DEN-14150 reversed the previous code changes. Calling resetBloodPumpRotorCount and resetDialInPumpRotorCount

DEN-14150 addressed a code review comment by making resetSignalFlags and resetAlarmSignalFlags functions to be called in pretreatment

DEN-14150 removed a blank line

DEN-14150 merged staging

  1. … 5 more files in changeset.
DEN-14099 Added pressure / release cycles to remove air bubbles during prime dialyzer.

Merged DEN-14175

  1. … 3 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
DEN-14175 addressed the code review comments. Updated the reservoir switch command in pre treatment and added the getHWConfig function in FPGA

  1. … 2 more files in changeset.
Merge branch 'staging' into DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 3 more files in changeset.
Merge branch 'DEN-14170-sw_dev_sprint_82_darren' into staging

  1. … 1 more file in changeset.
Merge branch 'DEN-14170-sw_dev_sprint_82_darren' into develop

  1. … 1 more file in changeset.
DEN-14175 merged staging and resolved the conflicts

  1. … 2 more files in changeset.
DEN-14175 addressed the code review comments

Merge branch 'DEN-14150-hd-dialysate-flow-estimation' into develop

  1. … 2 more files in changeset.
DEN-14170 Updated for code review comments.

  1. … 1 more file in changeset.
DEN-14150: Address code review comments

  1. … 2 more files in changeset.
Merged DEN-14175

  1. … 6 more files in changeset.
DEN-14170 Fix TODO with new fault.

  1. … 1 more file in changeset.
DEN-14175 updated the reservoir switch command to send the trimmer use the previous duty cycle or estimate the duty cycle

  1. … 5 more files in changeset.
DEN-14175: Added f/w support for alarm dialog blood stopped countdown while in rinseback state per DEN-10791.

DEN-14175: Stop syringe pump on transition to fault mode only. Continuous stopping of pump was preventing DAC write on force sensor check failure. DEN-14254.

Merge branch 'DEN-14175-dg_hd_dev-hd_dg_dvt-update-part-5' into develop

  1. … 2 more files in changeset.
DEN-14175 implemented the FGPA hardware configuration register, changed the trimmer heater start/stop command to HD command and not the log in command. Added a command to benign the polynomial calibration values of the temperature sensors

  1. … 7 more files in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
Merge branch 'DEN-13801-sw_dev-sprint-79_mg' into develop

DEN-13801: cosmetic code review changes.

DEN-13801: reverted priming changes to explicitly set active reservoir.

DEN-13801: fixed getpretreatmentfillvolume missing brackets for conditional statements.

DEN-13801: fixed rsvr mgmt causing r2 to fill to 700 instead of 1300.