Index: Alarms/Alarms.conf =================================================================== diff -u -r031ac493ce8214e599dbc0d06ce795c9b4033a93 -rd9dc3d70cd64985cc56e8a6d38f61ebfc95dabbb --- Alarms/Alarms.conf (.../Alarms.conf) (revision 031ac493ce8214e599dbc0d06ce795c9b4033a93) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision d9dc3d70cd64985cc56e8a6d38f61ebfc95dabbb) @@ -59,7 +59,7 @@ [10] Title = Message = - = ALARM_ID_HD___AVAILABLE_7 + = ALARM_ID_HD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE [11] Title = @@ -152,7 +152,7 @@ [27] Title = Message = - = ALARM_ID_HD___AVAILABLE_3 + = ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED [28] Title = @@ -375,22 +375,24 @@ HD venous pressure self-test failure alarm = ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE [72] -Title = -Message = +Title = No load cell data received +Message = Call service to schedule a repair. +HD No load cell data message receive at least once every 2 seconds = [73] -Title = -Message = +Title = No dialysate temperature data received +Message = Call service to schedule a repair. +HD No dialysate temperature data message receive at least once every 2 seconds = [72] Title = Message = - = ALARM_ID_HD___AVAILABLE_5 + = ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE [73] Title = Message = - = ALARM_ID_HD___AVAILABLE_6 + = ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE [74] Title = Self Test Error @@ -798,14 +800,14 @@ EEPROM operations (read write erase) failure write erase) failure = ALARM_ID_NVDATA_EEPROM_OPS_FAILURE [155] -Title = -Message = - = ALARM_ID_HD___AVAILABLE_9 +Title = No reservoir data received +Message = Call service to schedule a repair. +HD No reservoir data message receive at least once every 2 seconds = ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE [156] -Title = -Message = - = ALARM_ID_HD___AVAILABLE_10 +Title = No DG op mode received +Message = Call service to schedule a repair. +HD No DG op mode message receive at least once every 2 seconds = ALARM_ID_HD_NEW_DG_OPERATION_MODE_MESSAGE_NOT_RECEIVE [157] Title = @@ -859,7 +861,7 @@ [167] Title = Door Open -Message = Hemodialysis Delivery device door open. +Message = Hemodialysis Delivery device door open.\n - Close open HD door\n - Press RESUME button HD cartridge door opened alarm = ALARM_ID_CARTRIDGE_DOOR_OPENED [168] @@ -1165,7 +1167,7 @@ [228] Title = Maximum Prime Time Exceeded Message = Priming failure. Check blood and dialysate lines for clamps and kinks. -HD time out on prime sub-mode purge air state. = ALARM_ID_HD_PRIME_PURGE_AIR_TIME_OUT +HD time out on prime sub-mode purge air state. = ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT [229] Title = Maximum Prime Time Exceeded @@ -1477,3 +1479,28 @@ Message = DG RO flow too low while heating = ALARM_ID_DG_FLOW_TOO_LOW_WHILE_HEATER_ON +[291] +Title = +Message = + = ALARM_ID_HD_DISINFECT_FLUSH + +[292] +Title = +Message = + = ALARM_ID_HD_DISINFECT_HEAT + +[293] +Title = +Message = + = ALARM_ID_HD_DISINFECT_CHEM + +[294] +Title = +Message = + = ALARM_ID_HD_DISINFECT_CHEM_INSTRUCTION1 + +[295] +Title = +Message = + = ALARM_ID_UI_POST_FAILURE_CLOUDSYNC +