Index: Alarms/Alarms.conf =================================================================== diff -u -r801eb9280285a442d21be69da25d961638fbdf1f -rd0fe758eeff15a03d7b106b7db9cf824e20dafe4 --- Alarms/Alarms.conf (.../Alarms.conf) (revision 801eb9280285a442d21be69da25d961638fbdf1f) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision d0fe758eeff15a03d7b106b7db9cf824e20dafe4) @@ -1,959 +1,959 @@ [0] -Title = Indication for when no alarms is a possible situation +Title = ` Message = [1] -Title = Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up -Message = Stop or Off button are pressed for more than 1 sec during POST. +Title = HD POST: Stuck Button +Message = Stop or On button pressed during startup\n-Release button and power cycle\n-Call service to schedule a repair if problem persists [2] -Title = HD FPGA POST failure -Message = Minimum HD FPGA major revision is not compatible with this version of HD firmware or the FPGA Id is not correct. +Title = HD POST: FPGA Compatibility +Message = Call service to schedule a repair. [3] -Title = DG FPGA POST failure -Message = Minimum DG FPGA major and minor revision are not equal to the FPGA major and minor revision in FPGA header +Title = DG POST: FPGA Compatibility +Message = Call service to schedule a repair. SRSDG 602 SRSDG 814 SRSDG 1061 = [4] -Title = HD Watchdog POST failure -Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V 3. audio current >= 200mA If the following are not observed after 500ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V 3. audio current <= 10 mA +Title = HD POST: WD Safestate +Message = Call service to schedule a repair. [5] -Title = DG Watchdog POST failure -Message = If the following are not observed after 500ms of not petting the WD: 1. WD exp pin=low 2. 24V <= 5V If the following are not observed after 750ms of resumed WD petting at proper timing: 1. WD exp pin=high 2. 24V >= 22.6V +Title = DG POST: WD Safestate +Message = Call service to schedule a repair. SRSDG 53 = [6] -Title = UI communication POST failure. UI failed to communicate within a reasonable time after power up -Message = If the UI failed to checkin via the CAN bus within 40 seconds +Title = HD POST: Proc-UI Comm +Message = Call service to schedule a repair. [7] -Title = Real-time clock configuration error -Message = If the RTC is not functional as indicated by the bits in the first 3 control registers. +Title = HD Alarm (Low): RTC Bat Low +Message = Call service to schedule a repair. [8] -Title = HD accelerometer failed POST -Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. +Title = HD POST: Acc Vector +Message = Call service to schedule a repair. [9] -Title = DG accelerometer failed POST -Message = If the calculated vector length of the accelerometer x, y, z is not the nominal length. +Title = DG POST: Acc Vector +Message = Call service to schedule a repair. SRSDG 323 = [10] -Title = HD blood leak sensor embedded mode communication failed -Message = If the blood leak sensor embedded mode is unresponsive. +Title = HD POST: BLD Config Emb Mode +Message = Call service to schedule a repair. [11] -Title = Dialysate temperature is below the target temperature -Message = If the dialysate temperature is below the target temperature for a certain period of time. +Title = Dial. Gen: Dialysate Temp Low +Message = Dialysate temperature is low\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [12] -Title = HD firmware image integrity POST test failed -Message = If the calculated CRC of the HD image does not match the expected CRC. +Title = HD POST: HD Image +Message = Call service to schedule a repair. [13] -Title = DG firmware image integrity POST test failed -Message = If the calculated CRC of the DG firmware image does not matched the expected CRC +Title = DG POST: DG Image +Message = Call service to schedule a repair. SRSDG 813 = [14] -Title = DG non-volatile usage information group invalid record CRC -Message = If the DG usage information record's CRC does not match the calculated CRC in POST. +Title = DG POST: Usage Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 957 = [15] -Title = Dialysate temperature is out of high safety range -Message = If the dialysate temperature is > 46 C for a certain period of time. +Title = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high\n-Treatment is paused and will be ready to resume when \ndialysate is within target range. [16] -Title = HD alarm audio failed POST -Message = If alarm audio current is NOT in expected range for than 1 second while alarm tone is being output +Title = HD POST: Primary Alarm Current +Message = Call service to schedule a repair. [17] -Title = HD UI POST failed -Message = If UI reports POST failed but did not trigger fault OR UI has not sent POST results within 2 seconds. +Title = HD POST: UI POST Not Complete +Message = Call service to schedule a repair. [18] -Title = HD to DG comm error. -Message = HD message to DG that requires acknowledge was not acknowledged within a certain time. +Title = Loss of DG Comm (DG ACK) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [19] -Title = Dialysate temperature is above the target temperature -Message = If the dialysate temperature is above the target temperature for a certain period of time. +Title = Dial. Gen: Dialysate Temp High +Message = Dialysate temperature is high\n- Treatment is paused and will be ready to resume when \ndialysate is within target range. [20] -Title = DG conductivity sensors invalid calibration record -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Title = DG POST: Conductivity Cal +Message = Call service to schedule a repair. SRSDG 16 = [21] -Title = DG drain line volume invalid calibration record -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Title = DG POST: Drain Line Vol Cal +Message = Call service to schedule a repair. SRSDG 936 = [22] -Title = DG reservoirs invalid calibration record -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Title = DG POST: Reservoir Vol Cal +Message = Call service to schedule a repair. SRSDG 935 = [23] -Title = DG acid concentrate calibration record -Message = If the non-volatile data is valid and if not raise the alarm +Title = DG POST: Acid Conc Cal +Message = Call service to schedule a repair. SRSDG 934 = [24] -Title = DG bicarb concentrate calibration record -Message = If the non-volatile data is valid and if not raise the alarm +Title = DG POST: Bicarb Conc Cal +Message = Call service to schedule a repair. SRSDG 933 = [25] -Title = DG accelerometer invalid calibration record -Message = If the calibration data that was received from NVDataMgmt is NOT legitimate +Title = DG POST: Acc Cal +Message = Call service to schedule a repair. SRSDG 817 = [26] -Title = HD accelerometer invalid calibration record -Message = If the accel calibration data that was received from NVDataMgmt is NOT legitimate. +Title = HD POST: Acc Cal +Message = Call service to schedule a repair. [27] -Title = HD BLD sensor failed zero sequence -Message = BLD fails to complete zero sequence. +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test detected has a problem with the blood leak sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [28] -Title = DG two wire sensors FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = DG Fault: Conductivity RTD Comm +Message = Call service to schedule a repair. SRSDG 864 = [29] -Title = HD heparin force sensor invalid calibration record -Message = If the heparin force sensor calibration data that was received from NVDataMgmt is NOT legitimate. +Title = HD POST: SP DAC Cal +Message = Call service to schedule a repair. [30] -Title = HD Software fault. Software found itself in an unexpected state -Message = If the specific software faults listed in alarm management SW faults has been detected +Title = HD Fault: SW Invalid State +Message = Call service to schedule a repair. [31] -Title = Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = If the measured MC current is not greater than BP_MAX_CURR_WHEN_STOPPED_MA or is not greater than BP_MAX_CURR_WHEN_RUNNING_MA. +Title = HD Fault: BP Current +Message = Call service to schedule a repair. [32] -Title = Blood pump failed motor off check. Measured speed while commanded off -Message = If the measured motor speed is greater than BP_MAX_MOTOR_SPEED_WHILE_OFF_RPM +Title = HD Fault: BP Speed +Message = Call service to schedule a repair. [33] -Title = Blood pump failed motor direction check. Measured vs. commanded -Message = If the set direction versus hall sensors direction versus sign of motor speed DOES NOT agree. +Title = HD Fault: BP Direction +Message = Call service to schedule a repair. [34] -Title = Blood pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = If the delta of the measured motor speed and the commanded motor speed is not within tolerance. +Title = HD Fault: BP Rotor +Message = Blood pump rotor failure.\n- Check the pump rotor for secure connection. [35] -Title = Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = If the Motor Controller measured DPi pump current > 150.0 mA for more than 250.0 msec. +Title = HD Fault: DPi Current +Message = Call service to schedule a repair. [36] -Title = Dialysis inlet pump failed motor off check. Measured speed while commanded off -Message = If during command off state, the DPi pump motor speed is measured > 100.0 RMP. +Title = HD Fault: DPi Speed +Message = Call service to schedule a repair. [37] -Title = Dialysis inlet pump failed motor direction check. Measured vs. commanded -Message = If the set direction versus hall sensors direction DO NOT agree for more than 250.0 msec. +Title = HD Fault: DPi Direction +Message = Call service to schedule a repair. [38] -Title = Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = If the measured motor speed versus the measured motor speed while controlling the target > 5.0 for more than 12 seconds. +Title = HD Fault: DPi Rotor +Message = Dialysate in pump rotor failure.\n- Check the pump rotor for secure connection. [39] -Title = Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running -Message = If the DPo pump current > 150.0 mA while stopped or Dpo current > 2000.0 mA for more than 2 seconds while running. +Title = HD Fault: DPo Current +Message = Call service to schedule a repair. [40] -Title = Dialysis outlet pump failed motor off check. Measured speed while commanded off -Message = If the Dpo pump current > 100.0 RPM while commanded off for more than 5000 msec. +Title = HD Fault: DPo Speed +Message = Call service to schedule a repair. [41] -Title = Dialysis outlet pump failed motor direction check. Measured vs. commanded -Message = If the DPo pump set direction does not agree with hall sensors or set direction does not agree with the sign of the motor speed for more than 25 msec. +Title = HD Fault: DPo Direction +Message = Call service to schedule a repair. [42] -Title = Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds -Message = If DPo pump measured rotor speed does not agreed with the measured/commanded motor speed +Title = HD Fault: DPo Rotor +Message = Dialysate out pump rotor failure.\n- Check the pump rotor for secure connection. [43] -Title = Main primary heater FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = DG Fault: Iso ADC Comm +Message = Call service to schedule a repair. SRSDG 954 = [44] -Title = User interface communication timeout -Message = If the sub-system communication timeout has occurred. UI is 5 seconds. DG is 2 seconds. +Title = HD Fault: UI Comm +Message = Call service to schedule a repair. [45] -Title = Too many bad CRCs detected on received system messages -Message = If too many bad messages CRCs are received within a set period of time +Title = HD Fault: CAN CRC Com +Message = Call service to schedule a repair. [46] -Title = System message that required acknowledgment was not acknowledged -Message = If no retries left in the pending acknowlegement list. +Title = HD Fault: CAN ACK UI Com +Message = Call service to schedule a repair. [47] Title = Message = [48] -Title = Ultrafiltration volume accuracy error during treatment -Message = If the UF volume (refUFvolume - measureVolume) > 100.0 mL. +Title = HD Fault: UF Rate Tare Err +Message = Call service to schedule a repair. [49] -Title = HD FPGA communication down for too long -Message = If there are more than 3 comm errors in a 1 min period. +Title = HD Fault: FPGA Com +Message = Call service to schedule a repair. [50] -Title = DG FPGA not accepting commanded valve states -Message = If the read back FPGA valve states does not match the commanded valve states for a certain amount of time. +Title = DG Fault: Valve Control +Message = Call service to schedule a repair. SRSDG 26 = [51] -Title = Blood pump failed motor speed check. Measured vs. commanded -Message = If measured motor speed versus commanded rotor speed exceeded limits. +Title = HD Fault: BP Speed +Message = Call service to schedule a repair. [52] -Title = Dialysate inlet pump failed motor speed check. Measured vs. commanded -Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. +Title = HD Fault: DPi Speed +Message = Call service to schedule a repair. [53] -Title = Dialysate outlet pump failed motor speed check. Measured vs. commanded -Message = If measured motor speed versus commanded motor speed exceeded limits for more than 5 secs. +Title = HD Fault: DPo Speed +Message = Call service to schedule a repair. [54] -Title = HD critical data integrity check failed -Message = If the critical data failed range check. +Title = HD Fault: Critical Data Err +Message = Call service to schedule a repair. [55] -Title = DG critical data integrity check failed -Message = If the critical data failed range check. +Title = DG Fault: Critical Data Err +Message = Call service to schedule a repair. SRSDG 921 = [56] -Title = HD accelerometer error (no readings or FPGA reports error) -Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. +Title = HD Fault: Acc Com +Message = Call service to schedule a repair. [57] -Title = DG accelerometer error (no readings or FPGA reports error) -Message = If no fresh sample of accelerometer data has been detected for more than 100 msec. +Title = DG Fault: Acc Com +Message = Call service to schedule a repair. SRSDG 869 SRSDG 870 = [58] -Title = HD valve homing failed -Message = If the number of failed homing > = 3 counts. +Title = HD Fault: Valve Home +Message = Call service to schedule a repair. [59] -Title = HD valve transition time out -Message = if the valve's transition time has timed out (3000msec) +Title = HD Fault: Valve Transition Err +Message = Call service to schedule a repair. [60] -Title = HD valve not functional -Message = If the valve bit masks are not as expected in different valve control mode. +Title = HD Fault: Valve Control +Message = Call service to schedule a repair. [61] -Title = HD valve current out of range -Message = If the valve current is over the threshold of 1.0 A for 1000 msec. +Title = HD Fault: Valve Current +Message = Call service to schedule a repair. [62] -Title = HD valve position out of target -Message = If the current position has deviated from the expected position for more than 1 second. +Title = HD Fault: Valve Position +Message = Call service to schedule a repair. [63] -Title = FPGA reports error -Message = If the FPGA error counter increments for a certain period of time. +Title = HD Fault: BLD Com +Message = Call service to schedule a repair. [64] -Title = Barometric pressure is out of range -Message = If the barometric pressure is > 10.1 psi or < 15.4 psi for a certain period of time. +Title = DG Fault: Baro Range +Message = Call service to schedule a repair. SRSDG 874 = [65] -Title = HD requests DG command with invalid parameter fault -Message = If the latest DG command response indicated invalid parameter +Title = Loss of DG Comm (Reject) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [66] -Title = HD blood leak sensor set point set failure -Message = If the set point of the blood leak sensor does not match the set point in the calibration record. +Title = HD POST: BLD Config Set Point +Message = Call service to schedule a repair. [67] -Title = HD blood pump occlusion self-test failure alarm -Message = If the blood pump pressure > 5000 +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [68] -Title = HD active reservoir recirculation out of range -Message = If the recirculation level has exceeded the max allowed +Title = Dial. Gen: Dialysate not ready +Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. [69] -Title = HD blood leak sensor invalid calibration record -Message = If the BLD data that was received from NVDataMgmt is NOT legitimate. +Title = HD POST: BLD Cal +Message = Call service to schedule a repair. [70] -Title = HD arterial pressure self-test failure alarm -Message = If arterial pressure < -300 or > 100 mmHg +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [71] -Title = HD venous pressure self-test failure alarm -Message = If the venous pressure < -100 or > 600 mmHg +Title = Process: Pre-Tx Remove Cartridge +Message = Cartridge has been detected.\n- Open door\n- Remove cartridge\n- Close door\n- Resume [72] -Title = HD No load cell data message receive at least once every 2 seconds -Message = No new load cell data for 2 seconds. +Title = Loss of DG Comm (LC) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [73] -Title = HD No dialysate temperature data message received at least once every 2 seconds -Message = No new dialysate temperature data for 2 seconds. +Title = Loss of DG Comm (TDI/TRO) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [74] -Title = DG inlet UV reactor is on while there is no flow -Message = If the inlet UV reactor is on while VPi is closed for a certain period of time. +Title = DG Fault: UVi On No Flow +Message = Call service to schedule a repair. SRSDG 600 = [75] -Title = HD syringe pump self-test failure alarm -Message = If the calibration data that was received from NVDataMgmt and the DAC voltage with syringe retracted are not within tol. +Title = HD Fault: Syringe DAC +Message = Call service to schedule a repair. [76] -Title = HD monitored voltage is out of range -Message = If the HD monitored voltages are out of range for a certain period of time. +Title = HD Fault: Voltage Monitor +Message = Call service to schedule a repair. [77] -Title = DG monitored voltage is out of range -Message = If the DG monitored voltages are out of range for a certain period of time. +Title = DG Fault: Voltage Monitor +Message = Call service to schedule a repair. SRSDG 46 SRSDG 866 SRSDG 1088 = [78] -Title = HD syringe pump direction (from encoder) error -Message = If the measured encoder direction of the syringe pump does not agree with the expected direction +Title = HD Fault: Syringe Pump Enc Dir +Message = Call service to schedule a repair. [79] -Title = HD syringe pump direction (from controller) error -Message = If the measured direction of the syringe pump motor controller does not agree with the expected direction +Title = HD Fault: Syringe Pump Drv Dir +Message = Call service to schedule a repair. [80] -Title = HD syringe pump fault reported by FPGA -Message = If the FPGA reported the syringe pump fault +Title = HD Fault: Syringe Pump Driver +Message = Call service to schedule a repair. [81] -Title = HD syringe pump over travel error -Message = If the syringe pump position is deemed to have overtraveled +Title = HD Fault: Syringe Pump OT +Message = Call service to schedule a repair. [82] -Title = HD syringe pump DAC write failure -Message = If the syringe pump DAC write error is reported by the FPGA +Title = HD Fault: Syringe Pump DAC Err +Message = Call service to schedule a repair. [83] -Title = HD syringe pump is running while the BP is off -Message = If the syringe pump is on while the blood pump is off +Title = HD Fault: Syringe Pump On Err +Message = Call service to schedule a repair. [84] -Title = DG RTC set year is invalid -Message = If DG RTC set year is < 2021. +Title = DG POST: HD RTC Time +Message = Call service to schedule a repair. SRSDG 815 = [85] -Title = HD pump track lever is not fully latched -Message = If the pump track lever is detected as unlatched when it should be. +Title = Blood Stop: Cartridge Lever Open +Message = Cartridge Lock Lever is not closed.\n- Open front door and push lever all the way down.\n- Close door and press RESUME button to continue. [86] -Title = HD RTC set year is invalid -Message = If HD RTC set year is < 2021. +Title = HD POST: HD RTC Time +Message = Call service to schedule a repair. [87] -Title = DG heating parameters invalid calibration record -Message = If the heating paramters' non-volatile record's CRC does not match the calculated CRC. +Title = DG POST: Heat Cal +Message = Call service to schedule a repair. SRSDG 929 = [88] -Title = DG concentrate pump hall sensor out of range. -Message = Hall sensor pulse to pulse timing is < 1ms (speed too fast). +Title = DG Fault: CP1 CP2 Speed Range +Message = Call service to schedule a repair. [89] -Title = DG outlet UV reactor is on while there is no flow -Message = If the outlet UV reactor is on while the RO flow is less than 200 mL/min. +Title = DG Fault: UVo On No Flow +Message = Call service to schedule a repair. SRSDG 599 = [90] -Title = DG load cells A1/B1 FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time. +Title = DG Fault: LC A1/B1 Comm +Message = Call service to schedule a repair. SRSDG 604 = [91] -Title = HD No dialysate flow data received in the last 3 seconds -Message = If not receiving new dialysate flow readings in expected interval (3 seconds) +Title = Loss of DG Comm (FMD) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [92] Title = Message = [93] -Title = DG outlet primary conductivity sensor out of range -Message = If the outlet primary conductivity sensor, CPo is greater than 30 uS/cm for a certain period of time. +Title = DG Fault: RO Permeate Cond High +Message = Call service to schedule a repair. SRSDG 430 = [94] -Title = DG pressure sensor out of range -Message = If either of the pressure sensors is < 0 psia or > 200 psia for DG pressure sensors. +Title = DG Fault: Pressure Sensor Range +Message = Call service to schedule a repair. SRSDG 629 = [95] -Title = DG watchdog expired -Message = If the watchdog signal is low while not in POST. +Title = DG Fault: WD Error +Message = Call service to schedule a repair. SRSDG 907 = [96] -Title = DG inlet water temperature out of range -Message = If the inlet water temperature is > 37 C. +Title = Dial. Gen: Inlet Water Temp High +Message = Inlet water temperature is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 410 = [97] -Title = DG fill conductivity out of range -Message = If the average acid or bicard conductivity is > 5% of the normal acid and bicarb conductivity values. +Title = Dial. Gen: Dialysate Cond. +Message = Dialysate conductivity is out of range.\n- Check acid and bicarb concentrate is correct and not empty. SRSDG 397 = [98] -Title = HD battery or battery charger communication fault -Message = If battery I2C communication fault persistence count > 5. +Title = HD Alarm (Low): Battery Com +Message = Call service to schedule a repair. [99] -Title = HD syringe pump stall alarm -Message = If syringe pump measured rate < stall speed threshold (0.05) and the ramp up has been restarted 3 times after stall has been detected. +Title = HD Fault: Syringe Pump Stall +Message = Call service to schedule a repair. [100] -Title = HD no cartridge self-test timeout -Message = If during the no cartridge self-tests state, the elapsed self-test time > maximum no cartridge self test time (30secs) +Title = HD Fault: Timeout No Cart ST +Message = Call service to schedule a repair. [101] -Title = UI POST FW Communication failure -Message = If the HD is not communicating with the UI during POST. +Title = HD POST: UI-Proc Comm +Message = Call service to schedule a repair. [102] -Title = DG heat disinfection temperature gradient out of range -Message = If temperature gradient between Tpo and Thd exceeds maximum 15°C. +Title = Cleaning: Tpo THD Temp Delta +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 856 SRSDG 930 = [103] -Title = HD non-volatile calibration group invalid record CRC -Message = If the computed CRC of the calibration group record is not equal to the record CRC. +Title = HD POST: Cal NVRAM CRC +Message = Call service to schedule a repair. [104] -Title = Air trap level sensors reporting illegal combination of air/fluid -Message = If the air trap level readings are not valid for more than 2 seconds. +Title = HD Fault: Air Trap Sensor +Message = Call service to schedule a repair. [105] -Title = DG non-volatile system group invalid record CRC -Message = If the DG NVDATAMGMT CAL group calculated CRC does not mached the expected CRC. +Title = DG POST: Cal NVRAM CRC +Message = Call service to schedule a repair. SRSDG 608 = [106] -Title = DG flow sensors calibration error -Message = If the calculated CRC of the flow sensors calibration record does not match the calibration records' CRC. +Title = DG POST: Flow Sensor Cal +Message = Call service to schedule a repair. SRSDG 820 SRSDG 821 SRSDG 1087 = [107] -Title = HD reports DG restarted fault -Message = If DG reports being in any mode other than idle, drain, fill or fault during pre-tx, tx, or post-tx modes - means DG restarted unexpectedly and is no longer in sync w/ HD. +Title = DG Fault: DG Restarted +Message = Call service to schedule a repair. [108] -Title = HD syringe pump ADC error -Message = If the ADC read is not fresh after POST init and the ADC read counter has not changed. +Title = HD Fault: Syringe Pump ADC Com +Message = Call service to schedule a repair. [109] -Title = HD syringe pump volume check error -Message = If the volume delivered is off by more than 5% or 0.1mL. +Title = HD Fault: Syringe Pump Vol +Message = Call service to schedule a repair. [110] -Title = HD syringe pump speed check error -Message = If the syringe measured rate is off by more than 5% or 0.1mL. +Title = HD Fault: Syringe Pump Speed +Message = Call service to schedule a repair. [111] -Title = HD syringe pump not stopped in off state error -Message = If the syringe pump position is not equal to its last position. +Title = HD Fault: Syringe Pump Speed +Message = Call service to schedule a repair. [112] -Title = DG load cells A2/B2 FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time. +Title = DG Fault: LC A2/B2 Comm +Message = Call service to schedule a repair. SRSDG 953 = [113] -Title = DG drain pump current out of range -Message = If the DG drain pump current > 2.2 A when it is in open loop or if the current > 0.1 A when it is off. +Title = DG Fault: DRP Current +Message = Call service to schedule a repair. SRSDG 890 = [114] -Title = HD venous air bubble detector self-test failure -Message = If the venous air bubble self test did not detect bubble for more than 500 msec. +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem with the air bubble sensor\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [115] -Title = DG temperature sensor out of range -Message = If the measured temperature < 0 C or > 80 C for the non-fluid inline sensors and > 120 C for the fluid sensors. +Title = DG Fault: Temp Sensor Range +Message = Call service to schedule a repair. SRSDG 224 = [116] -Title = SD card is non-operational -Message = SD card is non-operational +Title = HD Alarm (Low): SD-Card Error +Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. [117] -Title = System message that required acknowledgment was not acknowledged -Message = If no retries left in the pending acknowlegement list. +Title = Loss of HD Comm (HD ACK) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. SRSDG 909 = [118] -Title = DG RTC configuration error -Message = If the RTC configuration registers report an error. +Title = DG Fault: RTC Config +Message = Call service to schedule a repair. SRSDG 43 = [119] -Title = DG RTC battery low -Message = If the RTC battery bit report a low alarm. +Title = DG Alarm (Low): RTC Bat Low +Message = Call service to schedule a repair. SRSDG 908 = [120] -Title = HD pre-treatment wet prime test failure -Message = If the HD pre-treatment wet self test failed. +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a bubble in the venous bubble detector\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and \nrestart with a new cartridge. [121] -Title = DG main primary heater voltage out of range -Message = If the main primary heater's voltage is out of range. +Title = DG Fault: Prim Heater Lg Control +Message = Call service to schedule a repair. SRSDG 611 = [122] -Title = DG small primary heater voltage out of range -Message = If the small primary heater's voltage is out of range. +Title = DG Fault: Prim Heater Sm Control +Message = Call service to schedule a repair. SRSDG 612 = [123] -Title = DG trimmer heater voltage out of range -Message = If the trimmer voltage is out of range. +Title = DG Fault: Trimmer Heater Control +Message = Call service to schedule a repair. SRSDG 613 = [124] -Title = HD end of treatment alarm (high priority - user still not acting to end treatment) -Message = In treatment end paused state for > 5 minutes. +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [125] -Title = Blood sitting too long after treatment stopped by user alarm (> 5 min) -Message = If in treatment stop sub mode when blood is in the tubing, the blood sitting counter exceeded 5 minutes. +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treatment must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [126] -Title = HD blood leak detected alarm -Message = If the blood leak is detected based on the blood leak status reported by the sensor. +Title = Blood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. Treatment must end. [127] -Title = Venous pressure too low during treatment -Message = If the venous pressure < TREATMENT_PARAM_VEN_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. +Title = Blood Stop: Venous Pres. Low +Message = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured \nproperly. [128] -Title = HD venous air bubble detected alarm -Message = If bubbles are detected during the treatment. +Title = Blood Stop: Venous Air Bubble +Message = Venous air bubble detected indicating failure of air trap \ncontrol system.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [129] Title = Message = [130] -Title = Venous pressure too high during treatment -Message = If the venous pressure > TREATMENT_PARAM_VEN_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. +Title = Blood Stop: Venous Pres. High +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [131] -Title = Arterial pressure too low during treatment -Message = If the arterial pressure < TREATMENT_PARAM_ART_PRESSURE_LOW_LIMIT. Or expanded pressure limits based on mode of operation. +Title = Blood Stop: Arterial Pres. Low +Message = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [132] -Title = Arterial pressure too high during treatment -Message = If the arterial pressure > TREATMENT_PARAM_ART_PRESSURE_HIGH_LIMIT. Or expanded pressure limits based on mode of operation. +Title = Blood Stop: Arterial Pres. High +Message = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [133] -Title = DG fluid leak detected alarm -Message = If the fluid leak is detected as reported by the fluid leak detector. +Title = DG Fault: Fluid Leak in DG +Message = Fluid leak detected in Dialysate Generation Delivery Device.\n-Treatment must be terminated.\n- Shut off water source and disconnect water and \nelectrical connections from system.\n- Call service to schedule a repair. SRSDG 328 = [134] -Title = HD fluid leak detected alarm -Message = If the fluid leak is detected as reported by the fluid leak detector. +Title = Blood Stop: Fluid Leak in HD +Message = Fluid leak detected in Hemodialysis Delivery device.\n- If in treatment, BLOOD STOPPED.\n-Identify and correct the cause of the spill.\n- Clean up the spill and dry leak sensor.\n- If in treatment, resume treatment. [135] -Title = HD experienced large acceleration -Message = If the system experienced excessive shock > 2.5 g. +Title = Blood Stop: HD Shock Detected +Message = Excessive movement detected with Hemodialysis Delivery \ndevice.\n- Confirm the device is stationary and level before continuing. [136] -Title = DG experienced large acceleration -Message = If the system experienced excessive shock > 2.5 g. +Title = Dial. Gen: Shock-Vibration +Message = Excessive movement detected with Dialysis Generation \ndevice.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing. SRSDG 327 = [137] -Title = HD tilt exceeds maximum -Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. +Title = Blood Stop: HD Tilted +Message = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. [138] -Title = DG tilt exceeds maximum -Message = If the system experienced excessive tilt x and y axis angle > 7 deg for more than 1 sec. +Title = Dial. Gen: Tilt +Message = Excessive tilt detected with Dialysis Generation device.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing. SRSDG 326 = [139] -Title = HD AC power lost alarm -Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in standby, service, or treatment parameters mode. +Title = Loss of AC Power +Message = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns. [140] -Title = Dialysate generator communication timeout -Message = If DG communication timeout is detected for more than 2 seconds. +Title = Loss of DG Comm (Timeout) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [141] -Title = Air trap fill timeout during treatment -Message = If the air trap fill timeout during treatment. +Title = Blood Stop: Air Trap Fill +Message = Failure to fill the Air Trap.\n- Treatment must be terminated without option for rinseback.\n- Check access for dislodgement. [142] -Title = Blood pump occlusion detected -Message = If blood occlusion is detected > OCCLUSION_THRESHOLD_OFFSET + bloodPumpOcclusionAfterCartridgeInstall +Title = Blood Stop: Blood Pump Occlusion +Message = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the \nvenous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. [143] -Title = DG dialysate temperature sensors drift timeout. -Message = If TDi and TRo have drifted for more 1 C for a certain period of time. +Title = DG Fault: TRo TDi Delta +Message = Call service to schedule a repair. SRSDG 912 = [144] -Title = DG cleaning mode inlet water conductivity out of high range -Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). In cleaning mode +Title = Cleaning: Inlet Water Cond High +Message = Cleaning Mode Inlet water conductivity too high.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] -Title = DG Concentrate conductivity after adding acid out of range alarm -Message = If the conductivity value for the acid concentrate is out of range for more than 45 secs. +Title = Dial. Gen: Acid Conc. Cond. +Message = Acid concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 846 = [146] -Title = DG RTC vs. Timer accuracy error -Message = If RTC and timer are not aligned on 1 second elapsed time. +Title = DG POST: DG RTC Timer +Message = Call service to schedule a repair. SRSDG 312 = [147] -Title = DG dialysate generation conductivity fault alarm -Message = If dialysate conductivity is not with 5% of the TCD +Title = Dial. Gen: Dialysate not ready +Message = Dialysate is not yet ready.\n- Treatment is paused and will be ready to resume when \ndialysate is ready for use. SRSDG 398 = [148] -Title = Cleaning Mode Inlet water temperature too high -Message = Inlet temperature > 37 C. +Title = Cleaning: Inlet Water Temp High +Message = Cleaning Mode Inlet water temperature too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] -Title = HD watchdog expired -Message = If the watchdog signal is low while not in POST. +Title = HD Fault: WD Error +Message = Call service to schedule a repair. [150] -Title = Inlet water conductivity too high -Message = Inlet water conductivity > 2000 uS/cm. Inlet water conductivity > 100 uS/cm (PURE WATER MODE). +Title = Dial. Gen: Inlet Water Cond High +Message = Inlet water conductivity is high.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 391 = [151] -Title = Inlet water conductivity too low -Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). +Title = Dial. Gen: Inlet Water Cond Low +Message = Inlet water conductivity is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 826 = [152] -Title = Inlet water pressure too low -Message = Inlet water pressure < 25 PSI. +Title = Dial. Gen: Inlet Water Pres Low +Message = Inlet water pressure is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 417 = [153] -Title = HD prime completed high priority alarm -Message = In Pre-Treatment Recirculation for 120 minutes (escalation of alarm 213) +Title = Process: Pre-Tx Terminated +Message = Pre-Treatment timeout.\n- Patient was not connected to device\n- Pre-Treatment has been terminated\n- Proceed to end treatment to remove cartridge from system. [154] -Title = Cleaning mode inlet water conductivity too low -Message = Inlet water conductivity < 200 uS/cm. Inlet water conductivity low disabled (PURE WATER MODE). In cleaning mode +Title = Cleaning: Inlet Water Cond Low +Message = Cleaning Mode Inlet water conductivity too low.