DEN-12224: Slowed blood side prime rates to 150 mL/min maximum.

did we want to keep these loose tolerances from the demo?

did we want to keep these loose tolerances from the demo?

We typically put #define like this in the source file. Is there a reason why this one is in the header file specifically?

We typically put #define like this in the source file. Is there a reason why this one is in the header file specifically?

This variable is defined but not used.

This variable is defined but not used.

I wanted to see how new #define worked. I think it's better, so we can remove this now.

I wanted to see how new #define worked. I think it's better, so we can remove this now.

Do we need to keep this define?

Do we need to keep this define?

or this alarm?

or this alarm?

Do we need to keep this alarm?

Do we need to keep this alarm?

HD-DEN-12224_DG HD Dev Switches Monitor
HD-DEN-12224_DG HD Dev Switches Monitor
Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

DEN-12224: Increased Heparin bolus rate to 720 mL/hr.

DEN-12224: Fixed alarm 109 part of bug DEN-12339.

DEN-12441 Added Alarm management for noBloodRecirc with init, set, call to check status.

DEN-12224 updated the payload data type

DEN_12224 renamed a data payload

DEN-12224 updated the publish data that did not have a structure

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Updated the build switches

Bamboo Commit: Updated HDCommon.h with build versions from Bamboo

Merge branch 'DEN-12224-dg_hd_dev-switches-monitor' into develop

DEN-12441 Added No Blood Recirculate flag to Alarms and messages.

DEN-12441 Added placeholder for Treatment Time NVRAM. Added Sean's changes to Alarms for No Recirculate Dialysate.

DEN-12224: Fixed bug DEN-12382.

DEN-12224: Added no dialysate recirc bit to alarm status message. Added some alarm properties to the alarm triggered message.

DEN-12224: Updated flow to RPM estimation for flow vs speed check.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.

RESOLVED in CODE WALKTHROUGH.