HD accelerometer invalid calibration record +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Acc Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [27] -Title = Process: Pre- Tx Wet Self- Test -ListTitle = BLD fails to complete zero sequence. -Message = HD BLD sensor failed zero sequence +Title = Self Test: Blood Leak +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. [28] -Title = DG Fault: Conductivity RTD Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = DG two wire sensors FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Conductivity RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 864 = [29] -Title = HD POST: SP DAC Cal -ListTitle = If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. -Message = HD heparin force sensor invalid calibration record +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: SP DAC Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [30] -Title = HD Fault: SW Invalid State -ListTitle = If the specific software faults listed in alarm management SW faults has been detected -Message = HD Software fault. Software found itself in an unexpected state +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: SW Invalid State +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [31] -Title = HD Fault: BP Current -ListTitle = If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. -Message = Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [32] -Title = HD Fault: BP Speed -ListTitle = If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM -Message = Blood pump failed motor off check. Measured speed while commanded off +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [33] -Title = HD Fault: BP Direction -ListTitle = If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. -Message = Blood pump failed motor direction check. Measured vs. commanded +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [34] -Title = HD Fault: BP Rotor -ListTitle = If the delta of the measured motor speed and the commanded motor speed is not within tolerance. -Message = Blood pump failed rotor speed check. Mismatch with rotor and motor speeds +Title = Blood Pump Error +ListTitle = HD Fault: BP Rotor +Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [35] -Title = HD Fault: DPi Current -ListTitle = If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. -Message = Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [36] -Title = HD Fault: DPi Speed -ListTitle = If during command off state, the DPi pump motor speed is measured > 100.0 RMP. -Message = Dialysis inlet pump failed motor off check. Measured speed while commanded off +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [37] -Title = HD Fault: DPi Direction -ListTitle = If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. -Message = Dialysis inlet pump failed motor direction check. Measured vs. commanded +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [38] -Title = HD Fault: DPi Rotor -ListTitle = If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. -Message = Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds +Title = Dialysate Pump Error +ListTitle = HD Fault: DPi Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [39] -Title = HD Fault: DPo Current -ListTitle = If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. -Message = Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [40] -Title = HD Fault: DPo Speed -ListTitle = If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. -Message = Dialysis outlet pump failed motor off check. Measured speed while commanded off +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [41] -Title = HD Fault: DPo Direction -ListTitle = If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. -Message = Dialysis outlet pump failed motor direction check. Measured vs. commanded +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [42] -Title = HD Fault: DPo Rotor -ListTitle = If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed -Message = Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds +Title = Dialysate Pump Error +ListTitle = HD Fault: DPo Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [43] -Title = DG Fault: Iso ADC Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = Main primary heater FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Iso ADC Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 954 = [44] -Title = HD Fault: UI Comm -ListTitle = If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. -Message = User interface communication timeout +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UI Comm +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [45] -Title = HD Fault: CAN CRC Com -ListTitle = If too many bad messages CRCs are received within a set period of time -Message = Too many bad CRCs detected on received system messages +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: CAN CRC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [46] -Title = HD Fault: CAN ACK UI Com -ListTitle = If no retries left in the pending acknowlegement list. -Message = System message that required acknowledgment was not acknowledged +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: CAN ACK UI Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [47] Title = ListTitle = Message = [48] -Title = HD Fault: UF Rate Tare Err -ListTitle = If the UF volume (refUFvolume - measureVolume) > 100.0 mL. -Message = Ultrafiltration volume accuracy error during treatment +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UF Rate Tare Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [49] -Title = HD Fault: FPGA Com -ListTitle = If there are more than 3 comm errors in a 1 min period. -Message = HD FPGA communication down for too long +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: FPGA Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [50] -Title = DG Fault: Valve Control -ListTitle = If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. -Message = DG FPGA not accepting commanded valve states +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Valve Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 26 = [51] -Title = HD Fault: BP Speed -ListTitle = If measured motor speed versus commanded rotor speed exceeded limits. -Message = Blood pump failed motor speed check. Measured vs. commanded +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [52] -Title = HD Fault: DPi Speed -ListTitle = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. -Message = Dialysate inlet pump failed motor speed check. Measured vs. commanded +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [53] -Title = HD Fault: DPo Speed -ListTitle = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. -Message = Dialysate outlet pump failed motor speed check. Measured vs. commanded +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [54] -Title = HD Fault: Critical Data Err -ListTitle = If the critical data failed range check. -Message = HD critical data integrity check failed +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Critical Data Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [55] -Title = DG Fault: Critical Data Err -ListTitle = If the critical data failed range check. -Message = DG critical data integrity check failed +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Critical Data Err +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 921 = [56] -Title = HD Fault: Acc Com -ListTitle = If no fresh sample of accelerometer data has been detected for more than 100 msec. -Message = HD accelerometer error (no readings or FPGA reports error) +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Acc Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [57] -Title = DG Fault: Acc Com -ListTitle = If no fresh sample of accelerometer data has been detected for more than 100 msec. -Message = DG accelerometer error (no readings or FPGA reports error) +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Acc Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 869 SRSDG 870 = [58] -Title = HD Fault: Valve Home -ListTitle = If the number of failed homing > = 3 counts. -Message = HD valve homing failed +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Home +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [59] -Title = HD Fault: Valve Transition Err -ListTitle = if the valve's transition time has timed out (3000msec) -Message = HD valve transition time out +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Transition Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [60] -Title = HD Fault: Valve Control -ListTitle = If the valve bit masks are not as expected in different valve control mode. -Message = HD valve not functional +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Control +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [61] -Title = HD Fault: Valve Current -ListTitle = If the valve current is over the threshold of 1.0 A for 1000 msec. -Message = HD valve current out of range +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [62] -Title = HD Fault: Valve Position -ListTitle = If the current position has deviated from the expected position for more than 1 second. -Message = HD valve position out of target +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Valve Position +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [63] Title = ListTitle = Message = [64] -Title = DG Fault: Baro Range -ListTitle = If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. -Message = Barometric pressure is out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 874 = [65] -Title = Loss of DG Comm (Reject) -ListTitle = If the latest DG command response indicated invalid parameter -Message = HD requests DG command with invalid parameter fault +Title = Communication Error +ListTitle = Loss of DG Comm (Reject) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [66] -Title = HD POST: BLD Config Set Point -ListTitle = If the set point of the blood leak sensor does not match the set point in the calibration record. -Message = HD blood leak sensor set point set failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Config Set Point +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [67] -Title = Process: Pre- Tx Remove Cartridge -ListTitle = If the blood pump pressure > 5000 -Message = HD blood pump occlusion self-test failure alarm +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [68] -Title = Dial. Gen: Dialysate not ready -ListTitle = If the recirculation level has exceeded the max allowed -Message = HD active reservoir recirculation out of range +Title = Dialysate Generation Error +ListTitle = Dial. Gen: Dialysate not ready +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. [69] -Title = HD POST: BLD Cal -ListTitle = If the BLD data that was received from NVDataMgmt is NOT legitimate. -Message = HD blood leak sensor invalid calibration record +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: BLD Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [70] -Title = Process: Pre- Tx Remove Cartridge -ListTitle = If arterial pressure < -300 or > 100 mmHg -Message = HD arterial pressure self-test failure alarm +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [71] -Title = Process: Pre- Tx Remove Cartridge -ListTitle = If the venous pressure < -100 or > 600 mmHg -Message = HD venous pressure self-test failure alarm +Title = Self Test: Cartridge +ListTitle = Process: Pre- Tx Remove Cartridge +Message = A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Press Resume/OK [72] -Title = Loss of DG Comm (LC) -ListTitle = No new load cell data for 2 seconds. -Message = HD No load cell data message receive at least once every 2 seconds +Title = Communication Error +ListTitle = Loss of DG Comm (LC) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [73] -Title = Loss of DG Comm (TDI/TRO) -ListTitle = No new dialysate temperature data for 2 seconds. -Message = HD No dialysate temperature data message received at least once every 2 seconds +Title = Communication Error +ListTitle = Loss of DG Comm (TDI/TRO) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [74] -Title = DG Fault: UVi On No Flow -ListTitle = If the inlet UV reactor is on while VPi is closed for a certain period of time. -Message = DG inlet UV reactor is on while there is no flow +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVi On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 600 = [75] -Title = HD Fault: Syringe DAC -ListTitle = If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. -Message = HD syringe pump self-test failure alarm +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe DAC +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [76] -Title = HD Fault: Voltage Monitor -ListTitle = If the HD monitored voltages are out of range for a certain period of time. -Message = HD monitored voltage is out of range +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Voltage Monitor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [77] -Title = DG Fault: Voltage Monitor -ListTitle = If the DG monitored voltages are out of range for a certain period of time. -Message = DG monitored voltage is out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Voltage Monitor +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] -Title = HD Fault: Syringe Pump Enc Dir -ListTitle = If the measured encoder direction of the syringe pump does not agree with the expected direction -Message = HD syringe pump direction (from encoder) error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Enc Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [79] -Title = HD Fault: Syringe Pump Drv Dir -ListTitle = If the measured direction of the syringe pump motor controller does not agree with the expected direction -Message = HD syringe pump direction (from controller) error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Drv Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [80] -Title = HD Fault: Syringe Pump Driver -ListTitle = If the FPGA reported the syringe pump fault -Message = HD syringe pump fault reported by FPGA +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Driver +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [81] -Title = HD Fault: Syringe Pump OT -ListTitle = If the syringe pump position is deemed to have overtraveled -Message = HD syringe pump over travel error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump OT +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [82] -Title = HD Fault: Syringe Pump DAC Err -ListTitle = If the syringe pump DAC write error is reported by the FPGA -Message = HD syringe pump DAC write failure +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump DAC Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [83] -Title = HD Fault: Syringe Pump On Err -ListTitle = If the syringe pump is on while the blood pump is off -Message = HD syringe pump is running while the BP is off +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump On Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [84] -Title = DG POST: DG RTC Time -ListTitle = If DG RTC set year is < 2021. -Message = DG RTC set year is invalid +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG RTC Time +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 815 = [85] -Title = Blood Stop: Cartridge Lever Open -ListTitle = If the pump track lever is detected as unlatched when it should be. -Message = HD pump track lever is not fully latched +Title = Cartridge Lever is Open +ListTitle = Blood Stop: Cartridge Lever Open +Message = A problem has been detected with the cartridge.\n- Open the device door and push lever all the way down.\n- Close the device door.\n- Press Resume/OK button to continue. [86] -Title = HD POST: HD RTC Time -ListTitle = If HD RTC set year is < 2021. -Message = HD RTC set year is invalid +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD RTC Time +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [87] -Title = DG POST: Heat Cal -ListTitle = If the heating paramters' non-volatile record's CRC does not match the calculated CRC. -Message = DG heating parameters invalid calibration record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Heat Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 929 = [88] -Title = DG Fault: CP1 CP2 Speed Range -ListTitle = Hall sensor pulse to pulse timing is < 1ms (speed too fast). -Message = DG concentrate pump hall sensor out of range. +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1 CP2 Speed Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [89] -Title = DG Fault: UVo On No Flow -ListTitle = If the outlet UV reactor is on while the RO flow is less than 200 mL/min. -Message = DG outlet UV reactor is on while there is no flow +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVo On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 599 = [90] -Title = DG Fault: LC A1/B1 Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time. -Message = DG load cells A1/B1 FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC A1/B1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 604 = [91] -Title = Loss of DG Comm (FMD) -ListTitle = If not receiving new dialysate flow readings in expected interval (3 seconds) -Message = HD No dialysate flow data received in the last 3 seconds +Title = Communication Error +ListTitle = Loss of DG Comm (FMD) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [92] Title = ListTitle = Message = [93] -Title = DG Fault: RO Permeate Cond High -ListTitle = If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. -Message = DG outlet primary conductivity sensor out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Permeate Cond High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 430 = [94] -Title = DG Fault: Pressure Sensor Range -ListTitle = If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. -Message = DG pressure sensor out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Pressure Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 629 = [95] -Title = DG Fault: WD Error -ListTitle = If the watchdog signal is low while not in POST. -Message = DG watchdog expired +Title = Service Required: Dialysate Device +ListTitle = DG Fault: WD Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 907 = [96] -Title = Dial. Gen: Inlet Water Temp High -ListTitle = If the inlet water temperature is > 37 C. -Message = DG inlet water temperature out of range +Title = Incoming Water Temperature High +ListTitle = Dial. Gen: Inlet Water Temp High +Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 410 = [97] -Title = Dial. Gen: Dialysate Cond. -ListTitle = If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. -Message = DG fill conductivity out of range +Title = Conductivity Out of Range +ListTitle = Dial. Gen: Dialysate Cond. +Message = Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 397 = [98] -Title = HD Alarm (Low): Battery Com -ListTitle = If battery I2C communication fault persistence count > 5. -Message = HD battery or battery charger communication fault +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Battery Com +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [99] -Title = HD Fault: Syringe Pump Stall -ListTitle = If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. -Message = HD syringe pump stall alarm +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Stall +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [100] Title = ListTitle = Message = [101] -Title = HD POST: UI- Proc Comm -ListTitle = If the HD is not communicating with the UI during POST. -Message = UI POST FW Communication failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI- Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [102] -Title = Cleaning: Tpo THD Temp Delta -ListTitle = If temperature gradient between Tpo and Thd exceeds maximum 15°C. -Message = DG heat disinfection temperature gradient out of range +Title = Disinfection Interrupted +ListTitle = Cleaning: Tpo THD Temp Delta +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 856 SRSDG 930 = [103] -Title = HD POST: Cal NVRAM CRC -ListTitle = If the computed CRC of the calibration group record is not equal to the record CRC. -Message = HD non-volatile calibration group invalid record CRC +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Cal NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [104] -Title = HD Fault: Air Trap Sensor -ListTitle = If the air trap level readings are not valid for more than 2 seconds. -Message = Air trap level sensors reporting illegal combination of air/fluid +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Air Trap Sensor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [105] -Title = DG POST: Cal NVRAM CRC -ListTitle = If the DG NVDATAMGMT CAL group calculated CRC does not matched the expected CRC. -Message = DG non-volatile system group invalid record CRC +Title = Service Required: Dialysate Device +ListTitle = DG POST: Cal NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 608 = [106] -Title = DG POST: Flow Sensor Cal -ListTitle = If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. -Message = DG flow sensors calibration error +Title = Service Required: Dialysate Device +ListTitle = DG POST: Flow Sensor Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] -Title = DG Fault: DG Restarted -ListTitle = If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. -Message = HD reports DG restarted fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DG Restarted +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [108] -Title = HD Fault: Syringe Pump ADC Com -ListTitle = If the ADC read is not fresh after POST init and the ADC read counter has not changed. -Message = HD syringe pump ADC error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump ADC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [109] -Title = HD Fault: Syringe Pump Vol -ListTitle = If the volume delivered is off by more than 5% or 0.1mL. -Message = HD syringe pump volume check error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Vol +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [110] -Title = HD Fault: Syringe Pump Speed -ListTitle = If the syringe measured rate is off by more than 5% or 0.1mL. -Message = HD syringe pump speed check error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [111] -Title = HD Fault: Syringe Pump Speed -ListTitle = If the syringe pump position is not equal to its last position. -Message = HD syringe pump not stopped in off state error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [112] -Title = DG Fault: LC A2/B2 Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time. -Message = DG load cells A2/B2 FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC A2/B2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 953 = [113] -Title = DG Fault: DRP Current -ListTitle = If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. -Message = DG drain pump current out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Current +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 890 = [114] -Title = Process: Pre- Tx Wet Self- Test -ListTitle = If the venous air bubble self test did not detect bubble for more than 500 msec. -Message = HD venous air bubble detector self-test failure +Title = Self Test: Venous Air Detector +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge. [115] -Title = DG Fault: Temp Sensor Range -ListTitle = If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. -Message = DG temperature sensor out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Temp Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 224 = [116] -Title = HD Alarm (Low): SD- Card Error -ListTitle = SD card is non-operational -Message = SD card is non-operational +Title = Service Required: Dialysate Device +ListTitle = HD Alarm (Low): SD- Card Error +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [117] -Title = Loss of HD Comm (HD ACK) -ListTitle = If no retries left in the pending acknowlegement list. -Message = System message that required acknowledgment was not acknowledged +Title = Communication Error +ListTitle = Loss of HD Comm (HD ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 909 = [118] -Title = DG Fault: RTC Config -ListTitle = If the RTC configuration registers report an error. -Message = DG RTC configuration error +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RTC Config +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 43 = [119] -Title = DG Alarm (Low): RTC Bat Low -ListTitle = If the RTC battery bit report a low alarm. -Message = DG RTC battery low +Title = Service Required: Dialysate Device +ListTitle = DG Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. SRSDG 908 = [120] -Title = Process: Pre- Tx Wet Self- Test -ListTitle = If the HD pre-treatment wet self test failed. -Message = HD pre-treatment wet prime test failure +Title = Self Test: Venous Air Detector +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [121] -Title = DG Fault: Prim Heater Lg Control -ListTitle = If the main primary heater's voltage is out of range. -Message = DG main primary heater voltage out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Prim Heater Lg Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 611 = [122] -Title = DG Fault: Prim Heater Sm Control -ListTitle = If the small primary heater's voltage is out of range. -Message = DG small primary heater voltage out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Prim Heater Sm Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 612 = [123] -Title = DG Fault: Trimmer Heater Control -ListTitle = If the trimmer voltage is out of range. -Message = DG trimmer heater voltage out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Trimmer Heater Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 613 = [124] -Title = Blood Stop: Blood Clot Risk -ListTitle = In treatment end paused state for > 5 minutes. -Message = HD end of treatment alarm (high priority - user still not acting to end treatment) +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [125] -Title = Blood Stop: Blood Clot Risk -ListTitle = If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. -Message = Blood sitting too long after treatment stopped by user alarm (> 5 min) +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [126] -Title = Blood Stop: Dialyzer Blood Leak -ListTitle = If the blood leak is detected based on the blood leak status reported by the sensor. -Message = HD blood leak detected alarm +Title = Blood Leak +ListTitle = Blood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. \n- Visually inspect dialysate return line for blood.\n- Press Resume/OK to proceed to the next step. Dialysate will\n remain in bypass. [127] -Title = Blood Stop: Venous Pres. Low -ListTitle = If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. -Message = Venous pressure too low during treatment +Title = Low Venous Pressure +ListTitle = Blood Stop: Venous Pres. Low +Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and\nterminate treatment without rinseback.\n- If access connections are properly secured, \npress Resume/OK to continue treatment. [128] -Title = Blood Stop: Venous Air Bubble -ListTitle = If bubbles are detected during the treatment. -Message = HD venous air bubble detected alarm +Title = Venous Air Detected +ListTitle = Blood Stop: Venous Air Bubble +Message = Air detected in the venous air trap.\n- Confirm arterial access placement and confirm\nconnections are secure.\n- Treatment must be terminated without rinseback. [129] -Title = Blood Stop: Dialyzer Blood Leak Bypass -ListTitle = If the blood leak is detected in treatment, the HD device transitions to treatment stop and runs in bypass until there is no blood detected. The user can hit end treatment anytime. -Message = HD blood leak detected and it is in the process of attempting to remove the bubbles/foam +Title = Blood Leak: Bypass +ListTitle = Blood Stop: Dialyzer Blood Leak Bypass +Message = The device is currently in bypass due to a potential blood leak.\n- If blood is present upon visual inspection, press\n END TREATMENT to terminate treatment without rinseback.\n- If blood is not visualized, check dialysate for blood using\n test strip.\n- Follow facility policy to Resume/OK or END TREATMENT. [130] -Title = Blood Stop: Venous Pres. High -ListTitle = If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. -Message = Venous pressure too high during treatment +Title = High Venous Pressure +ListTitle = Blood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [131] -Title = Blood Stop: Arterial Pres. Low -ListTitle = If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. -Message = Arterial pressure too low during treatment +Title = Low Arterial Pressure +ListTitle = Blood Stop: Arterial Pres. Low +Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [132] -Title = Blood Stop: Arterial Pres. High -ListTitle = If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. -Message = Arterial pressure too high during treatment +Title = High Arterial Pressure +ListTitle = Blood Stop: Arterial Pres. High +Message = Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback. [133] -Title = DG Fault: Fluid Leak in DG -ListTitle = If the fluid leak is detected as reported by the fluid leak detector. -Message = DG fluid leak detected alarm +Title = Dialysate Device Fluid Leak +ListTitle = DG Fault: Fluid Leak in DG +Message = Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 328 = [134] -Title = Blood Stop: Fluid Leak in HD -ListTitle = If the fluid leak is detected as reported by the fluid leak detector. -Message = HD fluid leak detected alarm +Title = Fluid Leak +ListTitle = Blood Stop: Fluid Leak in HD +Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, Resume/OK treatment. [135] -Title = Blood Stop: HD Shock Detected -ListTitle = If the system experienced excessive shock > 2.5 g. -Message = HD experienced large acceleration +Title = Blood Pump Stopped: Excessive Movement +ListTitle = Blood Stop: HD Shock Detected +Message = Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. [136] -Title = Dial. Gen: Shock- Vibration -ListTitle = If the system experienced excessive shock > 2.5 g. -Message = DG experienced large acceleration +Title = Dialysate Device: Excessive Movement +ListTitle = Dial. Gen: Shock- Vibration +Message = Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. SRSDG 327 = [137] -Title = Blood Stop: HD Tilted -ListTitle = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. -Message = HD tilt exceeds maximum +Title = Blood Pump Stopped: Excessive Tilt +ListTitle = Blood Stop: HD Tilted +Message = Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press Resume/OK. [138] -Title = Dial. Gen: Tilt -ListTitle = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. -Message = DG tilt exceeds maximum +Title = Dialysate Device: Excessive Tilt +ListTitle = Dial. Gen: Tilt +Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. SRSDG 326 = [139] -Title = Loss of AC Power -ListTitle = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. -Message = HD AC power lost alarm +Title = Power Loss +ListTitle = Loss of AC Power +Message = AC Power has been lost.