DEN-14150: Update DialInFLow.c

    • -7
    • +7
    /firmware/App/Controllers/DialInFlow.c
DEN-14150: Update DialInFLow.c

    • -13
    • +16
    /firmware/App/Controllers/DialInFlow.c
These inits are redundant as initPreTreatmentMode() is called just below which resets these flags already.

These inits are redundant as initPreTreatmentMode() is called just below which resets these flags already.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

I think this check is for over-fill. Counter is for illegal level sensor readings. Over-fill check should have it's own counter (or use new persistence style) and we should reset that separate coun...

I think this check is for over-fill. Counter is for illegal level sensor readings. Over-fill check should have it's own counter (or use new persistence style) and we should reset that separate counter in the "else" condition.

Corrected comment above. Why are we changing BP flow rate at air trap lower level? Should be set to slow or fast on entry to this state (depending on whether this is 1st pass or not) and stay at th...

Corrected comment above. Why are we changing BP flow rate at air trap lower level? Should be set to slow or fast on entry to this state (depending on whether this is 1st pass or not) and stay at that rate for duration of this pass.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Move up to line ~108 where the other shift #defines are.

Move up to line ~108 where the other shift #defines are.

Add a TODO or fix now.

Add a TODO or fix now.

Merge branch 'DEN-14150-hd-dialysate-flow-estimation' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-14150-hd-dialysate-flow-estimation

DEN-14150: Skip syringe force sensor check in standby mode if syringe pump disabled.

DEN-13801: fixed reservoir switching from happening after first flush when flags are not set.

DEN-13801: compiler fixes. docstrings.

    • -22
    • +24
    /firmware/App/Modes/ModePreTreat.c
DEN-13801: Implmeneted wait signal for fills to ensure a dry reservoir for dry self tests. Changed dpi speed and pressure tolerances to be in line with recent sets of data.

    • -38
    • +93
    /firmware/App/Modes/ModePreTreat.c
Merge DEN-14150

Merge DEN-14150

    • -1
    • +4
    /firmware/App/Controllers/DialInFlow.h
DEN-14150: Update DialInFlow.c

    • -7
    • +29
    /firmware/App/Controllers/DialInFlow.c
Commented out for testing?

Commented out for testing?

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'develop' of ssh://dvm-linux02:7999/hd/hdfirmware into develop

Disabled UI POST test

DEN-14150: Update DialInFlow.c

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
HD-DEN-14166_DG HD Dev HD DG Dvt Update Part 4
HD-DEN-14166_DG HD Dev HD DG Dvt Update Part 4
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo