Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Fixed merge conflicts with DEN-13801.

  1. … 2 more files in changeset.
DEN-13801: Updates to reservoir mgmt during dry cartridge pressure leak test.

  1. … 1 more file in changeset.
DEN-14166 updated the self tests

HD-DEN-13903_Sprint 80 MG Air Trap Optimization
HD-DEN-13903_Sprint 80 MG Air Trap Optimization
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
DEN-13903 merged staging

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-12744 Start adding code to preload position for Heparin despenser before user installs.

  1. … 2 more files in changeset.
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH

Done.

Done.

Done.

Done.

Fine for now. This one should probably just reset to zero on else instead of decrement. RESOLVED in CODE WALKTHROUGH.

Fine for now. This one should probably just reset to zero on else instead of decrement.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

Fixed. Thanks!

Fixed. Thanks!

Fixed. Using the old style for now. Will rework both alarms to use new persistence when we revisit the dialyzer timeout alarms in priming.

Fixed. Using the old style for now. Will rework both alarms to use new persistence when we revisit the dialyzer timeout alarms in priming.

DEN-13903: Reworked upper level persistence. Added todo for dialyizer timeout. code review changes.

  1. … 1 more file in changeset.
Fixed. Thanks!

Fixed. Thanks!

Seems like largest calculated t/o = 38.2 seconds vs. 60 second alarm t/o. Let's leave as is for now, but add a TODO comment here to remove alarm t/o and associated alarm.

Seems like largest calculated t/o = 38.2 seconds vs. 60 second alarm t/o.
Let's leave as is for now, but add a TODO comment here to remove alarm t/o and associated alarm.

DEN-13903: addressed code review comments.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.