\n- Check all electrical connections.\n- Press Resume/OK to return to welcome screen. [140] -Title = Loss of DG Comm (Timeout) -ListTitle = If DG communication timeout is detected for more than 2 seconds. -Message = Dialysate generator communication timeout +Title = Communication Error +ListTitle = Loss of DG Comm (Timeout) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [141] -Title = Blood Stop: Air Trap Fill -ListTitle = If the air trap fill timeout during treatment. -Message = Air trap fill timeout during treatment +Title = Blood Pump Stopped: Venous Air Detector +ListTitle = Blood Stop: Air Trap Fill +Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback. [142] -Title = Blood Stop: Blood Pump Occlusion -ListTitle = If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall -Message = Blood pump occlusion detected +Title = Blood Pump Stopped: Occlusion +ListTitle = Blood Stop: Blood Pump Occlusion +Message = Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [143] -Title = DG Fault: TRo TDi Delta -ListTitle = If TDi and TRo have drifted for more 1 C for a certain period of time. -Message = DG dialysate temperature sensors drift timeout. +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TRo TDi Delta +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 912 = [144] -Title = Cleaning: Inlet Water Cond High -ListTitle = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode -Message = DG cleaning mode inlet water conductivity out of high range +Title = Disinfection: Water Conductivity High +ListTitle = Cleaning: Inlet Water Cond High +Message = Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = Dial. Gen: Acid Conc. Cond. -ListTitle = If the conductivity value for the acid concentrate is out of range for more than 45 secs. -Message = DG Concentrate conductivity after adding acid out of range alarm +Title = Conductivity Error During Prime +ListTitle = Dial. Gen: Acid Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 846 = [146] -Title = DG POST: DG RTC Timer -ListTitle = If RTC and timer are not aligned on 1 second elapsed time. -Message = DG RTC vs. Timer accuracy error +Title = Service Required: Dialysate Device +ListTitle = DG POST: DG RTC Timer +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 312 = [147] -Title = Dial. Gen: Dialysate not ready -ListTitle = If dialysate conductivity is not with 5% of the TCD -Message = DG dialysate generation conductivity fault alarm +Title = Dialysate Generation Error +ListTitle = Dial. Gen: Dialysate not ready +Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will need to be resumed when \nthe dialysate is ready for use. SRSDG 398 = [148] -Title = Cleaning: Inlet Water Temp High -ListTitle = Inlet temperature > 37 C. -Message = Cleaning Mode Inlet water temperature too high +Title = Disinfection: Water Temperature High +ListTitle = Cleaning: Inlet Water Temp High +Message = Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] -Title = HD Fault: WD Error -ListTitle = If the watchdog signal is low while not in POST. -Message = HD watchdog expired +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: WD Error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [150] -Title = Dial. Gen: Inlet Water Cond High -ListTitle = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). -Message = Inlet water conductivity too high +Title = Incoming Water Conductivity High +ListTitle = Dial. Gen: Inlet Water Cond High +Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 391 = [151] -Title = Dial. Gen: Inlet Water Cond Low -ListTitle = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). -Message = Inlet water conductivity too low +Title = Incoming Water Conductivity Low +ListTitle = Dial. Gen: Inlet Water Cond Low +Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 826 = [152] -Title = Dial. Gen: Inlet Water Pres Low -ListTitle = Inlet water pressure < 25 PSI. -Message = Inlet water pressure too low +Title = Incoming Water Pressure Low +ListTitle = Dial. Gen: Inlet Water Pres Low +Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 417 = [153] -Title = Process: Pre- Tx Terminated -ListTitle = In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) -Message = HD prime completed high priority alarm +Title = Pre-Treatment Process Terminated +ListTitle = Process: Pre- Tx Terminated +Message = The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press Resume/OK for continue with next steps to remove \ncartridge from the device. [154] -Title = Cleaning: Inlet Water Cond Low -ListTitle = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode -Message = Cleaning mode inlet water conductivity too low +Title = Disinfection: Water Conductivity Low +ListTitle = Cleaning: Inlet Water Cond Low +Message = Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = Loss of DG Comm (Reservoir) -ListTitle = No new reservoir data for 2 seconds. -Message = HD No reservoir data message receive at least once every 2 seconds +Title = Communication Error +ListTitle = Loss of DG Comm (Reservoir) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [156] -Title = Loss of DG Comm (Op Mode) -ListTitle = No DG op mode for 2 seconds. -Message = HD No DG op mode message receive at least once every 2 seconds +Title = Communication Error +ListTitle = Loss of DG Comm (Op Mode) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [157] -Title = Cleaning: Chem Disinf PAA Prime -ListTitle = If the disinfectant acid line is not primed within a certain period of time. -Message = Chemical disinfect priming the acid line timeout +Title = Chemical Disinfectant Not Detected +ListTitle = Cleaning: Chem Disinf PAA Prime +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 988 = [158] -Title = Dial. Gen: Inlet Water Temp Low -ListTitle = Inlet temperature < 24 °C for 5 seconds. -Message = Inlet water temperature low +Title = Incoming Water Temperature Low +ListTitle = Dial. Gen: Inlet Water Temp Low +Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 408 = [159] -Title = Cleaning: Chem Disinf Connect Cap -ListTitle = Start of chemical disinfect flush. -Message = Close concentrate line be able to flush it +Title = Remove Chemical Disinfectant +ListTitle = Cleaning: Chem Disinf Connect Cap +Message = The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. SRSDG 1044 = [160] -Title = HD Fault: BP Rotor Max Speed -ListTitle = If blood pump motor speed exceeded 100.0 RPM. -Message = Blood pump rotor speed too high +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Rotor Max Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [161] -Title = DG Fault: HD- DG Comm -ListTitle = If >= 5 bad CRCs detected on received CAN messages within 10 minutes. -Message = Too many bad CRCs detected on received system messages +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HD- DG Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 122 = [162] -Title = DG Fault: Proc- FPGA Clock -ListTitle = If the FPGA clock speed is relatively close to the processor clock for a certain period of time. -Message = FPGA clock speed failure +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Proc- FPGA Clock +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 867 = [163] -Title = Dial. Gen: UF Rate Check -ListTitle = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). -Message = HD sees primary load cell for reservoir 1 change too much too fast +Title = Ultrafiltration Error +ListTitle = Dial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. [164] -Title = Dial. Gen: UF Rate Check -ListTitle = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). -Message = HD sees primary load cell for reservoir 2 change too much too fast +Title = Ultrafiltration Error +ListTitle = Dial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume/OK treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. [165] -Title = Blood Stop: Blood Clot Risk -ListTitle = If the maximum time inrinseback operation has been exceeded. -Message = HD in treatment rinseback operation timeout +Title = Blood Pump Stop Time Exceeded +ListTitle = Blood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback. [166] -Title = +Title = Test Alarm - Alarm 166 ListTitle = Message = [167] -Title = Blood Stop: Door Open -ListTitle = If the cartridge door is opened as reported by the FPGA. -Message = HD cartridge door opened alarm +Title = Blood Pump Stopped: Door Open +ListTitle = Blood Stop: Door Open +Message = The device door is open.\n Close the device door. \n- Press Resume/OK to continue. [168] -Title = +Title = Test Alarm - Alarm 168 ListTitle = Message = [169] -Title = DG Fault: Dialysate Flow Sens Rg -ListTitle = If the dialysate flow rate > 2.00 L/min. -Message = Dialysate flow rate out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Dialysate Flow Sens Rg +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 643 = [170] Title = Heparin Syringe Empty -ListTitle = If the syringe pump position is near empty position. -Message = HD syringe empty alarm +ListTitle = Heparin Syringe Empty +Message = Heparin syringe is empty.\n- Treatment will continue without heparin.