\n- Cleaning mode process has terminated \n- Check inlet water before restarting cleaning mode. SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] -Title = HD No reservoir data message receive at least once every 2 seconds -Message = No new reservoir data for 2 seconds. +Title = Loss of DG Comm (Reservoir) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [156] -Title = HD No DG op mode message receive at least once every 2 seconds -Message = No DG op mode for 2 seconds. +Title = Loss of DG Comm (Op Mode) +Message = Communication failure with Dialysis Generator.\n-Treatment must be terminated.\n- Ensure all cords are securely connected. [157] -Title = Chemical disinfect priming the acid line timeout -Message = If the disinfectant acid line is not primed within a certain period of time. +Title = Cleaning: Chem Disinf PAA Prime +Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 988 = [158] -Title = Inlet water temperature low -Message = Inlet temperature < 24 °C for 5 seconds. +Title = Dial. Gen: Inlet Water Temp Low +Message = Inlet water temperature is low.\n- Treatment is paused.\n- Check the incoming water supply. SRSDG 408 = [159] -Title = Close concentrate line be able to flush it -Message = Start of chemical disinfect flush. +Title = Cleaning: Chem Disinf Connect Cap +Message = Remove chemical disinfectant connection and close cap. SRSDG 1044 = [160] -Title = Blood pump rotor speed too high -Message = If blood pump motor speed exceeded 100.0 RPM. +Title = HD Fault: BP Rotor Max Speed +Message = Call service to schedule a repair. [161] -Title = Too many bad CRCs detected on received system messages -Message = If >= 5 bad CRCs detected on received CAN messages within 10 minutes. +Title = DG Fault: HD-DG Comm +Message = Call service to schedule a repair. SRSDG 122 = [162] -Title = FPGA clock speed failure -Message = If the FPGA clock speed is relatively close to the processor clock for a certain period of time. +Title = DG Fault: Proc-FPGA Clock +Message = Call service to schedule a repair. SRSDG 867 = [163] -Title = HD sees primary load cell for reservoir 1 change too much too fast -Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). +Title = Dial. Gen: UF Rate Check +Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system\n(shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a\nrepair. [164] -Title = HD sees primary load cell for reservoir 2 change too much too fast -Message = If the delta volume of the active reservoir exceeded 50.0 g (MAX_ACTIVE_LOAD_CELL_CHANGE_G). +Title = Dial. Gen: UF Rate Check +Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system \n(shock/vibration).\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a \nrepair. [165] -Title = HD in treatment rinseback operation timeout -Message = If the maximum time inrinseback operation has been exceeded. +Title = Blood Stop: Blood Clot Risk +Message = Blood pump has been off for too long.\n- Treament must be terminated. Blood clot risk.\n- NO RINSEBACK ALLOWED. [166] Title = Message = [167] -Title = HD cartridge door opened alarm -Message = If the cartridge door is opened as reported by the FPGA. +Title = Blood Stop: Door Open +Message = Device door is open.\n - Close door and press RESUME button to continue. [168] Title = Message = [169] -Title = Dialysate flow rate out of range -Message = If the dialysate flow rate > 2.00 L/min. +Title = DG Fault: Dialysate Flow Sens Rg +Message = Call service to schedule a repair. SRSDG 643 = [170] -Title = HD syringe empty alarm -Message = If the syringe pump position is near empty position. +Title = Heparin Syringe Empty +Message = Heparin syringe empty.\n- Treatment will continue\n- If additional heparin is required to complete treatment, \nadminister manually. [171] -Title = HD syringe pump occlusion alarm -Message = If the pressure sensor voltage reading > SYRINGE_FORCE_OCCLUSION_THRESHOLD_V. +Title = Process: Pre-Tx Hep. Pump Occ. +Message = Heparin pump occlusion detected.\n- Check heparin line is unclamped.\n- Check there are no kinks in the heparin line. [172] -Title = HD syringe pump not enough Heparin alarm -Message = If the estimated syringe volume needed for treatment is not sufficient +Title = Process: Pre-Tx Not Enough Hep. +Message = Not enough heparin is available for treatment.\n- Check if syringe pump is touching syringe plunger.\n- If syringe pump is touching syringe plunger, end treatment.\n- If syringe pump is not touching syringe plunger, reload \nsyringe with enough heparin. [173] -Title = HD RTC configuration error -Message = If the RTC configuration registers report an error. +Title = HD Fault: RTC Config +Message = Call service to schedule a repair. [174] -Title = HD RTC vs. Timer accuracy error -Message = If RTC and timer are not aligned on 1 second elapsed time. +Title = HD POST: HD RTC Timer +Message = Call service to schedule a repair. [175] -Title = HD pump commutation error -Message = If the set pump direction does not agree with the hall sensor direction. +Title = HD Fault: Pump Commutation Err +Message = Call service to schedule a repair. [176] Title = Message = [177] -Title = DG software fault. Software found itself in an unexpected state -Message = If the specific software faults listed in alarm management SW faults has been detected +Title = DG Fault: Software Invalid State +Message = Call service to schedule a repair. SRSDG 919 = [178] -Title = HD communication timeout -Message = If the HD stops communicating with the UI. +Title = HD Fault: UI-Proc Comm +Message = Call service to schedule a repair. = SRSUI 1066 [179] -Title = DG FPGA communication down for too long -Message = If retries for commands exceeds limit or FPGA reports comm error. +Title = DG Fault: FPGA Com +Message = Call service to schedule a repair. SRSDG 605 = [180] -Title = RO flow rate out of range -Message = If the RO flow rate > 2.00 L/min. +Title = DG Fault: RO Flow Sensor Range +Message = Call service to schedule a repair. SRSDG 883 = [181] -Title = DG load cells weight out of range for tare -Message = If the load cell weight is out of range for tare. +Title = DG Fault: Tare Out of Range +Message = Call service to schedule a repair. SRSDG 849 SRSDG 838 = [182] -Title = DG load cells invalid calibration -Message = If the CRC of load cells calibration data does not match the calculated CRC. +Title = DG POST: LC Cal +Message = Call service to schedule a repair. SRSDG 162 = [183] -Title = DG invalid load cell value -Message = If the half full reservoir is filled prior to filling the reservoir that is supposed to overflow to the half full reservoir. +Title = Cleaning: Invalid Resv Transfer +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 851 = [184] -Title = Inlet reactor not healthy -Message = If the inlet reactor are not healthy. +Title = DG Fault: UVi Error +Message = Call service to schedule a repair. SRSDG 559 = [185] -Title = DG fan RPM out of range -Message = If the fan RPM is out of Min/Max range. +Title = DG Alarm (Low): Fan Speed +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. SRSDG 628 = [186] -Title = DG concentrate pump fault -Message = If the concentrate pumps fault is reported by the FPGA. +Title = DG Fault: CP1/CP2 Fault +Message = Call service to schedule a repair. SRSDG 891 = [187] -Title = DG concentrate pump CP1 speed control error -Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. +Title = DG Fault: CP1 Speed Fault +Message = Call service to schedule a repair. SRSDG 905 = [188] -Title = DG concentrate pump CP2 speed control error -Message = If the concentrate pump CP1 speed > 0.02 tolerance for a certain period of time. +Title = DG Fault: CP2 Speed Fault +Message = Call service to schedule a repair. SRSDG 906 = [189] -Title = DG drain pump RPM out of range -Message = If the drain pump RPM > 100.0 of the target RPM. +Title = DG Fault: DRP Speed Fault +Message = Call service to schedule a repair. SRSDG 448 = [190] -Title = DG drain pump off fault -Message = If the pump is in off state and the RPM is greater than the minimum RPM of 300.0 RPM. +Title = DG Fault: DRP Speed Not Off +Message = Call service to schedule a repair. [191] -Title = DG flow rate out of upper range -Message = If the RO flow rate > 10% from the target flow rate. +Title = DG Fault: RO Flow High +Message = Call service to schedule a repair. SRSDG 885 = [192] -Title = DG flow rate out of lower range -Message = If the RO flow rate < 10% from the target flow rate. +Title = DG Fault: RO Flow Low +Message = Call service to schedule a repair. SRSDG 945 = [193] Title = Message = [194] -Title = DG RO pump duty cycle out of range -Message = If the DG RO pump duty cycle is out of range by 5% of the target duty cycle for a certain period of time. +Title = DG Fault: RO Driver Feedback +Message = Call service to schedule a repair. SRSDG 881 = [195] -Title = DG RO pump pressure out of range -Message = If the DG RO pump pressure > 135 psi for a certain period of time. +Title = DG Fault: RO Pump High Pressure +Message = Call service to schedule a repair. SRSDG 925 = [196] -Title = DG CPi/CPo conductivity sensors FPGA fault -Message = If the FPGA reports error for a certain period of time. +Title = DG Fault: CPi/CPo Comm +Message = Call service to schedule a repair. SRSDG 955 = [197] -Title = DG CD1/CD2 conductivity sensors FPGA fault -Message = If the FPGA reports error for a certain period of time. +Title = DG Fault: CD1/CD2 Comm +Message = Call service to schedule a repair. SRSDG 956 = [198] -Title = DG primary heater while RO flow is low -Message = If the RO flow is < 0.2 L/min for a certain period of time. +Title = DG Fault: HP On With No Flow +Message = Call service to schedule a repair. SRSDG 872 = [199] -Title = DG trimmer heater while dialysate flow is low -Message = If the FMD flow is ≤ 0 mL/min (raw) for a certain period of time. +Title = DG Fault: HR On With No Flow +Message = Call service to schedule a repair. SRSDG 871 = [200] -Title = DG thermistor or temperature sensor out of range -Message = If thermistor > 80.0 C or < 0.0 C. +Title = DG Fault: Thermistor Range +Message = Call service to schedule a repair. SRSDG 178 = [201] -Title = HD pre-treatment wet flow test failure -Message = If the pre-treatment wet flow test has failed. +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem in the dialysate flow \nmeasurement or there is a leak in the dialysate in/out lines.