HD-DEN-15858_NO Timer Countdown IS Visible For The Confirmation OF Chemical Flush Sample
HD-DEN-15858_NO Timer Countdown IS Visible For The Confirmation OF Chemical Flush Sample
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
HD-DEN-15886_Air Pump Pushes Air Into The Heparin Syringe During Dry Self Tests
HD-DEN-16016_Bug Unable TO Start The Saline Bolus 3
HD-DEN-16016_Bug Unable TO Start The Saline Bolus 3
HD-DEN-15512_DG HD Failing TO Set Calibration Record
HD-DEN-15512_DG HD Failing TO Set Calibration Record
This is a second, separate branch addressing this ticket, and these changes should only be in develop (for testing) at this point. Eventually, this branch should replace the branch you reverted out...

This is a second, separate branch addressing this ticket, and these changes should only be in develop (for testing) at this point. Eventually, this branch should replace the branch you reverted out of staging.

stopDialysis() almost exactly mirrors the four lines I commented out, except the pump "hardstop" routines also manage the control variables that start and stop the PI controllers. This should still make zero difference to the case Nick is investigating since these changes are local to a saline bolus, which is not run at all while active ultra filtration is being run, and only affect the initial bolus state (0), while stopping the pumps. State 1, puts the pumps back into closed loop control, once the valves separate the patient, the blood and the dialysate from the dialyzer.

These changes are in staging right now. Do we have to stop dialysis in this state?

These changes are in staging right now. Do we have to stop dialysis in this state?

HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
HD-DEN-15839_Alarm 199 Triggers During Mid Treatment Saline Bolus
HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed
HD-DEN-15893_Cartridge Lever IS Open Alarm When Cartridge Lever IS Closed
HD-DEN-15719_Blood Pump Stopped Alarms 211 212 124 125 Not Relevant For Alarms That Force TX TO
HD-DEN-15719_Blood Pump Stopped Alarms 211 212 124 125 Not Relevant For Alarms That Force TX TO
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
HD-DEN-15804_Blood Flow Rate Accuracy Results Fail OR Are Close TO Failing
RESOLVED IN CODE WALKTHROUGH

RESOLVED IN CODE WALKTHROUGH