\n- Refer to facility policy if additional heparin is needed to\ncomplete treatment. [171] -Title = Process: Pre- Tx Hep. Pump Occ. -ListTitle = If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. -Message = HD syringe pump occlusion alarm +Title = Heparin Pump Occlusion +ListTitle = Process: Pre- Tx Hep. Pump Occ. +Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped. [172] -Title = Process: Pre- Tx Not Enough Hep. -ListTitle = If the estimated syringe volume needed for treatment is not sufficient -Message = HD syringe pump not enough Heparin alarm +Title = Not Enough Heparin +ListTitle = Process: Pre- Tx Not Enough Hep. +Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. [173] -Title = HD Fault: RTC Config -ListTitle = If the RTC configuration registers report an error. -Message = HD RTC configuration error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: RTC Config +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [174] -Title = HD POST: HD RTC Timer -ListTitle = If RTC and timer are not aligned on 1 second elapsed time. -Message = HD RTC vs. Timer accuracy error +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: HD RTC Timer +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [175] -Title = HD Fault: Pump Commutation Err -ListTitle = If the set pump direction does not agree with the hall sensor direction. -Message = HD pump commutation error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Pump Commutation Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [176] -Title = +Title = Test Alarm - Alarm 176 ListTitle = Message = [177] -Title = DG Fault: Software Invalid State -ListTitle = If the specific software faults listed in alarm management SW faults has been detected -Message = DG software fault. Software found itself in an unexpected state +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Software Invalid State +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 919 = [178] -Title = HD Fault: UI- Proc Comm -ListTitle = If the HD stops communicating with the UI. -Message = HD communication timeout +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: UI- Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1066 [179] -Title = DG Fault: FPGA Com -ListTitle = If retries for commands exceeds limit or FPGA reports comm error. -Message = DG FPGA communication down for too long +Title = Service Required: Dialysate Device +ListTitle = DG Fault: FPGA Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 605 = [180] -Title = DG Fault: RO Flow Sensor Range -ListTitle = If the RO flow rate > 2.00 L/min. -Message = RO flow rate out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 883 = [181] -Title = DG Fault: Tare Out of Range -ListTitle = If the load cell weight is out of range for tare. -Message = DG load cells weight out of range for tare +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Tare Out of Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 849 SRSDG 838 = [182] -Title = DG POST: LC Cal -ListTitle = If the CRC of load cells calibration data does not match the calculated CRC. -Message = DG load cells invalid calibration +Title = Service Required: Dialysate Device +ListTitle = DG POST: LC Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 162 = [183] -Title = Cleaning: Invalid Resv Transfer -ListTitle = If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. -Message = DG invalid load cell value +Title = Disinfection Interrupted +ListTitle = Cleaning: Invalid Resv Transfer +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 851 = [184] -Title = DG Fault: UVi Error -ListTitle = If the inlet reactor are not healthy. -Message = Inlet reactor not healthy +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVi Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 559 = [185] -Title = DG Alarm (Low): Fan Speed -ListTitle = If the fan RPM is out of Min/Max range. -Message = DG fan RPM out of range +Title = Service Required: Dialysate Device +ListTitle = DG Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair. SRSDG 628 = [186] -Title = DG Fault: CP1/CP2 Fault -ListTitle = If the concentrate pumps fault is reported by the FPGA. -Message = DG concentrate pump fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1/CP2 Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 891 = [187] -Title = DG Fault: CP1 Speed Fault -ListTitle = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. -Message = DG concentrate pump CP1 speed control error +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP1 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 905 = [188] -Title = DG Fault: CP2 Speed Fault -ListTitle = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. -Message = DG concentrate pump CP2 speed control error +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CP2 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 906 = [189] -Title = DG Fault: DRP Speed Fault -ListTitle = If the drain pump RPM > 100.0 of the target RPM. -Message = DG drain pump RPM out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 448 = [190] -Title = DG Fault: DRP Speed Not Off -ListTitle = If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. -Message = DG drain pump off fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Speed Not Off +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [191] -Title = DG Fault: RO Flow High -ListTitle = If the RO flow rate > 10% from the target flow rate. -Message = DG flow rate out of upper range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 885 = [192] -Title = DG Fault: RO Flow Low -ListTitle = If the RO flow rate < 10% from the target flow rate. -Message = DG flow rate out of lower range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Flow Low +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 945 = [193] @@ -1058,116 +1058,116 @@ Message = [194] -Title = DG Fault: RO Driver Feedback -ListTitle = If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. -Message = DG RO pump duty cycle out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Driver Feedback +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 881 = [195] -Title = DG Fault: RO Pump High Pressure -ListTitle = If the DG RO pump pressure > 135 psi for a certain period of time. -Message = DG RO pump pressure out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Pump High Pressure +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 925 = [196] -Title = DG Fault: CPi/CPo Comm -ListTitle = If the FPGA reports error for a certain period of time. -Message = DG CPi/CPo conductivity sensors FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPi/CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 955 = [197] -Title = DG Fault: CD1/CD2 Comm -ListTitle = If the FPGA reports error for a certain period of time. -Message = DG CD1/CD2 conductivity sensors FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD1/CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 956 = [198] -Title = DG Fault: HP On With No Flow -ListTitle = If the RO flow is < 0.2 L/min for a certain period of time. -Message = DG primary heater while RO flow is low +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HP On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 872 = [199] -Title = DG Fault: HR On With No Flow -ListTitle = If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. -Message = DG trimmer heater while dialysate flow is low +Title = Service Required: Dialysate Device +ListTitle = DG Fault: HR On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 871 = [200] -Title = DG Fault: Thermistor Range -ListTitle = If thermistor > 80.0 C or < 0.0 C. -Message = DG thermistor or temperature sensor out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Thermistor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 178 = [201] -Title = Process: Pre- Tx Wet Self- Test -ListTitle = If the pre-treatment wet flow test has failed. -Message = HD pre-treatment wet flow test failure +Title = Self Test: Dialysate Flow Rate +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [202] -Title = Process: Pre- Tx Dry Self- Test -ListTitle = If the pre-treatment dry pressure normal test failed. -Message = HD pre-treatment dry pressure normal test failure +Title = Self Test: Pressure +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [203] Title = ListTitle = Message = [204] -Title = DG Fault: RO Rejection -ListTitle = If RO rejection ratio > 0.1 -Message = RO rejection ratio out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: RO Rejection +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 429 = [205] -Title = DG Fault: Cond Sensors Error -ListTitle = If conductivity sensor reports 0x11 for more than 2 consecutive reads. -Message = Conductivity sensor fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Cond Sensors Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 913 = [206] -Title = DG Fault: Reservoir Fill Timeout -ListTitle = If DG dialysate fill > 5 minutes. -Message = DG dialysate fill runs out of time +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Fill Timeout +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 437 = [207] -Title = DG Fault: FMP Flow Rate Error -ListTitle = If integrated Volume To Load Cell Reading Percent > 0.1. -Message = DG flow meter check failure alarm +Title = Service Required: Dialysate Device +ListTitle = DG Fault: FMP Flow Rate Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 240 = [208] -Title = Blood Stop: Venous Pres. High -ListTitle = If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. -Message = HD venous line occlusion fault +Title = High Venous Pressure +ListTitle = Blood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. [209] -Title = Cleaning: Flush Circ Error -ListTitle = If the integrated volume < 0.1 for more than 60 secs. -Message = DG drain circulation line timeout +Title = Disinfection Interrupted +ListTitle = Cleaning: Flush Circ Error +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 980 = [210] -Title = HD Alarm (Low): Battery Fault -ListTitle = If the battery pack status error is detected from the slave device I2C address 0x0B. -Message = HD battery pack detected an error +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Battery Fault +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [211] -Title = Blood Stopped: Warning -ListTitle = During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. -Message = Blood sitting too long warning (> 4 min) +Title = Approaching MAX Blood Pump Stop Time +ListTitle = Blood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback. [212] -Title = Blood Stopped: Warning -ListTitle = During treatment end paused state, if the blood is in the line for more than 4 minutes. -Message = Blood sitting too long warning (> 4 min) +Title = Approaching MAX Blood Pump Stop Time +ListTitle = Blood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must Resume/OK treatment before the blood pump stop \ntime timer expires.