Index: Alarms/Alarms.conf =================================================================== diff -u -rdc6f929465afd9338f4c9130206d3034e2a2de81 -rf09adad2cfbaa238cfedd09fe64df925a1a537b3 --- Alarms/Alarms.conf (.../Alarms.conf) (revision dc6f929465afd9338f4c9130206d3034e2a2de81) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision f09adad2cfbaa238cfedd09fe64df925a1a537b3) @@ -4,363 +4,363 @@ [1] Title = Stuck Button -Message = A button was pressed during start up and appears stuck. \n-Press and release the power and stop buttons on the front of the device. \n-Turn device off, then back on. \n-If problem continues, call service to schedule a repair. \n\nHD POST: Stuck Button +Message = A button was pressed during start up and appears stuck. \n- Press and release the power and stop buttons on the\nfront of the device.\n- Turn device off, then back on.\n- If problem continues, call service to schedule a repair.\n\nHD POST: Stuck Button [2] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: FPGA Compatibility +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: FPGA Compatibility [3] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: FPGA Compatibility +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: FPGA Compatibility SRSDG 602 SRSDG 814 SRSDG 1061 = [4] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: WD Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: WD Safestate [5] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: WD Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: WD Safestate SRSDG 53 = [6] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Proc-UI Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Proc- UI Comm [7] Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis\n device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): RTC Bat Low [8] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Acc Vector +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Acc Vector [9] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acc Vector +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Acc Vector SRSDG 323 = [10] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Emb Mode +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Emb Mode [11] Title = Low Dialysate Temperature -Message = Dialysate temperature is low. \n-Treatment is currently paused.\n-When dialysate temperature is within target range,\n treatment will need to be resumed.\n\nDial. Gen: Dialysate Temp Low +Message = Dialysate temperature is low. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed.\n\nDial. Gen: Dialysate Temp Low [12] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: HD Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: HD Image [13] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: DG Image +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: DG Image SRSDG 813 = [14] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Usage Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Usage Rec NVRAM CRC SRSDG 957 = [15] Title = High Dialysate Temperature -Message = Dialysate temperature is high. \n-Treatment is currently paused.\n-When dialysate temperature is within target range,\n treatment will need to be resumed.\n\nDial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed.\n\nDial. Gen: Dialysate Temp High [16] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Primary Alarm Current +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Primary Alarm Current [17] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI POST Not Complete +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI POST Not Complete [18] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (DG ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (DG ACK) [19] Title = High Dialysate Temperature -Message = Dialysate temperature is high. \n-Treatment is currently paused.\n-When dialysate temperature is within target range,\n treatment will need to be resumed. \n\nDial. Gen: Dialysate Temp High +Message = Dialysate temperature is high. \n- Treatment is currently paused.\n- When dialysate temperature is within target range,\n treatment will need to be resumed. \n\nDial. Gen: Dialysate Temp High [20] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Conductivity Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: Conductivity Cal SRSDG 16 = [21] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Drain Line Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Drain Line Vol Cal SRSDG 936 = [22] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Reservoir Vol Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Reservoir Vol Cal SRSDG 935 = [23] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acid Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Acid Conc Cal SRSDG 934 = [24] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Bicarb Conc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Bicarb Conc Cal SRSDG 933 = [25] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Acc Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Acc Cal SRSDG 817 = [26] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Acc Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Acc Cal [27] Title = Self Test: Blood Leak -Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Message = A problem has been detected with the blood leak sensor.\n- Confirm cartridge is properly installed.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre- Tx Wet Self- Test [28] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Conductivity RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Conductivity RTD Comm SRSDG 864 = [29] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: SP DAC Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: SP DAC Cal [30] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: SW Invalid State +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: SW Invalid State [31] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: BP Current [32] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed [33] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: BP Direction [34] Title = Blood Pump Error -Message = A problem has been detected with the blood pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue.\n\nHD Fault: BP Rotor +Message = A problem has been detected with the blood pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue.\n\nHD Fault: BP Rotor [35] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Current [36] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed [37] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Direction [38] Title = Dialysate Pump Error -Message = A problem has been detected with the dialysate pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue.\n\nHD Fault: DPi Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue.\n\nHD Fault: DPi Rotor [39] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Current [40] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed [41] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Direction +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Direction [42] Title = Dialysate Pump Error -Message = A problem has been detected with the dialysate pump.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue.\n\nHD Fault: DPo Rotor +Message = A problem has been detected with the dialysate pump.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue.