Index: Configurations/DataList.conf =================================================================== diff -u -r31aa4eb437f663553bf41098522ff9935d69fe2b -rfd36121a1a6f24e1c39957155a1b4a6b1ecb01c6 --- Configurations/DataList.conf (.../DataList.conf) (revision 31aa4eb437f663553bf41098522ff9935d69fe2b) +++ Configurations/DataList.conf (.../DataList.conf) (revision fd36121a1a6f24e1c39957155a1b4a6b1ecb01c6) @@ -1,24 +1,118 @@ -[Acid Concentrate] -# Fres. Naturalyte # it has been commented out regarding Sean's point on PRS35 +[Blood Flow Rate Ranges] # mL/min +Blood_Flow_Rate_Min = +100 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Max = +500 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Res = 25 # PRS 30, SRSUI 619, 639, 765 +Blood_Flow_Rate_Def = 100 # PRS 30, SRSUI 619, 639, 765 + +[Dialysate Flow Rate Ranges] # mL/min +Dialysate_Flow_Rate_Min = +100 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Max = +600 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Res = 50 # PRS 345, SRSUI 620, 640, 766 +Dialysate_Flow_Rate_Def = 100 # PRS 345, SRSUI 620, 640, 766 + +[Treatment Duration Ranges] # min +Treatment_Duration_Min = +60 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Max = +480 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Res = 15 # PRS 346, SRSUI 569, 767 +Treatment_Duration_Def = 60 # PRS 346, SRSUI 569, 767 + +[Heparin Dispensing Rate Ranges] # mL/hr (0 = OFF) +Heparin_Dispensing_Rate_Min = 0.2 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Max = 1.0 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Res = 0.1 # PRS 347, SRSUI 768 +Heparin_Dispensing_Rate_Def = 0.0 # PRS 347, SRSUI 768 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. + +[Heparin Bolus Volume Ranges] # mL (0 = OFF) +Heparin_Bolus_Volume_Min = 0.2 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Max = 2.0 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Res = 0.1 # PRS 348, SRSUI 769 +Heparin_Bolus_Volume_Def = 0.0 # PRS 348, SRSUI 769 # Since the minimum is (0.2 > 0.0), def = 0.0 means, the default value is OFF. + +[Heparin Stop Time Ranges] # min +Heparin_Stop_Time_Min = 0 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Max = 480 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Res = 10 # PRS 464. SRSUI 771 +Heparin_Stop_Time_Def = 0 # PRS 464. SRSUI 771 + +[Saline Bolus Volume Ranges] # mL +Saline_Bolus_Volume_Min = 100 # PRS 350 +Saline_Bolus_Volume_Max = 300 # PRS 350 +Saline_Bolus_Volume_Res = 100 # PRS 350 +Saline_Bolus_Volume_Def = 100 # PRS 350 + +[Heparin Type Options] # PRS ???, SRSUI ??? +UFH 1,000 IU/mL + +[Acid Concentrate Options] # PRS 35, SRSUI 772 08-1251-1 08-2251-0 08-3251-9 -[Bicarbonate Concentrate] +[Bicarbonate Concentrate Options] # PRS 351, SRSUI 773 Fres. Centrisol -[Dialyzer Type] +[Dialyzer Type Options] # PRS 352, SRSUI 774 BB Diacap Pro 13H BB Diacap Pro 16H BB Diacap Pro 19H F Optiflux F160NRe F Optiflux F180NRe F Optiflux F200NRe -[Heparin Type] -UFH 1,000 IU/mL +[Dialysate Temperature] # Celsius +Dialysate_Temperature_Min = 35 # PRS 353, SRSUI 775 +Dialysate_Temperature_Max = 37 # PRS 353, SRSUI 775 +Dialysate_Temperature_Res = 0.5 # PRS 353, SRSUI 775 +Dialysate_Temperature_Def = 37 # PRS 353, SRSUI 775 +[Arterial Blood Pressure Ranges] +# Adjustment, Arterial Low +Arterial_Adjust_Low_Min = -300 # PRS 354, SRSUI 776, 726 +Arterial_Adjust_Low_Max = +70 # PRS 354, SRSUI 776, 726 +Arterial_Adjust_Low_Res = 10 # PRS 354, SRSUI 776, 726 # this has to be the same for all since they are on the same slider with the same scale. +Arterial_Adjust_Low_Def = -300 # PRS 354, SRSUI 776, 726 +Arterial_Adjust_Low_Gap = 30 # PRS 354, SRSUI 728 +# Adjustment, Arterial High +Arterial_Adjust_High_Min = -270 # PRS 355, SRSUI 780, 727 +Arterial_Adjust_High_Max = +100 # PRS 355, SRSUI 780, 727 +Arterial_Adjust_High_Res = 10 # PRS 355, SRSUI 780, 727 # this has to be the same for all since they are on the same slider with the same scale. +Arterial_Adjust_High_Def = +100 # PRS 355, SRSUI 780, 727 +Arterial_Adjust_High_Gap = 30 # PRS 355, SRSUI 728 +#Monitoring +Arterial_Monitor_Res = 1 # PRS 362, SRS 617 +Arterial_Monitor_Low = -400 # PRS 362, SRS 617 +Arterial_Monitor_High = +100 # PRS 362, SRS 617 + +[Venous Blood Pressure Ranges] +# Adjustment, Venous Low +Venous_Adjust_Low_Min = +20 # PRS 356, SRSUI 777, 731 +Venous_Adjust_Low_Max = +370 # PRS 356, SRSUI 777, 731 +Venous_Adjust_Low_Res = 10 # PRS 356, SRSUI 777, 731 # this has to be the same for all since they are on the same slider with the same scale. +Venous_Adjust_Low_Def = +20 # PRS 356, SRSUI 777, 731 +Venous_Adjust_Low_Gap = 30 # PRS 356, SRSUI 1084 +# Adjustment, Venous High +Venous_Adjust_High_Min = +50 # PRS 357, SRSUI 781, 732 +Venous_Adjust_High_Max = +400 # PRS 357, SRSUI 781, 732 +Venous_Adjust_High_Res = 10 # PRS 357, SRSUI 781, 732 # this has to be the same for all since they are on the same slider with the same scale. +Venous_Adjust_High_Def = +400 # PRS 357, SRSUI 781, 732 +Venous_Adjust_High_Gap = 30 # PRS 357, SRSUI 1084 +##Monitoring +Venous_Monitor_Res = 1 # PRS 361, SRS 616 +Venous_Monitor_Low = -100 # PRS 361, SRS 616 +Venous_Monitor_High = +500 # PRS 361, SRS 616 + +[Blood Pressure Measure Interval Ranges] # min (0 = OFF) +Blood_Pressure_Measure_Interval_Min = 5 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Max = 60 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Res = 5 # PRS 358, SRSUI 778 +Blood_Pressure_Measure_Interval_Def = 0 # PRS 358, SRSUI 778 + +[Rinseback Flow Rate Ranges] # mL/min +Rinseback_Flow_Rate_Min = 50 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Max = 150 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Res = 25 # PRS 359, SRSUI 330, 770 +Rinseback_Flow_Rate_Def = 50 # PRS 359, SRSUI 330, 770 + [Water Sample Result] Fail Pass -