Modes

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
DEN-8103: Addressed code review

  1. … 2 more files in changeset.
Merge branch 'DEN-7605-hd_dg_dev-self-tests' into develop

  1. … 4 more files in changeset.
DEN-7605: Added POST test for alarm audio.

  1. … 4 more files in changeset.
Moved FPGA POST to last test.

DEN-7605: Added flow sensor comm errors.

  1. … 4 more files in changeset.
DEN-8103: Added mismatch check for alarm enum and alarm table

  1. … 2 more files in changeset.
DEN-8103: Initial alarm list implementation

  1. … 5 more files in changeset.
Merge branch 'DEN-7605-hd_dg_dev-self-tests' into develop

  1. … 2 more files in changeset.
DEN-7605: Added placeholder for s/w integrity POST test.

DEN-7605: Added reservoir mgmt changes to prevent UF overfill of reservoir.

  1. … 2 more files in changeset.
Fixed merge conflicts.

  1. … 6 more files in changeset.
DEN-7605: Added POST messaging.

  1. … 3 more files in changeset.
DEN-7860: Corrected spelling issue

DEN-8103: Added persistent to alarm no new treatment flag

  1. … 2 more files in changeset.
Addressed merge conflicts with staging

  1. … 4 more files in changeset.
DEN-7860: Addressed integration issues

  1. … 1 more file in changeset.
DEN-8103: Added DG restart check to sample water sub-mode

DEN-7605: Updated UF loadcell acceleration check.

  1. … 2 more files in changeset.
DEN-8103: Added dg restarted fault to sample water and reject treatment start if DG not ready

DEN-8103: Addressed code reviews

  1. … 5 more files in changeset.
DEN-8103: Added alarms for pre-treatment self-tests timeout

DEN-7860: Addressed code review comments

DEN-8103: Added treatment log data

  1. … 5 more files in changeset.
DEN-7860: Added air bubble detectors self-test and priming

  1. … 1 more file in changeset.
DEN-7860: Addressed code review comments

  1. … 3 more files in changeset.
DEN-8103: Added firmware image integrity check for POST

  1. … 4 more files in changeset.
DEN-8103: Removed Heparin concentration in treatment log

DEN-7605: Added pump direction error counter check and alarm. Added build switch to allow change to treatment duration to as low as 1 minute.

  1. … 5 more files in changeset.
DEN-7860: Finished implementing HD Air Bubble Detectors

  1. … 6 more files in changeset.
DEN-8103: Added battery communication fault

  1. … 2 more files in changeset.