\n- The cartridge should be primed\n- Check cartridge installation on to machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [202] -Title = HD pre-treatment dry pressure normal test failure -Message = If the pre-treatment dry pressure normal test failed. +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected blood circuit pressure cannot be \nrelieved.\n- Check PBo connection to HD for kinks\n -Check PBo transducer protector for occlusion\n- If problem persists, install a new cartridge and \nPBo transducer protector. [203] Title = Message = [204] -Title = RO rejection ratio out of range -Message = If RO rejection ratio > 0.1 +Title = DG Fault: RO Rejection +Message = Call service to schedule a repair. SRSDG 429 = [205] -Title = Conductivity sensor fault -Message = If conductivity sensor reports 0x11 for more than 2 consecutive reads. +Title = DG Fault: Cond Sensors Error +Message = Call service to schedule a repair. SRSDG 913 = [206] -Title = DG dialysate fill runs out of time -Message = If DG dialysate fill > 5 minutes. +Title = DG Fault: Reservoir Fill Timeout +Message = Call service to schedule a repair. SRSDG 437 = [207] -Title = DG flow meter check failure alarm -Message = If integrated Volume To Load Cell Reading Percent > 0.1. +Title = DG Fault: FMP Flow Rate Error +Message = Call service to schedule a repair. SRSDG 240 = [208] -Title = HD venous line occlusion fault -Message = If the venous pressure is > set limit + 50 mmHg if HD is in treatment or pre-treatment mode. +Title = Blood Stop: Venous Pres. High +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [209] -Title = DG drain circulation line timeout -Message = If the integrated volume < 0.1 for more than 60 secs. +Title = Cleaning: Flush Circ Error +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 980 = [210] -Title = HD battery pack detected an error -Message = If the battery pack status error is detected from the slave device I2C address 0x0B. +Title = HD Alarm (Low): Battery Fault +Message = Call service to schedule a repair. [211] -Title = Blood sitting too long warning (> 4 min) -Message = During treatment stop sub-mode, if the blood is in the line for more than 4 minutes. +Title = Blood Stopped: Warning +Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [212] -Title = Blood sitting too long warning (> 4 min) -Message = During treatment end paused state, if the blood is in the line for more than 4 minutes. +Title = Blood Stopped: Warning +Message = Blood pump is stopped.\n- Resume treatment before countdown timer expires. [213] -Title = HD prime completed medium priority alarm -Message = In Pre-Treatment Recirculation for 90 minutes (escalation of alarm 226) +Title = Process: Pre-Tx Prime Complete +Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [214] Title = @@ -964,114 +964,114 @@ Message = [216] -Title = HD end treatment sub-mode timeout alarm -Message = If the treatment ended successfully. +Title = Process: End of Dialysis Tx +Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [217] Title = Message = [218] -Title = HD syringe detected alarm -Message = If the syringe detection switch > 2.0 volts. +Title = Process: Remove Syringe +Message = Syringe not removed from machine.\n- Remove heparin syringe. [219] -Title = HD syringe pump syringe removed alarm -Message = If the syringe detection switch < 2.0 volts. +Title = Blood Stop: Syringe Not Detected +Message = Heparin is needed for treatment but syringe has been removed \nfrom syringe pump.\n- Install Heparin syringe. [220] -Title = DG THd sensor FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = DG Fault: THD RTD Comm +Message = Call service to schedule a repair. SRSDG 947 = [221] -Title = HD Saline bag is empty -Message = If the measured arterial pressure < -300.0 mmHg for more than 250 msec. +Title = Blood Stop: Saline Bag Empty +Message = Saline bag appears empty.\n- Check saline bag and line and replace if saline bag is \nempty. [222] -Title = HD occlusion sensor FPGA fault -Message = If the FPGA does not report fresh data or indicates an error for a certain period of time. +Title = HD Fault: BP Occ Com +Message = Call service to schedule a repair. [223] -Title = HD arterial sensor FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = HD Fault: Art. Pres. Com +Message = Call service to schedule a repair. [224] -Title = Treatment stopped by user action - pressed stop button -Message = If the user has pressed the stop button. +Title = Blood Stopped: Treatment Pause +Message = Treatment paused by user. [225] -Title = HD end of treatment warning -Message = If the treatment duration has exceeded the prescribed treatment time entered by the user. +Title = Process: End of Treatment +Message = Dialysis Treatment has completed\n- Proceed to end treatment and rinseback. [226] -Title = HD prime completed low priority alarm -Message = Start of Pre-Treatment Recirculation - notifiy user to proceed to patient connection +Title = Process: Pre-Tx Priming Complete +Message = Prime of the cartridge is complete.\n- Proceed to connect to patient. [227] Title = Message = [228] -Title = HD time out on prime sub-mode purge air state. -Message = During the pre-treatment prime state, if the prime saline purge air state exceeded 60 secs. +Title = Process: Pre-Tx Prime Timeout +Message = Blood line prime with saline timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [229] -Title = HD prime dialysate dialyzer time out alarm -Message = If the priming of the dialyzer dialysate flow path has exceeded the time limit. +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [230] -Title = HD prime dialysate bypass time out alarm -Message = If the priming of the dialyzer bypass flow path has exceeded the time limit. +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [231] -Title = HD pre-treatment mode dry self-tests failure -Message = If the following dry self tests have failed: 1. handleDrySelfTestPressureSensorsState 2. handleDrySelfTestPressureSensorsNormalState +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected a leak in the circuit\n- Check blood line connections\n- Check dialysate line connections\n- Check PBo and PBa transducer protectors for occlusion.\n- Check clamps.\n- If problem persists, install a new cartridge. [232] -Title = HD pre-treatment mode wet self-tests failure -Message = If the following wet self tests have failed: 1. handleWetSelfTestPrimeCheckState 2. handleWetSelfTestFirstDisplacementVerifyState 3. handleWetSelfTestSecondDisplacementVerifyState +Title = Process: Pre-Tx Wet Self-Test +Message = Self-test has detected a problem in the load cell \nmeasurement or there is a leak in the dialysate in/out lines\n- The cartridge should be primed\n- Check cartridge installation onto machine\n- If problem persists, terminate pre-treatment and restart \nwith a new cartridge. [233] -Title = DG barometric pressure sensor bad coefficients CRC -Message = If the following happens: 1. The stored CRC in the sensor does not match the calculated CRC. 2. The CRC and coefficients have not been received from the sensor after 20 seconds from the power up. +Title = DG Fault: Baro Cal CRC +Message = Call service to schedule a repair. SRSDG 964 SRSDG 949 SRSDG 1089 = [234] -Title = DG inlet water pressure is too high -Message = If the inlet water pressure is > 80 psig. +Title = Dial. Gen: Inlet Water Pres High +Message = Inlet water pressure is high.\n- Treatment is paused\n- Check the incoming water supply. SRSDG 1037 = [235] -Title = HD in treatment stopped sub-mode after rinseback completed (no escalation) -Message = If there is no active alarm and rinseback has completed. +Title = Process: Rinseback Complete +Message = Rinseback has been completed. [236] -Title = HD needs new cartridge to be installed -Message = If the air trap sensor reported AIR and bubble sensor detected bubbles. +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected fluid in the cartridge.\n- Cartridges are single-use only\n- Install new cartridge to resume. [237] -Title = HD prime saline dialyzer time out alarm -Message = If timeout, just transition to HD_PRIME_RESERVOIR_TWO_FILL_COMPLETE_STATE +Title = Process: Pre-Tx Prime Timeout +Message = Dialyzer prime timeout.