Merge branch 'DEN-14830-incorrect-dialysate-temperature-alarm-triggered' into develop

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
DEN-14834 updated the code to control the primary heater

  1. … 2 more files in changeset.
Merge branch 'DEN-13154-heparin-syringe-empty-position-incorrect' into develop

  1. … 1 more file in changeset.
DEN-13154: Further fix to DEN-14814 re: do not require syringe removal to re-do preload operation.

  1. … 1 more file in changeset.
RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

DEN-14834 changed one of the pre-treatment target fills from 600 to 750mL Changed the reservoir's upper limit in treatment from 1950 to 1900mL

  1. … 1 more file in changeset.
Fixed.

Fixed.

Fixed.

Fixed.

Done.

Done.

HD-DEN-14457_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Address comment

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm. And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm. D...

Why alarm if above low temp? I think >42 or > tgt+4 should be two ways to get same non-safety high temp alarm.
And I think <33 or < tgt-4 should be two ways to get same non-safety low temp alarm.
Did Systems want to separate these? My understanding was not to separate.

Add blank line before return.

Add blank line before return.

Fixed merge conflicts.

  1. … 1 more file in changeset.
DEN-13154: Removed dry self-test timeout alarm. Fixed bug DEN-14884.

  1. … 2 more files in changeset.
DEN-14834 added code to calculate the heating parameters in pre-treatment and send it to DG in the pre-treatment reservoir management

HD-DEN-14834_RO Flow Drops TO Zero And CP Measured Speed 1500 Temporarily During Pre Treatment
HD-DEN-14834_RO Flow Drops TO Zero And CP Measured Speed 1500 Temporarily During Pre Treatment
HD-DEN-14457-2_Reservoir Drains Before Treatment With Pre Treatment Build Flags Off

Checkpoint. Simplifying AIR_PUMP bug fix.

This will hard fault the HD. Please change to getTreatmentParameterU32()

This will hard fault the HD. Please change to getTreatmentParameterU32()

This will hard fault the HD. Please change to getTreatmentParameterU32()

This will hard fault the HD. Please change to getTreatmentParameterU32()

HD-DEN-14689_Messaging Bugs_MSG_ID_HD_POST_FINAL_TEST_RESULT_fix
HD-DEN-14689_Messaging Bugs_MSG_ID_HD_POST_FINAL_TEST_RESULT_fix
DEN-14689: DEN-14634: fixed HD_POST_FINAL_TEST_RESULT being sent to be sent 1 time instead of 40 times over 2s.

Merge branch 'DEN-14689-messaging-bugs' into develop

  1. … 4 more files in changeset.
Merge branch 'staging' into DEN-14689-messaging-bugs