hdfirmware

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge remote-tracking branch 'origin/DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into develop

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Changed indentation and wrapping

DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to

Corrected alarm block condition

HD-DEN-15536_FW Usability Dry Self Tests Did Not Repeat After Dry Self Test Failure
HD-DEN-15536_FW Usability Dry Self Tests Did Not Repeat After Dry Self Test Failure
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-15536-fw-usability-dry-self-tests-did-not-repeat-after-dry-self-test-failure' into develop

DEN-15536: Fix +/- 15 mmHg check on offset.

Do we need to be able to override broadcast interval for this? I wouldn't think so. But if we do, need Dialin updated to support that.

Do we need to be able to override broadcast interval for this? I wouldn't think so. But if we do, need Dialin updated to support that.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Updated with publish function and overrides.

Updated with publish function and overrides.

DEN-15664 Added publish function adn overrides for treatment end sub state.

    • -8
    • +103
    /firmware/App/Modes/TreatmentEnd.c
DENBUG-64 removed DPo from bypass detection in blood leak driver

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
Merge remote-tracking branch 'origin/DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus' into develop

DEN-15839-alarm-199-triggers-during-mid-treatment-saline-bolus

Tell DG to stop heating dialysate when saline bolus stops dialin pump.

Merge remote-tracking branch 'origin/DEN-15741-dpi-does-not-stop-when-recirculation-is-paused' into develop

Merge remote-tracking branch 'origin/DEN-15719-blood-pump-stopped-alarms-211-212-124-125-not-relevant-for-alarms-that-force-tx-to' into develop

    • -7
    • +11
    /firmware/App/Modes/TreatmentStop.c
Merge branch 'staging' into DENBUG-64-sys-dialyzer-blood-leak-alarm-triggered-during-treatment

    • -21
    • +5
    /firmware/App/Modes/ModeTreatment.c
    • -16
    • +2
    /firmware/App/Services/Reservoirs.c
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -1
    • +1
    /firmware/App/Controllers/DialInFlow.c
    • -1
    • +1
    /firmware/App/Controllers/DialOutFlow.c
Merged DEN-14934 to staging.

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

We should send this countdown msg from a publish function that gets called by exec function for treatment end only once per second (see how this msg is sent by rinseback mode for example). And shou...

We should send this countdown msg from a publish function that gets called by exec function for treatment end only once per second (see how this msg is sent by rinseback mode for example).
And should send the message all the time. If in the wait for rinseback state, just send zeroes (and UI won't show countdown). If in paused state, always send 300 seconds for timeout and 300-(bloodSittingTimerCtr / 20) for countdown (don't let it go negative or wrap though). If an alarm dialog is being displayed, it should show the countdown for the whole 5 minutes (not just the last minute after warning).

HD-DEN-15664_Blood Pump Stop Time Exceeded Alarms 124 125 165 And 212 Alarm Timer Feature IS
HD-DEN-15664_Blood Pump Stop Time Exceeded Alarms 124 125 165 And 212 Alarm Timer Feature IS
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo