DEN-15265 addressed a code review comment

DEN-15265, DEN-15236 changed the chemical disinfect monitor to check the dialysate cap not to be closed in during the disinfect since the PAA acid is connected to it.

Merged DEN-15265

DEN-15265, DEN-15246 changed the heat disinfect's primary target temperature to 60 C prior to filling with water for the actual heat disinfct. Updated the code to properly exit the bad fill state of the gen idle mode once the bad sequence is finished

Merge branch 'DEN-15246-active-reservoir-empty-post-concentrate-alarm-recovery' into develop

  1. … 2 more files in changeset.
DEN-15246 added more code to support different conductivity sensors calibration tables

  1. … 4 more files in changeset.
DEN-15246 changed the bad fill end of state

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged DEN-15231

  1. … 4 more files in changeset.
DEN-15231 changed the target flow in the mode drain to 300 mL/min and the drain pump's target RPM to 2200

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 7 more files in changeset.
Merged DEN-15199

  1. … 4 more files in changeset.
DEN-15199 added to code to check for dialin check in

  1. … 5 more files in changeset.
DEN-15199 addressed the code review comments

  1. … 1 more file in changeset.
DEN-15199 renamed the test configuration code

  1. … 4 more files in changeset.
DEN-15199 worked on the release software configuration

  1. … 4 more files in changeset.
DEN-15199 updated the code per the systems requirements. Initial work on the release software configurations

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 12 more files in changeset.
Merge branch 'staging' into DEN-15166-add-drain-pump-maxon-board-rpm

Merge branch 'DEN-15178-heat-disinfect-drain-timeout-transition-to-cancel-water-path-instead-of-cancel' into develop

# Conflicts:

# firmware/App/DGCommon.h

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

DEN-15182 removed the rejection ratio alarm from the standby mode

DEN-15182 added a pause state in standby mode for the inlet water conditions. Addressed the code review comments

  1. … 4 more files in changeset.
DEN-13949 updated heat disinfect valves actuator per SRSDG 504

DEN-15182 updated the code to handle the inlet water conditions in standby mode

DEN-15166, DEN-15182 updated the voltages code and fixed the filter flush code to be able to resume after an inlet condition alarm

  1. … 1 more file in changeset.
DEN-15178 DrainR1, DrainR2 on timeout transition to Cancel Water Path

Fixed merge conflicts from DEN-15105.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.