\n- Check blood and dialysate lines for clamps, kinks or leaks. [238] -Title = HD no cartridge loaded or installed improperly alarm -Message = If the measured blood pump occlucsion pressure reading is < 5000 or > 24500. +Title = Process: Pre-Tx Dry Self-Test +Message = Self-test has detected cartridge is not installed.\n- Check that a cartridge is properly installed. [239] -Title = HD fail to remove cartridge alarm -Message = In the post treatment state, if the cartridge has not been removed. +Title = Process: Post-Tx Remove Cartridge +Message = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. [240] -Title = Bicarb conductivity out of range during bicarb pump check alarm -Message = If the conductivity value for bicarbonate concentrate is out of range for more than 45 secs. +Title = Dial. Gen: Bicarb Conc. Cond. +Message = Bicarbonate conductivity during prime is out of range.\n- Check correct concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 482 = [241] -Title = DG reservoir drain time out -Message = If the DG reservoir drain is not complete in 120 secs. +Title = Cleaning: Resv Drain Timeout +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 848 SRSDG 90 SRSDG 561 @@ -1102,8 +1102,8 @@ SRSDG 1034 = [242] -Title = DG reservoir fill time out -Message = If the DG reservoir fill is not complete in 120 secs. +Title = Cleaning: Resv Fill Timeout +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 850 SRSDG 573 SRSDG 574 @@ -1124,67 +1124,67 @@ SRSDG 1024 = [243] -Title = DG reservoir leak time out -Message = If the reservoir level is drifted by 100 mL for more than 30 seconds. +Title = Cleaning: Fluid Loss Detected +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 853 = [244] -Title = DG temperature sensors difference out of range -Message = If any of the temperature sensors are out of the average of the sensors by more than 3 C. +Title = Cleaning: Temp Sensor Check +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 852 = [245] -Title = DG heat disinfect target temperature time out (could not reach to temperature) -Message = If heating up to minimum temperature for heat disinfect failed after 4 hours. +Title = Cleaning: Heat Disinfect Temp TO +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. SRSDG 854 = [246] -Title = DG chemical disinfect inlet conducitivty out of range -Message = If CPo and CD2 conductivity values are above 100 uS/cm. +Title = Cleaning: RO Water Cond Error +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [247] Title = Message = [248] -Title = DG chemical disinfect out of target temperature -Message = If TPo > 50 C after 1 second. +Title = Cleaning: Chem Disinfect Proc +Message = Cleaning mode process terminated unexpectedly.\n-Restart cleaning mode.\n-If problem persists, call service to schedule a repair. [249] -Title = DG chemical disinfect out of target conductivity -Message = If CD2 < 150 uS/cm or CD2 > 650 uS/cm after 10 seconds. +Title = Cleaning: Chem Disinf PAA Cond +Message = Unable to sense proper chemical disinfectant.\n- Cleaning mode process has terminated.\n- Check that proper chemical disinfectant is connected to the \nbicarb port.\n- Check chemical disinfectant bottle is not empty. SRSDG 992 = [250] -Title = Insert the acid to start chemical disinfect -Message = Start of chemical disinfect +Title = Cleaning: Chem Disinf Connect PAA +Message = Connect chemical disinfectant bottle to bicarb port. SRSDG 1046 = [251] -Title = HD invalid system record -Message = If the calculated CRC of the HD system record image does not match the expected CRC. +Title = HD POST: Sys Rec NVRAM CRC +Message = Call service to schedule a repair. [252] -Title = HD invalid service record -Message = If the calculated CRC of the HD service record image does not match the expected CRC. +Title = HD POST: Serv Rec NVRAM CRC +Message = Call service to schedule a repair. [253] -Title = DG invalid system record -Message = If the calculated CRC of the DG system record image does not match the expected CRC. +Title = DG POST: Sys Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 609 = [254] -Title = DG invalid service record -Message = If the calculated CRC of the DG service record image does not match the expected CRC. +Title = DG POST: Serv Rec NVRAM CRC +Message = Call service to schedule a repair. SRSDG 610 = [255] -Title = HD and UI software builds are not compatible -Message = If the compatibility revision field of the UI is not equal to SW_COMPATIBILITY_REV. +Title = HD POST: UI Compatibility +Message = Call service to schedule a repair. [256] -Title = DG flush started -Message = If DG flush successfully started from the UI. +Title = Cleaning: Chem Flush In Progress +Message = Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. = SRSUI 497 [257] @@ -1196,71 +1196,71 @@ Message = [259] -Title = HD temperatures out of range -Message = If thermistor reports temperature > 80.0 C or < 0.0 C for a certain period of time. +Title = HD Fault: Temp Sensors Error +Message = Call service to schedule a repair. [260] -Title = UI POST Application Integrity (Sha256Sum) failure -Message = If the UI POST reports a shasum faillure. +Title = HD POST: UI Image +Message = Call service to schedule a repair. = SRSUI 1067 [261] -Title = UI POST CANBus failure -Message = If the UI POST reports a can bus failure. +Title = HD POST: UI CAN Driver +Message = Call service to schedule a repair. = SRSUI 1068 [262] -Title = UI POST Display failure -Message = If the UI POST reports a display failure. +Title = HD POST: UI Disp Driver +Message = Call service to schedule a repair. = SRSUI 1069 [263] -Title = UI POST Touch failure -Message = If the UI POST reports a touch failure. +Title = HD POST: UI Touch Driver +Message = Call service to schedule a repair. = SRSUI 1070 [264] -Title = UI POST SD-Card failure -Message = If the UI POST reports SD card failure. +Title = HD POST: UI SD-Card +Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persists. = SRSUI 1071 [265] -Title = UI POST RTC failure -Message = If the UI POST reports RTC failure. +Title = HD POST: UI RTC +Message = Call service to schedule a repair. = SRSUI 1072 [266] -Title = UI POST WiFi failure -Message = If the UI POST reports WiFi failure. +Title = HD POST: UI WiFi +Message = Call service to schedule a repair. = SRSUI 1073 [267] -Title = UI POST Bluetooth failure -Message = If the UI POST reports bluetooth failure. +Title = HD POST: UI Bluetooth +Message = Call service to schedule a repair. = SRSUI 1074 [268] -Title = UI POST Ethernet failure -Message = If the UI POST reports ethernet failure. +Title = HD POST: UI Ethernet +Message = Call service to schedule a repair. = SRSUI 1075 [269] -Title = UI POST Sound failure -Message = If the UI POST reports sound failure. +Title = HD POST: UI Sound +Message = Call service to schedule a repair. = SRSUI 1076 [270] -Title = HD POST Safety Shutdown failure -Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. +Title = HD POST: Proc Safestate +Message = Call service to schedule a repair. [271] -Title = DG POST Safety Shutdown failure -Message = If the 24 volts power source is greater than 5.0 volts during safety shutdown self test state. If the 24 volts power source is less than 22.6 volts during safety shutdown self test recover state. +Title = DG POST: Proc Safestate +Message = Call service to schedule a repair. SRSDG 822 = [272] -Title = HD Fan RPM out of range -Message = If the fan RPM is out of Min/Max range. +Title = HD Alarm (Low): Fan Speed +Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. [273] Title = @@ -1271,226 +1271,226 @@ Message = [275] -Title = HD arterial pressure sensor is reading out of range -Message = If the arterial pressure > 2000.0 mmHg or < -1500.0 mmHg. +Title = HD Fault: Arterial Press Range +Message = Call service to schedule a repair. [276] -Title = HD venous pressure sensor is reading out of range -Message = If the venous pressure > 2000.0 mmHg or < -1500.0 mmHg. +Title = HD Fault: Venous Press Range +Message = Call service to schedule a repair. [277] -Title = HD BP occlusion sensor is reading out of range -Message = If the blood pump occlusion count is < 2000. +Title = HD Fault: Occlusion Sens Range +Message = Call service to schedule a repair. [278] -Title = HD active reservoir weight out of range. -Message = If the active reservoir weight is > 1900.0 g. +Title = DG Fault: Reservoir Overfill +Message = Call service to schedule a repair. [279] -Title = DG dialysate drain time out -Message = If the DG dialysate drain exceeded 120 seconds. +Title = DG Fault: Reservoir Drain +Message = Call service to schedule a repair. SRSDG 752 = [280] -Title = DG drain pump invalid direction -Message = If the drain pump's direction is not 1 when draining. +Title = DG Fault: DRP Direction Fault +Message = Call service to schedule a repair. SRSDG 888 = [281] -Title = DG Acid concentration bottle low volume alarm -Message = If the acid concentration bottle volume > 3240 (3340 - 100). +Title = Dial. Gen: Acid Conc. Low +Message = Acid concentrate low.