DENBUG-328: Venous Pressure alarm did not trigger when pressure dropped to 0 mmHg
- Added a new field in the institutional configurations t… Show more
DENBUG-328: Venous Pressure alarm did not trigger when pressure dropped to 0 mmHg

- Added a new field in the institutional configurations to display the value for the min venous pressure wide limit window

Show less