\n\nHD Fault: DPo Rotor [43] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Iso ADC Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Iso ADC Comm SRSDG 954 = [44] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UI Comm +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: UI Comm [45] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: CAN CRC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: CAN CRC Com [46] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: CAN ACK UI Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: CAN ACK UI Com [47] Title = Message = \n\n [48] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UF Rate Tare Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: UF Rate Tare Err [49] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: FPGA Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: FPGA Com [50] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Valve Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Valve Control SRSDG 26 = [51] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: BP Speed [52] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPi Speed [53] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: DPo Speed [54] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Critical Data Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Critical Data Err [55] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Critical Data Err +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Critical Data Err SRSDG 921 = [56] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Acc Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Acc Com [57] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Acc Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Acc Com SRSDG 869 SRSDG 870 = [58] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Home +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Home [59] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Transition Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Transition Err [60] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Control +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Control [61] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Current +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Current [62] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Position +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Valve Position [63] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BLD Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: BLD Com [64] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Range SRSDG 874 = [65] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reject) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reject) [66] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Set Point +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: BLD Config Set Point [67] Title = Self Test: Cartridge -Message = A cartridge has been detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre- Tx Remove Cartridge [68] Title = Dialysate Generation Error Message = A problem has been detected with dialysate generation. \n- Treatment is paused and will be need to be resumed when \ndialysate is ready for use. \n\nDial. Gen: Dialysate not ready [69] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: BLD Cal +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD POST: BLD Cal [70] Title = Self Test: Cartridge -Message = A cartridge has been detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge +Message = A cartridge has been detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre- Tx Remove Cartridge [71] Title = Self Test: Cartridge -Message = A cartridge was detected during self-test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre-Tx Remove Cartridge +Message = A cartridge was detected during self- test.\n- Open door\n- Remove cartridge\n- Close door\n- Resume\n\nProcess: Pre- Tx Remove Cartridge [72] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (LC) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (LC) [73] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (TDI/TRO) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (TDI/TRO) [74] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVi On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: UVi On No Flow SRSDG 600 = [75] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe DAC +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe DAC [76] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Voltage Monitor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Voltage Monitor [77] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Voltage Monitor +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Voltage Monitor SRSDG 46 SRSDG 866 SRSDG 1088 = [78] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Enc Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Enc Dir [79] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Drv Dir +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Drv Dir [80] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Driver +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Driver [81] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump OT +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump OT [82] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump DAC Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump DAC Err [83] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump On Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump On Err [84] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: HD RTC Time +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: HD RTC Time SRSDG 815 = [85] @@ -369,194 +369,194 @@ [86] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: HD RTC Time +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: HD RTC Time [87] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Heat Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Heat Cal SRSDG 929 = [88] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 CP2 Speed Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 CP2 Speed Range [89] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVo On No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: UVo On No Flow SRSDG 599 = [90] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC A1/B1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: LC A1/B1 Comm SRSDG 604 = [91] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nLoss of DG Comm (FMD) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nLoss of DG Comm (FMD) [92] Title = Message = [93] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Permeate Cond High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RO Permeate Cond High SRSDG 430 = [94] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Pressure Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Pressure Sensor Range SRSDG 629 = [95] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: WD Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: WD Error SRSDG 907 = [96] Title = Incoming Water Temperature High -Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the temperature is within\ntarget range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp High +Message = Incoming water temperature is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp High SRSDG 410 = [97] Title = Conductivity Out of Range -Message = Treatment is currently paused.\n-Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n-Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Dialysate Cond. +Message = Treatment is currently paused.