Index: Alarms/Alarms.conf =================================================================== diff -u -re05ad09b5e52f0a809acff53022dae881af05e55 -re6e895b2392181ca38ee2bf1c33d1321ae396820 --- Alarms/Alarms.conf (.../Alarms.conf) (revision e05ad09b5e52f0a809acff53022dae881af05e55) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision e6e895b2392181ca38ee2bf1c33d1321ae396820) @@ -724,7 +724,7 @@ [163] Title = Dial. Gen: UF Rate Check -Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system (shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a repair. +Message = Ultrafiltration control detected and incorrect rate.\n- This can be caused by excessive movement of system\n(shock/vibration)\n- Check system is stable and level\n- Resume and If problem persists, call service to schedule a\nrepair. [164] Title = Dial. Gen: UF Rate Check @@ -1066,7 +1066,7 @@ [240] Title = Dial. Gen: Bicarb Conc. Cond. -Message = Bicarbonate concentrate conductivity during prime is out of range.\n- Check correct acid concentrate is connected and full.\n- Resume to reprime concentrate. +Message = Bicarbonate conductivity during prime is out of range.\n- Check correct concentrate is connected and full.\n- Resume to reprime concentrate. SRSDG 482 = [241]