DEN-12847 Code review comment updates.

DEN12847 Code review comments update 1.

  1. … 1 more file in changeset.
This #define should be at the top of the file.

This #define should be at the top of the file.

Merge remote-tracking branch 'origin/DEN-13141-miscellaneous_hd_issues_bb' into develop

  1. … 1 more file in changeset.
DEN-12974 DEN-13067 Return execUITest to original state.

HD-DEN-12847_SW Dev Sprint 71 Darren
HD-DEN-12847_SW Dev Sprint 71 Darren
DEN-13141, DEN-13067, DEN013116 Correct POST messaging for UI. Continue to use #define to remove fault testing. Issue is a synchronization issue between HD and UI. Remove check for generating ALARM_ID_PRE_TREATMENT_DRY_PRESSURE_TEST_FAILURE which could never occur.

Merged DEN-12931

  1. … 5 more files in changeset.
DEN-13141, DEN-13067 Correct POST messaging for UI.

  1. … 1 more file in changeset.
DEN-12931 merged staging and resolved the conflicts

  1. … 3 more files in changeset.
DEN-12847 Remove Debug output. Change drift limit.

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 6 more files in changeset.
DEN-12961 addressed VectorCAST issues raised by PiSquare

  1. … 2 more files in changeset.
DEN-12847 Zero out the LoadCellBackupReadings. Updated ouput debug message.

  1. … 1 more file in changeset.
Merge branch 'DEN-12931-dg_hd_dev-hd_dg_dvt-update' into develop

  1. … 5 more files in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Removed the message.

Removed the message.

Should probably log ms - so need to multiply by task interval.

Should probably log ms - so need to multiply by task interval.

This appears to be a timer counter (task intervals, not ms).

This appears to be a timer counter (task intervals, not ms).

These are redundant with above. Keep above definitions as they have the 'F' suffix that we want. Remove these.

These are redundant with above. Keep above definitions as they have the 'F' suffix that we want. Remove these.

Add blank line between functions.

Add blank line between functions.

Not sure if this is an issue - but looks like this clear alarm condition check can happen even when occlusion level after install hasn't been set yet.

Not sure if this is an issue - but looks like this clear alarm condition check can happen even when occlusion level after install hasn't been set yet.

Add space after ifs.

Add space after ifs.

We got recommendation to monitor arterial pressure even when not dialyzing if patient is still connected (so all other treatment mode states other than dialysis when not delivering saline bolus), b...

We got recommendation to monitor arterial pressure even when not dialyzing if patient is still connected (so all other treatment mode states other than dialysis when not delivering saline bolus), but didn't want to apply the user provided limits - just want to monitor for max pressure situation.

Im assuming that this if statement covers sub modes like Saline and Blood Prime. How do we know during blood prime, specifically, that we would not be dealing with an edge case in treatment paramet...

Im assuming that this if statement covers sub modes like Saline and Blood Prime. How do we know during blood prime, specifically, that we would not be dealing with an edge case in treatment parameters for the patient (Ex: -300 mmHg to -270 mmHg) and triggering a false positive alarm?

Note: I dont know actual patient bp values but these are currently acceptable treatment parameters for arterial pressure limits.