DEN-13460: Fixed syringe home issues.

  1. … 1 more file in changeset.
DEN-13460: Added persistence to pump direction error count alarm. Fixed comments and pos A average calc for valves.

  1. … 4 more files in changeset.
DEN-13167 Code review comment updates 1.

Updated the self tests tolerance

Can we remove this now?

Can we remove this now?

I though cycle was an enum (start/final), not a boolean. Should we be looking for final (not TRUE)?

I though cycle was an enum (start/final), not a boolean. Should we be looking for final (not TRUE)?

Better name is reservoir use cycle steady sample points (during treatment, but we're in Dialysis.c here so treatment can be inferred).

Better name is reservoir use cycle steady sample points (during treatment, but we're in Dialysis.c here so treatment can be inferred).

HD-DEN-13167_SW Dev Sprint 74 Darren
HD-DEN-13167_SW Dev Sprint 74 Darren
Why removed?

Why removed?

Why removed? I thought we needed this for VC?

Why removed? I thought we needed this for VC?

F has been added.

F has been added.

No. Code removed.

No. Code removed.

If no actions handled in this mode, add a comment saying so.

If no actions handled in this mode, add a comment saying so.

Yes, at bottom of exec function above.

Yes, at bottom of exec function above.

Remove blank line.

Remove blank line.

Can we remove this now?

Can we remove this now?

Why?

Why?

Why?

Why?

HD-DEN-13141_Miscellaneous HD Issues BB
HD-DEN-13141_Miscellaneous HD Issues BB
Merged DEN-13460

  1. … 5 more files in changeset.
Merge branch 'DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2' of ssh://dvm-linux02:7999/hd/hdfirmware into DEN-13460-dg_hd_dev-hd_dg_dvt-update-part-2

  1. … 2 more files in changeset.
DEN-13460 updated the code with DVT changes

  1. … 6 more files in changeset.
DEN-13460: Updated thresholds for wet transfer test. Updated valve home function to just go to position A if already homed.

  1. … 1 more file in changeset.
DEN-13167 Update Load Cell drift to use Start to Final only.

Merge branch 'staging' into DEN-13141-miscellaneous_hd_issues_bb

DEN-13141 DEN-13149 Resolved alarm 219 (ALARM_ID_HD_SYRINGE_PUMP_SYRINGE_REMOVED). Removed check for presence of syringe pump.

DEN-13460: Updated arterial pressure limits range. Added actuators homing to standby start state.

  1. … 2 more files in changeset.
Check NEARLY_ZERO #define. Value should be 0.00000001F. I believe previous review commented indicated that all floating point #defines should have "F" at the end.

Check NEARLY_ZERO #define. Value should be 0.00000001F. I believe previous review commented indicated that all floating point #defines should have "F" at the end.

SW_CONFIG_ENABLE_VALUE should be first in the conditional.

SW_CONFIG_ENABLE_VALUE should be first in the conditional.

Need #define for minimum vRef

Need #define for minimum vRef