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Dialysate Cond. SRSDG 397 = [98] Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis\n device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Com +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Com [99] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Stall +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Stall [100] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Timeout No Cart ST +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Timeout No Cart ST [101] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI-Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI- Proc Comm [102] Title = Disinfection Interrupted -Message = Disinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Tpo THD Temp Delta +Message = Disinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Tpo THD Temp Delta SRSDG 856 SRSDG 930 = [103] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: Cal NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD POST: Cal NVRAM CRC [104] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Air Trap Sensor +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Air Trap Sensor [105] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: Cal NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: Cal NVRAM CRC SRSDG 608 = [106] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Flow Sensor Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: Flow Sensor Cal SRSDG 820 SRSDG 821 SRSDG 1087 = [107] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DG Restarted +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: DG Restarted [108] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump ADC Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump ADC Com [109] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Vol +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Vol [110] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed [111] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Syringe Pump Speed [112] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC A2/B2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: LC A2/B2 Comm SRSDG 953 = [113] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Current +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Current SRSDG 890 = [114] Title = Self Test: Venous Air Detector -Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Message = A problem has been detected with the venous air detector.\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n begin again using a new cartridge.\n\nProcess: Pre- Tx Wet Self- Test [115] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Temp Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Temp Sensor Range SRSDG 224 = [116] Title = Service Required: Dialysate Device -Message = A non-critical problem was detected with the hemodialysis\n device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): SD-Card Error +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): SD- Card Error [117] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of HD Comm (HD ACK) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of HD Comm (HD ACK) SRSDG 909 = [118] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RTC Config +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RTC Config SRSDG 43 = [119] Title = Service Required: Dialysate Device -Message = A non-critical problem was detected with the dialysate device.\n-Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): RTC Bat Low +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): RTC Bat Low SRSDG 908 = [120] Title = Self Test: Venous Air Detector -Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Message = A problem has been detected with the venous air detector\n- Check cartridge installation.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre- Tx Wet Self- Test [121] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Lg Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Lg Control SRSDG 611 = [122] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Sm Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Prim Heater Sm Control SRSDG 612 = [123] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Trimmer Heater Control +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Trimmer Heater Control SRSDG 613 = [124] Title = Blood Pump Stop Time Exceeded -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk [125] Title = Blood Pump Stop Time Exceeded -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk [126] Title = Blood Leak -Message = A blood leak has been detected. \n-Treatment must be terminated without rinseback.\n\nBlood Stop: Dialyzer Blood Leak +Message = A blood leak has been detected. \n- Treatment must be terminated without rinseback.\n\nBlood Stop: Dialyzer Blood Leak [127] Title = Low Venous Pressure -Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n-If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and terminate treatment\nwithout rinseback.\n-If access connections are properly secured, \npress RESUME to continue treatment.\n\nBlood Stop: Venous Pres. Low +Message = Venous pressure is low.\n- Ensure venous access connections are properly secured.\n- If there has been an access disconnection or needle\n dislodgement, follow facility policy to address and\nterminate treatment without rinseback.\n- If access connections are properly secured, \npress RESUME to continue treatment.\n\nBlood Stop: Venous Pres. Low [128] Title = Venous Air Detected @@ -568,79 +568,79 @@ [130] Title = High Venous Pressure -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n-Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Venous Pres. High [131] Title = Low Arterial Pressure -Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. Low +Message = Arterial pressure is low.\n- Inspect arterial bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. Low [132] Title = High Arterial Pressure -Message = Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n-Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. High +Message = Arterial pressure is high.\n- Ensure arterial access connections are properly secured.\n- Check needle placement and/or access position.\n- Resolve access issues if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Arterial Pres. High [133] Title = Dialysate Device Fluid Leak -Message = Fluid leak detected in the dialysate device.\n-Treatment must be terminated.\n- Turn off water source. \n-Disconnect water and electrical connections from the device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Fluid Leak in DG +Message = Fluid leak detected in the dialysate device.\n- Treatment must be terminated.\n- Turn off water source. \n- Disconnect water and electrical connections from the device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Fluid Leak in DG SRSDG 328 = [134] Title = Fluid Leak -Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n-Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, resume treatment.\n\nBlood Stop: Fluid Leak in HD +Message = Fluid leak detected in hemodialysis device.\n- If in treatment, blood pump is off.\n- Identify and correct the cause of the fluid leak.\n- Clean up fluid and dry leak sensor.\n- Once fluid has been cleaned up, resume treatment.\n\nBlood Stop: Fluid Leak in HD [135] Title = Blood Pump Stopped: Excessive Movement -Message = Excessive movement of the hemodialysis device has been\ndetected. \n-Treatment is currently paused. \n-When system movement has stablized, \ntreatment will need to be resumed.\n-To resume treatment, press OK.\n\nBlood Stop: HD Shock Detected +Message = Excessive movement of the hemodialysis device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press OK.\n\nBlood Stop: HD Shock Detected [136] Title = Dialysate Device: Excessive Movement -Message = Excessive movement of the dialysate device has been\ndetected. \n-Treatment is currently paused. \n-When system movement has stablized, \ntreatment will need to be resumed.\n-To resume treatment, press OK.\n\nDial. Gen: Shock-Vibration +Message = Excessive movement of the dialysate device has been\ndetected. \n- Treatment is currently paused. \n- When system movement has stablized, \ntreatment will need to be resumed.\n- To resume treatment, press OK.\n\nDial. Gen: Shock- Vibration SRSDG 327 = [137] Title = Blood Pump Stopped: Excessive Tilt -Message = Excessive tilt of the hemodialysis device has been detected. \n-Treatment is currently paused. \n-When the device is stationary and level, \ntreatment will need to be resumed.