Index: .gitignore =================================================================== diff -u --- .gitignore (revision 0) +++ .gitignore (revision f1537dd481acf785b57e85aa6c57ac5f53091fdb) @@ -0,0 +1,2 @@ + +*.*~ Index: Alarms/Alarms.conf =================================================================== diff -u -r42027ca5d01fb16232e2ded2aca8471af8c89c9f -rf1537dd481acf785b57e85aa6c57ac5f53091fdb --- Alarms/Alarms.conf (.../Alarms.conf) (revision 42027ca5d01fb16232e2ded2aca8471af8c89c9f) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision f1537dd481acf785b57e85aa6c57ac5f53091fdb) @@ -1,1540 +1,1604 @@ [0] Title = Message = -Indication for when no alarms is a possible situation = ALARM_ID_NO_ALARM [1] Title = Self Test Error Message = Call service to schedule a repair. -Stuck button POST failure. Stop or Off button detected to be pressed for at least 1 second during test shortly after power up = ALARM_ID_STUCK_BUTTON_TEST_FAILED +Self Test Error = Call service to schedule a repair. [2] Title = Self Test Error Message = Call service to schedule a repair. -HD FPGA POST failure = ALARM_ID_HD_FPGA_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [3] Title = Self Test Error Message = Call service to schedule a repair. -DG FPGA POST failure = ALARM_ID_DG_FPGA_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [4] Title = Self Test Error Message = Call service to schedule a repair. -HD Watchdog POST failure = ALARM_ID_HD_WATCHDOG_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [5] Title = Self Test Error Message = Call service to schedule a repair. -DG Watchdog POST failure = ALARM_ID_DG_WATCHDOG_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [6] Title = Self Test Error Message = Call service to schedule a repair. -UI communication POST failure. UI failed to communicate within a reasonable time after power up = ALARM_ID_UI_COMM_POST_FAILED +Self Test Error = Call service to schedule a repair. [7] Title = Machine Error Message = Call service to schedule a repair. -Real-time clock configuration error = ALARM_ID_HD_RTC_BATTERY_LOW +Machine Error = Call service to schedule a repair. [8] Title = Self Test Error Message = Call service to schedule a repair. -HD accelerometer failed POST = ALARM_ID_HD_ACCELEROMETER_SELF_TEST_FAILURE +Self Test Error = Call service to schedule a repair. [9] Title = Self Test Error Message = Call service to schedule a repair. -DG accelerometer failed POST = ALARM_ID_DG_ACCELEROMETER_SELF_TEST_FAILURE +Self Test Error = Call service to schedule a repair. [10] -Title = Self Test Error -Message = - = ALARM_ID_HD_BLOOD_LEAK_SENSOR_EMBEDDED_MODE_FAILURE +Title = Machine Error +Message = Call service to schedule a repair. +Machine Error = Call service to schedule a repair. [11] -Title = -Message = - = ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP +Title = Dialysate Temperature Error +Message = Call service to schedule a repair. +Dialysate Temperature Error = Call service to schedule a repair. [12] Title = Self Test Error Message = Call service to schedule a repair. -HD firmware image integrity POST test failed = ALARM_ID_HD_INTEGRITY_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [13] Title = Self Test Error Message = Call service to schedule a repair. -DG firmware image integrity POST test failed = ALARM_ID_DG_INTEGRITY_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [14] -Title = -Message = - = ALARM_ID_DG_INVALID_USAGE_RECORD_CRC +Title = Self Test Error +Message = Call service to schedule a repair. +Self Test Error = Call service to schedule a repair. [15] -Title = -Message = - = ALARM_ID_HD_INVALID_USAGE_RECORD +Title = Self Test Error +Message = Call service to schedule a repair. +Self Test Error = Call service to schedule a repair. [16] Title = Self Test Error Message = Call service to schedule a repair. -HD alarm audio failed POST = ALARM_ID_HD_ALARM_AUDIO_SELF_TEST_FAILURE +Self Test Error = Call service to schedule a repair. [17] Title = Self Test Error Message = Call service to schedule a repair. -HD UI POST failed = ALARM_ID_HD_UI_POST_FAILED +Self Test Error = Call service to schedule a repair. [18] Title = Calibration Error Message = Call service to schedule a repair. -DG load cells invalid calibration record = ALARM_ID_DG_PRESSURE_SENSORS_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [19] -Title = Calibration Error +Title = Dialysate Temperature Error Message = Call service to schedule a repair. -DG flow sensors invalid calibration record = ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP +Dialysate Temperature Error = Call service to schedule a repair. [20] Title = Calibration Error Message = Call service to schedule a repair. -DG conductivity sensors invalid calibration record = ALARM_ID_DG_COND_SENSORS_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [21] Title = Calibration Error Message = Call service to schedule a repair. -DG drain line volume invalid calibration record = ALARM_ID_DG_DRAIN_LINE_VOLUME_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [22] Title = Calibration Error Message = Call service to schedule a repair. -DG reservoirs invalid calibration record = ALARM_ID_DG_RESERVOIRS_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [23] Title = Calibration Error Message = Call service to schedule a repair. -DG acid concentrate calibration record = ALARM_ID_DG_ACID_CONCENTRATE_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [24] Title = Calibration Error Message = Call service to schedule a repair. -DG bicarb concentrate calibration record = ALARM_ID_DG_BICARB_CONCENTRATE_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [25] Title = Calibration Error Message = Call service to schedule a repair. -DG accelerometer invalid calibration record = ALARM_ID_DG_ACCELEROMETERS_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [26] Title = Calibration Error Message = Call service to schedule a repair. -HD accelerometer invalid calibration record = ALARM_ID_HD_ACCELEROMETERS_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [27] Title = BLD Sensor Error Message = Call service to schedule a repair. -HD BLD sensor failed zero sequence = ALARM_ID_HD_BLOOD_LEAK_SENSOR_ZERO_SEQUENCE_FAILED +BLD Sensor Error = Call service to schedule a repair. [28] -Title = -Message = - = ALARM_ID_DG_TWO_WIRE_SENSORS_FPGA_FAULT +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [29] Title = Calibration Error Message = Call service to schedule a repair. -HD heparin force sensor invalid calibration record = ALARM_ID_HD_HEPARIN_FORCE_SENSOR_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [30] Title = Machine Error Message = Call service to schedule a repair. -HD Software fault. Software found itself in an unexpected state = ALARM_ID_HD_SOFTWARE_FAULT +Machine Error = Call service to schedule a repair. [31] Title = Blood Pump Error Message = Call service to schedule a repair. -Blood pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running = ALARM_ID_BLOOD_PUMP_MC_CURRENT_CHECK +Blood Pump Error = Call service to schedule a repair. [32] Title = Blood Pump Error Message = Call service to schedule a repair. -Blood pump failed motor off check. Measured speed while commanded off = ALARM_ID_BLOOD_PUMP_OFF_CHECK +Blood Pump Error = Call service to schedule a repair. [33] Title = Blood Pump Error Message = Call service to schedule a repair. -Blood pump failed motor direction check. Measured vs. commanded = ALARM_ID_BLOOD_PUMP_MC_DIRECTION_CHECK +Blood Pump Error = Call service to schedule a repair. [34] Title = Blood Pump Error Message = Call service to schedule a repair. -Blood pump failed rotor speed check. Mismatch with rotor and motor speeds = ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_CHECK +Blood Pump Error = Call service to schedule a repair. [35] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis inlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running = ALARM_ID_DIAL_IN_PUMP_MC_CURRENT_CHECK +Dialysate Pump Error = Call service to schedule a repair. [36] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis inlet pump failed motor off check. Measured speed while commanded off = ALARM_ID_DIAL_IN_PUMP_OFF_CHECK +Dialysate Pump Error = Call service to schedule a repair. [37] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis inlet pump failed motor direction check. Measured vs. commanded = ALARM_ID_DIAL_IN_PUMP_MC_DIRECTION_CHECK +Dialysate Pump Error = Call service to schedule a repair. [38] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis inlet pump failed rotor speed check. Mismatch with rotor and motor speeds = ALARM_ID_DIAL_IN_PUMP_ROTOR_SPEED_CHECK +Dialysate Pump Error = Call service to schedule a repair. [39] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis outlet pump failed motor controller current check. Too high when pump should be off or out of range when pump should be running = ALARM_ID_DIAL_OUT_PUMP_MC_CURRENT_CHECK +Dialysate Pump Error = Call service to schedule a repair. [40] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis outlet pump failed motor off check. Measured speed while commanded off = ALARM_ID_DIAL_OUT_PUMP_OFF_CHECK +Dialysate Pump Error = Call service to schedule a repair. [41] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis outlet pump failed motor direction check. Measured vs. commanded = ALARM_ID_DIAL_OUT_PUMP_MC_DIRECTION_CHECK +Dialysate Pump Error = Call service to schedule a repair. [42] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysis outlet pump failed rotor speed check. Mismatch with rotor and motor speeds = ALARM_ID_DIAL_OUT_PUMP_ROTOR_SPEED_CHECK +Dialysate Pump Error = Call service to schedule a repair. [43] -Title = Machine Error +Title = FPGA Error Message = Call service to schedule a repair. -BLD failed its internal self-test after cartridge prime = ALARM_ID_DG_MAIN_PRIMARY_HEATER_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [44] Title = Machine Error Message = Call service to schedule a repair. -User interface communication timeout = ALARM_ID_UI_COMM_TIMEOUT +Machine Error = Call service to schedule a repair. [45] Title = Machine Error Message = Call service to schedule a repair. -Too many bad CRCs detected on received system messages = ALARM_ID_HD_COMM_TOO_MANY_BAD_CRCS +Machine Error = Call service to schedule a repair. [46] Title = Machine Error Message = Call service to schedule a repair. -System message that required acknowledgment was not acknowledged = ALARM_ID_CAN_MESSAGE_NOT_ACKED +Machine Error = Call service to schedule a repair. [47] Title = Ultrafiltration Error Message = Call service to schedule a repair. -Ultrafiltration rate is too high error during treatment = ALARM_ID_UF_RATE_TOO_HIGH_ERROR +Ultrafiltration Error = Call service to schedule a repair. [48] Title = Ultrafiltration Error Message = Call service to schedule a repair. -Ultrafiltration volume accuracy error during treatment = ALARM_ID_UF_VOLUME_ACCURACY_ERROR +Ultrafiltration Error = Call service to schedule a repair. [49] Title = Machine Error Message = Call service to schedule a repair. -HD FPGA communication down for too long = ALARM_ID_HD_FPGA_COMM_TIMEOUT +Machine Error = Call service to schedule a repair. [50] Title = Machine Error Message = Call service to schedule a repair. -DG FPGA not accepting commanded valve states = ALARM_ID_VALVE_CONTROL_FAILURE +Machine Error = Call service to schedule a repair. [51] Title = Blood Pump Error Message = Call service to schedule a repair. -Blood pump failed motor speed check. Measured vs. commanded = ALARM_ID_BLOOD_PUMP_MOTOR_SPEED_CHECK +Blood Pump Error = Call service to schedule a repair. [52] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysate inlet pump failed motor speed check. Measured vs. commanded = ALARM_ID_DIAL_IN_PUMP_MOTOR_SPEED_CHECK +Dialysate Pump Error = Call service to schedule a repair. [53] Title = Dialysate Pump Error Message = Call service to schedule a repair. -Dialysate outlet pump failed motor speed check. Measured vs. commanded = ALARM_ID_DIAL_OUT_PUMP_MOTOR_SPEED_CHECK +Dialysate Pump Error = Call service to schedule a repair. [54] Title = Machine Error Message = Call service to schedule a repair. -HD critical data integrity check failed = ALARM_ID_HD_CRITICAL_DATA_ERROR +Machine Error = Call service to schedule a repair. [55] -Title = Communication Failure +Title = Machine Error Message = Call service to schedule a repair. -DG critical data integrity check failed = ALARM_ID_DG_CRITICAL_DATA_ERROR +Machine Error = Call service to schedule a repair. [56] Title = Machine Error Message = Call service to schedule a repair. -HD accelerometer error (no readings or FPGA reports error) = ALARM_ID_HD_ACCELEROMETER_FAILURE +Machine Error = Call service to schedule a repair. [57] Title = Machine Error Message = Call service to schedule a repair. -DG accelerometer error (no readings or FPGA reports error) = ALARM_ID_DG_ACCELEROMETER_FAILURE +Machine Error = Call service to schedule a repair. [58] Title = Valve Error Message = Call service to schedule a repair. -HD valve homing failed = ALARM_ID_HD_VALVE_HOMING_FAILED +Valve Error = Call service to schedule a repair. [59] Title = Valve Error Message = Call service to schedule a repair. -HD valve transition time out = ALARM_ID_HD_VALVE_TRANSITION_TIMEOUT +Valve Error = Call service to schedule a repair. [60] Title = Valve Error Message = Call service to schedule a repair. -HD valve not functional = ALARM_ID_HD_VALVE_NOT_FUNCTIONAL +Valve Error = Call service to schedule a repair. [61] Title = Valve Error Message = Call service to schedule a repair. -HD valve current out of range = ALARM_ID_HD_VALVE_CURRENT_OUT_OF_RANGE +Valve Error = Call service to schedule a repair. [62] Title = Valve Error Message = Call service to schedule a repair. -HD valve position out of target = ALARM_ID_HD_VALVE_POSITION_OUT_OF_RANGE +Valve Error = Call service to schedule a repair. [63] -Title = Arterial Pressure Error +Title = FPGA Error Message = Call service to schedule a repair. -FPGA reports failure reading arterial pressure sensor = ALARM_ID_HD_BLOOD_LEAK_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [64] -Title = Venous Pressure Error +Title = Barometric Pressure Error Message = Call service to schedule a repair. -FPGA reports failure reading venous pressure sensor = ALARM_ID_DG_BARO_PRESSURE_OUT_OF_RANGE +Barometric Pressure Error = Call service to schedule a repair. [65] Title = Machine Error Message = Call service to schedule a repair. -HD requests DG command with invalid parameter fault = ALARM_ID_DG_COMMAND_INVALID_PARAMETER_FAULT +Machine Error = Call service to schedule a repair. [66] Title = Machine Error Message = Call service to schedule a repair. -HD blood leak sensor set point set failure = ALARM_ID_HD_BLOOD_LEAK_SENSOR_SET_POINT_SET_FAILURE +Machine Error = Call service to schedule a repair. [67] Title = Self Test Error Message = Call service to schedule a repair. -HD blood pump occlusion self-test failure alarm = ALARM_ID_HD_BP_OCCLUSION_SELF_TEST_FAILURE +Self Test Error = Call service to schedule a repair. [68] Title = Dialysate Recirculation Error Message = Dialysate has recirculated for too long. -HD active reservoir recirculation out of range = ALARM_ID_HD_ACTIVE_RESERVOIR_RECIRCULATION_OUT_OF_RANGE +Dialysate Recirculation Error = Dialysate has recirculated for too long. [69] Title = Self-Test Error Message = Call service to schedule a repair. -HD blood leak sensor invalid calibration record = ALARM_ID_HD_BLOOD_LEAK_INVALID_CAL_RECORD +Self-Test Error = Call service to schedule a repair. [70] Title = Self Test Error Message = Call service to schedule a repair. -HD arterial pressure self-test failure alarm = ALARM_ID_HD_ARTERIAL_PRESSURE_SELF_TEST_FAILURE +Self Test Error = Call service to schedule a repair. [71] Title = Self Test Error Message = Call service to schedule a repair. -HD venous pressure self-test failure alarm = ALARM_ID_HD_VENOUS_PRESSURE_SELF_TEST_FAILURE +Self Test Error = Call service to schedule a repair. [72] Title = No load cell data received Message = Call service to schedule a repair. -HD No load cell data message receive at least once every 2 seconds = ALARM_ID_HD_NEW_LOAD_CELL_DATA_MESSAGE_NOT_RECEIVE +No load cell data received = Call service to schedule a repair. [73] Title = No dialysate temperature data received Message = Call service to schedule a repair. -HD No dialysate temperature data message receive at least once every 2 seconds = ALARM_ID_HD_NEW_DIALYSATE_TEMPERATURE_DATA_MESSAGE_NOT_RECEIVE +No dialysate temperature data received = Call service to schedule a repair. [74] -Title = Self Test Error +Title = UV Reactor Error Message = Call service to schedule a repair. -HD blood leak detector self-test failure alarm = ALARM_ID_DG_INLET_UV_REACTOR_ON_WITH_NO_FLOW +UV Reactor Error = Call service to schedule a repair. [75] Title = Self Test Error Message = Call service to schedule a repair. -HD syringe pump self-test failure alarm = ALARM_ID_HD_SYRINGE_PUMP_SELF_TEST_FAILURE +Self Test Error = Call service to schedule a repair. [76] Title = Machine Error Message = Call service to schedule a repair. -HD monitored voltage is out of range = ALARM_ID_HD_VOLTAGE_OUT_OF_RANGE +Machine Error = Call service to schedule a repair. [77] Title = Machine Error Message = Call service to schedule a repair. -DG monitored voltage is out of range = ALARM_ID_DG_VOLTAGE_OUT_OF_RANGE +Machine Error = Call service to schedule a repair. [78] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump direction (from encoder) error = ALARM_ID_HD_SYRINGE_PUMP_ENCODER_DIRECTION_ERROR +Syringe Pump Error = Call service to schedule a repair. [79] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump direction (from controller) error = ALARM_ID_HD_SYRINGE_PUMP_CONTROLLER_DIRECTION_ERROR +Syringe Pump Error = Call service to schedule a repair. [80] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump fault reported by FPGA = ALARM_ID_HD_SYRINGE_PUMP_FAULT +Syringe Pump Error = Call service to schedule a repair. [81] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump over travel error = ALARM_ID_HD_SYRINGE_PUMP_OVER_TRAVEL_ERROR +Syringe Pump Error = Call service to schedule a repair. [82] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump DAC write failure = ALARM_ID_HD_SYRINGE_PUMP_DAC_WRITE_ERROR +Syringe Pump Error = Call service to schedule a repair. [83] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump is running while the BP is off = ALARM_ID_HD_SYRINGE_PUMP_RUNNING_WHILE_BP_OFF_ERROR +Syringe Pump Error = Call service to schedule a repair. [84] -Title = Venous Pressure Error +Title = RTC Error Message = Call service to schedule a repair. -HD venous pressure sensor not being read = ALARM_ID_DG_SET_RTC_YEAR_INVALID +RTC Error = Call service to schedule a repair. [85] Title = Pump Track Lever Not Fully Latched Message = Open front door and pull pump track lever all the way down. -HD pump track lever is not fully latched = ALARM_ID_PUMP_TRACK_LATCH_OPENED +Pump Track Lever Not Fully Latched = Open front door and pull pump track lever all the way down. [86] -Title = Occlusion Sensor Error +Title = RTC Error Message = Call service to schedule a repair. -HD BP occlusion sensor not being read = ALARM_ID_HD_SET_RTC_YEAR_INVALID +RTC Error = Call service to schedule a repair. [87] Title = Calibration Error Message = Call service to schedule a repair. - = ALARM_ID_DG_HEATING_INVALID_CAL_RECORD +Calibration Error = Call service to schedule a repair. [88] -Title = Self Test Error +Title = Concentrate Pumps Error Message = Call service to schedule a repair. - = ALARM_ID_DG_CONC_PUMP_HALL_SENSOR_OUT_OF_RANGE +Concentrate Pumps Error = Call service to schedule a repair. [89] -Title = Occlusion Sensor Error +Title = UV Reactor Error Message = Call service to schedule a repair. -HD BP occlusion sensor error = ALARM_ID_DG_OUTLET_UV_REACTOR_ON_WITH_NO_FLOW +UV Reactor Error = Call service to schedule a repair. [90] -Title = -Message = - = ALARM_ID_DG_LOAD_CELL_A1_B1_FPGA_FAULT +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [91] Title = Dialysate Measurement Error Message = Call service to schedule a repair. -HD No dialysate flow data receive in the last 3 seconds = ALARM_ID_HD_DIALYSATE_FLOW_DATA_NOT_RECEIVE +Dialysate Measurement Error = Call service to schedule a repair. [92] -Title = Self Test Error +Title = Conductivity Sensors Error Message = Call service to schedule a repair. - = ALARM_ID_DG_CONDUCTIVITY_SENSOR_BAD_STATUS +Conductivity Sensors Error = Call service to schedule a repair. [93] -Title = RO Rejection Ratio Error +Title = Conductivity Sensors Error Message = Call service to schedule a repair. - = ALARM_ID_DG_OUTLET_PRIMARY_CONDUCTIVITY_OUT_OF_RANGE +Conductivity Sensors Error = Call service to schedule a repair. [94] -Title = -Message = - = ALARM_ID_DG_PRESSURE_OUT_OF_RANGE +Title = Pressure Sensors Error +Message = Call service to schedule a repair. +Pressure Sensors Error = Call service to schedule a repair. [95] -Title = Self Test Error +Title = Watchdog Error Message = Call service to schedule a repair. - = ALARM_ID_DG_WATCHDOG_EXPIRED +Watchdog Error = Call service to schedule a repair. [96] Title = Inlet Water Temperature High Message = Inlet water temperature is high.\n- Check the incoming water for issues with cold water supply. - = ALARM_ID_INLET_WATER_TEMPERATURE_IN_HIGH_RANGE +Inlet Water Temperature High = Inlet water temperature is high.\n- Check the incoming water for issues with cold water supply. [97] Title = Dialysate Fill Error -Message = Dialysate Conductivity out of range. - = ALARM_ID_FILL_CONDUCTIVITY_OUT_OF_RANGE +Message = Call service to schedule a repair. +Dialysate Fill Error = Call service to schedule a repair. [98] Title = Battery Error Message = Call service to schedule a repair. -HD battery communication fault = ALARM_ID_HD_BATTERY_COMM_FAULT +Battery Error = Call service to schedule a repair. [99] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump stall alarm = ALARM_ID_HD_SYRINGE_PUMP_STALL +Syringe Pump Error = Call service to schedule a repair. [100] Title = Self Test Error Message = Call service to schedule a repair. -HD no cartridge self-test timeout = ALARM_ID_HD_NO_CART_SELF_TEST_TIMEOUT +Self Test Error = Call service to schedule a repair. [101] Title = Self Test Error Message = Call service to schedule a repair. -HD dry self-test timeout = ALARM_ID_HD_DRY_SELF_TEST_TIMEOUT +Self Test Error = Call service to schedule a repair. [102] Title = Self-Test Error Message = Call service to schedule a repair. -DG heat disinfection temperature gradient out of range = ALARM_ID_DG_HEAT_DISINFECT_TEMP_GRAD_OUT_OF_RANAGE +Self-Test Error = Call service to schedule a repair. [103] Title = Machine Error Message = Call service to schedule a repair. -HD non-volatile calibration group invalid record CRC = ALARM_ID_HD_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID +Machine Error = Call service to schedule a repair. [104] Title = Air Trap Error Message = Call service to schedule a repair. -Air trap level sensors reporting illegal combination of air/fluid = ALARM_ID_AIR_TRAP_ILLEGAL_LEVELS +Air Trap Error = Call service to schedule a repair. [105] Title = Machine Error Message = Call service to schedule a repair. -DG non-volatile system group invalid record CRC = ALARM_ID_DG_NVDATAMGMT_CAL_GROUP_RECORD_CRC_INVALID +Machine Error = Call service to schedule a repair. [106] -Title = -Message = -This alarm ID is available for use = ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD +Title = Calibration Error +Message = Call service to schedule a repair. +Calibration Error = Call service to schedule a repair. [107] Title = Machine Error Message = Call service to schedule a repair. -HD reports DG restarted fault = ALARM_ID_DG_RESTARTED_FAULT +Machine Error = Call service to schedule a repair. [108] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump ADC error = ALARM_ID_HD_SYRINGE_PUMP_FPGA_ADC_FAULT +Syringe Pump Error = Call service to schedule a repair. [109] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump volume check error = ALARM_ID_HD_SYRINGE_PUMP_VOLUME_ERROR +Syringe Pump Error = Call service to schedule a repair. [110] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump speed check error = ALARM_ID_HD_SYRINGE_PUMP_SPEED_ERROR +Syringe Pump Error = Call service to schedule a repair. [111] Title = Syringe Pump Error Message = Call service to schedule a repair. -HD syringe pump not stopped in off state error = ALARM_ID_HD_SYRINGE_PUMP_NOT_STOPPED_ERROR +Syringe Pump Error = Call service to schedule a repair. [112] -Title = Blood Leak Detector Error +Title = FPGA Error Message = Call service to schedule a repair. -HD blood leak detector fault = ALARM_ID_DG_LOAD_CELL_A2_B2_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [113] -Title = Blood Leak Detector Error +Title = Drain Pump Error Message = Call service to schedule a repair. -HD blood leak zero command fault = ALARM_ID_DG_DRAIN_PUMP_CURRENT_OUT_OF_RANGE +Drain Pump Error = Call service to schedule a repair. [114] Title = Air Bubble Detector Error Message = Call service to schedule a repair. -HD venous air bubble detector self-test failure = ALARM_ID_HD_VENOUS_BUBBLE_SELF_TEST_FAILURE +Air Bubble Detector Error = Call service to schedule a repair. [115] Title = Dialysate Temperature Error Message = Call service to schedule a repair. -DG temperature sensor out of range = ALARM_ID_DG_TEMPERATURE_SENSOR_OUT_OF_RANGE +Dialysate Temperature Error = Call service to schedule a repair. [116] Title = Dialysate Temperature Error Message = Call service to schedule a repair. -DG temperature sensor ADC out of range = ALARM_ID_DG_TEMPERATURE_SENSORS_ADC_FAULT +Dialysate Temperature Error = Call service to schedule a repair. [117] -Title = Dialysate Temperature Error +Title = Machine Error Message = Call service to schedule a repair. -DG primary heater internal temperature out of range = ALARM_ID_DG_CAN_MESSAGE_NOT_ACKED +Machine Error = Call service to schedule a repair. [118] -Title = Dialysate Temperature Error +Title = RTC Error Message = Call service to schedule a repair. -DG primary heater cold junction temperature out of range = ALARM_ID_DG_RTC_CONFIG_ERROR +RTC Error = Call service to schedule a repair. [119] -Title = Dialysate Temperature Error +Title = RTC Error Message = Call service to schedule a repair. -DG trimmer heater internal temperature out of range = ALARM_ID_DG_RTC_BATTERY_LOW +RTC Error = Call service to schedule a repair. [120] -Title = Dialysate Temperature Error +Title = Pre Treament Error Message = Call service to schedule a repair. -DG trimmer heater cold junction temperature out of range = ALARM_ID_PRE_TREATMENT_WET_PRIME_TEST_FAILURE +Pre Treament Error = Call service to schedule a repair. [121] -Title = Dialysate Temperature Error +Title = Heaters Error Message = Call service to schedule a repair. -DG main primary heater voltage out of range = ALARM_ID_DG_MAIN_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE +Heaters Error = Call service to schedule a repair. [122] -Title = Dialysate Temperature Error +Title = Heaters Error Message = Call service to schedule a repair. -DG small primary heater voltage out of range = ALARM_ID_DG_SMALL_PRIMARY_HEATER_VOLTAGE_OUT_OF_RANGE +Heaters Error = Call service to schedule a repair. [123] -Title = Dialysate Temperature Error +Title = Heaters Error Message = Call service to schedule a repair. -DG trimmer heater voltage out of range = ALARM_ID_DG_TRIMMER_HEATER_VOLTAGE_OUT_OF_RANGE +Heaters Error = Call service to schedule a repair. [124] Title = Blood Stagnant Message = Blood has sat too long in the blood tubing set. No rinseback allowed. -HD end of treatment alarm (high priority - user still not acting to end treatment) = ALARM_ID_END_OF_TREATMENT_HIGH +Blood Stagnant = Blood has sat too long in the blood tubing set. No rinseback allowed. [125] Title = Blood Stagnant Message = Blood has sat too long in the blood tubing set. Treatment must end. -Blood sitting too long after treatment stopped by user alarm (> 5 min) = ALARM_ID_TREATMENT_STOPPED_NO_RINSEBACK +Blood Stagnant = Blood has sat too long in the blood tubing set. Treatment must end. [126] Title = Dialyzer Blood Leak Detected Message = A blood leak has been detected. Treatment must end. -HD blood leak detected alarm = ALARM_ID_HD_BLOOD_LEAK_DETECTED +Dialyzer Blood Leak Detected = A blood leak has been detected. Treatment must end. [127] Title = Venous Pressure Low Message = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured properly. -Venous pressure too low during treatment = ALARM_ID_VENOUS_PRESSURE_LOW +Venous Pressure Low = Venous pressure is low.\n- Check to be sure your blue venous connections are secure.\n- Check that your venous access is connected and secured properly. [128] Title = Venous Air Bubble Detected Message = Venous air bubble detected.\n- Check for air bubbles in the venous line.\n- If bubbles are present, tap the tubing to move the bubbles up to the air trap.\n- If bubbles cannot be removed, treatment must end. -HD venous air bubble detected alarm = ALARM_ID_HD_VENOUS_BUBBLE_DETECTED +Venous Air Bubble Detected = Venous air bubble detected.\n- Check for air bubbles in the venous line.\n- If bubbles are present, tap the tubing to move the bubbles up to the air trap.\n- If bubbles cannot be removed, treatment must end. [129] Title = Venous Air Bubble Detected During Rinseback Message = Venous air bubble detected.\n- Check for air bubbles in the venous line.\n- If bubbles are present, tap the tubing to move the bubbles up to the air trap.\n- If bubbles cannot be removed, treatment must end. -HD venous air bubble detected rinseback alarm = ALARM_ID_HD_VENOUS_BUBBLE_DETECTED_RINSEBACK +Venous Air Bubble Detected During Rinseback = Venous air bubble detected.\n- Check for air bubbles in the venous line.\n- If bubbles are present, tap the tubing to move the bubbles up to the air trap.\n- If bubbles cannot be removed, treatment must end. [130] Title = Venous Pressure High -Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present." -Venous pressure too high during treatment = ALARM_ID_VENOUS_PRESSURE_HIGH +Message = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. +Venous Pressure High = Venous pressure is high.\n- Check for access or patient discomfort.\n- Resolve access issue if applicable.\n- Check that there are no kinks or clots in the tubing.\n- Unkink the lines if applicable.\n- End treatment and do not return blood if clots are present. [131] Title = Arterial Pressure Low Message = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. -Arterial pressure too low during treatment = ALARM_ID_ARTERIAL_PRESSURE_LOW +Arterial Pressure Low = Arterial pressure is low.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure the red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [132] Title = Arterial Pressure High Message = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. -Arterial pressure too high during treatment = ALARM_ID_ARTERIAL_PRESSURE_HIGH +Arterial Pressure High = Arterial pressure is high.\n- Check that the access is properly connected and secured.\n- Resolve access issue if applicable.\n- Check to be sure your red arterial connections are secure.\n- Check that there are no kinks or clots in the tubing.\n- Unkink lines if applicable.\n- End treatment and do not return blood if clots are present. [133] Title = Fluid Leak in DG Message = Fluid leak detected in Dialysate Generation Delivery Device. -DG fluid leak detected alarm = ALARM_ID_DG_FLUID_LEAK_DETECTED +Fluid Leak in DG = Fluid leak detected in Dialysate Generation Delivery Device. [134] Title = Fluid Leak in HD Message = Fluid leak detected in Hemodialysis Delivery device.\n- Identify and correct the cause of the spill.\n- Clean up the spill with a clean paper towel.\n- Resume treatment. -HD fluid leak detected alarm = ALARM_ID_HD_FLUID_LEAK_DETECTED +Fluid Leak in HD = Fluid leak detected in Hemodialysis Delivery device.\n- Identify and correct the cause of the spill.\n- Clean up the spill with a clean paper towel.\n- Resume treatment. [135] Title = Excessive Acceleration Detected in the HD Message = Excessive movement detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. -HD experienced large acceleration = ALARM_ID_HD_SHOCK +Excessive Acceleration Detected in the HD = Excessive movement detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing. [136] Title = Excessive Acceleration Detected in the DG Message = Excessive movement detected with Dialysis Generation device.\n- Confirm the device is stationary and level before continuing. -DG experienced large acceleration = ALARM_ID_DG_SHOCK +Excessive Acceleration Detected in the DG = Excessive movement detected with Dialysis Generation device.\n- Confirm the device is stationary and level before continuing. [137] Title = Excessive Tilt Detected in the Hemodialysis Device Message = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing -HD tilt exceeds maximum = ALARM_ID_HD_EXCESSIVE_TILT +Excessive Tilt Detected in the Hemodialysis Device = Excessive tilt detected with Hemodialysis Delivery device.\n- Confirm the device is stationary and level before continuing [138] Title = Excessive Tilt Detected in the Dialysate Generation Device Message = Excessive tilt detected with Dialysis Generation device.\n-Confirm the device is stationary and level before continuing. -DG tilt exceeds maximum = ALARM_ID_DG_EXCESSIVE_TILT +Excessive Tilt Detected in the Dialysate Generation Device = Excessive tilt detected with Dialysis Generation device.\n-Confirm the device is stationary and level before continuing. [139] -Title = Power Lost +Title = HD Power Lost Message = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until power returns. -HD AC power lost alarm = ALARM_ID_HD_AC_POWER_LOST +HD Power Lost = Power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until power returns. [140] Title = Communication Failure with Dialysate Generator Message = Communication failure with Dialysis Generator.\n- Ensure all cords are securely connected. -Dialysate generator communication timeout = ALARM_ID_DG_COMM_TIMEOUT +Communication Failure with Dialysate Generator = Communication failure with Dialysis Generator.\n- Ensure all cords are securely connected. [141] Title = Air Trap Unable to Fill Message = Failure to fill the Air Trap.\n- Check the air trap vent line and the transducer protector.\n- If the transducer protector is wet, replace it. -Air trap fill timeout during treatment = ALARM_ID_AIR_TRAP_FILL_DURING_TREATMENT +Air Trap Unable to Fill = Failure to fill the Air Trap.\n- Check the air trap vent line and the transducer protector.\n- If the transducer protector is wet, replace it. [142] Title = Blood Pump Occlusion Detected Message = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the venous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. -Blood pump occlusion detected = ALARM_ID_OCCLUSION_BLOOD_PUMP +Blood Pump Occlusion Detected = Occlusion detected after the blood pump.\n- Check for clots or kinks around the dialyzer and in the venous line.\n- Resolve access issue if applicable.\n- End treatment and do not return blood if clots are present.\n- Check for access or patient discomfort.\n- Unkink the lines if applicable. [143] Title = Dialysate Temperature Error Message = Call service to schedule a repair. -DG dialysate temperature sensors drift timeout. = ALARM_ID_DG_DIALYSATE_TEMPERATURE_SENSORS_OUT_OF_RANGE +Dialysate Temperature Error = Call service to schedule a repair. [144] -Title = -Message = -This alarm ID is available for use = ALARM_ID_DG_RO_FLOW_SENSOR_INVALID_CAL_RECORD +Title = Cleaning Mode Inlet Water Conductivity Too High +Message = Inlet water conductivity too high.\n- Check the incoming water for issues with cold water supply. +Cleaning Mode Inlet Water Conductivity Too High = Inlet water conductivity too high.\n- Check the incoming water for issues with cold water supply. [145] Title = Acid Concentrate Conductivity Out of Range Message = Dialysate conductivity is out of range.\n- Check correct acid concentrate is connected. -DG Concentrate conductivity after adding acid out of range alarm = ALARM_ID_ACID_CONDUCTIVITY_OUT_OF_RANGE +Acid Concentrate Conductivity Out of Range = Dialysate conductivity is out of range.\n- Check correct acid concentrate is connected. [146] Title = Self-Test Error Message = Call service to schedule a repair. -DG RTC vs. Timer accuracy error = ALARM_ID_DG_RTC_OR_TIMER_ACCURACY_FAILURE +Self-Test Error = Call service to schedule a repair. [147] Title = Dialysate Conductivity Error Message = Call service to schedule a repair. -DG dialysate generation conductivity fault alarm = ALARM_ID_CREATING_DIALYSATE_PLEASE_WAIT +Dialysate Conductivity Error = Call service to schedule a repair. [148] -Title = Inlet Water Temperature High -Message = Inlet water temperature is high.\n- Check the incoming water for issues with cold water supply. -Inlet water temperature too high = ALARM_ID_INLET_WATER_HIGH_TEMPERATURE +Title = Cleaning Mode Inlet Water Temperature Too High +Message = Cleaning Mode Inlet water temperature too high.\n- Check the incoming water for issues with cold water supply. +Cleaning Mode Inlet Water Temperature Too High = Cleaning Mode Inlet water temperature too high.\n- Check the incoming water for issues with cold water supply. [149] -Title = Inlet Water Temperature Low -Message = Inlet water temperature is low.\n- Check the incoming water for issues with hot water supply. -Inlet water temperature too low = ALARM_ID_HD_WATCHDOG_EXPIRED +Title = Watchdog Error +Message = Call service to schedule a repair. +Watchdog Error = Call service to schedule a repair. [150] Title = Inlet Water Conductivity High Message = Inlet water conductivity is high.\n- Check the incoming water for issues. -Inlet water conductivity too high = ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_HIGH_RANGE +Inlet Water Conductivity High = Inlet water conductivity is high.\n- Check the incoming water for issues. [151] Title = Inlet Water Conductivity Low Message = Inlet water conductivity is low.\n- Check the incoming water for issues. -Inlet water conductivity too low = ALARM_ID_INLET_WATER_CONDUCTIVITY_IN_LOW_RANGE +Inlet Water Conductivity Low = Inlet water conductivity is low.\n- Check the incoming water for issues. [152] Title = Inlet Water Pressure Low Message = Inlet water pressure is low.\n- Check the incoming water supply. -Inlet water pressure too low = ALARM_ID_INLET_WATER_PRESSURE_IN_LOW_RANGE +Inlet Water Pressure Low = Inlet water pressure is low.\n- Check the incoming water supply. [153] Title = Prime Complete Message = Prime of the bloodlines is complete. -HD prime completed high priority alarm = ALARM_ID_PRIME_COMPLETED_HIGH +Prime Complete = Prime of the bloodlines is complete. [154] -Title = Communication Failure -Message = Communication has been lost. Ensure all cords are securely connected. -EEPROM operations (read write erase) failure write erase) failure = ALARM_ID_NVDATA_EEPROM_OPS_FAILURE +Title = Cleaning Mode Inlet Water Conductivity Low +Message = Cleaning Mode Inlet water conductivity is low.\n- Check the incoming water. +Cleaning Mode Inlet Water Conductivity Low = Cleaning Mode Inlet water conductivity is low.\n- Check the incoming water. [155] Title = No reservoir data received Message = Call service to schedule a repair. -HD No reservoir data message receive at least once every 2 seconds = ALARM_ID_HD_NEW_RESERVOIRS_DATA_MESSAGE_NOT_RECEIVE +No reservoir data received = Call service to schedule a repair. [156] Title = No DG op mode received Message = Call service to schedule a repair. -HD No DG op mode message receive at least once every 2 seconds = ALARM_ID_HD_NEW_DG_OPERATION_MODE_MESSAGE_NOT_RECEIVE +No DG op mode received = Call service to schedule a repair. [157] -Title = -Message = -This alarm ID is available for use = ALARM_ID_DG_CHEM_DISINFECT_PRIME_ACID_LINE_TIME_OUT +Title = Chemical Disinfection Error +Message = Ensure the acid is connected to the bicarb line +Chemical Disinfection Error = Ensure the acid is connected to the bicarb line [158] Title = Inlet water temperature low -Message = -Inlet water temperature low = ALARM_ID_INLET_WATER_TEMPERATURE_IN_LOW_RANGE +Message = Inlet water temperature is low.\n- Check the incoming water supply. +Inlet water temperature low = Inlet water temperature is low.