Index: Alarms/Alarms.conf =================================================================== diff -u -ra0b3e71e3bb0ba51509b2370dbd9b20e4e526af5 -rf24ecf486ae12326e176bc88b3fe7f956111388f --- Alarms/Alarms.conf (.../Alarms.conf) (revision a0b3e71e3bb0ba51509b2370dbd9b20e4e526af5) +++ Alarms/Alarms.conf (.../Alarms.conf) (revision f24ecf486ae12326e176bc88b3fe7f956111388f) @@ -59,12 +59,12 @@ [10] Title = Message = - = ALARM_ID_HD_BLOOD_LEAK_FAULT + = ALARM_ID_HD___AVAILABLE_7 [11] Title = Message = - = ALARM_ID_HD_BLOOD_LEAK_SELF_TEST_FAILURE + = ALARM_ID_HD_DIALYSATE_TEMP_BELOW_TARGET_TEMP [12] Title = Self Test Error @@ -79,12 +79,12 @@ [14] Title = Message = - = ALARM_ID_HD_BLOOD_FLOW_INVALID_CALIBRATION + = ALARM_ID___AVAILABLE_1 [15] Title = Message = - = ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CALIBRATION + = ALARM_ID___AVAILABLE_2 [16] Title = Self Test Error @@ -104,7 +104,7 @@ [19] Title = Calibration Error Message = Call service to schedule a repair. -DG flow sensors invalid calibration record = ALARM_ID_DG_FLOW_SENSORS_INVALID_CAL_RECORD +DG flow sensors invalid calibration record = ALARM_ID_HD_DIALYSATE_TEMP_ABOVE_TARGET_TEMP [20] Title = Calibration Error @@ -152,12 +152,12 @@ [27] Title = Message = - = ALARM_ID_HD_BLOOD_FLOW_INVALID_CAL_RECORD + = ALARM_ID_HD___AVAILABLE_3 [28] Title = Message = - = ALARM_ID_HD_DIALYSATE_FLOW_INVALID_CAL_RECORD + = ALARM_ID_HD___AVAILABLE_4 [29] Title = Calibration Error @@ -385,12 +385,12 @@ [72] Title = Message = - = ALARM_ID_HD_BLOOD_FLOW_STATUS_SELF_TEST_FAILURE + = ALARM_ID_HD___AVAILABLE_5 [73] Title = Message = - = ALARM_ID_HD_DIALYSATE_FLOW_STATUS_SELF_TEST_FAILURE + = ALARM_ID_HD___AVAILABLE_6 [74] Title = Self Test Error @@ -800,12 +800,12 @@ [155] Title = Message = - = ALARM_ID_DG_TEMPERATURE_SENSOR_ADC_OUT_OF_RANGE + = ALARM_ID_HD___AVAILABLE_9 [156] Title = Message = - = ALARM_ID_DG_TEMPERATURE_SENSOR_FAULT + = ALARM_ID_HD___AVAILABLE_10 [157] Title = @@ -905,7 +905,7 @@ [176] Title = Message = -// TODO remove this alarm HD active reservoir is full during treatment and DG not ready to switch reservoirs. = ALARM_ID_HD_RESERVOIR_FULL_AND_DG_NOT_READY_TO_SWITCH +// TODO remove this alarm HD active reservoir is full during treatment and DG not ready to switch reservoirs. = ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_HIGH_SAFETY_RANGE [177] Title = Machine Error @@ -990,7 +990,7 @@ [193] Title = Message = - = ALARM_ID__AVAILABLE_13 + = ALARM_ID_HD_DIALYSATE_TEMP_OUT_OF_LOW_SAFETY_RANGE [194] Title = RO Pump Error @@ -1190,7 +1190,7 @@ [233] Title = Message = - = ALARM_ID__AVAILABLE_8 + = ALARM_ID_DG_BAROMETRIC_SENSOR_COEFFS_BAD_CRC [234] Title = Communication Failure @@ -1344,7 +1344,7 @@ [264] Title = Machine Error -Message = Call service to schedule a repair. +Message = Power cycle the system.\nCall service to schedule a repair if problem persist. UI POST SD-Card failure = ALARM_ID_UI_POST_FAILURE_SDCARD [265]