\n-To resume treatment, press OK.\n\nBlood Stop: HD Tilted +Message = Excessive tilt of the hemodialysis device has been detected. \n- Treatment is currently paused. \n- When the device is stationary and level, \ntreatment will need to be resumed.\n- To resume treatment, press OK.\n\nBlood Stop: HD Tilted [138] Title = Dialysate Device: Excessive Tilt -Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n-Confirm the device is stationary and level before continuing.\n\nDial. Gen: Tilt +Message = Excessive tilt of the dialysate device has been detected.\n- Treatment is paused.\n- Confirm the device is stationary and level before continuing.\n\nDial. Gen: Tilt SRSDG 326 = [139] Title = Power Loss -Message = AC Power has been lost.\n-Check all electrical connections.\n- Press OK to return to welcome screen. \n\nLoss of AC Power +Message = AC Power has been lost.\n- Check all electrical connections.\n- Press OK to return to welcome screen. \n\nLoss of AC Power [140] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Timeout) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Timeout) [141] Title = Blood Pump Stopped: Venous Air Detector Message = A problem was detected with the venous air detector.\n- Confirm arterial access placement and ensure\nconnections are secure.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Air Trap Fill [142] Title = Blood Pump Stopped: Occulusion -Message = Occlusion detected after the blood pump.\n-Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Blood Pump Occlusion +Message = Occlusion detected after the blood pump.\n- Inspect bloodlines for kinks or clamps.\n- Check needle placement or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback.\n\nBlood Stop: Blood Pump Occlusion [143] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TRo TDi Delta +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: TRo TDi Delta SRSDG 912 = [144] Title = Disinfection: Water Conductivity High -Message = Incoming water conductivity is too high.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond High +Message = Incoming water conductivity is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond High SRSDG 1066 SRSDG 1054 SRSDG 970 = [145] Title = Conductivity Error During Prime -Message = Conductivity out of range during prime.\n-Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n-Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Acid Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Acid Conc. Cond. SRSDG 846 = [146] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: DG RTC Timer +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: DG RTC Timer SRSDG 312 = [147] @@ -650,105 +650,105 @@ [148] Title = Disinfection: Water Temperature High -Message = Incoming water temperature is too high.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp High +Message = Incoming water temperature is too high.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp High SRSDG 1065 SRSDG 1053 SRSDG 969 = [149] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: WD Error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: WD Error [150] Title = Incoming Water Conductivity High -Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the conductivity is within\ntarget range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond High +Message = Conductivity of incoming water is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond High SRSDG 391 = [151] Title = Incoming Water Conductivity Low -Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the conductivity is within\ntarget range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond Low +Message = Conductivity of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the conductivity is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Cond Low SRSDG 826 = [152] Title = Incoming Water Pressure Low -Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the pressure is within\ntarget range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres Low +Message = Water pressure of incoming water is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres Low SRSDG 417 = [153] Title = Pre-Treatment Process Terminated -Message = The pre-treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n-Press OK for continue with next steps to remove cartridge\nfrom the device.\n\nProcess: Pre-Tx Terminated +Message = The pre- treatment process has been terminated because \ntreatment did not begin within two hours of priming. \n- Press OK for continue with next steps to remove cartridge\nfrom the device.\n\nProcess: Pre- Tx Terminated [154] Title = Disinfection: Water Conductivity Low -Message = Incoming water conductivity is too low.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond Low +Message = Incoming water conductivity is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Cond Low SRSDG 1067 SRSDG 1055 SRSDG 971 = [155] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reservoir) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Reservoir) [156] Title = Communication Error -Message = A communication problem has been detected.\n-Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Op Mode) +Message = A communication problem has been detected.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nLoss of DG Comm (Op Mode) [157] Title = Chemical Disinfectant Not Detected -Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Prime +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet. \n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Prime SRSDG 988 = [158] Title = Incoming Water Temperature Low -Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the temperature is within\ntarget range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp Low +Message = Incoming water temperature is low.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the temperature is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Temp Low SRSDG 408 = [159] Title = Remove Chemical Disinfectant -Message = The chemical disinfectant is no longer needed. \n-Disconnect chemical disinfectant from the device. \n-Place the concentrate connection cap on the device.\n-Keep dialysate cap on the device.\n\nCleaning: Chem Disinf Connect Cap +Message = The chemical disinfectant is no longer needed. \n- Disconnect chemical disinfectant from the device. \n- Place the concentrate connection cap on the device.\n- Keep dialysate cap on the device.\n\nCleaning: Chem Disinf Connect Cap SRSDG 1044 = [160] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Rotor Max Speed +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: BP Rotor Max Speed [161] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HD-DG Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: HD- DG Comm SRSDG 122 = [162] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Proc-FPGA Clock +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Proc- FPGA Clock SRSDG 867 = [163] Title = Ultrafiltration Error -Message = A problem was detected with the dialysate device. \n-Confirm the device is level with minimal movement. \n-Resume treatment. \n-If problem persists, call service to report the issue and\nschedule a repair. \n\nDial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. \n\nDial. Gen: UF Rate Check [164] Title = Ultrafiltration Error -Message = A problem was detected with the dialysate device. \n-Confirm the device is level with minimal movement. \n-Resume treatment. \n-If problem persists, call service to report the issue and\nschedule a repair. \n\nDial. Gen: UF Rate Check +Message = A problem was detected with the dialysate device. \n- Confirm the device is level with minimal movement. \n- Resume treatment. \n- If problem persists, call service to report the issue and\nschedule a repair. \n\nDial. Gen: UF Rate Check [165] Title = Blood Pump Stop Time Exceeded -Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk +Message = The blood pump has been off for too long.