\n- Check the incoming water supply. [159] -Title = Dialysate Pump Failure -Message = Dialysate pump failure.\n- Check there are no kinks in the dialysate tubing. -Dialysate inlet pump failed flow vs. motor speed check. = ALARM_ID_DIAL_IN_PUMP_FLOW_VS_MOTOR_SPEED_CHECK +Title = Chemical Disinfect Flush +Message = Remove Acid Bottle and Close Concentrate Cap. +Chemical Disinfect Flush = Remove Acid Bottle and Close Concentrate Cap. [160] Title = Blood Pump Speed High Message = Blood pump failure.\n- Check the pump rotor for secure connection. -Blood pump rotor speed too high = ALARM_ID_BLOOD_PUMP_ROTOR_SPEED_TOO_HIGH +Blood Pump Speed High = Blood pump failure.\n- Check the pump rotor for secure connection. [161] Title = Machine Error Message = Call service to schedule a repair. -Too many bad CRCs detected on received system messages = ALARM_ID_DG_COMM_TOO_MANY_BAD_CRCS +Machine Error = Call service to schedule a repair. [162] -Title = -Message = - = ALARM_ID_DG_FPGA_CLOCK_SPEED_CHECK_FAILURE +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [163] Title = Ultrafiltration Measurement Error Message = Ultrafiltration error. -HD sees primary load cell for reservoir 1 change too much too fast = ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_1_ALARM +Ultrafiltration Measurement Error = Ultrafiltration error. [164] Title = Ultrafiltration Measurement Error Message = Ultrafiltration error. -HD sees primary load cell for reservoir 2 change too much too fast = ALARM_ID_HD_LOAD_CELL_ACCELERATION_RES_2_ALARM +Ultrafiltration Measurement Error = Ultrafiltration error. [165] Title = Rinseback Timeout Message = Maximum Rinseback time exceeded. Rinseback must end. -HD in treatment rinseback operation timeout = ALARM_ID_TREATMENT_RINSEBACK_TIMEOUT_ALARM +Rinseback Timeout = Maximum Rinseback time exceeded. Rinseback must end. [166] Title = Recirculation Timeout Message = Maximum Recirculation time exceeded. Treatment must end. -HD in treatment re-circ sub-mode for too long = ALARM_ID_TREATMENT_RECIRC_TIMEOUT_ALARM +Recirculation Timeout = Maximum Recirculation time exceeded. Treatment must end. [167] Title = Door Open Message = Hemodialysis Delivery device door open.\n - Close open HD door\n - Press RESUME button -HD cartridge door opened alarm = ALARM_ID_CARTRIDGE_DOOR_OPENED +Door Open = Hemodialysis Delivery device door open.\n - Close open HD door\n - Press RESUME button [168] Title = Dialysate Recirculation Error Message = Dialysate recirculation error. -HD active reservoir depletion timeout. = ALARM_ID_HD_ACTIVE_RESERVOIR_DEPLETION_TIME_OUT +Dialysate Recirculation Error = Dialysate recirculation error. [169] Title = Dialysate Generation Error Message = Call service to schedule a repair. -Dialysate flow rate out of range = ALARM_ID_DG_DIALYSATE_FLOW_RATE_OUT_OF_MAX_RANGE +Dialysate Generation Error = Call service to schedule a repair. [170] Title = Syringe Empty Message = Heparin syringe empty.\n- Option to resume treatment without heparin. -HD syringe empty alarm = ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_EMPTY +Syringe Empty = Heparin syringe empty.\n- Option to resume treatment without heparin. [171] Title = Heparin Occlusion Message = Heparin pump occlusion detected. Check heparin line is unclamped and there are no kinks in the heparin line. -HD syringe pump occlusion alarm = ALARM_ID_HD_SYRINGE_PUMP_OCCLUSION +Heparin Occlusion = Heparin pump occlusion detected. Check heparin line is unclamped and there are no kinks in the heparin line. [172] Title = Not Enough Heparin Message = Not enough heparin is available for treatment. Check if syringe pump is touching syringe plunger. If syringe pump is touching syringe plunger, end treatment. If syringe pump is not touching syringe plunger, reload syringe with enough heparin. -HD syringe pump not enough Heparin alarm = ALARM_ID_HD_SYRINGE_PUMP_NOT_ENOUGH_HEPARIN_ALARM +Not Enough Heparin = Not enough heparin is available for treatment. Check if syringe pump is touching syringe plunger. If syringe pump is touching syringe plunger, end treatment. If syringe pump is not touching syringe plunger, reload syringe with enough heparin. [173] -Title = Self-Test Error +Title = RTC Error Message = Call service to schedule a repair. -HD RTC configuration error = ALARM_ID_HD_RTC_CONFIG_ERROR +RTC Error = Call service to schedule a repair. [174] Title = Self-Test Error Message = Call service to schedule a repair. -HD RTC vs. Timer accuracy error = ALARM_ID_HD_RTC_OR_TIMER_ACCURACY_FAILURE +Self-Test Error = Call service to schedule a repair. [175] Title = Blood Pump Error Message = Blood pump failure. \n- Check the pump rotor for secure connection. -HD pump direction status error = ALARM_ID_HD_PUMP_DIRECTION_STATUS_ERROR +Blood Pump Error = Blood pump failure. \n- Check the pump rotor for secure connection. [176] -Title = -Message = -// TODO remove this alarm HD active reservoir is full during treatment and DG not ready to switch reservoirs. = ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_HIGH_SAFETY_RANGE +Title = Dialysate Temperature Error +Message = Call service to schedule a repair. +Dialysate Temperature Error = Call service to schedule a repair. [177] Title = Machine Error Message = Call service to schedule a repair. -DG software fault. Software found itself in an unexpected state = ALARM_ID_DG_SOFTWARE_FAULT +Machine Error = Call service to schedule a repair. [178] -Title = -Message = -HD communication timeout = ALARM_ID_HD_COMM_TIMEOUT +Title = Communication Failure +Message = Call service to schedule a repair. +Communication Failure = Call service to schedule a repair. [179] Title = Communication Failure Message = Call service to schedule a repair. -DG FPGA communication down for too long = ALARM_ID_DG_FPGA_COMM_TIMEOUT +Communication Failure = Call service to schedule a repair. [180] -Title = Ultrafiltration Measurement Error +Title = RO Flow Error Message = Call service to schedule a repair. -DG load cell ADC error = ALARM_ID_DG_RO_FLOW_RATE_OUT_OF_MAX_RANGE +RO Flow Error = Call service to schedule a repair. [181] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. -DG load cells weight out of range for tare = ALARM_ID_DG_LOAD_CELLS_TARE_WEIGHT_OUT_OF_RANGE +Ultrafiltration Measurement Error = Call service to schedule a repair. [182] Title = Ultrafiltration Measurement Error Message = Call service to schedule a repair. -DG load cells invalid calibration = ALARM_ID_DG_LOAD_CELLS_INVALID_CAL_RECORD +Ultrafiltration Measurement Error = Call service to schedule a repair. [183] -Title = Ultrafiltration Measurement Error +Title = Heat Disinfection Error Message = Call service to schedule a repair. -DG invalid load cell value = ALARM_ID_DG_INVALID_LOAD_CELL_VALUE +Heat Disinfection Error = Call service to schedule a repair. [184] -Title = Machine Error +Title = UV Reactor Error Message = Call service to schedule a repair. -Inlet or Outlet UV reactor not healthy = ALARM_ID_UV_REACTOR_NOT_HEALTHY +UV Reactor Error = Call service to schedule a repair. [185] Title = Fan Error Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. -DG fan RPM out of range = ALARM_ID_DG_FAN_RPM_OUT_OF_RANGE +Fan Error = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. [186] Title = Dialysate Concentrate Pump Error Message = Call service to schedule a repair. -DG concentrate pump fault = ALARM_ID_DG_CONCENTRATE_PUMP_FAULT +Dialysate Concentrate Pump Error = Call service to schedule a repair. [187] Title = Dialysate Concentrate Pump Error Message = Call service to schedule a repair. -DG concentrate pump CP1 speed control error = ALARM_ID_CP1_SPEED_CONTROL_ERROR +Dialysate Concentrate Pump Error = Call service to schedule a repair. [188] Title = Dialysate Concentrate Pump Error Message = Call service to schedule a repair. -DG concentrate pump CP2 speed control error = ALARM_ID_CP2_SPEED_CONTROL_ERROR +Dialysate Concentrate Pump Error = Call service to schedule a repair. [189] -Title = Dialysate Drain Pump Error +Title = Drain Pump Error Message = Call service to schedule a repair. -DG drain pump RPM out of range = ALARM_ID_DRAIN_PUMP_RPM_OUT_OF_RANGE +Drain Pump Error = Call service to schedule a repair. [190] -Title = Dialysate Drain Pump Error +Title = Drain Pump Error Message = Call service to schedule a repair. -DG drain pump off fault = ALARM_ID_DRAIN_PUMP_OFF_FAULT +Drain Pump Error = Call service to schedule a repair. [191] -Title = Dialysate Flowrate Error +Title = RO Flow Error Message = Call service to schedule a repair. -DG flow rate out of upper range = ALARM_ID_FLOW_RATE_OUT_OF_UPPER_RANGE +RO Flow Error = Call service to schedule a repair. [192] -Title = Dialysate Flowrate Error +Title = RO Flow Error Message = Call service to schedule a repair. -DG flow rate out of lower range = ALARM_ID_FLOW_RATE_OUT_OF_LOWER_RANGE +RO Flow Error = Call service to schedule a repair. [193] -Title = -Message = - = ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE +Title = Dialysate Temperature Error +Message = Call service to schedule a repair. +Dialysate Temperature Error = Call service to schedule a repair. [194] Title = RO Pump Error Message = Call service to schedule a repair. -DG RO pump duty cycle out of range = ALARM_ID_RO_PUMP_DUTY_CYCLE_OUT_OF_RANGE +RO Pump Error = Call service to schedule a repair. [195] Title = RO Pump Error Message = Call service to schedule a repair. -DG RO pump pressure out of range = ALARM_ID_RO_PUMP_PRESSURE_OUT_OF_RANGE +RO Pump Error = Call service to schedule a repair. [196] -Title = Dialysate Temperature Error +Title = FPGA Error Message = Call service to schedule a repair. -DG temperature sensor error flag fault. = ALARM_ID_DG_CPI_CPO_SENSORS_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [197] -Title = Dialysate Temperature Error +Title = FPGA Error Message = Call service to schedule a repair. -DG temperature sensors ADC fault = ALARM_ID_DG_CD1_CD2_SENSORS_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [198] -Title = Dialysate Heater Error +Title = Heaters Error Message = Call service to schedule a repair. -DG heaters cold junction temperature out of range = ALARM_ID_RO_FLOW_TOO_LOW_WHILE_PRIMARY_HEATER_IS_ON +Heaters Error = Call service to schedule a repair. [199] -Title = -Message = - = ALARM_ID_DIALYSATE_FLOW_TOO_LOW_WHILE_TRIMMER_HEATER_IS_ON +Title = Heaters Error +Message = Call service to schedule a repair. +Heaters Error = Call service to schedule a repair. [200] -Title = Dialysate Temperature Error +Title = Thermistors Error Message = Call service to schedule a repair. -DG thermistors/sensors temperature out of range = ALARM_ID_DG_THERMISTORS_TEMPERATURE_OUT_OF_RANGE +Thermistors Error = Call service to schedule a repair. [201] -Title = Inlet Water Error -Message = Inlet water quality is bad. -Inlet water quality is bad (temperature conductivity pressure) conductivity pressure) = ALARM_ID_PRE_TREATMENT_WET_FLOW_TEST_FAILURE +Title = Pre-Treatment Error +Message = Call service to schedule a repair. +Pre-Treatment Error = Call service to schedule a repair. [202] -Title = Inlet Water Error -Message = Inlet water quality is bad. -Inlet water pressure fault = ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_NORMAL_TEST_FAILURE +Title = Pre-Treatment Error +Message = Call service to schedule a repair. +Pre-Treatment Error = Call service to schedule a repair. [203] -Title = Dialysate Pressure Error +Title = Pressure Sensors Error Message = Call service to schedule a repair. -Pressure sensor fault = ALARM_ID_PRESSURE_SENSOR_FAULT +Pressure Sensors Error = Call service to schedule a repair. [204] Title = RO Rejection Error Message = Call service to schedule a repair. -RO rejection ratio out of range = ALARM_ID_RO_REJECTION_RATIO_OUT_OF_RANGE +RO Rejection Error = Call service to schedule a repair. [205] -Title = Dialysate Conductivity Error +Title = Conductivity Sensors Error Message = Call service to schedule a repair. -Conductivity sensor fault = ALARM_ID_CONDUCTIVITY_SENSOR_FAULT +Conductivity Sensors Error = Call service to schedule a repair. [206] Title = Dialysate Production Error Message = Call service to schedule a repair. -DG dialysate fill runs out of time = ALARM_ID_DG_DIALYSATE_FILL_OUT_OF_TIME +Dialysate Production Error = Call service to schedule a repair. [207] Title = Dialysate Flowrate Error Message = Call service to schedule a repair. -DG flow meter check failure alarm = ALARM_ID_DG_FLOW_METER_CHECK_FAILURE +Dialysate Flowrate Error = Call service to schedule a repair. [208] -Title = Dialysate Temperature Error +Title = Venous Occlusion Error Message = Call service to schedule a repair. -DG primary heater on with no flow time out = ALARM_ID_HD_VENOUS_LINE_OCCLUSION +Venous Occlusion Error = Call service to schedule a repair. [209] -Title = Dialysate Production Error +Title = Flush Mode Error Message = Call service to schedule a repair. -DG drain circulation line timeout = ALARM_ID_DG_DRAIN_CIRCULATION_LINE_TIMEOUT +Flush Mode Error = Call service to schedule a repair. [210] Title = Battery Error Message = Battery pack error. -HD battery pack detected an error = ALARM_ID_HD_BATTERY_PACK_ERROR_DETECTED +Battery Error = Battery pack error. [211] Title = Blood Stagnant Message = Maximum time for blood stoppage has been exceeded. -Blood sitting too long warning (> 4 min) = ALARM_ID_BLOOD_SITTING_WARNING +Blood Stagnant = Maximum time for blood stoppage has been exceeded. [212] Title = Blood Stagnant Message = Maximum time for blood stoppage has been exceeded. -Blood sitting too long warning (> 4 min) = ALARM_ID_END_OF_TREATMENT_ALARM +Blood Stagnant = Maximum time for blood stoppage has been exceeded. [213] Title = Prime Complete Message = Prime of the bloodlines is complete. -HD prime completed medium priority alarm = ALARM_ID_PRIME_COMPLETED_MEDIUM +Prime Complete = Prime of the bloodlines is complete. [214] Title = Saline Bolus Error Message = Inability to deliver saline bolus.\n- Check for kinks or clamped lines in the saline line. -HD Saline bolus volume check failure = ALARM_ID_SALINE_BOLUS_VOLUME_CHECK_FAILURE +Saline Bolus Error = Inability to deliver saline bolus.\n- Check for kinks or clamped lines in the saline line. [215] Title = Rinseback Error Message = Unable to complete Rinseback.\n- Check for kinks or clamped blood lines. -HD rinseback volume check failure = ALARM_ID_RINSEBACK_VOLUME_CHECK_FAILURE +Rinseback Error = Unable to complete Rinseback.\n- Check for kinks or clamped blood lines. [216] Title = Treatment Ended Message = Treatment has ended -HD end treatment sub-mode timeout alarm = ALARM_ID_END_TREATMENT_TIMEOUT_ALARM +Treatment Ended = Treatment has ended [217] Title = Prime Error Message = Blood prime error. Check for kinks or clamped blood lines. -HD blood prime volume check failure = ALARM_ID_BLOOD_PRIME_VOLUME_CHECK_FAILURE +Prime Error = Blood prime error. Check for kinks or clamped blood lines. [218] -Title = Syringe Not Detected -Message = Heparin syringe not detected. -HD syringe detected alarm = ALARM_ID_HD_SYRINGE_DETECTED +Title = Syringe Detected +Message = Remove Heparin syringe. +Syringe Detected = Remove Heparin syringe. [219] Title = Syringe Removed -Message = Heparin syringe removed. -HD syringe pump syringe removed alarm = ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED +Message = Install Heparin syringe. +Syringe Removed = Install Heparin syringe. [220] -Title = Battery Error -Message = Battery pack error. -HD battery pack SOC is too low. = ALARM_ID_DG_THD_SENSORS_FPGA_FAULT +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [221] Title = Saline Bag Empty Message = Saline back appears empty.\n- Check saline bag and line and replace if saline bag is empty. -HD Saline bag is empty = ALARM_ID_EMPTY_SALINE_BAG +Saline Bag Empty = Saline back appears empty.\n- Check saline bag and line and replace if saline bag is empty. [222] -Title = Dialysate Temperature High -Message = Dialysate temperature high. -HD dialysate temperature too high alarm = ALARM_ID_HD_OCCLUSION_SENSOR_FPGA_FAULT +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [223] -Title = Dialysate Temperature Low -Message = Dialysate temperature low. -HD dialysate temperature too low alarm = ALARM_ID_HD_ARTERIAL_SENSOR_FPGA_FAULT +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [224] Title = Treatment Stopped Message = Treatment paused. -Treatment stopped by user action - pressed stop button = ALARM_ID_TREATMENT_STOPPED_BY_USER +Treatment Stopped = Treatment paused. [225] Title = End of Treatment Message = Treatment has ended -HD end of treatment warning = ALARM_ID_END_OF_TREATMENT_WARNING +End of Treatment = Treatment has ended [226] Title = Prime Complete Message = Prime of the bloodlines is complete. -HD prime completed low priority alarm = ALARM_ID_PRIME_COMPLETED_LOW_PRIORITY +Prime Complete = Prime of the bloodlines is complete. [227] Title = Maximum Prime Time Exceeded Message = Maximum time for blood prime has been exceeded. -HD prime out of time alarm = ALARM_ID_PRIME_OUT_OF_TIME +Maximum Prime Time Exceeded = Maximum time for blood prime has been exceeded. [228] Title = Maximum Prime Time Exceeded Message = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. -HD time out on prime sub-mode purge air state. = ALARM_ID_HD_PRIME_SALINE_PURGE_AIR_TIME_OUT +Maximum Prime Time Exceeded = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. [229] Title = Maximum Prime Time Exceeded Message = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. -HD prime dialysate dialyzer time out alarm = ALARM_ID_PRIME_DIALYSATE_DIALYZER_TIME_OUT +Maximum Prime Time Exceeded = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. [230] Title = Maximum Prime Time Exceeded Message = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. -HD prime dialysate bypass time out alarm = ALARM_ID_PRIME_DIALYSATE_BYPASS_TIME_OUT +Maximum Prime Time Exceeded = Priming failure.\n- Check blood and dialysate lines for clamps and kinks. [231] Title = Self Test Error Message = Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. -HD pre-treatment mode dry self-tests failure = ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE +Self Test Error = Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. [232] Title = Self Test Error Message = Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. Check for closed clamps or kinks in the tubing. -HD pre-treatment mode wet self-tests failure = ALARM_ID_PRE_TREATMENT_WET_LC_TEST_FAILURE +Self Test Error = Self Test Failure. Please ensure the disposable cartridge is completely installed and the handle moved to the down position. Check for closed clamps or kinks in the tubing. [233] -Title = -Message = - = ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC +Title = Pressure Sensors Error +Message = Call service to schedule a repair. +Pressure Sensors Error = Call service to schedule a repair. [234] -Title = Communication Failure -Message = Communication failure.\n- Ensure all cords are securely connected. -RTC (i.e read time) or RAM operations failure (read or write) = ALARM_ID_RTC_RAM_OPS_ERROR +Title = Inlet Water Pressure High +Message = Inlet water pressure is high.\n- Check the incoming water for issues. +Inlet Water Pressure High = Inlet water pressure is high.\n- Check the incoming water for issues. [235] Title = Rinseback Complete Message = Rinseback has been completed. -HD in treatment stopped sub-mode after rinseback completed (no escalation) = ALARM_ID_TREATMENT_STOPPED_AFTER_RINSEBACK +Rinseback Complete = Rinseback has been completed. [236] Title = New Cartridge Needed Message = New cartridge needs to be installed. -HD needs new cartridge to be installed = ALARM_ID_INSTALL_NEW_CARTRIDGE +New Cartridge Needed = New cartridge needs to be installed. [237] -Title = -Message = -HD prime saline dialyzer time out alarm = ALARM_ID_PRIME_SALINE_DIALYZER_TIME_OUT +Title = Prime Error +Message = Call service to schedule a repair. +Prime Error = Call service to schedule a repair. [238] Title = Cartridge Needed Message = Cartridge not loaded or loaded improperly.\n- Check that a cartridge is properly installed. -HD no cartridge loaded or installed improperly alarm = ALARM_ID_NO_CARTRIDGE_LOADED +Cartridge Needed = Cartridge not loaded or loaded improperly.\n- Check that a cartridge is properly installed. [239] Title = Cartridge Removal Error Message = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. -HD fail to remove cartridge alarm = ALARM_ID_CARTRIDGE_REMOVAL_FAILURE +Cartridge Removal Error = Cartridge not unloaded, or unloaded improperly.\n- Check that the cartridge has been removed. [240] Title = Dialysate Conductivity Error Message = Dialysate conductivity out of range.\n- Check to be sure the correct bicarbonate dialysate concentrate is connected. -Bicarb conductivity out of range during bicarb pump check alarm = ALARM_ID_BICARB_CONDUCTIVITY_OUT_OF_RANGE +Dialysate Conductivity Error = Dialysate conductivity out of range.\n- Check to be sure the correct bicarbonate dialysate concentrate is connected. [241] -Title = Dialysate Drain Pump Error +Title = Cleaning Mode Error Message = Call service to schedule a repair. -DG reservoir drain time out = ALARM_ID_DG_RESERVOIR_DRAIN_TIMEOUT +Cleaning Mode Error = Call service to schedule a repair. [242] -Title = Dialysate Production Error +Title = Cleaning Mode Error Message = Call service to schedule a repair. -DG reservoir fill time out = ALARM_ID_DG_RESERVOIR_FILL_TIMEOUT +Cleaning Mode Error = Call service to schedule a repair. [243] -Title = Dialysate Production Error +Title = Heat Disinfection Error Message = Call service to schedule a repair. -DG reservoir leak time out = ALARM_ID_DG_RESERVOIR_LEAK_TIMEOUT +Heat Disinfection Error = Call service to schedule a repair. [244] -Title = Dialysate Temperature Error +Title = Cleaning Mode Error Message = Call service to schedule a repair. -DG temperature sensors difference out of range = ALARM_ID_DG_TEMP_SENSORS_DIFF_OUT_OF_RANGE +Cleaning Mode Error = Call service to schedule a repair. [245] Title = Heat Disinfection Error Message = Call service to schedule a repair. -DG heat disinfect target temperature time out (could not reach to temperature) = ALARM_ID_DG_HEAT_DISINFECT_TARGET_TEMP_TIMEOUT +Heat Disinfection Error = Call service to schedule a repair. [246] -Title = Heat Disinfection Error +Title = Chemical Disinfection Error Message = Call service to schedule a repair. -DG heat disinfect inlet pressure and temperature sensors out of range = ALARM_ID_DG_HEAT_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT +Chemical Disinfection Error = Call service to schedule a repair. [247] -Title = Heat Disinfection Error +Title = Heat Disinfection Active Cool Error Message = Call service to schedule a repair. -DG heat disinfect inlet conductivity and temperature out of range = ALARM_ID_DG_HEAT_DISINFECT_INLET_COND_AND_TEMP_OUT +Heat Disinfection Active Cool Error = Call service to schedule a repair. [248] Title = Chemical Disinfection Error Message = Call service to schedule a repair. -DG chemical disinfect could not reach to target temperature = ALARM_ID_DG_CHEM_DISINFECT_TARGET_TEMP_TIMEOUT +Chemical Disinfection Error = Call service to schedule a repair. [249] Title = Chemical Disinfection Error Message = Call service to schedule a repair. -DG chemical disinfect inlet pressure and temperature sensors out of range = ALARM_ID_DG_CHEM_DISINFECT_INLET_PRES_AND_TEMP_SNSRS_OUT +Chemical Disinfection Error = Call service to schedule a repair. [250] -Title = Chemical Disinfection Error -Message = Call service to schedule a repair. -DG chemical disinfect inlet conductivity and temperature out of range = ALARM_ID_DG_CHEM_DISINFECT_INLET_COND_AND_TEMP_OUT +Title = Chemical Disinfect +Message = Insert the Acid Bottle +Chemical Disinfect = Insert the Acid Bottle [251] Title = Machine Error Message = Call service to schedule a repair. -HD invalid system record = ALARM_ID_HD_INVALID_SYSTEM_RECORD_CRC +Machine Error = Call service to schedule a repair. [252] Title = Machine Error Message = Call service to schedule a repair. -HD invalid service record = ALARM_ID_HD_INVALID_SERVICE_RECORD_CRC +Machine Error = Call service to schedule a repair. [253] Title = Machine Error Message = Call service to schedule a repair. -DG invalid system record = ALARM_ID_DG_INVALID_SYSTEM_RECORD_CRC +Machine Error = Call service to schedule a repair. [254] Title = Machine Error Message = Call service to schedule a repair. -DG invalid service record = ALARM_ID_DG_INVALID_SERVICE_RECORD_CRC +Machine Error = Call service to schedule a repair. [255] Title = Machine Error Message = Call service to schedule a repair. -HD and UI software builds are not compatible = ALARM_ID_HD_UI_COMPATIBILITY_ERROR +Machine Error = Call service to schedule a repair. [256] Title = Self-Test Error Message = Call service to schedule a repair. -HD and DG software builds are not compatible = ALARM_ID_HD_DG_COMPATIBILITY_ERROR +Self-Test Error = Call service to schedule a repair. [257] Title = Machine Error Message = Call service to schedule a repair. -DG FPGA power out timeout = ALARM_ID_DG_FPGA_POWER_OUT_TIMEOUT +Machine Error = Call service to schedule a repair. [258] -Title = -Message = -HD FPGA power out timeout = ALARM_ID_HD_FPGA_POWER_OUT_TIMEOUT +Title = Machine Error +Message = Call service to schedule a repair. +Machine Error = Call service to schedule a repair. [259] -Title = Dialysate Temperature Error +Title = Temperature Sensors Error Message = Call service to schedule a repair. -HD temperatures out of range = ALARM_ID_HD_TEMPERATURES_OUT_OF_RANGE +Temperature Sensors Error = Call service to schedule a repair. [260] Title = Machine Error Message = Call service to schedule a repair. -UI POST Application Integrity (Sha256Sum) failure = ALARM_ID_UI_POST_FAILURE_SHASUM +Machine Error = Call service to schedule a repair. [261] Title = Machine Error Message = Call service to schedule a repair. -UI POST CANBus failure = ALARM_ID_UI_POST_FAILURE_CANBUS +Machine Error = Call service to schedule a repair. [262] Title = Machine Error Message = Call service to schedule a repair. -UI POST Display failure = ALARM_ID_UI_POST_FAILURE_DISPLAY +Machine Error = Call service to schedule a repair. [263] Title = Machine Error Message = Call service to schedule a repair. -UI POST Touch failure = ALARM_ID_UI_POST_FAILURE_TOUCH +Machine Error = Call service to schedule a repair. [264] Title = Machine Error Message = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persist. -UI POST SD-Card failure = ALARM_ID_UI_POST_FAILURE_SDCARD +Machine Error = SD-Card Failure \n- Power cycle the system.\n- Call service to schedule a repair if problem persist. [265] Title = Machine Error Message = Call service to schedule a repair. -UI POST RTC failure = ALARM_ID_UI_POST_FAILURE_RTC +Machine Error = Call service to schedule a repair. [266] Title = Machine Error Message = Call service to schedule a repair. -UI POST WiFi failure = ALARM_ID_UI_POST_FAILURE_WIFI +Machine Error = Call service to schedule a repair. [267] Title = Machine Error Message = Call service to schedule a repair. -UI POST Bluetooth failure = ALARM_ID_UI_POST_FAILURE_BLUETOOTH +Machine Error = Call service to schedule a repair. [268] Title = Machine Error Message = Call service to schedule a repair. -UI POST Ethernet failure = ALARM_ID_UI_POST_FAILURE_ETHERNET +Machine Error = Call service to schedule a repair. [269] Title = Machine Error Message = Call service to schedule a repair. -UI POST Sound failure = ALARM_ID_UI_POST_FAILURE_SOUND +Machine Error = Call service to schedule a repair. [270] Title = Self Test Error Message = Call service to schedule a repair. -HD POST Safety Shutdown failure = ALARM_ID_HD_SAFETY_SHUTDOWN_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [271] Title = Self Test Error Message = Call service to schedule a repair. -DG POST Safety Shutdown failure = ALARM_ID_DG_SAFETY_SHUTDOWN_POST_TEST_FAILED +Self Test Error = Call service to schedule a repair. [272] -Title = +Title = Fans Error Message = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. -HD Fan RPM out of range = ALARM_ID_HD_FAN_RPM_OUT_OF_RANGE +Fans Error = Fan failure.\n- Check that the fans inlets are not blocked or obstructed. [273] Title = Blood Pump Error Message = Call service to schedule a repair. -HD measured blood pump flow rate is out of range = ALARM_ID_HD_BLOOD_FLOW_OUT_OF_RANGE +Blood Pump Error = Call service to schedule a repair. [274] Title = Dialysate Pump Error Message = Call service to schedule a repair. -HD measured dialysate inlet pump flow rate is out of range = ALARM_ID_HD_DIAL_IN_FLOW_OUT_OF_RANGE +Dialysate Pump Error = Call service to schedule a repair. [275] Title = Arterial Pressure Error Message = Call service to schedule a repair. -HD arterial pressure sensor is reading out of range = ALARM_ID_HD_ARTERIAL_PRESSURE_OUT_OF_RANGE +Arterial Pressure Error = Call service to schedule a repair. [276] Title = Venous Pressure Error Message = Call service to schedule a repair. -HD venous pressure sensor is reading out of range = ALARM_ID_HD_VENOUS_PRESSURE_OUT_OF_RANGE +Venous Pressure Error = Call service to schedule a repair. [277] Title = Occlusion Sensor Error Message = Call service to schedule a repair. -HD BP occlusion sensor is reading out of range = ALARM_ID_HD_BP_OCCLUSION_OUT_OF_RANGE +Occlusion Sensor Error = Call service to schedule a repair. [278] Title = Dialysate Production Error Message = Ultrafiltration error. -HD active reservoir weight out of range. = ALARM_ID_HD_ACTIVE_RESERVOIR_WEIGHT_OUT_OF_RANGE +Dialysate Production Error = Ultrafiltration error. [279] Title = Dialysate Drain Pump Error Message = Call service to schedule a repair. -DG dialysate drain time out = ALARM_ID_DG_DIALYSATE_DRAIN_TIME_OUT +Dialysate Drain Pump Error = Call service to schedule a repair. [280] -Title = Arterial Pressure Error +Title = Drain Pump Error Message = Call service to schedule a repair. -HD arterial pressure sensor read timeout error = ALARM_ID_DG_DRAIN_PUMP_DIRECTION_INVALID +Drain Pump Error = Call service to schedule a repair. [281] -Title = Acid Concentrate Low +Title = Acid Concentrate Volume Low Message = Acid concentrate low.\n- Replace acid concentrate bottle. -DG Acid concentration bottle low volume alarm = ALARM_ID_DG_ACID_BOTTLE_LOW_VOLUME +Acid Concentrate Volume Low = Acid concentrate low.