Lines of code count not available
I don't see where this alarm is being triggered.
Should this be a DEBUG_DENALI comment so Bamboo will prevent a build?
Should we keep the TODO for this?
Does 2400 RPM get us 2400 mL/min drain rate? It's the rate that we want to be 2400.
Make a private definitions section above private data and move this there.
Should this module be removed (replaced by ModeGenIdle)?
Added new alarm ID for dialysate flow out of range.
Handling of passive notification is removed in my branch.
Debug code is removed in my branch. Should s/w fault here I think.
This is removed in my branch.
Make a build switch to disable ACK alarms.
Can the two datas be memcpy'd as EVENT_DATA_Ts instead of U32 and EVENT_DATAS_T separately?
Let's finish optimizing broadcasts.
There are now 2 DG flow sensors (FMP and FMD).
RESOLVED in CODE WALKTHROUGH.