Services

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'DEN-5963-dg-heat-disinfection-2-of-2' into develop

  1. … 4 more files in changeset.
Merge branch 'DEN-7091-pre-treatment-mode' of ssh://dvm-linux02:7999/dg/dgfirmware into DEN-7091-pre-treatment-mode

DEN-7091: removed load cell filter test code.

  1. … 1 more file in changeset.
DEN-7091: Addressed code review

  1. … 1 more file in changeset.
DEN-7091: Added test code to test alpha filter for load cell readings.

  1. … 1 more file in changeset.
DEN-5963 tested the code in the lab

  1. … 3 more files in changeset.
DEN-7091: Added handler for alarm clear

DEN-7091: Added new CAN channel for UI to DG comm

  1. … 3 more files in changeset.
DEN-6890: Removed unnecessary accelerometer calibration parameter message

DEN-6890: Implemented first pass of DG Fluid Leak Detector driver

DEN-6890: Implemented first pass of DG Fluid Leak Detector driver

  1. … 5 more files in changeset.
DEN-5963 work in progress. tested cancellation path with water

  1. … 4 more files in changeset.
DEN-5963 added other inlet and outlet fans RPM

  1. … 2 more files in changeset.
DEN-5963 tested heat disinfect

  1. … 4 more files in changeset.
DEN-7091: Fixed FPGA revision

DEN-7091: Expanded fill command to accept start/stop flag

DEN-7091: Addressed code review

DEN-5963 added Ilimit to the PIControllers.c, added feedback temperature sensor to the heaters.c, worked on heat disinfect mode

  1. … 4 more files in changeset.
DEN-5963 worked on RO pump and the pre heat disinfect states

  1. … 4 more files in changeset.
DEN-5963 merged staging. Worked on RO pump

  1. … 9 more files in changeset.
DEN-5963 worked on RO pump and the drain states of heat disinfect

  1. … 5 more files in changeset.
DEN-7091: Changed startup active reservoir to 1

  1. … 1 more file in changeset.
DEN-5963 changed the RO pump control scheme from pressure control to flow control

  1. … 8 more files in changeset.
DEN-7091: Added filter flush progress data broadcast

  1. … 1 more file in changeset.
DEN-7091: Added water sample command handler

  1. … 2 more files in changeset.
DEN-5980 added or removed some of the states

  1. … 4 more files in changeset.
DEN-5980 removed old calibration commands

  1. … 2 more files in changeset.
DEN-6402: Fixed load cell tare reset between tare requests

  1. … 1 more file in changeset.
DEN-5980 added more commands to send and receive NV data

  1. … 4 more files in changeset.
DEN-6402: Added flow meter check and filters for load cells

  1. … 6 more files in changeset.