Modes

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-12224 fixed the switches enable high/low

  1. … 3 more files in changeset.
DEN-12224, DEN-12536 updated the post treatment mode

DEN-12224, DEN-12536 updated the post treatment's reservoir states

DEN-12224, DEN-12536 updated post treatment mode to drain both reservoirs

Dara removed this line of code after lab test w/o this change could not replicate the bug.

Dara removed this line of code after lab test w/o this change could not replicate the bug.

Merged DEN-12224

DEN-12224 updated the consumables self test

Disabled code for testing

Why is this necessary. This is already called somewhere else.

Why is this necessary. This is already called somewhere else.

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

DEN-12224 Fixed ticket DEN-12335 System stops during acid and bicarb check in pre-treatment

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 1 more file in changeset.
DEN-12224: fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-12224: Explicitly set actuators on pre-treatment transition to functions. Testing longer reservoir settle time.

  1. … 1 more file in changeset.
DEN-12224 updated the code

  1. … 1 more file in changeset.
DEN-12224 updated the code

  1. … 3 more files in changeset.
DEN-12224 Fixed DEN-12346 HD prime out of time alarm [ 227] Systems agreed to increase MAX_PRIME_TIME from 10 min to 15 minutes

DEN-12224 Fixed ticket DEN-12533 HD should accepting only value for arterial pressure high from -270 to 0

DEN-12224 Added build switch to disable syringe pump removed alarm[219] on system without syringe pump installed

DEN-12441 Reset blood sitting timer when turning on blood circulation.

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

  1. … 1 more file in changeset.
DEN-12224: Fixed merge error in ConsumableSelfTests.c

DEN-12224: Added explicit actuator sets to pre-treatment sub-mode transition functions. Renamed no cartridge self-test for pressure sensors. Removed no cartridge self-tests for blood and dialysate flow sensors and board temperature.

DEN-12441 Handle blood sitting timer when blood is not circulating. Update broadcast info.

DEN-12441 Added Recirc Dialysate Only and Blood Only sub states to Treatment End.

DEN-12224: Command DG to standby when HD faults. Removed redundant code.

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-12224-dg_hd_dev-switches-monitor

DEN-12224: only allow 3 options for acid treatment param.