\n- Replace acid concentrate bottle. [282] -Title = Bicarbonate Concentrate Low +Title = Bicarbonate Concentrate Volume Low Message = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. -DG Bicarbonate concentration bottle low volume alarm = ALARM_ID_DG_BICARB_BOTTLE_LOW_VOLUME +Bicarbonate Concentrate Volume Low = Bicarbonate concentrate low.\n- Replace bicarbonate concentrate bottle. [283] -Title = Ultrafiltration Measurement Error +Title = Load Cells Error Message = Call service to schedule a repair. -DG load cells weight out of range = ALARM_ID_DG_LOAD_CELL_WEIGHT_OUT_OF_RANGE +Load Cells Error = Call service to schedule a repair. [284] -Title = Self-Test Error +Title = Load Cells Error Message = Call service to schedule a repair. -DG load cells primary/back up drift out of range = ALARM_ID_DG_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE +Load Cells Error = Call service to schedule a repair. [285] Title = Dialysate Recirculation Error Message = Dialysate recirculation error. -HD treatment recirculate timeout warning = ALARM_ID_HD_TREATMENT_RECIRC_TIMEOUT_WARNING +Dialysate Recirculation Error = Dialysate recirculation error. [286] Title = Rinseback Complete Message = Rinseback complete. -HD treatment rinseback complete timeout warning = ALARM_ID_HD_TREATMENT_RINSEBACK_TIMEOUT_WARNING +Rinseback Complete = Rinseback complete. [287] Title = Machine Error Message = Call service to schedule a repair. -HD processor clock speed checks against FPGA clock failure = ALARM_ID_HD_FPGA_CLOCK_SPEED_CHECK_FAILURE +Machine Error = Call service to schedule a repair. [288] -Title = Dialysate Temperature Error +Title = Load Cells Error Message = Call service to schedule a repair. -HD load cell inconsistency = ALARM_ID_HD_LOAD_CELL_PRIMARY_BACKUP_DRIFT_OUT_OF_RANGE +Load Cells Error = Call service to schedule a repair. [289] -Title = DG Cap State Error +Title = Cleaning Mode Error Message = Check dialysate/concentrate caps. -DG cap in wrong state = ALARM_ID_DG_DIALYSATE_OR_CONC_CAP_NOT_IN_PROPER_POSITION +Cleaning Mode Error = Check dialysate/concentrate caps. [290] -Title = RO Flow Low -Message = -DG RO flow too low while heating = ALARM_ID_HD_VENOUS_SENSOR_FPGA_FAULT +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [291] -Title = DG Flush Failure -Message = Dialsyate Generation Device flush failure. - = ALARM_ID_HD_DISINFECT_FLUSH +Title = Cleaning Modes Start Alarm +Message = No Instructions +Cleaning Modes Start Alarm = No Instructions [292] -Title = DG Heat Disinfection Failure -Message = DG Heat Disinfection Failure. - = ALARM_ID_HD_DISINFECT_HEAT +Title = Cleaning Modes Start Alarm +Message = No Instructions +Cleaning Modes Start Alarm = No Instructions [293] -Title = DG Chemical Disinfection Failure -Message = DG Chemical Disinfection Failure - = ALARM_ID_HD_DISINFECT_CHEM +Title = Cleaning Modes Start Alarm +Message = No Instructions +Cleaning Modes Start Alarm = No Instructions [294] -Title = -Message = - = ALARM_ID_HD_DISINFECT_CHEM_INSTRUCTION1 +Title = Cleaning Modes Start Alarm +Message = No Instructions +Cleaning Modes Start Alarm = No Instructions [295] -Title = Self-Test Error +Title = Machine Error Message = Call service to schedule a repair. - = ALARM_ID_UI_POST_FAILURE_CLOUDSYNC +Machine Error = Call service to schedule a repair. [296] -Title = Self-Test Error +Title = FPGA Error Message = Call service to schedule a repair. - = ALARM_ID_DG_TDI_SENSORS_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [297] -Title = Self-Test Error +Title = FPGA Error Message = Call service to schedule a repair. - = ALARM_ID_DG_TRO_SENSORS_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [298] -Title = Self-Test Error +Title = FPGA Error Message = Call service to schedule a repair. - = ALARM_ID_DG_BARO_SENSOR_FPGA_FAULT +FPGA Error = Call service to schedule a repair. [299] -Title = -Message = - = ALARM_ID_DG_INVALID_SERIAL_NUMBER +Title = Self Test Error +Message = Call service to schedule a repair. +Self Test Error = Call service to schedule a repair. [300] -Title = -Message = - = ALARM_ID_HD_INVALID_SERIAL_NUMBER +Title = Self Test Error +Message = Call service to schedule a repair. +Self Test Error = Call service to schedule a repair. [301] -Title = Self-Test Error +Title = Heaters Error Message = Call service to schedule a repair. - = ALARM_ID_DG_PRIMARY_HEATER_POWER_VOLTAGE_OUT_OF_RANGE +Heaters Error = Call service to schedule a repair. [302] -Title = Processore RAM error +Title = Processor RAM error Message = Call service to schedule a repair. -HD_CPU RAM error, or DG_CPU RAM error = ALARM_ID_DG_CPU_RAM_ERROR +Processor RAM error = Call service to schedule a repair. [303] -Title = -Message = - = ALARM_ID__AVAILABLE_4 +Title = FPGA Error +Message = Call service to schedule a repair. +FPGA Error = Call service to schedule a repair. [304] Title = Self-Test Error Message = Call service to schedule a repair. - = ALARM_ID_HD_INVALID_USAGE_RECORD_CRC +Self-Test Error = Call service to schedule a repair. [305] -Title = Self-Test Error +Title = Processor RAM error Message = Call service to schedule a repair. - = ALARM_ID_HD_CPU_RAM_ERROR +Processor RAM error = Call service to schedule a repair. [306] Title = Air Trap Error Message = Air Trap Error.\n- Check air trap fill level. - = ALARM_ID_HD_AIR_TRAP_OVERFILL +Air Trap Error = Air Trap Error.\n- Check air trap fill level. [307] -Title = -Message = - = ALARM_ID_HD_AC_POWER_LOST_SECOND +Title = HD AC power loss +Message = HD power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until power returns. +HD AC power loss = HD power has been lost. Check all electrical connections.\n- The display and CPU will run off of the battery until power returns. +[308] +Title = Conductivity Sensors Error +Message = Call service to schedule a repair. +Conductivity Sensors Error = Call service to schedule a repair. + +[309] +Title = Conductivity Sensors Error +Message = Call service to schedule a repair. +Conductivity Sensors Error = Call service to schedule a repair. + +[310] +Title = Conductivity Sensors Error +Message = Call service to schedule a repair. +Conductivity Sensors Error = Call service to schedule a repair. + +[311] +Title = Conductivity Sensors Error +Message = Call service to schedule a repair. +Conductivity Sensors Error = Call service to schedule a repair. + +[312] +Title = Dialysate Flowrate Error +Message = Call service to schedule a repair. +Dialysate Flowrate Error = Call service to schedule a repair. + +[313] +Title = Cleaning Mode Inlet Water Temperature Too Low +Message = Inlet water temperature too low.\n- Check the incoming water for issues. +Cleaning Mode Inlet Water Temperature Too Low = Inlet water temperature too low.\n- Check the incoming water for issues. + +[314] +Title = Cleaning Mode Inlet Water Pressure Too High +Message = Inlet water pressure too high.\n- Check the incoming water for issues. +Cleaning Mode Inlet Water Pressure Too High = Inlet water pressure too high.\n- Check the incoming water for issues. + +[315] +Title = Cleaning Mode Inlet Water Pressure Too Low +Message = Inlet water pressure too low.\n- Check the incoming water for issues. +Cleaning Mode Inlet Water Pressure Too Low = Inlet water pressure too low.\n- Check the incoming water for issues. + +[316] +Title = Acid Concentrate Pump Park Fault +Message = Call service to schedule a repair. +Acid Concentrate Pump Park Fault = Call service to schedule a repair. + +[317] +Title = Bicarb Concentrate Pump Park Fault +Message = Call service to schedule a repair. +Bicarb Concentrate Pump Park Fault = Call service to schedule a repair. + +[318] +Title = Air Pump Operation Timeout +Message = Call service to schedule a repair. +Air Pump Operation Timeout = Call service to schedule a repair. + +[319] +Title = Chemical Disinfect Flush +Message = Collect Flush Sample. +Chemical Disinfect Flush = Collect Flush Sample. + +[320] +Title = Chemical Disinfect Flush +Message = Call service to schedule a repair. +Chemical Disinfect Flush = Call service to schedule a repair. + Index: Confirm/Confirm.conf =================================================================== diff -u -r8faf38d02db36b4fde0683da7f1b47eb377d45c5 -rf1537dd481acf785b57e85aa6c57ac5f53091fdb --- Confirm/Confirm.conf (.../Confirm.conf) (revision 8faf38d02db36b4fde0683da7f1b47eb377d45c5) +++ Confirm/Confirm.conf (.../Confirm.conf) (revision f1537dd481acf785b57e85aa6c57ac5f53091fdb) @@ -1,18 +1,23 @@ [1] Title = Water Flush -Message = Are sure you want to stop water flush? +Message = Are you sure want to stop water flush? Confirm = CONFIRM Cancel = CANCEL [2] Title = Heat Disinfect -Message = Are sure you want to stop heat disinfection? +Message = Are you sure want to stop heat disinfection? Confirm = CONFIRM Cancel = CANCEL [3] Title = Chemical Disinfect -Message = Are sure you want to stop chemical disinfection? +Message = Are you sure want to stop chemical disinfection? Confirm = CONFIRM Cancel = CANCEL +[4] +Title = Chemical Flush +Message = Are you sure want to stop chemical flush? +Confirm = CONFIRM +Cancel = CANCEL Index: Instructions/Instructions.conf =================================================================== diff -u -r822daa04828216cb1adf94bd12a28c728b27043d -rf1537dd481acf785b57e85aa6c57ac5f53091fdb --- Instructions/Instructions.conf (.../Instructions.conf) (revision 822daa04828216cb1adf94bd12a28c728b27043d) +++ Instructions/Instructions.conf (.../Instructions.conf) (revision f1537dd481acf785b57e85aa6c57ac5f53091fdb) @@ -94,7 +94,8 @@ Disinfection3=logo.png [Chemical Disinfect^Chemical Disinfect] -Remove the concentrate fluid cap and connect the chemical disinfectant = 17_connect PAA_v4.png -Disconnect the chemical disinfectant and replace concentrate cap. = 19_replace concentrate cap-01.png -Test the system for residual chemicals using a test strip. = 43_test residual sample-01.png +Chemical Disinfect 1 = logo.png +[Chemical Flush^Chemical Flush] +Chemical Flush 1 = logo.png + Fisheye: Tag f1537dd481acf785b57e85aa6c57ac5f53091fdb refers to a dead (removed) revision in file `ui.config'. Fisheye: No comparison available. Pass `N' to diff?