\n- If treatment is not Resume/OK before the timer expires,\ntreatment will be terminated without rinseback. [213] -Title = Process: Pre- Tx Prime Complete -ListTitle = In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) -Message = HD prime completed medium priority alarm +Title = Prime Complete - 30 MIN WARNING +ListTitle = Process: Pre- Tx Prime Complete +Message = The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press Resume/OK to continue. [214] Title = @@ -1180,140 +1180,140 @@ Message = [216] -Title = Process: End of Dialysis Tx -ListTitle = If the treatment ended successfully. -Message = HD end treatment sub-mode timeout alarm +Title = Treatment Complete +ListTitle = Process: End of Dialysis Tx +Message = Dialysis treatment is complete.\n- Press RINSEBACK to end treatment and rinseback. [217] Title = ListTitle = Message = [218] -Title = Process: Remove Syringe -ListTitle = If the syringe detection switch > 2.0 volts. -Message = HD syringe detected alarm +Title = Syringe Detected +ListTitle = Process: Remove Syringe +Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press Resume/OK to continue. [219] -Title = Blood Stop: Syringe Not Detected -ListTitle = If the syringe detection switch < 2.0 volts. -Message = HD syringe pump syringe removed alarm +Title = Blood Stopped: Syringe Not Detected +ListTitle = Blood Stop: Syringe Not Detected +Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Resume/OK to resume treatment. [220] -Title = DG Fault: THD RTD Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = DG THd sensor FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: THD RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 947 = [221] -Title = Blood Stop: Saline Bag Empty -ListTitle = If the measured arterial pressure < -300.0 mmHg for more than 250 msec. -Message = HD Saline bag is empty +Title = Blood Pump Stopped: Saline Bag Empty +ListTitle = Blood Stop: Saline Bag Empty +Message = The saline bag appears empty.\n- Check the saline bag and replace if empty. [222] -Title = HD Fault: BP Occ Com -ListTitle = If the FPGA does not report fresh data or indicates an error for a certain period of time. -Message = HD occlusion sensor FPGA fault +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: BP Occ Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [223] -Title = HD Fault: Art. Pres. Com -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = HD arterial sensor FPGA fault +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Art. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [224] -Title = Blood Stopped: Treatment Pause -ListTitle = If the user has pressed the stop button. -Message = Treatment stopped by user action - pressed stop button +Title = Blood Pump Stopped: Treatment Paused +ListTitle = Blood Stopped: Treatment Pause +Message = Treatment has been stopped by user.\n- Press Resume/OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. [225] -Title = Process: End of Treatment -ListTitle = If the treatment duration has exceeded the prescribed treatment time entered by the user. -Message = HD end of treatment warning +Title = Treatment Complete +ListTitle = Process: End of Treatment +Message = Dialysis treatment is complete.\n- If blood sampling is required, draw blood at\n the arterial sample port\n- Press RINSEBACK to end treatment and rinseback. [226] -Title = Process: Pre- Tx Priming Complete -ListTitle = Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection -Message = HD prime completed low priority alarm +Title = Priming Complete +ListTitle = Process: Pre- Tx Priming Complete +Message = Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press Resume/OK to continue. [227] Title = ListTitle = Message = [228] -Title = Process: Pre- Tx Prime Timeout -ListTitle = During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. -Message = HD time out on prime sub-mode purge air state. +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [229] -Title = Process: Pre- Tx Prime Timeout -ListTitle = If the priming of the dialyzer dialysate flow path has exceeded the time limit. -Message = HD prime dialysate dialyzer time out alarm +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [230] -Title = Process: Pre- Tx Prime Timeout -ListTitle = If the priming of the dialyzer bypass flow path has exceeded the time limit. -Message = HD prime dialysate bypass time out alarm +Title = Priming Error +ListTitle = Process: Pre- Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. [231] -Title = Process: Pre- Tx Dry Self- Test -ListTitle = If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState -Message = HD pre-treatment mode dry self-tests failure +Title = Self Test: Cartridge Check +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press Resume/OK to resume. \n- If problem persists, install a new cartridge. [232] -Title = Process: Pre- Tx Wet Self- Test -ListTitle = If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState -Message = HD pre-treatment mode wet self-tests failure +Title = Self Test: Dialysate Flow Rate +ListTitle = Process: Pre- Tx Wet Self- Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge. [233] -Title = DG Fault: Baro Cal CRC -ListTitle = If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. -Message = DG barometric pressure sensor bad coefficients CRC +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Cal CRC +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = Dial. Gen: Inlet Water Pres High -ListTitle = If the inlet water pressure is > 80 psig. -Message = DG inlet water pressure is too high +Title = Incoming Water Pressure High +ListTitle = Dial. Gen: Inlet Water Pres High +Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nResume/OK button will be available.\n- Press Resume/OK to continue treatment. SRSDG 1037 = [235] -Title = Process: Rinseback Complete -ListTitle = If there is no active alarm and rinseback has completed. -Message = HD in treatment stopped sub-mode after rinseback completed (no escalation) +Title = Rinseback Complete +ListTitle = Process: Rinseback Complete +Message = Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press Resume/OK to continue. [236] -Title = Process: Pre- Tx Dry Self- Test -ListTitle = If the air trap sensor reported AIR and bubble sensor detected bubbles. -Message = HD needs new cartridge to be installed +Title = Self Test: Fluid Present +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume. [237] Title = ListTitle = Message = [238] -Title = Process: Pre- Tx Dry Self- Test -ListTitle = If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. -Message = HD no cartridge loaded or installed improperly alarm +Title = Cartridge Not Detected +ListTitle = Process: Pre- Tx Dry Self- Test +Message = A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed. [239] -Title = Process: Post- Tx Remove Cartridge -ListTitle = In the post treatment state, if the cartridge has not been removed. -Message = HD fail to remove cartridge alarm +Title = Cartridge Still Detected +ListTitle = Process: Post- Tx Remove Cartridge +Message = A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed. [240] -Title = Dial. Gen: Bicarb Conc. Cond. -ListTitle = If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. -Message = Bicarb conductivity out of range during bicarb pump check alarm +Title = Conductivity Error During Prime +ListTitle = Dial. Gen: Bicarb Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range. SRSDG 482 = [241] -Title = Cleaning: Resv Drain Timeout -ListTitle = If the DG reservoir drain is not complete in 120 secs. -Message = DG reservoir drain time out +Title = Disinfection Interrupted +ListTitle = Cleaning: Resv Drain Timeout +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 848 SRSDG 90 SRSDG 561 @@ -1344,9 +1344,9 @@ SRSDG 1034 = [242] -Title = Cleaning: Resv Fill Timeout -ListTitle = If the DG reservoir fill is not complete in 120 secs. -Message = DG reservoir fill time out +Title = Disinfection Interrupted +ListTitle = Cleaning: Resv Fill Timeout +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 850 SRSDG 573 SRSDG 574 @@ -1367,81 +1367,81 @@ SRSDG 1024 = [243] -Title = Cleaning: Fluid Loss Detected -ListTitle = If the reservoir level is drifted by 100 mL for more than 30 seconds. -Message = DG reservoir leak time out +Title = Disinfection Interrupted +ListTitle = Cleaning: Fluid Loss Detected +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 853 = [244] -Title = Cleaning: Temp Sensor Check -ListTitle = If any of the temperature sensors are out of the average of the sensors by more than 3 C. -Message = DG temperature sensors difference out of range +Title = Disinfection Interrupted +ListTitle = Cleaning: Temp Sensor Check +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 852 = [245] -Title = Cleaning: Heat Disinfect Temp TO -ListTitle = If heating up to minimum temperature for heat disinfect failed after 4 hours. -Message = DG heat disinfect target temperature time out (could not reach to temperature) +Title = Disinfection Interrupted +ListTitle = Cleaning: Heat Disinfect Temp TO +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 854 = [246] -Title = Cleaning: RO Water Cond Error -ListTitle = If CPo and CD2 conductivity values are above 100 uS/cm. -Message = DG chemical disinfect inlet conducitivty out of range +Title = Disinfection Interrupted +ListTitle = Cleaning: RO Water Cond Error +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. [247] Title = ListTitle = Message = [248] -Title = Cleaning: Chem Disinfect Proc -ListTitle = If TPo > 50 C after 1 second. -Message = DG chemical disinfect out of target temperature +Title = Disinfection Interrupted +ListTitle = Cleaning: Chem Disinfect Proc +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. [249] -Title = Cleaning: Chem Disinf PAA Cond -ListTitle = If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. -Message = DG chemical disinfect out of target conductivity +Title = Chemical Disinfectant Not Detected +ListTitle = Cleaning: Chem Disinf PAA Cond +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair. SRSDG 992 = [250] -Title = Cleaning: Chem Disinf Connect PAA -ListTitle = Start of chemical disinfect -Message = Insert the acid to start chemical disinfect +Title = Connect Chemical Disinfectant +ListTitle = Cleaning: Chem Disinf Connect PAA +Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press Resume/OK to continue. SRSDG 1046 = [251] -Title = HD POST: Sys Rec NVRAM CRC -ListTitle = If the calculated CRC of the HD system record image does not match the expected CRC. -Message = HD invalid system record +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Sys Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [252] -Title = HD POST: Serv Rec NVRAM CRC -ListTitle = If the calculated CRC of the HD service record image does not match the expected CRC. -Message = HD invalid service record +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Serv Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [253] -Title = DG POST: Sys Rec NVRAM CRC -ListTitle = If the calculated CRC of the DG system record image does not match the expected CRC. -Message = DG invalid system record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Sys Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 609 = [254] -Title = DG POST: Serv Rec NVRAM CRC -ListTitle = If the calculated CRC of the DG service record image does not match the expected CRC. -Message = DG invalid service record +Title = Service Required: Dialysate Device +ListTitle = DG POST: Serv Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 610 = [255] -Title = HD POST: UI Compatibility -ListTitle = If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. -Message = HD and UI software builds are not compatible +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Compatibility +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [256] -Title = Cleaning: Chem Flush In Progress -ListTitle = If DG flush successfully started from the UI. -Message = DG flush started +Title = Chemical Flush in Progress +ListTitle = Cleaning: Chem Flush In Progress +Message = Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. = SRSUI 497 [257] @@ -1455,84 +1455,84 @@ Message = [259] -Title = HD Fault: Temp Sensors Error -ListTitle = If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. -Message = HD temperatures out of range +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Temp Sensors Error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [260] -Title = HD POST: UI Image -ListTitle = If the UI POST reports a shasum faillure. -Message = UI POST Application Integrity (Sha256Sum) failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1067 [261] -Title = HD POST: UI CAN Driver -ListTitle = If the UI POST reports a can bus failure. -Message = UI POST CANBus failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI CAN Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1068 [262] Title = -ListTitle = If the UI POST reports a display failure. -Message = UI POST Display failure +ListTitle = +Message = [263] -Title = HD POST: UI Touch Driver -ListTitle = If the UI POST reports a touch failure. -Message = UI POST Touch failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Touch Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1070 [264] -Title = HD POST: UI SD- Card -ListTitle = If the UI POST reports SD card failure. -Message = UI POST SD-Card failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI SD- Card +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1071 [265] -Title = HD POST: UI RTC -ListTitle = If the UI POST reports RTC failure. -Message = UI POST RTC failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI RTC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1072 [266] -Title = HD POST: UI WiFi -ListTitle = If the UI POST reports WiFi failure. -Message = UI POST WiFi failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI WiFi +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1073 [267] -Title = HD POST: UI Bluetooth -ListTitle = If the UI POST reports bluetooth failure. -Message = UI POST Bluetooth failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Bluetooth +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1074 [268] -Title = HD POST: UI Ethernet -ListTitle = If the UI POST reports ethernet failure. -Message = UI POST Ethernet failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Ethernet +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1075 [269] -Title = HD POST: UI Sound -ListTitle = If the UI POST reports sound failure. -Message = UI POST Sound failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI Sound +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1076 [270] -Title = HD POST: Proc Safestate -ListTitle = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. -Message = HD POST Safety Shutdown failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Proc Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [271] -Title = DG POST: Proc Safestate -ListTitle = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. -Message = DG POST Safety Shutdown failure +Title = Service Required: Dialysate Device +ListTitle = DG POST: Proc Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 822 = [272] -Title = HD Alarm (Low): Fan Speed -ListTitle = If the fan RPM is out of Min/Max range. -Message = HD Fan RPM out of range +Title = Service Required: Hemodialysis Device +ListTitle = HD Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair. [273] Title = @@ -1545,279 +1545,279 @@ Message = [275] -Title = HD Fault: Arterial Press Range -ListTitle = If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. -Message = HD arterial pressure sensor is reading out of range +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Arterial Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [276] -Title = HD Fault: Venous Press Range -ListTitle = If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. -Message = HD venous pressure sensor is reading out of range +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Venous Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [277] -Title = HD Fault: Occlusion Sens Range -ListTitle = If the blood pump occlusion count is < 2000. -Message = HD BP occlusion sensor is reading out of range +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Occlusion Sens Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [278] -Title = DG Fault: Reservoir Overfill -ListTitle = If the active reservoir weight is > 1900.0 g. -Message = HD active reservoir weight out of range. +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Overfill +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [279] -Title = DG Fault: Reservoir Drain -ListTitle = If the DG dialysate drain exceeded 120 seconds. -Message = DG dialysate drain time out +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Reservoir Drain +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 752 = [280] -Title = DG Fault: DRP Direction Fault -ListTitle = If the drain pump's direction is not 1 when draining. -Message = DG drain pump invalid direction +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Direction Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 888 = [281] -Title = Dial. Gen: Acid Conc. Low -ListTitle = If the acid concentration bottle volume > 3240 (3340 - 100). -Message = DG Acid concentration bottle low volume alarm +Title = Acid Concentrate Low +ListTitle = Dial. Gen: Acid Conc. Low +Message = The acid concentrate is low.\n- Replace acid concentrate container. SRSDG 836 = [282] -Title = Dial. Gen: Bicarb Conc. Low -ListTitle = If the bicarbonate concentration bottle volume > 3680 (3780 - 100). -Message = DG Bicarbonate concentration bottle low volume alarm +Title = Bicarb Concentrate Low +ListTitle = Dial. Gen: Bicarb Conc. Low +Message = Bicarbonate concentrate is low.\n- Replace the bicarbonate concentrate container. SRSDG 837 = [283] -Title = DG Fault: LC Range -ListTitle = If the weight of the DG load cells > 4500.0 g or < 0.0 g. -Message = DG load cells weight out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 615 = [284] -Title = DG Fault: Redundant LC Check -ListTitle = If redundant load cells deviate by more than 60g after tare for 5 seconds. -Message = DG load cells primary/back up drift out of range +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Redundant LC Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 383 = [285] -Title = Cleaning: RO Permeate Sample In Progress -ListTitle = If DG RO permeate sample successfully started from UI. -Message = DG RO permeate sample started +Title = RO Permeate Sample in Progress +ListTitle = Cleaning: RO Permeate Sample In Progress +Message = RO Permeate Sample is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [286] -Title = Cleaning: Conc Cap Removed -ListTitle = If the concentrate cap is not in the right position in the cleaning modes. -Message = DG cap in wrong state +Title = Concentrate Connection Cap Removed +ListTitle = Cleaning: Conc Cap Removed +Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. [287] -Title = HD Fault: Proc- FPGA Clock -ListTitle = If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. -Message = HD processor clock speed checks against FPGA clock failure +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Proc- FPGA Clock +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [288] -Title = DG Fault: LC Drift Check -ListTitle = If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. -Message = HD load cell inconsistency +Title = Service Required: Dialysate Device +ListTitle = DG Fault: LC Drift Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [289] -Title = Cleaning: Dialysate Cap Removed -ListTitle = If the dialysate cap is not in the right position in the cleaning modes. -Message = DG cap in wrong state +Title = Dialysate Connection Cap Removed +ListTitle = Cleaning: Dialysate Cap Removed +Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. SRSDG 928 = [290] -Title = HD Fault: Ven. Pres. Com -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = HD venous sensor FPGA fault +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Ven. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [291] -Title = Cleaning: Flush In Progress -ListTitle = If DG flush successfully started from the UI. -Message = DG flush started +Title = Flush In Progress +ListTitle = Cleaning: Flush In Progress +Message = System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [292] -Title = Cleaning: Heat Disinf In Progress -ListTitle = If DG heat disinfect successfully started from the UI. -Message = DG heat disinfect started +Title = Heat Disinfection in Progress +ListTitle = Cleaning: Heat Disinf In Progress +Message = Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [293] -Title = Cleaning: Chem Disinfect In Progress -ListTitle = If DG flush successfully started from the UI. -Message = DG chem disinfect started +Title = Chemical Disinfection in Progress +ListTitle = Cleaning: Chem Disinfect In Progress +Message = Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [294] -Title = Cleaning: Active Cool In Progress -ListTitle = If DG active cool successfully started. -Message = DG active cool started +Title = Active Cool in Progress +ListTitle = Cleaning: Active Cool In Progress +Message = Active Cool is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device. [295] -Title = HD POST: UI CloudSync -ListTitle = If the UI POST reports cloud sync failure. -Message = UI cloud sync POST failure +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: UI CloudSync +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. = SRSUI 1238 [296] -Title = DG Fault: TDi RTD Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = DG TDi sensor FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TDi RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 951 = [297] -Title = DG Fault: TRo RTD Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = DG TRo sensor FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: TRo RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 950 = [298] -Title = DG Fault: Baro Pres Comm -ListTitle = If the FPGA does not report fresh data for a certain period of time -Message = DG barometric sensor FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Baro Pres Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 948 = [299] -Title = DG POST: Serial Number Check -ListTitle = If the DG serial number is empty and does not have any characters. -Message = DG invalid serial number +Title = Service Required: Dialysate Device +ListTitle = DG POST: Serial Number Check +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 952 = [300] -Title = HD POST: Serial Number Check -ListTitle = If the HD serial number is empty and does not have any characters. -Message = HD invalid serial number +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Serial Number Check +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [301] -Title = Cleaning: RO Permeate Sample Connect Cap -ListTitle = RO permeate sample collection state -Message = Open dialysate cap to be able to collecte the dispensing fluid +Title = Remove Dialysate Cap +ListTitle = Cleaning: RO Permeate Sample Connect Cap +Message = Remove the dialysate cap\n - Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device. [302] -Title = DG Fault: Processor RAM error -ListTitle = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. -Message = DG_CPU RAM error +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Processor RAM error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 126 = [303] -Title = DG Fault: DRP Hall Commutation -ListTitle = If the FPGA indicates an error for a certain period of time. -Message = DG drain pump direction FPGA fault +Title = Service Required: Dialysate Device +ListTitle = DG Fault: DRP Hall Commutation +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 958 = [304] -Title = HD POST: Usage Rec NVRAM CRC -ListTitle = If the CRC of the HD usage record does not match the calculated CRC. -Message = HD invalid usage reocord +Title = Service Required: Hemodialysis Device +ListTitle = HD POST: Usage Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [305] -Title = HD Fault: Processor RAM error -ListTitle = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. -Message = HD_CPU RAM error +Title = Service Required: Hemodialysis Device +ListTitle = HD Fault: Processor RAM error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [306] -Title = DG Fault: CD1 CD2 Check -ListTitle = CD1 and CD2 conductivity measurement during acid check is not within 5% of each other -Message = +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD1 CD2 Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [307] -Title = Blood Stop: Loss of AC Power -ListTitle = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. -Message = HD AC power lost alarm +Title = Power Loss +ListTitle = Blood Stop: Loss of AC Power +Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed. [308] -Title = DG Fault: CPi Comm -ListTitle = If the CPi conductivity sensor transmits invalid characters for a certain period of time. -Message = DG CPi conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPi Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 959 = [309] -Title = DG Fault: CPo Comm -ListTitle = If the CPo conductivity sensor transmits invalid characters for a certain period of time. -Message = DG CPo conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 960 = [310] -Title = DG Fault: CD1 Comm -ListTitle = If the CD1 conductivity sensor transmits invalid characters for a certain period of time. -Message = DG CD1 conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 962 = [311] -Title = DG Fault: CD2 Comm -ListTitle = If the CD2 conductivity sensor transmits invalid characters for a certain period of time. -Message = DG CD2 conductivity sensor transmits invalid characters +Title = Service Required: Dialysate Device +ListTitle = DG Fault: CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 961 = [312] -Title = Dial. Gen: Dialysate In Flow -ListTitle = If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. -Message = The dialysate inlet flow is too low. +Title = Low Dialysate Flow +ListTitle = Dial. Gen: Dialysate In Flow +Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nResume/OK button will be available.\n- Press Resume/OK to continue. [313] -Title = Cleaning: Inlet Water Temp Low -ListTitle = If inlet water temperature is < 24 C. -Message = DG cleaning mode inlet water temperature too low +Title = Disinfection: Water Temperature Low +ListTitle = Cleaning: Inlet Water Temp Low +Message = Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = Cleaning: Inlet Water Press High -ListTitle = If inlet water temperature is > 80 psig. -Message = DG cleaning mode inlet water pressure too high +Title = Disinfection: Water Pressure High +ListTitle = Cleaning: Inlet Water Press High +Message = Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = Cleaning: Inlet Water Press Low -ListTitle = If inlet water pressure is < 14 psig. -Message = DG cleaning mode inlet water pressure too low +Title = Disinfection: Water Pressure Low +ListTitle = Cleaning: Inlet Water Press Low +Message = Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water initiating disinfection again. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = DG Fault: Acid CP Pump Park -ListTitle = If the park took more than 1 second. -Message = Acid concentrate pump did not park successfully +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Acid CP Pump Park +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 983 = [317] -Title = DG Fault: Bicarb CP Pump Park -ListTitle = If the park took more than 1 second. -Message = Bicarb concentrate pump did not park successfully +Title = Service Required: Dialysate Device +ListTitle = DG Fault: Bicarb CP Pump Park +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 984 = [318] -Title = HD Fault: Air Pump Timeout -ListTitle = If air pump has been turned on for greater than 10s. -Message = HD Air Pump operation timeout +Title = Air Trap Control Error +ListTitle = HD Fault: Air Pump Timeout +Message = A problem has been detected with the air trap level\ncontrol system.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue. [319] -Title = Cleaning: Chem Residual Sampling -ListTitle = Chemical disinfect flush sample alarm is prompted. -Message = Collect sample to check the chemical disinfectant acid +Title = Residual Chemical Sampling +ListTitle = Cleaning: Chem Residual Sampling +Message = A water sample is needed to confirm the \nabsence of residual chemical following disininfection. \n- Test sample according to facility policy. \n- Confirm absence of residual chemical disinfectant. SRSDG 1040 = [320] -Title = Cleaning: Chem Residual Timeout -ListTitle = Chemical disinfect flush sample was not collected form more than 30 minutes. -Message = The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes +Title = Chemical Residual Timeout +ListTitle = Cleaning: Chem Residual Timeout +Message = A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush. SRSDG 1043 = [321] -Title = DG Fault: UVo Error -ListTitle = If the outlet reactor are not healthy. -Message = Outlet reactor not healthy +Title = Service Required: Dialysate Device +ListTitle = DG Fault: UVo Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. SRSDG 1069 = [322] -Title = HD POST Alarm: UI Invalid Year -ListTitle = If the UI POST reports a invalid year. -Message = UI POST Invalid Year +Title = Service Required: Hemodialysis Device +ListTitle = HD POST Alarm: UI Invalid Year +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. [323] -Title = HD POST Alarm: UI Bad Settings -ListTitle = If the UI POST reports a bad settings. (e.g. no settings read, empty settings file, invalid/empty parms, CRC failure) -Message = UI POST Bad Settings +Title = Service Required: Hemodialysis Device +ListTitle = HD POST Alarm: UI Bad Settings +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.