For DG faults, let's be consistent. DG fault = TRUE Clear Immediate = TRUE (alarm condition moot because DG going to fault mode) Stops = TRUE (HD pauses treatment for alarm) No Clear = FALSE (so HD...
For DG faults, let's be consistent.
DG fault = TRUE
Clear Immediate = TRUE (alarm condition moot because DG going to fault mode)
Stops = TRUE (HD pauses treatment for alarm)
No Clear = FALSE (so HD can clear alarm when user selects option)
No Resume = TRUE (cannot resume treatment w/o functioning DG)
No Rinseback = FALSE (no reason to prevent patient getting their blood back just because DG failed)
No End Treatment = FALSE (should try to end treatment as normally as possible w/o DG - won't be able to drain reservoirs though)
No Blood Recirc = FALSE (why not recirc blood?)
No Dialysate Recirc = TRUE (may not be possible depending on what's wrong w/ DG)
Clear Only = FALSE