\n- Treatment must be terminated without rinseback.\n\nBlood Stop: Blood Clot Risk [166] Title = Message = \n\n [167] Title = Blood Pump Stopped: Door Open -Message = The device door is open.\n Close the device door. \n-Press RESUME to continue.\n\nBlood Stop: Door Open +Message = The device door is open.\n Close the device door. \n- Press RESUME to continue.\n\nBlood Stop: Door Open [168] Title = Message = \n\n [169] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Dialysate Flow Sens Rg +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Dialysate Flow Sens Rg SRSDG 643 = [170] @@ -757,106 +757,106 @@ [171] Title = Heparin Pump Occlusion -Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped.\n\nProcess: Pre-Tx Hep. Pump Occ. +Message = A problem has been detected with the heparin pump.\n- Ensure heparin line is not kinked or clamped.\n\nProcess: Pre- Tx Hep. Pump Occ. [172] Title = Not Enough Heparin -Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \n- ensure the correct amount of heparin has been drawn\nand reload the syringe. \n\nProcess: Pre-Tx Not Enough Hep. +Message = There is not enough heparin available for treatment.\n- Ensure the syringe pump is touching the syringe plunger.\n- If the syringe pump is touching the syringe plunger,\nend treatment.\n- If syringe pump is not touching the syringe plunger, \nensure the correct amount of heparin has been drawn and\nreload the syringe. \n\nProcess: Pre- Tx Not Enough Hep. [173] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: RTC Config +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: RTC Config [174] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: HD RTC Timer +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD POST: HD RTC Timer [175] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Pump Commutation Err +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Pump Commutation Err [176] Title = Message = \n\n [177] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Software Invalid State +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Software Invalid State SRSDG 919 = [178] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: UI-Proc Comm +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: UI- Proc Comm = SRSUI 1066 [179] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: FPGA Com +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: FPGA Com SRSDG 605 = [180] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Sensor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Sensor Range SRSDG 883 = [181] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Tare Out of Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Tare Out of Range SRSDG 849 SRSDG 838 = [182] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nDG POST: LC Cal +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nDG POST: LC Cal SRSDG 162 = [183] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Invalid Resv Transfer +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Invalid Resv Transfer SRSDG 851 = [184] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVi Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: UVi Error SRSDG 559 = [185] Title = Service Required: Dialysate Device -Message = A non-critical problem was detected with the dialysate device.\n-Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the dialysate device.\n- Call service to report the issue and schedule a repair.\n\nDG Alarm (Low): Fan Speed SRSDG 628 = [186] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1/CP2 Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CP1/CP2 Fault SRSDG 891 = [187] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CP1 Speed Fault SRSDG 905 = [188] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CP2 Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CP2 Speed Fault SRSDG 906 = [189] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Fault SRSDG 448 = [190] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Not Off +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Speed Not Off [191] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow High +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow High SRSDG 885 = [192] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Low +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RO Flow Low SRSDG 945 = [193] @@ -865,95 +865,95 @@ [194] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Driver Feedback +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RO Driver Feedback SRSDG 881 = [195] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Pump High Pressure +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RO Pump High Pressure SRSDG 925 = [196] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPi/CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CPi/CPo Comm SRSDG 955 = [197] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1/CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CD1/CD2 Comm SRSDG 956 = [198] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HP On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: HP On With No Flow SRSDG 872 = [199] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: HR On With No Flow +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: HR On With No Flow SRSDG 871 = [200] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Thermistor Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Thermistor Range SRSDG 178 = [201] Title = Self Test: Dialysate Flow Rate -Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre- Tx Wet Self- Test [202] Title = Self Test: Pressure -Message = A problem has been detected during self-test. \n-Confirm the cartridge is properly installed. \n-Ensure arterial and venous pressure sensor lines are not kinked.\n-Press OK to resume. \n-If problem persists, install a new cartridge.\n\nProcess: Pre-Tx Dry Self-Test +Message = A problem has been detected during self- test. \n- Confirm the cartridge is properly installed. \n- Ensure arterial and venous pressure sensor lines are\nnot kinked.\n- Press OK to resume. \n- If problem persists, install a new cartridge.\n\nProcess: Pre- Tx Dry Self- Test [203] Title = Message = \n\n [204] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: RO Rejection +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: RO Rejection SRSDG 429 = [205] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Cond Sensors Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Cond Sensors Error SRSDG 913 = [206] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Fill Timeout +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Fill Timeout SRSDG 437 = [207] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: FMP Flow Rate Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: FMP Flow Rate Error SRSDG 240 = [208] Title = High Venous Pressure -Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n-Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. \n\nBlood Stop: Venous Pres. High +Message = Venous pressure is high. \n- Check venous bloodlines for kinks or clamps.\n- Check needle placement and/or access position.\n- Resolve access issues, if applicable.\n- If clots are present, end treatment without rinseback. \n\nBlood Stop: Venous Pres. High [209] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Flush Circ Error +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Flush Circ Error SRSDG 980 = [210] Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis\n device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Fault +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Battery Fault [211] Title = Approaching MAX Blood Pump Stop Time -Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback.\n\nBlood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback.\n\nBlood Stopped: Warning [212] Title = Approaching MAX Blood Pump Stop Time -Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback.\n\nBlood Stopped: Warning +Message = The blood has been stopped for nearly 5 minutes.\n- You must resume treatment before the blood pump stop time\ntimer expires.\n- If treatment is not resumed before the timer expires,\ntreatment will be terminated without rinseback.\n\nBlood Stopped: Warning [213] Title = Prime Complete - 30 MIN WARNING -Message = The priming process completed 90 minutes ago. \n-Treatment must begin within 30 minutes or the\npre-treatment process will be terminated.\n-Press OK to continue. \n\nProcess: Pre-Tx Prime Complete +Message = The priming process completed 90 minutes ago. \n- Treatment must begin within 30 minutes or the\npre- treatment process will be terminated.\n- Press OK to continue. \n\nProcess: Pre- Tx Prime Complete [214] Title = @@ -965,23 +965,23 @@ [216] Title = Treatment Complete -Message = Dialysis treatment is complete.\n-Press OK to end treatment and rinseback.\n\nProcess: End of Dialysis Tx +Message = Dialysis treatment is complete.\n- Press OK to end treatment and rinseback.\n\nProcess: End of Dialysis Tx [217] Title = Message = \n\n [218] Title = Syringe Detected -Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n-Press OK to continue.\n\nProcess: Remove Syringe +Message = Heparin syringe was not removed from device.\n- Remove heparin syringe.\n- Press OK to continue.\n\nProcess: Remove Syringe [219] Title = Blood Stopped: Syringe Not Detected -Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n-Press Ok to resume treatment.\n\nBlood Stop: Syringe Not Detected +Message = There is a problem detecting a syringe in the heparin pump.\n- Confirm the heparin syringe is properly installed. \n- Press Ok to resume treatment.\n\nBlood Stop: Syringe Not Detected [220] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: THD RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: THD RTD Comm SRSDG 947 = [221] @@ -990,88 +990,88 @@ [222] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: BP Occ Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: BP Occ Com [223] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Art. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Art. Pres. Com [224] Title = Blood Pump Stopped: Treatment Paused -Message = Treatment has been stopped by user.\n- Press OK to resume treatment. \n-If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. \n\nBlood Stopped: Treatment Pause +Message = Treatment has been stopped by user.\n- Press OK to resume treatment. \n- If treatment is not resumed within five minutes,\ntreatment will be terminated without rinseback. \n\nBlood Stopped: Treatment Pause [225] Title = Treatment Complete -Message = Dialysis treatment is complete.\n-Press OK to end treatment and rinseback.\n\nProcess: End of Treatment +Message = Dialysis treatment is complete.\n- Press OK to end treatment and rinseback.\n\nProcess: End of Treatment [226] Title = Priming Complete -Message = Priming is complete. \n-Treatment must begin within 120 minutes \nor the pre-treatment process will be terminated.\n-Press OK to continue. \n\nProcess: Pre-Tx Priming Complete +Message = Priming is complete. \n- Treatment must begin within 120 minutes \nor the pre- treatment process will be terminated.\n- Press OK to continue. \n\nProcess: Pre- Tx Priming Complete [227] Title = Message = \n\n [228] Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout [229] Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout [230] Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout [231] Title = Self Test: Pressure -Message = A problem has been detected during self-test. \n- Confirm cartridge is properly installed without kinks or clamps. \n-Ensure all connections are secure. \n-Press OK to resume. \n- If problem persists, install a new cartridge.\n\nProcess: Pre-Tx Dry Self-Test +Message = A problem has been detected during self- test. \n- Confirm cartridge is properly installed without kinks or clamps.\n- Ensure all connections are secure. \n- Press OK to resume. \n- If problem persists, install a new cartridge.\n\nProcess: Pre- Tx Dry Self- Test [232] Title = Self Test: Dialysate Flow Rate -Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre-Tx Wet Self-Test +Message = A problem has been detected when measuring dialysate flow.\n- Confirm cartridge is properly installed and primed.\n- If problem persists, terminate treatment preparation and \n start over using a new cartridge.\n\nProcess: Pre- Tx Wet Self- Test [233] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Cal CRC +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Cal CRC SRSDG 964 SRSDG 949 SRSDG 1089 = [234] Title = Incoming Water Pressure High -Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n-When the pressure is within\ntarget range RESUME button will be available. \n-Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres High +Message = Incoming water pressure is high.\n- Treatment is currently paused.\n- Check the incoming water supply.\n- When the pressure is within target range\nRESUME button will be available.\n- Press RESUME to continue treatment.\n\nDial. Gen: Inlet Water Pres High SRSDG 1037 = [235] Title = Rinseback Complete -Message = Rinseback is complete. \n-Follow facililty policy prior to disconnecting patient. \n-Press OK to continue. \n\nProcess: Rinseback Complete +Message = Rinseback is complete. \n- Follow facililty policy prior to disconnecting patient. \n- Press OK to continue. \n\nProcess: Rinseback Complete [236] Title = Self Test: Fluid Present -Message = A problem has been detected during self-test. \n-Cartridges are single-use only and fluid is present.\n- Install new cartridge to resume.\n\nProcess: Pre-Tx Dry Self-Test +Message = A problem has been detected during self- test. \n- Cartridges are single- use only and fluid is present.\n- Install new cartridge to resume.\n\nProcess: Pre- Tx Dry Self- Test [237] Title = Priming Error -Message = A problem has been detected during prime. \n- Prime was unable to complete. \n -Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre-Tx Prime Timeout +Message = A problem has been detected during prime. \n- Prime was unable to complete. \n - Check blood and dialysate lines for clamps, kinks or leaks. \n\nProcess: Pre- Tx Prime Timeout [238] Title = Cartridge Not Dectected -Message = A problem has been detected during self-test. \n-Cartridge is not detected.\n- Confirm cartridge is properly installed.\n\nProcess: Pre-Tx Dry Self-Test +Message = A problem has been detected during self- test. \n- Cartridge is not detected.\n- Confirm cartridge is properly installed.\n\nProcess: Pre- Tx Dry Self- Test [239] Title = Cartridge Still Detected -Message = A problem has been detected during self-test. \n-Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed.\n\nProcess: Post-Tx Remove Cartridge +Message = A problem has been detected during self- test. \n- Cartridge is still present or was unloaded incorrectly.\n- Confirm the cartridge has been fully removed.\n\nProcess: Post- Tx Remove Cartridge [240] Title = Conductivity Error During Prime -Message = Conductivity out of range during prime.\n-Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n-Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Bicarb Conc. Cond. +Message = Conductivity out of range during prime.\n- Confirm that acid and bicarbonate concentrates\nare securely connected.\n- If empty, replace. \n- Treatment will need to be resumed when conductivity\nis within target range.