DEN-9416: Implementations - Added a newly added parameter of the BloodPrime (Safety Vol). - UI POST is done. - Out of 10 tests 1 - CANBus 2 - Touch Screen 3 - SD-Card 4 - RTC tests are done.
- The 5 - File System integrity will be done as part of CyberSecurity.
- The 6 - Bluetooth will be added when the feature itself added to the application. - The 7 - WiFi is in progress and having issue correctly detect it.
- Some tests are listed to be added later in case required, like 8 - Display 9 - Ethernet 10- Sound test couldn't find a way to test it yet.
- On each unsuccessful test an alarm will be triggered. - UI alarms are in range of [250 - 259]