Checked with Nick. The DG does not expose motors to the user. The straw door switch should be open in Fill Mode to allow access to straw connection. The straw door switch should be closed in modes ...
Checked with Nick. The DG does not expose motors to the user.
The straw door switch should be open in Fill Mode to allow access to straw connection.
The straw door switch should be closed in modes such as (heat disinfection, chem disinfection) to prevent user from exposure to hot water or chemical.
Added TODO.