DEN-6372: Fixes per lab testing. Added function to AlarmMgmt to determine whether any alarms w/ stop property are active. Changed treatment mode handling of alarm stop from signal to modes approach to modes polling alarm mgmt approach. Added f/w support for override of treatment time remaining.

    • -149
    • +29
    ./Services/SystemCommMessages.c
DEN-6078 merged staging and resolved merge conflicts

DEN-6078 commented the count to PWM conversion macros

DEN-6372: Addressed code review feedback.

DEN-6372: Addressed code review feedback.

DEN-6372: Fixed bug in treatment recirc progress message broadcast.

DEN-6402: Switch self-test for flow meters

DEN-5980 removed old calibration functions from the drivers

DEN-6372: Fixes per lab testing results.

DEN-5887: No adjustment to DPo initial PWM at this time. Restored P and I control terms to original settings.

DEN-5980 added more NV comm messages

DEN-5980 added comm messages to be able to set and get NV records

  1. … 1 more file in changeset.
DEN-6372: Fixes per test results of lab testing.

DEN-6402: Set final and start reservoir volume at the same time

DEN-6372: Fixed signal handling per lab test results.

DEN-6402: Fixed DG commands

DEN-6372: doxygen fixes.

DEN-6372: Addressed code review feedback.

DEN-6372: reoved redundant flag resets.

DEN-6402: Used enums from DG defs

DEN-6372: consolidated the two treatment end sub-modes into one.

DEN-6372: Reset rinseback timer whenver BP is running. Set rinseback timeout to 5 minutes.

DEN-6372: set alarm responses in treatment stop sub-mode.

DEN-6372: Added timeout & countdown to rinseback and recirc status messages to UI.

DEN-6402: Added DG valve setting change command

DEN-6402: Stored command response for each possible command

DEN-6402: Added wet self-test for FMD and load cell

DEN-5887: Refactored load cell acceleration alarm. Tuned UF control.

DEN-6372: consolidated dialysis transition and start functions. Re-initialize treatment modes before starting a new treatment workflow.

DEN-6402: Added alarm signal action handlers