\n\nDial. Gen: Bicarb Conc. Cond. SRSDG 482 = [241] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Resv Drain Timeout +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Resv Drain Timeout SRSDG 848 SRSDG 90 SRSDG 561 @@ -1103,7 +1103,7 @@ [242] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Resv Fill Timeout +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Resv Fill Timeout SRSDG 850 SRSDG 573 SRSDG 574 @@ -1125,66 +1125,66 @@ [243] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Fluid Loss Detected +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Fluid Loss Detected SRSDG 853 = [244] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Temp Sensor Check +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Temp Sensor Check SRSDG 852 = [245] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Heat Disinfect Temp TO +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Heat Disinfect Temp TO SRSDG 854 = [246] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: RO Water Cond Error +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: RO Water Cond Error [247] Title = Message = \n\n [248] Title = Disinfection Interrupted -Message = Dinsinfection terminated unexpectedly.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinfect Proc +Message = Dinsinfection terminated unexpectedly.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinfect Proc [249] Title = Chemical Disinfectant Not Detected -Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n-Restart disinfection.\n-If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Cond +Message = Unable to sense proper chemical disinfectant.\n- Disinfection process has terminated.\n- Confirm chemical disinfectant is connected to the\nconcentrate inlet.\n- Confirm chemical disinfectant container is not empty.\n- Restart disinfection.\n- If problem persists, call service to schedule a repair.\n\nCleaning: Chem Disinf PAA Cond SRSDG 992 = [250] Title = Connect Chemical Disinfectant -Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n-Keep dialysate cap on the device.\n-Place disinfectant straw in the disinfectant bottle. \n-Attach straw to concentrate connection and ensure\nconnection is secure.\n-Press OK to continue.\n\nCleaning: Chem Disinf Connect PAA +Message = Connect chemical disinfectant to the device. \n- Remove concentrate cap.\n- Keep dialysate cap on the device.\n- Place disinfectant straw in the disinfectant bottle. \n- Attach straw to concentrate connection and ensure\nconnection is secure.\n- Press OK to continue.\n\nCleaning: Chem Disinf Connect PAA SRSDG 1046 = [251] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Sys Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Sys Rec NVRAM CRC [252] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Serv Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Serv Rec NVRAM CRC [253] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Sys Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: Sys Rec NVRAM CRC SRSDG 609 = [254] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Serv Rec NVRAM CRC +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: Serv Rec NVRAM CRC SRSDG 610 = [255] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Compatibility +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI Compatibility [256] Title = Chemical Flush in Progress -Message = Chemical flush in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Flush In Progress +Message = Chemical flush in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Flush In Progress = SRSUI 497 [257] @@ -1197,16 +1197,16 @@ [259] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Temp Sensors Error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Temp Sensors Error [260] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Image +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI Image = SRSUI 1067 [261] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI CAN Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI CAN Driver = SRSUI 1068 [262] @@ -1216,51 +1216,51 @@ [263] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Touch Driver +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI Touch Driver = SRSUI 1070 [264] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI SD-Card +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI SD- Card = SRSUI 1071 [265] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI RTC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI RTC = SRSUI 1072 [266] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI WiFi +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI WiFi = SRSUI 1073 [267] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Bluetooth +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI Bluetooth = SRSUI 1074 [268] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Ethernet +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI Ethernet = SRSUI 1075 [269] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: UI Sound +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: UI Sound = SRSUI 1076 [270] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Proc Safestate +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Proc Safestate [271] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Proc Safestate +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: Proc Safestate SRSDG 822 = [272] Title = Service Required: Hemodialysis Device -Message = A non-critical problem was detected with the hemodialysis\n device.\n-Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Fan Speed +Message = A non- critical problem was detected with the hemodialysis\n device.\n- Call service to report the issue and schedule a repair.\n\nHD Alarm (Low): Fan Speed [273] Title = @@ -1272,28 +1272,28 @@ [275] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Arterial Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Arterial Press Range [276] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Venous Press Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Venous Press Range [277] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Occlusion Sens Range +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Occlusion Sens Range [278] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Overfill +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Overfill [279] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Drain +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Reservoir Drain SRSDG 752 = [280] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Direction Fault +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Direction Fault SRSDG 888 = [281] @@ -1308,12 +1308,12 @@ [283] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC Range +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: LC Range SRSDG 615 = [284] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Redundant LC Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Redundant LC Check SRSDG 383 = [285] @@ -1322,158 +1322,158 @@ [286] Title = Concentrate Connection Cap Removed -Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n-Replace cap before attempting disinfection again. \n\nCleaning: Conc Cap Removed +Message = The concentrate connection cap was removed from the \ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. \n\nCleaning: Conc Cap Removed [287] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Proc-FPGA Clock +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Proc- FPGA Clock [288] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: LC Drift Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: LC Drift Check [289] Title = Dialysate Connection Cap Removed -Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n-Replace cap before attempting disinfection again. \n\nCleaning: Dialysate Cap Removed +Message = The dialysate connection cap was removed from the\ndialysate device during disinfection.