DENBUG-287 ( DEN-16823 ) : Tx Log Export Success Message via USB is Hidden Behind REQUEST_REJECT_REASON_DRAIN_NOT_COMPLETE Msg - DEN-16823: Tx Log Export Success Message via USB is Hidden Behind REQUEST_REJECT_REASON_DRAIN_NOT_COMPLETE Msg - Cherry Picked from Dara's commit
DEN-17161: Incorrect Calcium Value recorded in TX logs
- Cherry picked from Dara's commit 8fa28af55b1a183b75fb6ddd800192e88ac8d258.
DENBUG-294: UI: Log export between different devices on the same USB drive has conflict of names.
DENBUG-294: UI: Log export between different devices on the same USB drive has conflict of names.
- Added functionality to properly backup the BootPOST log which is the log before getting the HDSerial. - It will get backed up if it is in the current day and system just has changed to the new log file due to receiving of the serial. - When trying to back up the normal logs (logs with serial), will try for its BootPOST log, if exists will get backed up.
DENBUG-294: UI: Log export between different devices on the same USB drive has conflict of names.
- renamed the BootPOST log before backing up so won't remain Serial-less on the device.
DENBUG-294: UI: Log export between different devices on the same USB drive has conflict of names.
- now using the old "denali" added log format for update/manufacturing - and delete them with the lockdown.sh script.
DENBUG-294: UI: Log export between different devices on the same USB drive has conflict of names.
- after the reboot the newly created BootPOST could not get renamed because of the same file already exists, therefore application is now append the new one to the oldone with serial and deletes the new one. - since the BootPOST log files are small in size like 5~10K this should affect the performance of the system.