\n- Replace acid concentrate bottle. SRSDG 836 = [282] -Title = DG Bicarbonate concentration bottle low volume alarm -Message = If the bicarbonate concentration bottle volume > 3680 (3780 - 100). +Title = Dial. Gen: Bicarb Conc. Low +Message = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. SRSDG 837 = [283] -Title = DG load cells weight out of range -Message = If the weight of the DG load cells > 4500.0 g or < 0.0 g. +Title = DG Fault: LC Range +Message = Call service to schedule a repair. SRSDG 615 = [284] -Title = DG load cells primary/back up drift out of range -Message = If redundant load cells deviate by more than 60g after tare for 5 seconds. +Title = DG Fault: Redundant LC Check +Message = Call service to schedule a repair. SRSDG 383 = [285] Title = Message = [286] -Title = DG cap in wrong state -Message = If the concentrate cap is not in the right position in the cleaning modes. +Title = Cleaning: Conc Cap Removed +Message = Concentrate cap removed from DG during cleaning.\n- Cleaning Process Terminated. [287] -Title = HD processor clock speed checks against FPGA clock failure -Message = If during initialization mode, the difference between the processor clock speed and the FPGA clock speed > 1. +Title = HD Fault: Proc-FPGA Clock +Message = Call service to schedule a repair. [288] -Title = HD load cell inconsistency -Message = If redundant load cells deviate from start to end of a reservoir use during treatment by more than 10g. +Title = DG Fault: LC Drift Check +Message = Call service to schedule a repair. [289] -Title = DG cap in wrong state -Message = If the dialysate cap is not in the right position in the cleaning modes. +Title = Cleaning: Dialysate Cap Removed +Message = Dialysate cap removed from DG during cleaning.\n- Cleaning Process Terminated. SRSDG 928 = [290] -Title = HD venous sensor FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = HD Fault: Ven. Pres. Com +Message = Call service to schedule a repair. [291] -Title = DG flush started -Message = If DG flush successfully started from the UI. +Title = Cleaning: Flush In Progress +Message = System flush in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [292] -Title = DG heat disinfect started -Message = If DG heat disinfect successfully started from the UI. +Title = Cleaning: Heat Disinf In Progress +Message = Heat disinfect in progress\n- Please wait for the process to complete.\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [293] -Title = DG flush started -Message = If DG flush successfully started from the UI. +Title = Cleaning: Chem Flush In Progress +Message = Chemical disinfect flush in progress\n- Please wait for the process to complete\n- Treatment will not be able to be started until process \ncompletes.\n- Process may be terminated by pressing the HD stop button. [294] Title = Message = [295] -Title = UI cloud sync POST failure -Message = If the UI POST reports cloud sync failure. +Title = HD POST: UI CloudSync +Message = Call service to schedule a repair. = SRSUI 1238 [296] -Title = DG TDi sensor FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = DG Fault: TDi RTD Comm +Message = Call service to schedule a repair. SRSDG 951 = [297] -Title = DG TRo sensor FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = DG Fault: TRo RTD Comm +Message = Call service to schedule a repair. SRSDG 950 = [298] -Title = DG barometric sensor FPGA fault -Message = If the FPGA does not report fresh data for a certain period of time +Title = DG Fault: Baro Pres Comm +Message = Call service to schedule a repair. SRSDG 948 = [299] -Title = DG invalid serial number -Message = If the DG serial number is empty and does not have any characters. +Title = DG POST: Serial Number Check +Message = Call service to schedule a repair. SRSDG 952 = [300] -Title = HD invalid serial number -Message = If the HD serial number is empty and does not have any characters. +Title = HD POST: Serial Number Check +Message = Call service to schedule a repair. [301] Title = Message = [302] -Title = DG_CPU RAM error -Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. +Title = DG Fault: Processor RAM error +Message = Call service to schedule a repair. SRSDG 126 = [303] -Title = DG drain pump direction FPGA fault -Message = If the FPGA indicates an error for a certain period of time. +Title = DG Fault: DRP Hall Commutation +Message = Call service to schedule a repair. SRSDG 958 = [304] -Title = HD invalid usage reocord -Message = If the CRC of the HD usage record does not match the calculated CRC. +Title = HD POST: Usage Rec NVRAM CRC +Message = Call service to schedule a repair. [305] -Title = HD_CPU RAM error -Message = If the RAM status indicates any of the following: Double bit error, Address decode failure, Address decode logic element failure, Read address parity failure, Write address parity failure. +Title = HD Fault: Processor RAM error +Message = Call service to schedule a repair. [306] -Title = -Message = CD1 and CD2 conductivity measurement during acid check is not within 5% of each other +Title = DG Fault: CD1 CD2 Check +Message = Call service to schedule a repair. [307] -Title = HD AC power lost alarm -Message = The HD Software shall trigger an alarm if an AC power loss is detected while the HD system is in pre-treatment, treatment, or post-treatment mode. +Title = Blood Stop: Loss of AC Power +Message = HD power has been lost. \n- Treatment must be terminated.\n-Check all electrical connections.\n- The display and CPU will run off of the battery until \npower returns.\n-If power is returned prior to blood stopped countdown elapses, \nrinseback will be allowed. [308] -Title = DG CPi conductivity sensor transmits invalid characters -Message = If the CPi conductivity sensor transmits invalid characters for a certain period of time. +Title = DG Fault: CPi Comm +Message = Call service to schedule a repair. SRSDG 959 = [309] -Title = DG CPo conductivity sensor transmits invalid characters -Message = If the CPo conductivity sensor transmits invalid characters for a certain period of time. +Title = DG Fault: CPo Comm +Message = Call service to schedule a repair. SRSDG 960 = [310] -Title = DG CD1 conductivity sensor transmits invalid characters -Message = If the CD1 conductivity sensor transmits invalid characters for a certain period of time. +Title = DG Fault: CD1 Comm +Message = Call service to schedule a repair. SRSDG 962 = [311] -Title = DG CD2 conductivity sensor transmits invalid characters -Message = If the CD2 conductivity sensor transmits invalid characters for a certain period of time. +Title = DG Fault: CD2 Comm +Message = Call service to schedule a repair. SRSDG 961 = [312] -Title = The dialysate inlet flow is too low. -Message = If the dialysate inlet flow < 10 mL/min (raw) while in treatment mode or pre-treatment mode after prime and DPi is running. +Title = Dial. Gen: Dialysate In Flow +Message = Dialysate in flow rate is low\n- Check dialysate line connection to the DG. [313] -Title = DG cleaning mode inlet water temperature too low -Message = If inlet water temperature is < 24 C. +Title = Cleaning: Inlet Water Temp Low +Message = Cleaning Mode Inlet water temperature too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] -Title = DG cleaning mode inlet water pressure too high -Message = If inlet water temperature is > 80 psig. +Title = Cleaning: Inlet Water Press High +Message = Cleaning Mode Inlet water pressure too high.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] -Title = DG cleaning mode inlet water pressure too low -Message = If inlet water pressure is < 14 psig. +Title = Cleaning: Inlet Water Press Low +Message = Cleaning Mode Inlet water pressure too low.\n- Cleaning mode process has terminated.\n- Check inlet water before restarting cleaning mode. SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] -Title = Acid concentrate pump did not park successfully -Message = If the park took more than 1 second. +Title = DG Fault: Acid CP Pump Park +Message = Call service to schedule a repair. SRSDG 983 = [317] -Title = Bicarb concentrate pump did not park successfully -Message = If the park took more than 1 second. +Title = DG Fault: Bicarb CP Pump Park +Message = Call service to schedule a repair. SRSDG 984 = [318] -Title = HD Air Pump operation timeout -Message = If air pump has been turned on for greater than 10s. +Title = HD Fault: Air Pump Timeout +Message = Unable to reduce fluid level.\n- Air trap may be overfilled.\n- Pressure transducer protector may be contaminated.\n- Treatment must be terminated without option for rinseback.\n- Call service to schedule a repair. [319] -Title = Collect sample to check the chemical disinfectant acid -Message = Chemical disinfect flush sample alarm is prompted. +Title = Cleaning: Chem Residual Sampling +Message = Collect chemical residuals water sample from drain line. SRSDG 1040 = [320] -Title = The user did not proceed with collecting chemical disinfect flush sample for more than 30 minutes -Message = Chemical disinfect flush sample was not collected form more than 30 minutes. +Title = Cleaning: Chem Residual Timeout +Message = During Chemical Disinfection Flush\n-user residual sampling timed out. SRSDG 1043 = [321] -Title = Outlet reactor not healthy -Message = If the outlet reactor are not healthy. +Title = DG Fault: UVo Error +Message = Call service to schedule a repair. SRSDG 1069 = [322] -Title = UI POST Invalid Year -Message = If the UI POST reports a invalid year. +Title = HD POST Alarm: UI Invalid Year +Message = Call service to schedule a repair.