\n- Disinfection terminated.\n- Replace cap before attempting disinfection again. \n\nCleaning: Dialysate Cap Removed SRSDG 928 = [290] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Ven. Pres. Com +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Ven. Pres. Com [291] Title = Flush In Progress -Message = System flush is in progress\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Flush In Progress +Message = System flush is in progress\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Flush In Progress [292] Title = Heat Disinfection in Progress -Message = Heat disinfection is in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Heat Disinf In Progress +Message = Heat disinfection is in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Heat Disinf In Progress [293] Title = Chemical Disinfection in Progress -Message = Chemical disinfection in progress.\n-Please wait for the process to complete. \n-You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Disinfect In Progress +Message = Chemical disinfection in progress.\n- Please wait for the process to complete. \n- You will not be able to create a treatment until the process\nis finished.\n- To end this process, press STOP on the front of the device.\n\nCleaning: Chem Disinfect In Progress [294] Title = Message = \n\n [295] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: UI CloudSync +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD POST: UI CloudSync = SRSUI 1238 [296] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TDi RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: TDi RTD Comm SRSDG 951 = [297] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: TRo RTD Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: TRo RTD Comm SRSDG 950 = [298] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Pres Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Baro Pres Comm SRSDG 948 = [299] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG POST: Serial Number Check +Message = A problem was detected with the dialysate device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG POST: Serial Number Check SRSDG 952 = [300] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST: Serial Number Check +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST: Serial Number Check [301] Title = Message = \n\n [302] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Processor RAM error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Processor RAM error SRSDG 126 = [303] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Hall Commutation +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: DRP Hall Commutation SRSDG 958 = [304] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD POST: Usage Rec NVRAM CRC +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD POST: Usage Rec NVRAM CRC [305] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nHD Fault: Processor RAM error +Message = A problem was detected with the hemodialysis device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nHD Fault: Processor RAM error [306] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 CD2 Check +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 CD2 Check [307] Title = Power Loss -Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n-Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed.\n\nBlood Stop: Loss of AC Power +Message = Treatment must be terminated due to a power loss.\n- Follow clinic instructions on when to initiate manual rinseback.\n- Check all electrical connections.\n- If power returns within 5 minutes of outage, \n- rinseback will be allowed.\n\nBlood Stop: Loss of AC Power [308] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPi Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CPi Comm SRSDG 959 = [309] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CPo Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CPo Comm SRSDG 960 = [310] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CD1 Comm SRSDG 962 = [311] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: CD2 Comm +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: CD2 Comm SRSDG 961 = [312] Title = Low Dialysate Flow -Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nRESUME button will be available.\n-Press RESUME to continue.\n\nDial. Gen: Dialysate In Flow +Message = Dialysate flow rate is low. \n- Confirm dialysate concentrate connections are secure. \n- When the flow rates are within target range \nRESUME button will be available.\n- Press RESUME to continue.\n\nDial. Gen: Dialysate In Flow [313] Title = Disinfection: Water Temperature Low -Message = Incoming water temperature is too low.\n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp Low +Message = Incoming water temperature is too low.\n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Temp Low SRSDG 1064 SRSDG 1052 SRSDG 968 = [314] Title = Disinfection: Water Pressure High -Message = Incoming water pressure is too high. \n-Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Press High +Message = Incoming water pressure is too high. \n- Disinfection has been terminated.\n- Check incoming water before initiating disinfection again. \n\nCleaning: Inlet Water Press High SRSDG 1063 SRSDG 1051 SRSDG 967 = [315] Title = Disinfection: Water Pressure Low -Message = Incoming water pressure is too low. \n-Disinfection has been terminated.\n- Check incoming water initiating disinfection again. \n\nCleaning: Inlet Water Press Low +Message = Incoming water pressure is too low. \n- Disinfection has been terminated.\n- Check incoming water initiating disinfection again. \n\nCleaning: Inlet Water Press Low SRSDG 1068 SRSDG 1056 SRSDG 966 = [316] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Acid CP Pump Park +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Acid CP Pump Park SRSDG 983 = [317] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: Bicarb CP Pump Park +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: Bicarb CP Pump Park SRSDG 984 = [318] Title = Air Trap Control Error -Message = A problem has been detected with the air trap level\ncontrol system.\n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue.\n\nHD Fault: Air Pump Timeout +Message = A problem has been detected with the air trap level\ncontrol system.\n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue.\n\nHD Fault: Air Pump Timeout [319] Title = Residual Chemical Sampling @@ -1482,15 +1482,15 @@ [320] Title = Chemical Residual Timeout -Message = A chemical residual sample was not collected \n-within 30min of chemical flush.\n-To collect a sample for chemical residual,\nyou must repeat the chemical flush.\n\nCleaning: Chem Residual Timeout +Message = A chemical residual sample was not collected \n- within 30min of chemical flush.\n- To collect a sample for chemical residual,\nyou must repeat the chemical flush.\n\nCleaning: Chem Residual Timeout SRSDG 1043 = [321] Title = Service Required: Dialysate Device -Message = A problem was detected with the dialysate device. \n-Treatment must be terminated.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair.\n\nDG Fault: UVo Error +Message = A problem was detected with the dialysate device. \n- Treatment must be terminated.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair.\n\nDG Fault: UVo Error SRSDG 1069 = [322] Title = Service Required: Hemodialysis Device -Message = A problem was detected with the hemodialysis device.\n-Locate the ID code found in the bottom left corner\nof the alarm screen.\n-Call service to report the issue and schedule a repair. \n\nHD POST Alarm: UI Invalid Year +Message = A problem was detected with the hemodialysis device.\n- Locate the ID code found in the bottom left corner\nof the alarm screen.\n- Call service to report the issue and schedule a repair. \n\nHD POST Alarm: UI Invalid Year