Clone
Behrouz NematiPour
committed
on 03 Aug 21
DEN-9416: Implementations
- Added a newly added parameter of the BloodPrime (Safety Vol).
- UI POST is done.
- Out of 10 tests
1 - C… Show more
DEN-9416: Implementations

- Added a newly added parameter of the BloodPrime (Safety Vol).

- UI POST is done.

         - Out of 10 tests

                1 - CANBus

                2 - Touch Screen

                3 - SD-Card

                4 - RTC

                tests are done.

        

         - The

                5 - File System integrity

                will be done as part of CyberSecurity.

        

         - The

                6 - Bluetooth

                will be added when the feature itself added to the application.

         - The

                7 - WiFi

                is in progress and having issue correctly detect it.

                

         - Some tests are listed to be added later in case required, like

                8 - Display

                9 - Ethernet

                10- Sound

                test couldn't find a way to test it yet.

                

- On each unsuccessful test an alarm will be triggered.

- UI alarms are in range of [250 - 259] Show less

master + 213 more